Oi Lei Wong1, Jing Yuan1, Yihang Zhou1, Siu Ki Yu1, and Kin Yin Cheung1
1Medical Physics and Research Department, Hong Kong Sanatorium & Hosptial, Hong Kong, Hong Kong
Synopsis
Geometric accuracy is critical
for radiotherapy and is one major concern in the application of MRI
in radiotherapy. Since the geometric distortion and B0
inhomogeneity are related, we aim to evaluate the geometric
distortion due to ΔB0
variation at inhalation and exhalation breath-hold. Based on our
results, larger geometric distortion was noted during inhalation than
exhalation.
Purpose
Geometric accuracy is one major concern when applying MRI for radiotherapy purpose. MR geometric distortion is related to various factors, and B0 inhomogeneity (ΔB0) is one of them. Besides ΔB0 due to magnet imperfection, B0 inhomogeneity can be also related to organ shape, tissue-specific magnetic susceptibility and respiration. Recently, change in the ΔB0 due to respiratory states was reported in the breast [1]. Dynamic geometric distortion in the lung has been investigated via ΔB0 simulation in another study [2]. However, in vivo ΔB0 and the associated geometric distortion in the liver have rarely been reported. In this study, we aimed to evaluate the geometric distortion of liver in vivo due to ΔB0 variation at inhalation and exhalation breath-hold.Method
Liver imaging was conducted in 7 healthy volunteers using a 1.5T Siemens MR scanner (Aera, Siemens Healthineers, Erlangen, Germany). Coronal images at inhalation breath-hold and exhalation breath-hold were acquired using a dual echo GE sequence (TR=70-137ms, TE1/TE2 = 4.76/9.52ms, slice thickness = 8-10mm, 5-11 slices, 38-48cm FOV, matrix=192x192 or 110x110, Bandwidth=260 or 1000Hz/pixel, phase encoding (PE) direction = left-right). The corresponding ΔB0 map was first calculated from phase images using customized Matlab script (Mathworks, Natick, MA) and then geometric distortion along frequency encoding (FE) direction (ΔdFE) was derived. For each image set, three square ROIs, with two ROIs at the right liver lobe (proximal and distal to the lung, hereafter RLp and RLd) and one ROI at the left liver lobe (LL), were drawn (Figure 1). To avoid phase wrapping, all ROIs were chosen away from the interface with large susceptibility difference. A Wilcoxon signed-rank test was performed to compare the ΔdFE at inhalation and exhalation breath-hold in three ROIs.Results
As illustrated in Figure 2, the
measured ΔdFE
at inhalation in the RLd
was significantly higher than that at exhalation (0.21±0.16mm and
0.13±0.08mm, P<0.05).
Larger ΔdFE
was also observed in the RLp (inhalation: 0.64±0.44mm and
exhalation: 0.53±0.39mm, P=0.11)
and LL (inhalation: 0.16±0.11mm and exhalation: 0.13±0.16mm,
P=0.30).
Significantly higher ΔdFE
was also observed in
RLp (inhalation: 0.64±0.44mm and exhalation: 0.53±0.39mm) when
comparing to LL (inhalation: 0.16±0.11mm, P<0.05
and exhalation: 0.13±0.16mm, P<0.05)
and RLd (inhalation: 0.21±0.16mm, P<0.05
and exhalation: 0.13±0.08mm, P<0.05).
When comparing between RLd and LL, no significant difference in ΔdFE
was
observed at inhalation (RLd: 0.21±0.16mm and LL: 0.16±0.11mm,
P=0.14)
and at exhalation (RLd: 0.13±0.08mm and LL: 0.13±0.16mm, P
> 0.05).Discussion
In this study, we have presented
the geometric distortion pattern in the liver due to ΔB0
variation at inhalation and exhalation. For
all ROIs, larger ΔdFE
was observed at inhalation breath-hold than at exhalation
breath-hold, which may be related to the amount of air presence in
the lung.
As illustrated in Figure 3, the
largest geometric distortion was observed mostly at the location with
large magnetic susceptibility difference (e.g. liver-lung interface).
The largest ΔdFE
was observed at the RLp because RLp was chosen at a location close to
the liver-lung interface. On the other hand, ΔdFE
of LL was smaller than ΔdFE
of RLp though ROIs of LL and RLp were chosen at a similar proximity
to the lung. The smaller ΔdFE
observed in LL than in RLp might be explained by the lower tissue
magnetic susceptibility difference between the heart and the liver.
Note that LL is more prone to cardiac motion but this passive motion
of LL should be negligible in the ΔTE
of 4.76ms. There are
several limitations in this study: (1) the mismatch in the each ROI
location between inhalation and exhalation breath-hold was ignored;
(2) geometric distortion could only be assessed along FE direction
and (3) small sample size used in this study. Nonetheless, compared
to the reported 12 to 26 mm of the liver respiratory motion along SI
by Rohlfing et al
[3], our observed ΔdFE
was generally small (mostly less than 1mm) for both inhalation and
exhalation breath-hold for all ROIs, indicating the small
contribution of geometric distortion due to ΔB0
variation to the positional uncertainty of the liver during
respiration.Conclusion
The geometric distortion in liver due to ΔB0 variation at inhalation and exhalation has been obtained in this study. For all ROIs and both respiratory states, the calculated ΔdFE (along SI direction) was generally less than 1mm.Acknowledgements
No acknowledgement found.References
[1] Bolan PJ, Henry P, Baker EH,
et al. Measurement and correction of respiration-induced B0
variations in breast 1H MRS at 4 Tesla. MRM 52: 1239-1245 (2004)
[2] Stanescu T and Jaffray D.
Investigation of the 4D composite MR image distortion field
associated with tumor motion for MR-guided radiotherapy. Medical
Physics 43: 1550-62 (2016)
[3] Rohlfing T, Maurer C, O’Dell
W, et al. Modeling liver motion and deformation during the
respiratory cycle using intensity-based registration of gated MR
images. Med. Phys. 32: 427-32 (2004)