Eugene Milshteyn1,2, Galen D. Reed3, Cornelius von Morze1, Zihan Zhu1,2, Jeremy W. Gordon1, and Daniel B. Vigneron1,2
1Radiology and Biomedical Imaging, UCSF, San Francisco, CA, United States, 2UC Berkeley-UCSF Graduate Program in Bioengineering, UCSF and University of California, Berkeley, San Francisco, CA, United States, 3HeartVista Inc., Los Altos, CA, United States
Synopsis
Hyperpolarized 13C
probe development has allowed in vivo
monitoring of different physiological processes relating to various diseases,
including cancer and diabetes. Each new probe is typically characterized with polarization and T1
measurements, but T2 is also an important parameter for optimal
sequence design, including progressive flip angle schemes. To improve the
spatiotemporal resolution of T2 mapping sequences and subsequent
multi-exponential analysis, this project investigated using a local low rank
plus sparse reconstruction for 2-fold acceleration of in vivo T2 mapping with the bSSFP sequence.
Purpose
New advances in probe
development for hyperpolarized (HP) 13C imaging has enabled monitoring of
various physiological processes, such as metabolism and perfusion, of a variety
of diseases.1,2 With each new developed probe, several measures
are typically performed for the proper characterization of the probe to
determine potential in vivo use, with
the primary measures being T1 and polarization.3 T1 is important for proper sequence
design due to the effects on achievable spatial and temporal resolution, as
well as optimization of progressive flip angle schemes, and has been well
studied for many current preclinical and clinical probes.2,3 The T2 values of hyperpolarized
probes have recently also been investigated, especially in vivo, as they often are a critical component of optimal sequence
design.4–7 T2 mapping using the balanced
steady-state free precession (bSSFP) sequence has provided high spatial
resolution maps for multi-exponential fitting of several compounds, including [13C]urea,
[13C,15N2]urea, [2-13C]pyruvate,
and [1-13C]lactate.5,8 However, T2 mapping acquisitions are
slow and would benefit from acceleration methods such as compressed sensing9 to provide improvements in both spatial
resolution and temporal resolution. In this study, we demonstrated 2-fold
acceleration of in vivo bSSFP T2
mapping using a local low rank plus sparse (LLR+S) reconstruction10 that provided significant spatiotemporal
improvements over existing methods.Methods
Retrospective
simulations using LLR+S were performed by undersampling previously acquired [13C,15N2]urea
and [1-13C]lactate datasets, which were acquired as 2D coronal
projections with parameters listed in Reed et al.5 and Milshteyn et al.,8 respectively. For each time-point, a
different 50% undersampling variable-density pattern was used, and the structural
similarity index (SSIM) was used to assess the reconstruction as well as
comparison of multi-exponential fits. The initial in vivo studies were performed with both a fully sampled
acquisition and an accelerated acquisition (HP001 and [13C,15N2]urea). The parameters for
the fully sampled acquisition were: 14x7 cm2 FOV, 140x70 matrix
size, 1.6ms TBW 4 sinc pulse, 8.5-12.5ms TR, 4.25-6.25ms TE, 90°-180° θ/2-θ
flip angles, 20 time-points, total scan time of 11.9-17.5s. The parameters for
the HP001 (bis-1,1-(hydroxymethyl)-[1-13C]cyclopropane-d8)
and [2-13C]pyruvate accelerated acquisitions were similar to the
fully sampled acquisition with an 8.5/4.25ms TR/TE, but with 2-fold
acceleration, leading to a total scan time of 5.95s. The parameters for the [13C,15N2]urea
undersampled acquisition were: 14x7 cm2 FOV, 280x140 matrix size,
1.6ms TBW 4 sinc pulse, 15ms TR, 7.5ms TE, 90°-180° θ/2-θ flip angles, 20 time-points,
total scan time of 11.9s. The resulting data was reconstructed using the LLR+S
algorithm and fit using a nonnegative least squares algorithm, as described in Reed et al.6 The experiments were conducted on a 3T
GE MR scanner and DNP experiments used a HyperSense polarizer. The scans
started at 20s (HP001) or 30s ([13C,15N2]urea
and [2-13C]pyruvate) after beginning of injection and 3mL of 100mM
HP001, 80mM [2-13C]pyruvate, and
110mM [13C,15N2]urea was injected over 12s via
tail vein catheters in three Sprague-Dawley rats.Results and Discussion
The LLR+S
reconstruction of the retrospectively simulated data, which can be seen in
Figure 1 for [13C,15N2]urea, was in good
agreement with the fully sampled data in kidney and vasculature regions, based
on the SSIM (urea: 0.992, lactate: 0.947). The prospective LLR+S reconstructed acquisitions
also matched up well with the fully sampled acquisitions, which can be seen in
three of the first five time-points of the HP001 acquisitions in Figure 2. The
subsequent reconstructed T2 maps were also in good agreement as
evidenced by the ratio maps of the fully sampled and accelerated acquisitions
for HP001 (Figure 3) and [13C,15N2]urea
(Figure 4), which had an average value of 1.12±0.22 and 0.94±0.21 in the
kidneys, respectively. The urea acquisition also demonstrates the capability of
sub-millimeter in-plane resolution acquisitions for HP probes with sufficient acceleration.
