Valery Ozenne1, Solenn Toupin1,2, Pierre Bour1, Baudouin Denis de Senneville3, Alexis Vaussy2, Matthieu Lepetit-Coiffé2, Pierre Jaïs4, Hubert Cochet4, and Bruno Quesson1
1Electrophysiology and Heart Modeling Institute, Bordeaux, France, 2Siemens Healthcare, Paris, France, 3Mathematical Institute of Bordeaux, Bordeaux, France, 4Department of Cardiac Electrophysiology, Hôpital Cardiologique de Haut-Lévêque, Bordeaux, France
Synopsis
Recent studies have proposed to monitor
radiofrequency ablation on the heart using real-time MR-thermometry. Methods
rely on ECG triggering which can fail in presence of arrhythmia. This study
evaluates the precision of MR-thermometry on patients (N=15) even in presence
of cardiac arrhythmia. Phase images were
acquired using a single-shot multi-slice echo planar imaging and temperature
maps were calculated and displayed on the fly. ECG was recorded simultaneously
for further analysis of cardiac rhythm and post-processing of temperature
images. Stability of temperature
mapping without RF-heating was evaluated in each pixel and correlated to the
prevalence of arrhythmia.
INTRODUCTION
Catheter-based radiofrequency ablation (RFA) is
a widely accepted method for clinical treatment of cardiac arrhythmia, with 200
000 procedures performed in Europe in 2014 [1]. However, both
safety and efficacy could be improved by monitoring lesion formation in real
time, thereby reducing recurrences which often require redo procedures [2]. MR-thermometry has the
potential to visualize lesion formation in real-time during RFA [3]. Recent studies
reported reliable cardiac thermometry [4, 5] during RFA by
combining ECG-triggered imaging and online correction of respiratory motion and
associated susceptibility artifacts. However, variations in cardiac cycle
duration may lead to erroneous temperature estimate. The purpose of the present
study is to evaluate the precision of MR-thermometry without RF-heating, including
in patients with cardiac arrhythmia and to demonstrate feasibility of the
method in clinical environment.METHODS
Patients: the study was approved
by the Institutional Review Board and all subjects (N=15, 58.5 ± 16.2 years old and eleven
male) gave informed consent to be included in the study. Five patients were in
sinus rhythm and did not show substantial variations of the RR duration during
scanning, whereas ten patients displayed irregular rhythm. MR Imaging: 4
to 5 temperature slices in coronal orientation were acquired sequentially under
free breathing at each heartbeat during approximately 3’30 minutes on a 1.5 T MRI
(Avanto, Siemens Healthcare). The sequence was a single shot gradient EPI
(TE=20ms, TR=85ms, Grappa=2) with 110x110 voxels corresponding to a 1.6x1.6x3mm3
voxel size (zero filled to 0.8x0.8x3mm3). Image reconstruction,
correction of residual in-plane respiratory motion and associated
susceptibility variations, compensation of spatial-temporal phase drift and low
pass temporal filtering were implemented
in the Gadgetron framework [6], ensuring online
visualization of temperature images [5]. To assess the
precision of cardiac MR-thermometry, the temporal standard deviation of
temperature (σT) over time was computed in each voxel from all
slices during the procedure. The distribution of σT values was
analyzed on a manually drawn ROI surrounding the left ventricle. Voxels where σT
was higher than 5°C were removed from the statistical analysis. Categorization of Beats: ECG was
recorded using standard 3-lead (ECG) acquisition. An algorithm was designed to
identify, synchronize and categorize beats based on two successive RR durations,
as proposed by Contijoch et al [7].RESULTS
The
mean ± SD heart rate across all patients was 69 ± 10 bpm, ranging from 54 to 90
bpm (Table 1). Data were first sorted in three groups: 1) patients in sinus rhythm,
