0015

MRI based RF safety characterization of implants using the implant response matrix: a simulation study.
Janot P. Tokaya1, Alexander J.E. Raaijmakers2,3, Peter R. Luijten2, and Cornelis A.T. van den Berg1

1Radiotherapy, UMC Utrecht, Utrecht, Netherlands, 2Radiology, UMC Utrecht, Utrecht, Netherlands, 3Biomedical Image Analysis, Eindhoven University of Technology, Netherlands

Synopsis

We introduce a general description of the RF response of an implant, defined as the implant response matrix (IRM). An analytical expression for the IRM is derived through basis functions that depend on a limited number of parameters. This analytical model is validated with a simulation study (Pearson correlation coefficients with simulations R: 0.9979-0.9996). This description allows a significant reduction in unknowns enabling IRM assessment by MRI measurements without hardware modifications to scanner or implant. The feasibility of MRI based IRM/TF measurement is shown in silico. With the simulated complex B1+ fields the IRM is accurately reconstructed (R: 0.974).

Purpose

Transfer functions (TFs) describe the electric field enhancement at the tip of an implant due to an electric field exposure1. Dedicated bench setups1-3 are required to measure TFs in phantom experiments. Recently, an alternative MRI based approach was presented4:A coax cable is soldered to the implant which is subsequently used as a transceive coil. Through MR imaging the current distribution along the implant is determined, which directly reflects its TF. This method has advantages in terms of measurement speed and versatility. However, the necessary galvanic contact to the implant limits the practical applicability of this approach. We now present a feasible method to determine TFs from MR acquisitions without this galvanic contact. It requires the introduction of a more general description of the RF response of an implant, defined as the implant response matrix (IRM) which contains the TF in its first column. An analytical expression for the IRM is derived through basis functions that depend on a limited number of parameters. This reduction in unknowns enables IRM assessment by MRI measurements without hardware modifications to scanner or implant. The feasibility of this MRI based IRM/TF measurement is shown by electromagnetic simulations mimicking the envisioned MRI experiments.

Theory

We define the implant response matrix $$$M\,$$$as$$$\,I_{ind}=ME_{inc}\,$$$where$$$\,I_{ind}\,$$$is a vector containing the induced current distribution along the implant and $$$E_{inc}$$$ is a vector with the incident, tangential electric field. This field induces an RF current in the implant as shown in figure 1. If this exposure is a point-excitation at the tip of the implant, the resultant current distribution in the implant directly reflects the TF of the implant3. If this excitation occurs at a different location,$$$z_j$$$, it will induce another current pattern. This scattering can be described by the IRM, $$$M$$$.

By definition, the induced current patterns in figure 1 are its columns. The parameter space of this matrix is greatly reduced when the current patterns are modeled by damped waves originating at the location of excitation. They repeatedly reflect at both ends of the implant. The summation of these waves leads to a converging geometric series. The IRM then becomes:$$\tag{1}M_{ij}=\sigma_{eff}E_{0}e^{-c|z_i-z_j|}+\frac{\sigma_{eff}E_{0}}{1-r_{0}r_{L}e^{-2cL}}[\sqrt{r_{0}r_{L}}e^{-cL}\cosh(c(z_i+z_j-L)-\frac{1}{2}\ln(\frac{r_{0}}{r_{l}}))+r_{0}r_{L}e^{-2cL}\cosh(c(z_i-z_j))],$$ $$$c$$$ is the complex propagation constant of the current wave in the implant of length $$$L$$$. $$$r_0$$$, and $$$r_L$$$, are the reflection coefficients. Hereby the IRM of an implant is described by four complex parameters. With additional parameters,$$$\{c_2,c_3,...,t_{12},t_{23},..\}$$$, similar equations can be derived for implants that consist of multiple segments resulting in additional transmissions and reflections.

Methods

Electromagnetic simulations (Sim4Life,ZMT, Zurich) were used to validate the theoretical model to describe the IRM. $$$M$$$ is determined by applying narrow (5mm) plane wave excitations with positively interfering electric fields at various positions $$$z_j$$$ along the implant and monitoring the induced current distribution in the elongated implant. The resultant $$$M$$$ is considered the ground truth representation of the IRM.

To demonstrate the feasibility of MRI-based assessment of the IRM, an MRI measurement is designed that was tested in silico: we simulated an MRI experiment of a large phantom, with $$$\sigma = 0.47S/m$$$ and $$$\epsilon_r=78$$$, with and without an implant present. The simulated B1+ distribution without the implant, which is accessible from MRI acquisitions5 assuming the transceive phase approximaton6 to be valid, was used to determine $$$E_{inc}$$$ using the method from Buchenau et al7.

In a separate simulation with the implant in situ $$$I_{ind}$$$ is determined from the distortion in the B1+ field8. $$$M$$$ is subsequently determined through:$$\tag{2}\underset{\vec{c}}{\text{min}}\|M(z_i,z_j;\vec{c})E_{inc}-I_{ind}\|,$$

where $$$\vec{c}\,$$$is the set of parameters,$$$\{\sigma_{eff}E_0,r_0,r_L,c_1,c_2,...,t_{12},t_{23},..\}$$$ describing the implant.

