Synopsis
Echo Planar Imaging (1), or EPI, is a
prototype for pulse sequences that sample two dimensions of K-space after a
single excitation. 2D scanning after a single excitation means that signal
modulations unrelated to applied gradient fields, such as transverse relaxation
and resonance offsets, distribute across two dimensions in k-space and image
space. EPI is highly demanding of gradient performance and fidelity. Gradient
hardware advances have enabled the implementation of EPI, and continue to
improve the utility and robustness of EPI. There are complex tradeoffs involved
in the design of EPI pulse sequences and selection of EPI parameters with
regard to gradient performance, SNR, image artifacts, ramp sampling, and other
pulse sequence features.Introduction to EPI
In a conventional spinwarp imaging
pulse sequence (2), lines of
k-space are acquired one at a time, with a new RF excitation pulse, and newly
generated coherent transverse magnetization for each line. This creates two
distinct in-plane directions, usually referred to as ‘frequency encoding’ and
‘phase encoding’ directions. In the frequency encoding direction, adjacent data
points in k-space are acquired in rapid succession, microseconds apart in time
under the influence of the readout or frequency encoding gradient. Thus this
direction is traversed and sampled in ‘real time’. In contrast, data points
that are adjacent to one another along the phase encoding direction are
collected after separate excitations, and at identical times after the
excitation pulse.
In EPI, a large oscillating readout
gradient is applied to traverse kx in a back and forth manner, while small
‘blip’ gradients create incremental steps in ky. A basic EPI (1) pulse
sequence and corresponding k-space trajectory are shown in Figure 1.
In EPI, both directions are ‘real time’
directions, in that all points in the 2D plane are sampled at different times
after the excitation pulse as the trajectory winds down the zig-zag k-space
path. However, there is still a strong anisotropy between directions, as the
readout direction is sampled quickly (per line), and the blip direction is
sampled much slower.
EPI is defined by the type of
trajectory in k-space, rather than by the methods used for RF excitation or
slice selection. An EPI pulse sequence can be a gradient echo sequence as in Figure 1, or a spin echo or stimulated
echo, and can be preceded by any magnetization preparation scheme, such as
inversion recovery, magnetization transfer, diffusion weighting, arterial spin
labeling, etc.
Signal Modulation across K-space
Following excitation, transverse
magnetization evolves under three main influences:
1) Applied gradient fields.
2) Transverse relaxation.
3) Resonant frequency offsets unrelated
to gradient fields, such as those caused by chemical shift or magnetic
susceptibility inhomogeneity.
These can be incorporated into an MRI
signal equation as three exponential terms:
$$S(t)\propto\int M_0(r)e^{ik(t)\cdot r}e^{-t/T2^*(r)}e^{i\omega(r)t}dr$$
where M0 is the transverse magnetization immediately after the
excitation pulse. The first exponential term describes the effects of the
gradient fields, and with this term alone, the signal in k-space and the transverse
magnetization are strictly related to one another by Fourier Transform. The next
two terms are modulations arising from relaxation and resonance offsets,
respectively. Note that relaxation and resonance offsets both produce simple
multiplicative modulations in the data domain. By the Fourier Convolution
Theorem (3), this implies
that these effects are described by convolutions with point spread functions
(PSFs) in the image domain, and that these PSFs are the Fourier Transforms of
the modulation terms. While these modulations are naturally functions of time,
the k-space trajectory maps these modulations onto k-space.
Relaxation produces amplitude
modulation across k-space, while resonance offset produces phase modulation. In
spinwarp imaging, these effects only produce modulation in the frequency
encoding direction. In the phase encoding direction, the signal is not
modulated by these effects, and the PSF in that direction is ideal. In EPI,
modulations appear in both directions, but because the traversal of k-space in
the blip direction is much slower, the effects of modulation is greater in the
blip direction, and the non-ideality of the PSF is dominated by blurring and
distortion in that direction.
Note that the modulation functions are
dependent upon the relaxation time T2* and the resonance offset ω, both of
which are local properties and typically vary in space. The PSFs are therefore
dependent upon the local relaxation and resonance offsets, and result in spatially
varying blurring and image distortion, respectively.
