Petrice M. Cogswell1, Jeroen C.W. Siero2,3, Guillaume Gilbert4, Taylor Davis1, Allison O. Scott1, Katie Lants1, Helen B. Mahany1, Jennifer M. Watchmaker1, Jeroen Hendrikse2, and Manus J. Donahue1
1Vanderbilt University Medical Center, Nashville, TN, United States, 2Radiology, University Medical Center Utrecht, Netherlands, 3Spinoza Center for Neuroimaging, Amsterdam, Netherlands, 4MR Clinical Science, Philips Healthcare Canada, Markham, Canada
Synopsis
This
work compares vessel wall SNR and CSF suppression from multiple approaches to
determine optimal imaging parameters for intracranial VWI at the
clinically-available field strength of 3T. T1-weighted TSE acquisition using
variable refocusing angle pulse-train and DANTE preparation provides for blood
and CSF suppression while maintaining adequate vessel wall SNR. The use of a
variable refocusing pulse train with sweep of 40-120° provides improved
performance compared to a sweep of 50-120°. Variation of the DANTE flip angle showed
that a flip angle of 8° provides good CSF suppression with minimal SNR loss
compared to flip angles of 10 and 12°.
Introduction
Vessel
wall imaging (VWI) of the intracranial vasculature has recently become
available and may be of clinical use for differentiating types of vascular
disease1 or evaluating
atherosclerotic lesions that are not yet hemodynamically significant, as shown
in other vessels2,3. To accurately
assess the vessel wall, imaging must provide adequate suppression of intraluminal
blood and surrounding cerebrospinal fluid (CSF) as well as maintain high
signal-to-noise ratio (SNR) at the spatial resolution of the intracranial
vessel walls (0.2-0.6 mm). Several methods have been implemented to null blood
and CSF signal including variable flip angle turbo-spin-echo readouts4,5, inversion
recovery preparations6, post-readout
anti-DRIVE modules7 and DANTE preparations8–10. However, while a
broad consensus on the relevance of intracranial VWI exists11, a consensus on
optimal clinical sequence parameters is currently unavailable. Therefore, the goal of this work is to compare
vessel wall SNR and CSF suppression from the above candidate approaches to determine
optimal imaging parameters for intracranial VWI at the clinically-available
field strength of 3T. Materials and Methods
Imaging:
Scans were performed on a 3T whole body scanner (Philips Medical Systems, Best,
The Netherlands) using body coil transmission and a 32 channel receiver head
coil. The base T1-weighted
3D-TSE acquisition was axial orientation with TSE factor =56, TR/TE=1500/33ms, field-of-view=200x166x45
mm3, spatial resolution=0.5x0.5x1.0mm3, and duration=4min42s.
Blood and CSF suppression methods tested were motivated by Bloch equation
simulations and prior experimental reports and included: (i) T1-weighted TSE
with variable refocusing angle pulse-train using sweep=40-120° and sweep=50-120° (ii) T1-weighted
TSE with DANTE preparation8,9 (repetitions=300,
interval=1.1 ms, gradient strength in three directions=22.5mT/m) using 8, 10,
and 12º flip angles, (iii) T2-weighted
TSE readout (TSE factor=15, TR/TE=4000/83ms, field-of-view=200x166x45 mm3,
spatial resolution=0.5x0.4x1.0 mm3, duration=5min12s), and (iv)
T1-weighted inversion recovery TSE readout6 (TSE factor=25,
TR/TI/TE=800/400/36ms, field-of-view=250x250x170 mm3, spatial
resolution 0.5x0.5x1.0 mm3, duration 8min48s). Subjects: Ten healthy adults (mean age 39±17 years, range 22-60
years) were imaged with each of the above sequences. Inclusion criterion was
age 18-65 years. Exclusion criteria included prior stroke, TIA, heart attack,
diabetes, smoking, and BMI > 40. Analysis:
Analysis was performed using Osirix (Pixmeo, Bernex, Switzerland). SNR of the
CSF (measured in the prepontine cistern), parenchyma (central pons), and vessel
wall (basilar artery) were calculated as the ratio of the mean signal to
standard deviation of thermal noise. ROIs with areas approximately 10 mm2
were used for the background, CSF, and pons (Figure 1). Vessel wall ROIs were
hand drawn. The ROIs were drawn on acquisition (i) using sweep=40-120°and
copied to the others series in that exam. Mean and standard deviations of the signal
and SNR were calculated for each CSF suppression technique. A Wilcoxon signed-rank
