Synopsis
RF transmit/receive gain and transmit/receive
frequency must be adjusted for each patient exam. The problem is similar to
parameter mapping but over a smaller volume. Transmit gain accuracy ideally produces the
desired flip angle, however B1 field non-uniformity prevents this in practice. The
receive gain is ideally set so that the maximum signal does not saturate the
A/D converters which would produce shading, and the noise standard deviation is
at least one bit to avoid quantization error.
Transmit/receive frequency accuracy is required for accurate
localization, good EPI and spiral image quality and for fat suppression pulses
to work optimally.Highlights
·
Settings for RF transmit gain (TG),
transmit/receive frequency (center frequency or CF) and receiver gain (RG) are patient-dependent
and are critical to good image quality.
·
Finding optimal settings for these parameters is
similar to the problem of mapping the parameters but over a smaller volume,
usually one slice near the magnet isocenter.
·
The parameters must be set robustly, quickly and
without operator intervention.
·
Different vendors use different methods for
measuring and calibrating these parameters.
·
Non-optimal RF transmit gain can result in loss
of signal and contrast, poor image uniformity, and poor fat saturation.
·
Non-optimal transmit/receive frequency can
result in banding and artifacts in SSFP, blurring with spirals, phase encode
shifting in EPI, poor image intensity homogeneity, localization errors, signal
loss, and poor fat saturation.
·
Non-optimal receive gain can cause shading,
intensity distortion, loss of SNR and artifacts.
Target Audience
Scientists
and clinicians interested in patient-dependent scanner calibration methods and
the effects on image quality.
Objectives
·
Understand ways of setting the RF
transmit/receive gain and transmit/receive frequency.
· Understand
possible artifacts that result if the values are not set reasonably well.
·
Understand situations in which the calibrations
could be prone to being incorrectly set.
Purpose
Scanners
are calibrated using phantoms for patient-independent effects such as large
scale B0 inhomogeneity, eddy currents, RF output power, gradient amplifier/coil
gain and gradient non-linearity. Once
these calibrations are completed the scanner is ready for clinical use. The RF transmit power, RF receiver
attenuation and transmit/receive frequency must be set for each patient to obtain
optimal image quality. Various methods
are available to set the parameters but in general the methods must be fast and
robust to patient and protocol variation.
A variety of artifacts can also arise depending on how the parameters
are set.
Methods
Differences in patient size and
composition within the sensitive volume of the transmit coil change RF energy
absorption. As a result the RF power
needed to obtain the desired flip angle must be calibrated for each patient at
the desired scanning location. The same
methods that are used for B1 mapping can be used for this purpose. Examples of such methods include the double
angle method (DAM) (1), actual flip angle imaging (AFI) (2), phase sensitive
imaging (3), dual refocusing echo acquisition mode (DREAM) (4), and
Bloch-Siegert (5). However for TG
estimation, only a small subset of the patient imaging volume is mapped,
usually within a slice at the magnet isocenter.
The TG calibration sequence is used to find the flip angle that results
from a given TG for some standard RF pulse.
The TG needed for the flip angle for any other pulse can then be
obtained by scaling. The process need
not be repeated unless patient loading is substantially changed.
The attenuation of the analog
signal or equivalently the receive gain (RG) must be adjusted prior to
digitization by the A/D converter. If
the RG is too high the signal will saturate the A/D converter resulting in
shading and intensity distortion. If the
RG is too low, the noise will be dominated by the step size of the A/D
converter instead of the noise itself (quantization error), resulting in noise
and artifacts (6,7). Since the highest
signal occurs at the center of k-space, the imaging pulse sequence can be used
to acquire all or part of the imaging volume without phase encoding. If the system is well designed, the RG can be
adjusted so that the signal maximum is somewhat less than the amplitude that
saturates the A/D while still avoiding quantization error. The system design requirements depend on the
dynamic range of the analog signal (8).
