Synopsis
Fundamentals of transmission lines and power transfer are
presented to help in the understanding, design, implementation and performance evaluation of MRI hardware.Target Audience
Those interested in understanding some of the RF concepts behind
common interconnections in the MRI hardware setup.
Outline
This talk will cover the following topics:
- Distributed-circuit analysis and wave equation of a terminated transmission line (TL)
- Type of TLs commonly used in MRI
- Scattering parameters and the Schmitt chart, TL impedance and reflection coefficient
- TL applications in MRI: Impedance matching, balanced to unbalanced connection, power splitting and monitoring, phase shifting, preamplifier decoupling, and coil design.
- TL practical considerations and TL lumped element equivalents
- Non-conventional power transfer in MRI
Introduction
During both the excitation and acquisition phases of the MRI experiment, RF signals need to
travel efficiently between the scanner system and the MR coils. In the conventional MRI setup, the coil and the scanner are
interconnected through coaxial cables, which is a type of shielded transmission
line that allows power to be transferred from one point to the other, ideally
without radiation losses. As an example,
Figure
1 shows a local transmit (TX) receive (RX) setup in the MRI scanner. Power
is transferred from the power amplifier (RFPA) located in the equipment room to
the coil plug located in the scanner bore through a long coaxial connection. There are
multiple RF interconnections between the system and the coil, built from
transmission line segments (with lengths of a fraction of the signal wavelength)
or their lumped element equivalent implementation (LC network). The impedance
of interconnecting devices and coil (coupled to a nominal load) are matched to
the characteristic impedance of the coaxial cable (MRI standard 50 Ω) for
maximum power transfer. An RF switch (Transmit/Receive switch) is used to
isolate the low-power receiver’s electronics, mainly the receive amplifier (or preamplifier),
from the high-power RF transmit signal. An RF coupler can be inserted in the transmit line to have information about the actual power
delivered to the coil (for example under a change in load impedance). A balun
(
balanced to
unbalanced) is a device used in coaxial connections to suppress RF
currents on the cable’s shield, which can reduce system performance and
compromise safety.
TL Basics
A differential segment of a
lossless transmission line can be modeled as a series inductance connected to a
shunt capacitance as shown in Figure 2. L and
C are the distributed (in units
per length) inductance and capacitance respectively. The TL (wave) equations
(shown in the figure) can be formulated from this differential length model
applying Kirchhoff’s current and voltage laws (KCL and KVL).
A lossless TL
can transport electromagnetic energy from one point to the other in the form of
a pure transverse electromagnetic (TEM) wave. This means the electrical and
magnetic field components are both zero in the direction along the line (z). The termination at both ends of the
line (source and load impedance) will determine how energy is absorbed and
reflected along the lines (Figure
3).
From the wave solution, the
impedance at the input of the line can be calculated as:
$$ Z_{i n}=Z_{0}\frac{Z_{L}+jZ_{0}\tan\frac{2\pi}{\lambda}l}{Z_{0}+jZ_{L}\tan\frac{2\pi}{\lambda}l}, Z_{0}=\sqrt{\frac{L}{C}}, \beta=\frac{2\pi}{\lambda}=\frac{2\pi f}{v_{p}}, $$
where Z0 is the characteristic impedance of the line, ZL the load impedance, β the
propagation constant and λ the signal
wavelength. When the source and the load are matched to the characteristic
impedance of the line (RS =RL=Z0 ), all transmitted power is
absorbed by the load. In contrast, when one or both impedances are different
than Z0, power is
reflected back and forth along the line (standing wave pattern) and less power
is delivered to the load. Normally, in
MRI we use coaxial connections with Z0
= 50 Ω, therefore to ensure maximum
power transfer to the loaded coil we perform 50 Ω matching almost everywhere along the transmit and receive chains.
In practice, coaxial cables are lossy TLs (they have an attenuation factor) and
this cannot be ignored as frequency and/or length of the connection increase.
This is the case for the long high power coaxial connection between the RFPA
and the coil plug. For example in a 7T system, even with impedance matching, roughly
50 % of the transmit power reaches the coil.