2) patients with low frequency of event of arrhythmia (less than 15%
prevalence) and 3) patients with high frequency of event of arrhythmia (more
than 15% prevalence). In each group, a representative case study is detailed in
Figures 1, 2 and 3, respectively. The statistical analysis of temperature
distribution over the patients is presented in Figure 4 using a box and whisker
plot. Over the patients in group 1, σT was 1.57 ± 0.34 °C with a
total of 4343 ± 1817 voxels per slice. Only 3 % of the voxels in the ROI were
excluded due to residual phase unwraps or σT higher than 5°C. In group 2,
σT was 1.83 ± 0.70 °C (7 % of voxels excluded) and 2.17 ±
0.46 °C in group 3 (19 % of voxels excluded).DISCUSSION
Temperature
uncertainty remains below 2°C in more than 60% of the voxels of the left
ventricle in 12/15 patients (Fig 4C). In patient
with sinus rhythm, temperature uncertainty was consistent across subjects, with
variation below 0.5°C. Although significant
cardiac motion inconsistency during arrhythmia can lead to image artifacts, in
most cases phase images showed good quality, as illustrated in Figures 2D, 3D,
4D, resulting in stable temperature all over the left myocardium with the
exception of small area. Nevertheless, both temperature uncertainty and number
of voxel with imprecise temperature increased with prevalence of arrhythmia. Additional
method to automatically identify and discard artifacted images in the time
series might help to improve cardiac thermometry in presence of arrhythmia.
CONCLUSION
This study presents the first
evaluation of cardiac MR-thermometry on patients during free-breathing. The
precision of temperature estimate was found of sufficient quality to monitor
catheter-based RF ablation procedures in most cases.Acknowledgements
This work received the financial support from the French National Founding Agency (ANR) within the context of the Investments for the Future Program: referenced ANR-10-LABX-57 and named TRAIL and referenced ANR-10-IAHU-04 and named IHU LIRYC. This study was also supported by public grants from the French ANR: program TACIT ANR-11-TecSan-003-01; Equipex MUSIC ANR-11-EQPX-0030; and program MIGAT ANR-13-PRTS-0014-01References
1. Raatikainen MJ, Arnar DO, Zeppenfeld
K, Merino JL, Levya F, Hindriks G, et al. Statistics on the use of cardiac
electronic devices and electrophysiological procedures in the European Society
of Cardiology countries: 2014 report from the European Heart Rhythm Association.
Europace. 2015;17 Suppl 1:i1-75.
2. Tanner H,
Hindricks G, Volkmer M, Furniss S, Kuhlkamp V, Lacroix D, et al. Catheter
ablation of recurrent scar-related ventricular tachycardia using electroanatomical
mapping and irrigated ablation technology: results of the prospective
multicenter Euro-VT-study. J Cardiovasc Electrophysiol. 2010;21(1):47-53.
3. Kolandaivelu A,
Zviman MM, Castro V, Lardo AC, Berger RD, Halperin HR. Noninvasive assessment
of tissue heating during cardiac radiofrequency ablation using MRI
thermography. Circ Arrhythm Electrophysiol. 2010;3(5):521-9.
4. de Senneville BD,
Roujol S, Jaïs P, Moonen CTW, Herigault G, Quesson B. Feasibility of fast
MR-thermometry during cardiac radiofrequency ablation. NMR in Biomedicine.
2012;25(4):556-62.
5. Ozenne V, Toupin
S, Bour P, de Senneville BD, Lepetit-Coiffe M, Boissenin M, et al. Improved
cardiac magnetic resonance thermometry and dosimetry for monitoring lesion
formation during catheter ablation. Magn Reson Med. 2016.
6. Hansen MS,
Sorensen TS. Gadgetron: an open source framework for medical image
reconstruction. Magn Reson Med. 2013;69(6):1768-76.
7. Contijoch F, Rogers K, Rears H, Shahid M, Kellman P,
Gorman J, 3rd, et al. Quantification of Left Ventricular Function With
Premature Ventricular Complexes Reveals Variable Hemodynamics. Circ Arrhythm
Electrophysiol. 2016;9(4).