Uday Krishnamurthy1,2, Brijesh K Yadav1, Pavan K Jella2, Swati Mody2, Edgar Hernandez-Andrade3,4, FeiFei Qu2, Anabela Trifan2, Ewart M Haacke1,2, Sonia S Hassan3,4, Roberto Romero4, and Jaladhar Neelavalli1,2
1Biomedical Engineering, Wayne State University, Detroit, MI, United States, 2Radiology, Wayne State University, Detroit, MI, United States, 3Obstetrics and Gynecology, Wayne State University, Detroit, MI, United States, 4Perinatology Research Branch, NICHD/NIH/DHHS, Detroit, MI, United States
Synopsis
To show that the isotropic gradient delay issue can be addressed by a simple shift of the readout-window . We also report the use of a fully flow-compensated, readout-shifted 2D radial gradient echo sequence to perform non-contrast MRA of the human fetus in-utero
Introduction
Recent
studies have shown the feasibility of using non-contrast time of flight (TOF)
MRA in third trimester fetuses1. However, predominantly due to concerns
of motion, it is challenging to consistently obtain diagnostic quality and
replicate the same in younger gestation fetuses. On the other hand, it is known
that oversampling the central k-space can provide immunity to motion related artifacts.
However, radial trajectories are very susceptible to gradient delays causing
relative shifts between the sampling window and the read-out gradient (shift in
the center of the echo). Manufacturers address this delay by modifying the
moment of the dephasing lobe which ensures that the echo-center occurs at the
center of the sampling window. This approach is robust for a non-flow
compensated gradient structure; addition of a bipolar gradient for flow
compensation (FC), as required for MRA (or) SWI, makes it difficult to correct
for the gradient delay by moment compensation because both the zeroth and first
moment of the readout need to be simultaneously zero at the echo coinciding with the read-out
window. In this work, we show that a simple shift of the readout-window to
account for the isotropic gradient delays can address this issue. We report the
use of a fully FC, readout-shifted 2D radial gradient echo sequence to perform
non-contrast MRA of the human fetus in-utero. The added advantage of using
radial sampling is the ability to do FC in all dimensions without additional
time. With long TRs used in fetal angiography for reduced SAR1, multi
echo readouts were also possible, allowing for T2* quantification of fetal
blood at the same time. Methods
All MRI scans were performed on a Siemens Verio 3.0T.
The quality of the gradient shifting concept was tested on adult volunteers. The
isotropic gradient delay was calculated as Δt = 0.42*Tos
– 3.7μs 2 and the readout was shifted by the same time (fig 1). A
2D TOF MRA1 sequence was first acquired followed by a 2D radial sampling
scheme (rMRA), covering almost identical volumes in the fetuses. The same sequence
timing parameters and field-of-view as the Cartesian were used for the radial
acquisition with number of radial spokes = 361 (over 360o) and 720
points per readout with ~30% asymmetry in the readout for both cases. The in-plane
resolution of the radial scans was higher (0.42mm vs. 0.57mm) compared to
Cartesian data due to the differences in the number of points per readout (720
vs. 512). 3D rendering and maximum intensity projections (MIP) of the fetal
vessels were created for visualization and anatomical landmarks. In a subset of
the cases, T2* was estimated by fitting the dual echo data to a mono-exponential
function. T2* of the fetal blood was measured by placing the ROI in the
descending Aorta. Results
Shifting the sampling window to compensate for
the gradient delay improves the image quality as demonstrated in Figure 1a. It
is important to note the improvement in the edge definition of the superior
saggital sinus (SSS). Fetal MRA data was acquired from 7 subjects (4 third & 3 second trimester) with
uncomplicated singleton pregnancies. In cases of no fetal motion, particularly
in the third trimester, the rMRAs were as good as those obtained from regular Cartesian
MRA (Figure 2), however, in cases of moderate fetal motion, the rMRAs performed
much better (figure 3). The T2* of blood measured from the descending aorta in a
27 week fetus was estimated to be 21.24 ± 3.23 ms. Discussion & Conclusion
Radially sampled data provided the touted advantage of
motion-insenstivity in fetal angiography. Radial MRA, with 361
spokes however, took 7.9 sec/slice, compared to 4.73
sec/slices for Cartesian. This was because the Cartesian images used both
partial Fourier (factor 1.33) and GRAPPA (factor 2). Radial GRAPPA or other
non-Cartesian parallel imaging approaches were not taken advantage of for
radial sampling in this work. The
long TR allowed for fully FC multi-echo readouts that offers: (a) T2* mapping
of the fetal vasculaure; b) enhancement of the SNR of the first echo through
data averaging (figure 2c); and (c) using the phase from the second echo for MR
susceptometry3. Higher order acceleration and turbulence terms might be
underestimating the T2* and needs to be further optimized. In conclusion, the radial
sampling scheme holds great promise for fetal angiography applications. Future
works that can incorporate FC multi-spoke readout within the same TR could
further accelerate such acquisitions, bringing the acquisition time under 2 or 3 seconds per slice, with the ability for simultaneous T2* quantification. Acknowledgements
No acknowledgement found.References
[1] Neelavalli, J., Krishnamurthy, U., Jella, P. K., Mody, S. S., Yadav, B. K., Hendershot, K., ... & Hassan, S. S. (2016). Magnetic resonance angiography of fetal vasculature at 3.0 T. European radiology, 1-7.
[2] Speier P, Trautwein F. Robust radial imaging with predetermined isotropic gradient delay correction. In: Proc 14th Annual Meeting ISMRM, Seattle; 2006: 2379.
[3] Neelavalli, J., Jella, P. K., Krishnamurthy, U., Buch, S., Haacke, E. M., Yeo, L., ... & Romero, R. (2014). Measuring venous blood oxygenation in fetal brain using susceptibility-weighted imaging. Journal of Magnetic Resonance Imaging, 39(4), 998-1006.