For
a quarter-wavelength transmission line,
$$Z_{i n}(n\frac{\lambda}{4})=Z_{0}\frac{Z_{L}+jZ_{0}\tan n\frac{\pi}{2}}{Z_{0}+jZ_{L}\tan n\frac{\pi}{2}}=\frac{Z_0^2}{Z_{L}}$$
The
transmission line behaves as an open circuit when ZL=0 and as a short circuit when ZL=∞. This impedance transformation is used in many of
the interconnecting devices and an example is presented in the next section. On the other hand for a half-wavelength
transmission line there is not impedance transformation:
$$Z_{i n}(n\frac{\lambda}{2})=Z_{0}\frac{Z_{L}+jZ_{0}\tan n \pi}{Z_{0}+jZ_{L}\tan n \pi}=Z_{L}$$
We will discuss these and other interesting
cases of impedance transformation for MRI applications.
Examples of TL applications in MRI
As an
introduction to this presentation, and because of their importance to ensure RF safety
and system performance, I will discuss briefly two of the aforementioned interconnecting
subsystems: the balun and the directional coupler.
A balun is a
device that transforms a balanced signal to an unbalanced one. The quarter wavelength balun (or bazooka
balun) is a nice example to show the practicality of the impedance
transformation property of a TL. It can be placed, for example, on the coaxial cable
that connects the coil (balanced port) to the T/R switch (unbalanced port) to suppress
common mode currents in the outside of the coaxial shield. This is extremely
important to ensure safety of coaxial connections placed near the patient as
well as to avoid cable cross talking when other coaxial connections are nearby.
An outer sleeve (normally implemented by a connection to the shield of a coaxial
segment) is connected to the shield of the main coaxial connection at a
distance of λ/4 forming a short circuited TL. Because of the impedance
transformation (Zin(λ/4)),
a current circulating in the coaxial shield will see a very high impedance at
the balanced connection point as it is shown in Figure 4.
To ensure RF
safety and system performance, power delivered to the coil can be monitored through
a directional coupler located in the
transmit chain. A MR compatible coupler can be built from a pair of coupled microstrip lines
as shown in Figure 5a. Using one
more time the impedance equation (Zin) and performing a differential and
common-mode excitation analysis we can derive the coupling factor and the
equivalent characteristic impedance of the microstrip coupler modeled as two TLs
as shown in Figure 5b.
$$C(\frac{\lambda}{4})=\frac{Z_{0cm}-Z_{0d}}{Z_{0cm}+Z_{0d}}, Z_{0}=\sqrt{Z_{0cm}Z_{0d}}, Z_{0cm}=f(w,h,s,\epsilon_{r}), Z_{0d}=g(w,h,s,\epsilon_{r})$$
C is the coupling factor between the
input and the coupled port (normally above 10 dB),
Z0d and
Z0cm
are the differential and common-mode impedances respectively which are
(non-linear) functions of the geometry of the coupled microstrip lines (
Figure 5a). A figure of merit of
the directional coupler is its directivity which is a measure of how much power
is coupled when power is injected in the output port (reversed connection). Microstrip
couplers are tuned couplers and have poor directivity in general. As
alternative, transformer based couplers present higher directivity (> 15 dB)
and insertion losses are minimal (<2 dB) across the MR spectrum. However,
these are magnetic and should be left out the MRI
room. During my presentation I will discuss further these and other devices for
power measurement.
Discussion
This presentation will focus on TL
concepts and how this concepts are used in the design of the different
interconnecting devices located in the transmit and the receive chains.
Conventionally in MRI the RF signal travels along these chains through coaxial
connections. However, the complexity of these connections (cabling layout,
coupling and balun construction) increases with the number of transmit and/or
receive channels. Therefore there are ongoing research efforts to replace
coaxial connections by optical links or wireless communications. Miniaturization
of the electronics is another important factor at the time of construction of high
density arrays. To this end, transmission line segments are replaced by their
lumped element equivalent. However a discrete component implementation becomes
difficult when the MR frequency falls in the ultra-high frequency range (300
MHz-3 GHz). Finally, new RF power amplification approaches have been presented
that eliminate the need of some of the conventional interconnections inserted between the
amplifier and the coil to ensure efficient power transfer. I will discuss these
non-conventional approaches further during the last part of my presentation.
Acknowledgements
To my colleagues from the LFMI
group for feedback during the preparation of this syllabus.References
Recommended books:
R.
E. Collin, Foundations for Microwave Engineering, 2nd ed. IEEE Wiley-Interscience, 2000
J.
C. Freeman, Fundamentals of Microwave Transmission Lines, John Wiley & Sons, Inc., 1996.
T. H. Lee, Planar Microwave Engineering.
Cambridge University Press, 2004.
B. C. Wadell, Transmission Line Design
Handbook. Artech House, Boston, 1991.
D. M. Pozar, Microwave RF Engineering. 3rd ed, John Wiley & Sons, Inc., New York, 2005.