Synopsis
Having a
homogeneous magnetic field is an essential requirement to ensure high
image quality in MRI. Significant field inhomogeneity can result in
severe signal losses or geometric distortions. To achieve the desired
uniformity, effective and efficient shimming strategies are needed.
Specifically, passive and active shimming strategies have been
developed to correct for both the intrinsic and extrinsic magnetic
field inhomogeneities. Advantages and disadvantages of these various
solutions are reviewed. In addition, specific applications in imaging
experiments for some advanced shimming strategies are discussed, when
the conventional shimming solutions are inadequate.The impact of magnetic field
uniformity on the MR signal
MR images are generated based on the
assumption that the magnetic field at any given spatial location is
precisely known. With this knowledge, the MR signal can be quantified
as:
$$S(t) = \int_x \int_y \int_z M_{xy0}(x,y,z) e^{-i\gamma \int_0^t (G_x(\tau)x + G_y(\tau)y + G_z(\tau)z + \Delta B_0(x,y,z))d\tau} dx dy dz$$
However, any magnetic field inhomogeneity, $$$\Delta B_0$$$,
could have spatial gradient components along the $$$x$$$, $$$y$$$,
and $$$z$$$ directions, either globally or locally, resulting in
geometric distortions and/or signal losses. To minimize this
inhomogeneity, the magnetic field needs to be “shimmed”. The
recent advent of high field MRI scanners further exacerbated the
image artifacts due to field inhomogeneity, demanding even greater
shimming capabilities.
How do we achieve a uniform magnetic
field?
There are multiple layers of shimming
technologies embedded in a typical MRI scanner to address both the
intrinsic (e.g., arising from the magnet itself) and extrinsic (e.g.,
arising from the human subject inside the bore) magnetic field
inhomogeneities.
Correcting intrinsic magnetic
field inhomogeneity
Both passive shimming and active
superconducting shimming approaches are typically used to
compensate for the intrinsic inhomogeneities arising from the magnet
itself.
Passive shimming
Passive shimming, judging by its name,
is a field-correction method that relies on the use of passive
ferromagnetic materials such as iron rods or patches placed in
specific configurations inside the magnet (Hoult et al., 1985). For
example, iron rods can be inserted into or iron patches can be lined
along the magnet bore in specific patterns. The passive shimming
process can be time-consuming and tedious, as a large number of
iterations (mapping the magnetic field and tweaking the shim
elements) may be needed to determine the optimal pattern of the
spatial arrangements for the passive shim elements. Although passive
shimming is an effective and economical solution, the shimming
materials are usually heat-sensitive. Fortunately, many modern
scanners have liquid-cooled gradient coils, mitigating the
heating-induced magnetic field drifts.
Active superconducting shimming
Many MRI manufacturers also implement
active superconducting shim coils within the cryogen chamber, based
on spherical harmonics (SH) functions of various orders (Romeo et
al., 1985). The mathematical foundation is based on the solution of
Laplace’s equation, $$$\nabla^2B=0$$$, which is derived from Maxwell’s equations
in regions with no current sources. This solution is a linear
combination of SH functions. Typical 3T scanners are equipped with SH
shim coils up to the 3rd order ($$$X, Y, Z, ZX, ZY, XY,
X^2-Y^2, Z^2, Z^2X, Z^2Y, ZXY, Z(X^2-Y^2), Z^3$$$), while scanners with higher
magnetic fields such as 7 Tesla may have SH shim coils up to the 5th
order or even higher. The advantages of active superconducting
shimming are two-fold: first, it is more stable, and second, it
allows for an efficient shimming procedure as all currents can be
determined based on the known orders of SH.
Together, passive shimming and active
superconducting shimming coils ensure a uniform magnetic field under
the factory setting, typically reaching a magnetic field uniformity <
0.1 ppm.
Correcting extrinsic magnetic
field inhomogeneity
As effective as they are, one common
and main shortcoming of passive and active superconducting shims is
that they are usually permanent and only address static magnetic
field inhomogeneity. When a human subject enters the MRI scanner, the
magnetic field inhomogeneity inside the human body is dependent on
that particular person, so that it cannot be easily corrected by the
permanent shims. Instead, adjustable shims are necessary, giving rise
to the advent of active resistive shimming coils and local
shimming solutions.
