Synopsis
· Basic understanding of how an MRI works can be
achieved by comprehending its major functional subsystems.
· The subsystem currently experiencing the
greatest innovation is RF transmission.
· Software defines the look and feel of the
system and is the most important differentiator between systems.
Introduction
MRI has
reached a high level of maturity in the past four decades. Although many of the
system components can be considered to be “black boxes” that most users will
never have to modify, it is beneficial to have an understanding of the various
operational blocks. These can be categorized based on the functionality
necessary to acquire and reconstruct an MR image. To produce an image, three
basic groups of electromagnetic fields have to be generated: the static
magnetic field, the gradient magnetic fields, and the radiofrequency (RF)
fields; in addition, RF fields need to be detected and processed by the system.
The subsystems necessary to generate and manipulate these fields are briefly
described below as well as some further subsystems that are part of most
clinical MRI scanners.
Magnet
The most
obvious subsystem of any MRI system is the magnet used to generate the static
magnetic field B0. This field is required to orient the nuclear spins and
separate them into discrete energy levels. The stronger the magnet, the larger
the overall spin polarization and thus the larger the achievable
signal-to-noise ratio (SNR). As this magnetic field is static with no time
variation, there is no associated electric field. Typical field strengths of
modern clinical MRI systems are 1.5 and 3 Tesla. To achieve higher SNR, there
has been a continuing trend to higher magnetic fields throughout the history of
MRI. Currently, 7 Tesla is being considered for introduction into routine clinical
use, but there are also human research systems operating at 9.4 and 10.5 Tesla,
with 11.7 Tesla in planning.
Lower
magnetic fields can be generated by electromagnets or by permanent magnets.
Above about 1 Tesla, it is more efficient to generate the field using superconducting
technology, even though this approach requires cooling the magnet down to
temperatures typically around 4 K and immersing the superconducting coils in a
bath of liquid helium. Niobium titanium is the superconductor of choice for
most MRI magnets, but this technology reaches its technical limits at 11.7
Tesla. For higher magnetic fields, new superconductors such as niobium tin will
have to be utilized.
Two major
criteria when designing magnets for MRI are spatial uniformity and stability
over time. Minor departures from ideal uniformity are corrected with various
shims, which can be generated using superconducting coils, passive shim
material, or resistive coils.
Gradient Amplifiers
and Coils
The
gradient amplifiers and coils are responsible for generating magnetic fields in
addition to the main static magnetic field B0 provided by the magnet. These
fields are fairly weak, i.e. on the order of a few mT, and ideally they are
designed to vary linearly in the three orthogonal spatial directions (Gx, Gy,
Gz). Thus, there are usually three gradient amplifiers driving three separate
gradient coils. Due to switching of the magnetic fields in pulse sequences,
fields are produced with a broad spectrum of frequency components in the audio
band.
The application
of these fields to body tissue leads to a minor change in the Larmor resonant
frequency of the precessing nuclear spins, as the precession frequency is
linearly proportional to the external magnetic field experienced by the spins.
Without these fields, all of the spins would precess at the same nominal
frequency, and it would not be possible to determine from which position in the
body a particular signal is originating. By applying a Fourier transform to the
received time-signal, it is possible to assign signal strengths along each of
the spatial axes and reconstruct spatially resolved images. Note that by
spectroscopically assigning the spatial origin of the signal, the resolution of
MR images is not limited by the wavelength of the RF radiation emanated by the
excited spins. This makes MR quite different than optical imaging.
If the
gradient amplitude is too high, eddy currents can be generated in the
conducting tissue that provoke peripheral nerve stimulation (PNS). The
amplitude at which PNS occurs is dependent on the switching time of the
gradient, which has led to limits on the performance of the gradient subsystem
based on physiological considerations. Many systems have a dedicated
stimulation monitor to shut down the system if excessive gradient waveforms are
applied.
To circumvent these performance limits, one
strategy is to reduce the linearity of the gradients, so that the field
amplitudes at the periphery of the field of view (FOV) of the gradient coil are
less; such nonlinearities can be corrected to a certain extent during image
reconstruction, since the non-idealities can be well characterized during
system calibration. Another strategy is to reduce the target FOV of the coil itself.
Note that this means reducing the maximum physical FOV of the gradient system
and cannot be achieved just by acquiring images at reduced FOV. Thus, head-only
gradient systems can provide higher performance than whole-body gradient
systems.
RF Transmission
The goal of
the RF transmission subsystem is to excite the polarized spins and drive them
into the other allowed energy level. Viewed from the perspective of classical
physics, the RF energy tips the spins from alignment along B0 to an orientation
with components orthogonal to B0. Only these transverse components can be
detected by the RF reception system. The frequency of the RF energy must be
resonant with the precession frequency of the spins and ranges from about 64
MHz for hydrogen at 1.5 Tesla up to 500 MHz for hydrogen at 11.7 Tesla. Other
nuclei possess lower resonant frequencies than hydrogen due to their lower
gyromagnetic ratios.
An initial small-signal
RF pulse is generated in the system exciter, which consists of an oscillator
and a phase/amplitude modulator. This pulse is augmented to high power levels by
an RF amplifier and applied to the transmit RF coil. Typically, the goal of RF
excitation is to generate a uniform magnetic RF field B1 that is transverse to
B0. Spatial uniformity ensures that the flip angle does not vary across the
subject, and thus the image contrast is predictable and constant throughout the
entire excited volume. At lower fields, birdcage coil geometries are commonly
used, since these coils produce a very uniform transverse magnetic field in the
central portion of the coil. They can also be easily driven to produce a
circularly polarized field, which is the only component useful for spin
excitation. The reverse circularly polarized component that is inherently present
in a linear polarized field produces unwanted tissue heating due its electric
fields but no spin excitation.