The highest dynamic range results from the highest SNR scans. These are usually high field, 3D scans with
receive coils that are sensitive to a relatively large patient volume, as well
as scan protocols that generate high signal.
The transmit/receive frequency or center
frequency (CF) must also be adjusted for each patient and sometimes several
times during the exam depending on the scanned location. A large CF error can result in localization
errors in the slice direction. With EPI,
because of the low effective bandwidth in the phase encoding direction, a large
CF error can shift the object in the phase encoding direction. With balanced SSFP, CF errors can cause
banding artifacts (9). With spirals, CF
errors cause blurring artifacts (10). With
some fat saturation methods, a CF shift can result in poor fat saturation and
loss of water signal.
Results
Finding the optimal TG for the
diagnostically relevant anatomy can be difficult if the transmit B1 varies strongly
over the patient, for example at high field where there is strong dielectric
shading or when the anatomical shape is complicated. In these cases a compromise must be made
between image quality and having a fast, automated way to set TG.
When adjusting the RG, some
receive coils can have a large enough phase variation over the imaging volume
that the maximum signal does not occur at zero phase encoding. In this case, a phase encoding value near
zero can partially cancel the receive coil phase and produce the maximum
signal. Therefore it is prudent to set
the RG to produce somewhat less than the maximum signal that results when zero
phase encoding is used.
The CF can be particularly difficult to set in
some areas of the body with poor B0 homogeneity such as the neck due to rapid local
susceptibility variations. Therefore it
is advantageous to set CF over a small localized volume and after or in
conjunction with some sort of local shimming procedure. CF can also be difficult to set in areas that
have multiple, complicated spectral peaks, or when the main fat peak is
actually larger than the water peak, for example in breast imaging. Therefore a robust fat elimination method can
be helpful in finding the water peak.
Conclusion
RF
transmit gain, receive gain and transmit/receive frequency must be adjusted for
optimal image quality for each patient exam and sometimes further adjusted
within the patient exam. Calibration
must be robust, fast and automated. Different
vendors use different methods. Non-optimal
settings can result in artifacts, poor contrast, SNR loss, and poor fat
suppression.
Acknowledgements
No acknowledgement found.References
1. Cunningham CH, Pauly JM, Nayak KS. Saturated double angle method for rapid B1+
mapping. Magn Reson Med 2006;55:1326-1333.
2. Yarnykh VL. Actual flip-angle imaging in the
pulsed steady state: a method for rapid
three-dimensional mapping of transmitted radiofrequency field. Magn Reson Med 2007;57:192-200.
3. Morrell GR.
A phase-sensitive method of flip angle mapping. Magn Reson Med 2008;60:889-894.
4. Nehrke K, Börnert P. DREAM – a novel approach for robust,
ultrafast, Multislice B1 mapping.
Magn Reson Med 2012;68:1517-1526.
5. Sacolick LI, Wiesinger F, Hancu I, Vogel
MW. B1 mapping by Bloch-Siegert
shift. Magn Reson Med 2010;63:1315-1322.
6. Behin
R, Bishop J, Henkelman RM. Dynamic range requirements for MRI. Concepts Magn.
Reson. Part B. 2005;26B:28–35.
7. Cadiz RF, Munoz C, Tejos C, Andia ME, Uribe S,
Irarrazaval P. Quantization Error in
Magnetic Resonance Imaging. Concepts
Magn Reson Part A. 2014;43A:79-89.
8. Gabr RE, Schär M, Edelstein AD, Kraitchman DL,
Bottomley PA, Edelstein WA. MRI Dynamic
Range and its Compatibility with Signal Transmission Media. J Magn Reson 2009;198:137-145.
9. Deshpande VS, Shea SM, Li D. Artifact Reduction in True-FISP Imaging of
the Coronary Arteries by Adjusting Imaging Frequency. Magn Reson Med 2003;49:803-809.
10. Meyer CH, Hu BS, Nishimura DG, Macovski A. Fast spiral coronary artery imaging. Magn Reson Med 1992;28:202-213.