test was used to evaluate measurements from different scan sequences applied in
the same subjects.Results
Representative
images from one subject are shown in Figure 2. Table 1 summarizes the SNR
results. The inversion recovery
acquisition was of insufficient quality to make vessel wall measurements and
was therefore not included in analysis. CSF suppression was improved (one-sided
p=0.034) when using sweep=40-120° compared to sweep=45-120°. CSF suppression
was significantly improved (two-sided p<0.01) with the addition of the DANTE
preparation with approximately 3x reduction in mean CSF signal between variants
(i) and (ii). However, the DANTE
preparation also resulted in significantly reduced (p<0.01) vessel wall signal
of approximately 2x between variants (i) and (ii). With the DANTE scans, CSF
suppression improved (two-sided p<0.01) from DANTE flip angle 8° to DANTE 10
and 12° but not between DANTE 10 and 12°. Vessel wall signal was also reduced
going from DANTE flip angle 8° to DANTE 12° (two sided p<0.01) but not from
DANTE 10 to 12°. The T2-weighted TSE readout (iii) provided similar CSF
suppression and vessel wall signal as the DANTE preparation with 8° flip angle.Conclusions
The
variable flip angle T1-weighted TSE acquisition provides blood and CSF
suppression that allows for visualization of the vessel wall in healthy
subjects, and the CSF suppression is improved with sweep=40-120° versus
sweep=50-120°. Addition of the DANTE preparation results in further reduction
in CSF signal, eradicating CSF signal that may obscure or simulate vessel wall
lesions. However, the improved CSF suppression comes at the cost of a decrease
in the vessel wall signal, which is likely in part due to the pulsation of the
vessel wall, affected by the motion sensitive DANTE pulse10. Addition of the
DANTE preparation with a flip angle of 8° minimizes SNR losses. Based on these
comparisons, we recommend a 0.5 mm variable refocusing angle (40-120°) TSE
acquisition and DANTE preparation with flip angle 8° in patients with
intracranial stenosis.
Acknowledgements
No acknowledgement found.References
1. Mossa-Basha,
M. et al. Multicontrast high-resolution vessel wall magnetic resonance
imaging and its value in differentiating intracranial vasculopathic processes. Stroke
J. Cereb. Circ. 46, 1567–1573 (2015).
2. Babiarz, L. S., Astor, B., Mohamed, M. A. & Wasserman,
B. A. Comparison of gadolinium-enhanced cardiovascular magnetic resonance
angiography with high-resolution black blood cardiovascular magnetic resonance
for assessing carotid artery stenosis. J. Cardiovasc. Magn. Reson. Off. J.
Soc. Cardiovasc. Magn. Reson. 9, 63–70 (2007).
3. Glagov, S., Weisenberg, E., Zarins, C. K., Stankunavicius,
R. & Kolettis, G. J. Compensatory enlargement of human atherosclerotic
coronary arteries. N. Engl. J. Med. 316, 1371–1375 (1987).
4. Qiao, Y. et al. Intracranial arterial wall imaging
using three-dimensional high isotropic resolution black blood MRI at 3.0 Tesla.
J. Magn. Reson. Imaging Jmri 34, 22–30 (2011).
5. Hennig, J., Weigel, M. & Scheffler, K. Multiecho
sequences with variable refocusing flip angles: optimization of signal behavior
using smooth transitions between pseudo steady states (TRAPS). Magn. Reson.
Med. 49, 527–535 (2003).
6. Kolk, A. G. van der et al. Intracranial Vessel Wall
Imaging at 7.0-T MRI. Stroke 42, 2478–2484 (2011).
7. Harteveld, A. A. et al. High-resolution intracranial
vessel wall MRI in an elderly asymptomatic population: comparison of 3T and 7T.
Eur. Radiol. (2016). doi:10.1007/s00330-016-4483-3
8. Li, L., Miller, K. L. & Jezzard, P. DANTE-prepared pulse
trains: a novel approach to motion-sensitized and motion-suppressed
quantitative magnetic resonance imaging. Magn. Reson. Med. 68,
1423–1438 (2012).
9. Wang, J. et al. Joint blood and cerebrospinal fluid
suppression for intracranial vessel wall MRI. Magn. Reson. Med. 75,
831–838 (2016).
10. Viessmann, O., Li, L., Benjamin, P. &
Jezzard, P. T2-Weighted intracranial vessel wall imaging at 7 Tesla using a
DANTE-prepared variable flip angle turbo spin echo readout (DANTE-SPACE). Magn.
Reson. Med. (2016). doi:10.1002/mrm.26152
11. Mandell, D. M. et al. Intracranial
Vessel Wall MRI: Principles and Expert Consensus Recommendations of the
American Society of Neuroradiology. Am. J. Neuroradiol. (2016).
doi:10.3174/ajnr.A4893