Tissue
heating is in general the result of the RF electrical fields that compulsorily
accompany all RF magnetic fields through Maxwell’s equations; RF power
deposition into the tissue is characterized by the specific absorption rate
(SAR) given in W/kg. MRI systems require monitoring of the RF power being
transmitted to make sure that the RF exposure of the patient remains within
regulatory limits. Such a SAR or RF exposure monitoring system should ideally
be independent of other system components, and its correct function should be
regularly verified.
Unlike
low-frequency gradient magnetic fields, for which the body tissue is basically
transparent, RF fields do interact extensively with the tissue through which
they propagate. These interactions, which include absorption, refraction and
reflection, are dependent on the distribution of tissue conductivity and
permittivity, and they are more pronounced at higher RF frequencies. This
implies that it is quite challenging to generate a uniform B1 field in MR systems
with higher B0.
To account
for this challenge, it is now common to excite the tissue with a set of
independently controllable coil elements rather than a single monolithic coil.
Such parallel transmit systems require at least the ability to apply RF power
with arbitrary amplitude and phase weightings to each of the coil elements;
this approach is often called RF shimming, since it is similar to the shimming
strategy pursued for B0 using multiple shim coils. For more complex
applications, the complex weightings are changed at each support point of the
RF pulse, implying that different RF pulse forms are applied to each coil
element. The effect of such parallel transmit pulses (Transmit SENSE) can only
be understood in conjunction with knowledge about the trajectory through
k-space, which is determined by the accompanying gradient magnetic fields.
Accurate monitoring of RF exposure in parallel transmit systems is very
challenging.
RF Reception
Once the
spins have been excited using the RF transmission system, the RF reception
system is utilized to detect the very weak magnetic fields issued by the spins
precessing in the plane transverse to B0. In principle, the same RF coil as
used for transmission could also be used for this purpose, but in practice
separate coils are used as the design criteria differ for the two functions.
During RF reception, sensitivity is more important than uniformity. Thus, RF
coils for reception are often composed of multiple, smaller coil elements that
are placed as close as possible to the volume of interest.
Not only do
the smaller elements enhance the sensitivity (SNR) of the coils, they also
enable use of parallel imaging techniques. Since each coil element only receives
signal from a limited portion of the volume of interest, the coils inherently
provide information about the spatial localization of the signal that they
receive. This information can be used to supplement the spatial encoding
performed using magnetic field gradients, thereby reducing the acquisition time
significantly. It is important to have sufficient RF coil encoding capability along
the phase-encoding direction or directions. For simultaneous multislice (SMS) techniques
that rely on RF coil encoding, it is important to have coil elements
distributed along the slice select direction.
The RF
signals received by the RF coils are amplified by low-noise amplifiers followed
by sampling and digitization. In some systems, the digitization occurs in the
examination room very close to the coil itself. In other systems, the
digitization is performed in the equipment room. In cases where the receiver
may process signals from different-sized coil elements, the receiver must be
able to cover a large dynamic range, since in some cases signal might be
acquired from a large volume of tissue and in other cases only from a very
small volume. If the receiver is not able to cover this range at fixed signal
amplification, the system may have to provide multiple levels of gain between
the coil and the receiver.
Patient Table,
Physiological Monitoring and Other Subsystems
In addition
to the major subsystems mentioned above, a clinically useful MR imager relies
on multiple other hardware components to enable patient examinations. These
include the patient table, which is usually motor-driven and able to move the
patient to the appropriate longitudinal positon so that the images are acquired
in the central portion of the magnet. For multi-station acquisitions, the table
must also be able to quickly and accurately transport the patient from station
to station.
Another set
of important components are the physiological monitoring devices. In many
instances it is necessary to synchronize the MR data acquisition or
reconstruction to the ECG, peripheral pulse, and/or respiratory signals to
reduce motion artifacts or produce CINE movies of periodic motion.
Computer Processing
The part of
a MR system that a user is most familiar with is the computer front-end.
Software represents the main source of system complexity in clinical MRIs. It
allows the user to control all of the other hardware by selecting pulse
sequences and changing the parameters of the acquisition to achieve the desired
coverage, resolution, and contrast. Imaging parameters include TR, TE, flip
angle, FOV, readout bandwidth, imaging matrix and many others. After the data
are acquired, they have to be reconstructed, which nowadays often occurs on
fairly standard PC hardware, possibly in conjunction with GPUs (Graphics
Processing Units). Beyond basic image reconstruction, advanced post-processing
and analysis techniques are increasingly important in clinical routine and
research. Most MR systems have dedicated computers for the front-end user
interface, system control, and image reconstruction.
Conclusion
MR imaging
depends on the interaction between several hardware subsystems, including the
magnet to generate spin polarization, the magnetic field gradients to provide
spatial localization, and the RF coils to achieve both spin excitation and
signal reception. In particular the available field of view depends on the
interplay of the static field homogeneity, the linearity of the gradients, and
the transmit volume of the body coil. The performance limits of the gradients
and the RF transmit coils are set by physiological considerations, i.e. PNS and
SAR. The limits of the static magnetic field, on the other hand, are still governed
by technological and economic considerations. The physiological upper limit of
B0 acceptable for MRI is still unknown.
Acknowledgements
No acknowledgement found.References
No reference found.