Figure 5 shows the [2-13C]pyruvate acquisition T2 map
with 1x1 mm2 in-plane resolution. The average value in the kidneys,
0.786±0.13, agreed well with previously acquired, lower resolution T2
maps.8 The LLR+S reconstruction allowed ~6-fold improvement
in resolution even with [2-13C]pyruvate having a relatively short in vivo average T2.Conclusion
We demonstrated the
ability to acquire undersampled T2 mapping data of various HP probes
that correlated well with both fully sampled acquisitions and literature
values. The LLR+S reconstruction allowed for improvements in spatiotemporal
resolution, achieving 1x1 mm2 and sub-millimeter in-plane
resolution. Further development of the method, such as increasing the amount of
undersampling, may improve multi-exponential fitting via detection of
short-lived components, potentially allow for 3D T2 mapping of
hyperpolarized probes, and be translated to the clinic to optimize current HP
acquisitions.Acknowledgements
The authors would
like to thank Mark Van Criekinge, Lucas Carvajal, and Dr. Robert A. Bok for all
their help and funding from the NIH (P41EB013598 and R01EB017449).References
1. Kurhanewicz, J. et al. Analysis
of cancer metabolism by imaging hyperpolarized nuclei: prospects for
translation to clinical research. Neoplasia 13, 81–97 (2011).
2. Brindle,
K. M., Bohndiek, S. E., Gallagher, F. A. & Kettunen, M. I. Tumor Imaging
Using Hyperpolarized Resonance Spectroscopy 13C Magnetic. Magn.
Reson. Med. 66, 505–519 (2011).
3. Chaumeil,
M. M., Najac, C. & Ronen, S. M. in Methods in Enzymology 561,
(2015).
4. Yen,
Y. F. et al. T2 relaxation times of 13C
metabolites in a rat hepatocellular carcinoma model measured in vivo using 13C-MRS
of hyperpolarized [1-13C]pyruvate. NMR Biomed. 23,
414–423 (2010).
5. Reed,
G. D. et al. High Resolution 13C MRI With Hyperpolarized
Urea: In Vivo T2 Mapping and 15N Labeling Effects. IEEE
Trans. Med. Imaging 33, 362–371 (2014).
6. Reed,
G. D. et al. Imaging Renal Urea Handling in Rats at Millimeter
Resolution using Hyperpolarized Magnetic Resonance Relaxometry. Tomography
2, 125–137 (2016).
7. Shang,
H. et al. Variable Flip Angle Design for Balanced SSFP Transient State
Imaging to Improve HP 13C MRI. in Proceedings of the 23rd Annual
Meeting of ISMRM, Toronto, Canada 3768 (2015).
8. Milshteyn,
E. et al. Development of High Resolution 3D Hyperpolarized 13C
Imaging Techniques. in Proceedings of the International Society for Magnetic
Resonance in Medicine 23 (2015) 4599.
9. Feng,
L. et al. Accelerated Cardiac T2 Mapping using Breath-hold
Multiecho Fast Spin-Echo Pulse Sequence with k-t FOCUSS. Magn. Reson. Med.
65, 1661–1669 (2011).
10. Milshteyn,
E. et al. Using a Low Rank plus Sparse Reconstruction Approach to
Accelerate 3D Dynamic bSSFP Hyperpolarized Carbon-13 MR Imaging. in Proceedings
of the International Society for Magnetic Resonance in Medicine 24 (2016) 2329.