Active resistive shimming
Active resistive shimming is based on
the same SH functions as used in active superconducting shimming,
with the difference being that coils are located within the scanner bore
at room temperature. Note that since the onboard $$$X$$$, $$$Y$$$, and $$$Z$$$ gradient
coils already produce magnetic field gradients of the first order,
there are no separate $$$X$$$, $$$Y$$$, and $$$Z$$$ shim coils. As such, resistive
coils are almost always used for higher-order shimming. Structurally,
these coils can be placed along the cylindrical surface in the space
between the magnet and the gradient coils, and oftentimes they are
integrated with the gradient coils. Ring-shaped resistive coils
correct for spherical harmonics in the z-direction, for example, the
$$$Z^2$$$ and $$$Z^3$$$ terms, while saddle-shaped coils correct for crossing terms
such as $$$XY, ZY, ZXY, Z^2Y$$$, etc.
As mentioned earlier, one main
advantage of resistive shimming over passive and superconducting
shimming is that the currents through resistive shims can be changed
dynamically. This allows shimming to be performed on a
subject-by-subject basis, or even dynamically on a slice-by-slice
basis within the same subject (Morrell et al., 1997).
Local shimming
Although SH shimming has been the
standard practice for the past three decades of human MRI, local
shimming methodologies have been gaining momentum in recent years.
Since the magnetic field inhomogeneities arising from the human
anatomy can be highly localized and are thus not always best solved
by combinations of SH functions, it may be more effective to design
and implement local coils that are close to the sources of magnetic
field perturbation, such as the nasal sinus, the oral cavity, or the
ear canal, which contain (paramagnetic) air spaces.
The simplest solutions for local
shimming may be the use of specially-shaped paramagnetic materials
that can be placed near the anatomical regions causing the magnetic
field disturbance. For example, a mouth piece made of paramagnetic
material can effectively compensate for the field inhomogeneity
arising from the oral cavity (Wilson et al., 2002). Similar to the
previously mentioned whole-body passive shimming solutions, these
local passive shimming materials cannot be readily adjusted for
different subjects. In addition, they can be uncomfortable to use.
Recent effort has been made in
developing active local shimming solutions (Hsu et al., 2005).
Specifically, a set of direct current (DC) loops, placed closely to
the imaging sample, showed promise in achieving a highly uniform
magnetic field (Juchem et al., 2010). While effective, one limitation
in this approach is that it requires a separate array of coils, which
takes up additional space and could reduce either SNR or shimming
efficiency depending on its spatial arrangement with respect to the
RF array.
The most recent, and perhaps the most
effective and efficient, solution for local shimming is based on the
newly proposed framework known as integrated parallel reception,
excitation, and shimming (iPRES) (Han et al., 2013; Truong et al.,
2014; Stockmann et al., 2016). Specifically, a single coil array can
simultaneously and inherently accommodate both alternating currents
(AC) for parallel RF reception/excitation and DC currents for local
$$$B_0$$$ shimming, thereby providing an integrated solution to
acquire MR images with high spatial fidelity. The main advantage is
that, by inherently enabling $$$B_0$$$ shimming within the same
RF coil array, one can achieve greatly improved magnetic field
homogeneity over existing shimming technologies, while still
maintaining the same high RF performance.
Auto-shimming procedures
To provide fast and effective shimming
solutions across all subjects, auto-shimming procedures (Webb et al.,
1991; Gruetter et al., 1993) are essential in routine MRI exams to
ensure high image quality, especially in diffusion MRI, functional
MRI, and MR spectroscopy. Whether it is for linear shimming or
higher-order shimming, most auto-shimming procedures require fast and
accurate magnetic field mapping, as well as effective determination
of required currents in either SH coils or local shimming coils.
While most modern scanners are equipped with adequate auto-shimming
routines, some applications such as MR spectroscopy may still require
manual shimming over specific regions.
Advanced applications in MRI
While conventional shimming
methodologies can usually ensure a magnetic field homogeneity <
0.1 ppm and are adequate for anatomical scans in human subjects, at
times they may not be sufficient for functional or diffusion MRI,
which can result in signal losses (in the case of $$$B_0$$$
inhomogeneity along the slice-encoding direction) or geometric
distortions (in the case of $$$B_0$$$ inhomogeneity in-plane).
For example, there are still marked signal losses in gradient-echo
echo-planar imaging (EPI) (e.g., for functional MRI application) and
geometric distortions in spin-echo EPI (e.g., for
diffusion MRI application) in the ventral frontal region influenced
by the air cavity in the nasal sinus, even after high-order shimming,
as illustrated in Fig. 1.