Imaging Speed and Ramp Sampling
The hallmark feature of EPI is the
ability to acquire a full plane of k-space after a single excitation. The goal
is to cover a sufficiently large patch of k-space with sufficient density to
produce a useful image. However, as noted above, the image can be blurred and
distorted, and these effects increase with acquisition time. There is therefore
a strong motivation to move through k-space as quickly as possible, which in turn
places strong demands on gradient performance. The gradient amplitude and slew
rate, which are proportional to velocity and acceleration in k-space,
explicitly limit the minimum time for complete sampling of 2D k-space. The
relative importance of gradient amplitude and slew rate varies with the desired
resolution and matrix size, and with the gradient hardware, but for modern
whole body human imaging systems, where typical amplitude and slew rate limits
are around 50mT/m and 200mT/m/ms, the slew rate is typically more limiting. In
this regime, as much or more time may be spent on the ramps of the trapezoidal
readout gradient as on the flat portions. Under these conditions, if data is
acquired only on the flat portions of the readout gradient, then the SNR
suffers both because the duty cycle of the ADC is reduced, and because the TE
is lengthened, causing increased T2/T2* decay. Blurring and off-resonance
related distortion are increased as well. In response to these problems, many
implementations of EPI now acquire data during the ramp portions of the readout
gradient as well as the flat portions, This is termed ramp sampling. This can
greatly improve the overall time efficiency of the coverage of k-space, but
introduces additional complications as well. With ramp sampling, data is
typically acquired with uniform sampling in time, and must be resampled to a
uniform grid in Kx prior to Fourier Transform. While this step is in principle
relatively straightforward, it places much more stringent requirements on the
accuracy of the gradient waveform and timing of data acquisition than does a
flat readout gradient. In addition, with sharply peaked readout gradients, the
readout bandwidth is highest at the center of k-space, reducing the SNR of the
estimation of those points in k-space that contain the bulk of the image
energy.
Thus, while the basic concept of EPI
scanning is in principle quite simple, the interdependence between resolution,
matrix size, SNR, image artifacts, and hardware requirements generate a
relatively complex set of tradeoffs that should be understood in order to
optimize the EPI measurement.
Relationship to Other Methods
Single shot EPI represents a prototype
2D pulse sequence in which all of 2D k-space is covered in a single excitation,
in a single echo (spin echo or FID). 2D k-space can also be covered using
multiple excitations and/or multiple spin echos (4), and the
relationship between these approaches is shown schematically in
Figure 2. Several of these methods are
considered hybrid EPI methods. Two other imaging features that are now commonly
used with EPI are coil based parallel acceleration, using approaches in the
SENSE (5) or GRAPPA (6) families of
methods, and multiband excitation (7), in which
multiple slices are excited simultaneously and are disentangled using a combination
of parallel acceleration methods and through-slice gradient modulation.
Acknowledgements
No acknowledgement found.References
1. Mansfield P. Multi-planar image
formation using NMR spin echoes. J Phys 1977; C10:L55-L58.
2. Edelstein
WA, Hutchison JM, Johnson G, Redpath T. Spin warp NMR imaging and applications
to human whole-body imaging. Phys Med Biol 1980;25(4):751-756.
3. Bracewell
RN. The Fourier Transform and its Applications. New York: McGraw-Hill; 1965.
4. Hennig
J, Nauerth A, Friedburg H. RARE imaging: A fast imaging method for clinical MR.
Magn Reson Med 1986;3:823-833.
5. Pruessmann
KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast
MRI. Magn Reson Med 1999;42(5):952-962.
6. Griswold
MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A.
Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn
Reson Med 2002;47(6):1202-1210.
7. Setsompop
K, Gagoski BA, Polimeni JR, Witzel T, Wedeen VJ, Wald LL. Blipped-controlled
aliasing in parallel imaging for simultaneous multislice echo planar imaging
with reduced g-factor penalty. Magn Reson Med 2012;67(5):1210-1224.