Results

Both the simulated IRM (considered as ground truth) and the analytical fits using formula (1) are shown in figure 3. These are in good agreement (Pearson correlation coefficient $$$R:0.9979-0.9996$$$). This demonstrates the validity of expression (1). In figure 4 it is shown that the TFs are directly proportional to the first column of .

The feasibility of MRI based IRM determination is shown in figure 5. With merely B1+ distributions from simulations with and without the implant the IRM was accurately reconstructed ($$$R:0.974$$$).

Discussion and conclusion

The IRM is an extended version of the TF, which actually is its first column. The feasibility of MRI based measurement of the IRM/TF has been demonstrated in silica. The actual experimental evalution of the IRM will be explored in the near future.

The IRM has been presented to characterize an implant’s response to impinging electric fields. With a model of converging infinite sums of reflected and transmitted damped waves an accurate analytical description of the IRM is derived. The analytical description of the RF response of an implant makes TF determination with regular MRI experiments feasible.

Acknowledgements

This work was supported by the DeNeCor project being part of the ENIAC Joint Undertaking.

References

1. S.M. Park, et al., "Calculation of MRI-induced heating of an implanted medical lead wire with an electric eld transfer function". JMRI, 2007.

2.ISO/TS 10974, International Standards Organization Technical Speci cation, "Requirements for the safety of magnetic resonance imaging for patients withan active implantable medical device", 2011.

3. S. Feng, et al., "A technique to evaluate MRI-Induced Electric Fields at the Ends of Practical Implanted Lead", IEEE, 2015.

4. Tokaya, et al., "Ultra-Fast MRI Based Transfer Function Determination for the Assessment of Implant Safety", ISMRM proc, 2016.

5. V.L. Yarnykh, "Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field", Magn Reson Med, 2007.

6. A.L. van Lier, et al., "B+1phase mapping at 7 T and its application for in vivo electrical conductivity mapping", Magn Reson Med, 2012.

7. S. Buchenau, et al., "Iterative separation of transmit and receive phase contributions and B1+-based estimation of the specific absorption rate for transmit arrays", Magn Reson Mater Phy, 2013

8. M. R. van den Bosch, et al., "New method to quantify RF induced currents inside conductive wires", Proc. Int. Soc. Mag. Reson., 2009

Figures

The principle behind the implant response matrix. The upper figure with the red axis shows how the TF is proportional to the current distribution in the wire if it is excited by an Dirac delta-like electric field at the tip of the implant3. Other excitation locations give other current distributions that together form the matrix $$$M$$$. The matrix entries $$$M_{ij}$$$ describe the current at location $$$z_j$$$ induced by an electric field excitation at $$$z_i$$$.

An overview of the simulations that are performed to determine $$$M$$$ and to mimic a realistic experiment. Figure A shows the plane wave boxes with destructively interfering magnetic and constructively interfering electric fields. Figure B shows the model of implant in the phantom inside an Ingenia 1.5T birdcage coil (IQ-feed). From a separate simulation without the implant present the incident electric field distribution can be determined. The induced current is accesible from the simulation with the wire. Nonetheless, $$$M$$$ was deduced using only the complex $$$B_1^+$$$ distribution from the latter simulations since this information is accessible with MR experiments.

The IRM from simulations for different implants compared to their analytical counterparts. Implant A is a bare copper wire of 20 cm length with a 1.25cm radius. Implant B is a 40 cm insulated wire (copper radius 1.25cm and outer radius 1.75mm) with 1cm of insulation removed at both ends. Implant C is a 20 cm wire with 8 cm of its length insulated (same radii). The top row (in red) shows the $$$M$$$s that result from the plane wave excitation simulations and the bottom row shows their analytical counterparts. All implants can be accurately characterized with formula 1.

The first column of the IRM is directly proportional to the TF of the implant. The TFs are normalized to M(z,0) instead of having a unit integral value to better capture the correspondence between both quantities. This result shows that if $$$M$$$ is measured, the TF is simultaneously determined. The TF and the IRM are shown for an insulated and bare wire of 40cm length. The other implants showed similar agreement between the TF and the first column of $$$M$$$.

In silica IRM/TF determination. The $$$E_{inc}$$$ field can be reconstructed from the B1+ distribution without the wire. The induced current can be determined from the B1+ distribution with the wire and especially the distortion this current creates. The simulation results are shown in red. The reconstructed electric field distribution closely resembles the simulated electric field distribution. With the reconstructed $$$E_{inc}$$$ and $$$I_{ind}$$$, the IRM $$$M$$$ can be determined accurately (with a Pearson correlation coefficient $$$R: 0.974$$$ and a cumulative mean error of 0.3%).

Proc. Intl. Soc. Mag. Reson. Med. 25 (2017)
0015