In this section, we discuss some
examples of advanced shimming applications that can overcome these
remaining challenges. One solution to recover signal losses is a
technique known as z-shimming (Frahm et al., 1988; Yang et al.,
1998). However, one cannot simply set the z shim to compensate for
signal loss, for example, in the ventral frontal regions, as it will
distort other regions that are otherwise homogeneous. Instead, two
(or more) images (one with z-shim, and one without) need to be
combined (example shown in Fig. 2) to generate a uniform image
(Glover et al., 1999), leading to a somewhat reduced throughput, even
in the case of the more efficient single-shot solutions (Song, 2001,
Truong and Song, 2008).
Because of its local
nature, a natural solution to correct the magnetic field
inhomogeneity such as that in the ventral brain region is local
shimming. Shown in Fig. 3 is an example illustrating the promise of
the aforementioned iPRES technique. By using the local shimming
inherently embedded in the RF array, the magnetic field uniformity
was greatly improved from 55.3 Hz (root mean square error) to 9.2 Hz,
and geometric distortions were corrected to restore the spatial
fidelity of the frontal lobe and the anterior horns of the
ventricles. Green contours were derived from a $$$T_1$$$-weighted anatomical
image.
Summary
Having a homogeneous magnetic field is
an essential requirement to ensure high image quality in MRI.
Significant field inhomogeneity can result in severe signal losses or
geometric distortions. To achieve the desired uniformity, effective
and efficient shimming strategies are needed. Specifically, passive
and active shimming strategies have been developed to correct for
both the intrinsic and extrinsic magnetic field inhomogeneities.
Advantages and disadvantages of these various solutions are reviewed.
In addition, specific applications in imaging experiments for some
advanced shimming strategies are discussed, when the conventional
shimming solutions are inadequate. With the ever increasing demand
for higher magnetic field and higher spatial resolution, magnetic
field shimming will remain a highly active research field for many
years to come.
Acknowledgements
No acknowledgement found.References
Frahm J, Merboldt KD, Hanicke W.
Direct FLASH MR imaging of magnetic field inhomogeneities by gradient
compensation. Magn Reson Med. 6(4):474-80, 1988.
Glover GH. 3D z-shim method for
reduction of susceptibility effects in BOLD fMRI. Magn Reson Med.
42(2):290-9, 1999.
Gruetter R. Automatic, localised in
vivo adjustment of all first- and second-order shim coils. Magn Reson
Med. 29:804-811, 1993.
Han H, Song
AW, Truong TK, Integrated parallel reception, excitation, and
shimming (iPRES), Magn. Reson. Med. 70:241-7, 2013.
Hoult DI, Lee D. Shimming a
superconducting nuclear-magnetic-resonance imaging magnet with steel.
Rev Sci Instrum. 56(1):131-5, 1985.
Hsu JJ, Glover GH. Mitigation of
susceptibility-induced signal loss in neuroimaging using localized
shim coils. Magn Reson Med. 53(2):243-8. Erratum in: Magn Reson Med.
2005 53(4):992, 2005.
Juchem C, Nixon TW, McIntyre S,
Rothman DL, de Graaf RA. Magnetic field modeling with a set of
individual localized coils. J Magn Reson 204:281–289, 2010.
Morrell G, Spielman D. Dynamic
shimming for multi-slice magnetic resonance imaging. Magn Reson Med.
38(3):477-83, 1997.
Romeo F, Hoult DI. Magnet field
profiling: analysis and correcting coil design. Magn Reson Med. 1984
1(1):44-65, 1984.
Song
AW, Single-shot EPI with signal recovery from susceptibility induced
losses, Magn. Reson. Med. 46, 407-411, 2001.
Stockmann
JP, Witzel T, Keil B, Polimeni JR, Mareyam A, LaPierre C, Setsompop
K, Wald LL. A 32-channel combined RF and B0 shim array for 3T
brain imaging. Magn Reson Med. 75(1):441-51, 2016.
Truong TK, Song AW, Single-shot z-shimmed sensitivity-encoded spiral-in/out imaging for fMRI with reduced susceptibility artifacts. Magn Reson Med 59(1):221-7,
2008.
Truong TK,
Darnell D, Song AW.
Integrated RF/Shim coil array for parallel reception and localized B0
shimming in the human brain, NeuroImage 103:235-40, 2014.
Webb P, Macovski A. Rapid, fully
automatic, arbitrary-volume in vivo shimming. Magn Reson Med.
20:113-122, 1991.
Wilson JL, Jenkinson M, Jezzard P.
Optimization of static field homogeneity in human brain using
diamagnetic passive shims. Magn Reson Med. 48(5):906-14, 2002.
Yang QX, Williams GD, Demeure RJ,
Mosher TJ, Smith MB. Removal of local field gradient artifacts in
T2*-weighted images at high fields by gradient-echo slice excitation
profile imaging. Magn Reson Med. 39(3):402-9, 1998.