Synopsis
The NMR (Nuclear Magnetic Resonance) signal can be described classically by considering the motion of the net magnetisation (the vector sum of magnetic moments of individual nuclei). By considering individual isochromats – i.e. subsets of the spins that are behaving identically– we can visualise how the received signal will decay away due to T1, T2 and T2* relaxation. By additionally considering the effects of magnetic field gradients, we can spatially localise the signal, extending NMR to MRI (Magnetic Resonance Imaging). All these effects can be described by the Bloch equations, which give complete classical description of the behaviour of magnetisation.Highlights
- The production of an NMR (Nuclear Magnetic Resonance)
signal can be described classically by considering motion (and in particular precession)
of the net magnetisation (the vector sum of magnetic moments produced by
individual nuclei).
- By considering how precession is affected by the
application of linear magnetic field gradients, we can envisage
ways to spatially localise the signal, leading to MRI (Magnetic Resonance
Imaging).
- By considering individual isochromats – i.e. subsets of the spins that are behaving
identically in terms of their orientation and precession – we can also
visualise how the received signal will decay away. This decay (described
mathematically in terms of the relaxation times T1 and T2
(and T2*) differs between tissues, allowing us to sensitize
our images to particular pathologies.
- The Bloch
equations give complete classical description of the behaviour of
magnetisation under influence of an applied field, and of relaxation.
Target Audience
MR
physicists and engineers, pulse sequence developers and clinicians who want to
deepen their understanding of the classical description of the NMR/MRI process.
Purpose and Objectives
The lecture aims to provide the information necessary to allow audience members to appraise the applicability of the classical model, and to apply this – where appropriate - in their own clinical or research work.
Following this lecture, audience members
should be able to:
Discuss the basic concepts of the classical
description of NMR, in terms of:
- Magnetic moments
- Isochromats
- Net magnetisation
- Precession
- Nutation
Understand how the use of magnetic field
gradients allow us to extend the concepts of NMR to MRI
Understand how relaxation can be described is
the classical model, and how relaxation times affect image contrast.
Be aware of the Bloch Equations, and their use
to calculate the evolution of the net magnetisation during an NMR/MRI
experiment.
From a Quantum Mechanical to a Classical Description
Background and Quantum Mechanics
Nuclear Magnetic Resonance (NMR) was first described (independently) by
Purcell, Pound and Torrey[1]
and Bloch, Hansen and Packard[2], in 1946. It is based on the resonance (i.e. matching) effect between properties of (some)
nuclei and a magnetic field. In medical fields it is common to also hear the
terms MRS (Magnetic Resonance Spectroscopy) and MRI (Magnetic Resonance
Imaging), both of which utilize the same underlying NMR phenomenon.
While a full description of NMR requires a quantum
mechanical approach, many aspects of NMR (and thus MRI/MRS) can be explained in
classical terms, by considering the nuclei as simple magnetic dipoles with a magnetic moment (μ).
Quantum mechanics tells us that in
a magnetic field B0, nuclei with an angular momentum I will have
2I+1 energy levels (labelled by a
value m, running from -I to +I). For “spin half” nuclei (I = ½) such as
hydrogen, there are just two levels (m = +/- ½), which are often referred to as
‘spin up’ (m = -½) and ‘spin down’ (m = +½),
which can be thought of as representing "spins” which are either “aligned with”
or “aligned against” the magnetic field. The energy levels differ by
$$\triangle E = \frac{\mu B_{0}}{I} =\gamma B_{0}\frac{h}{2 \pi}$$
where γ is the gyromagnetic ratio and h
is Planck’s constant. The
number of spins in each energy level is governed by a Boltzmann distribution:
$$\frac{N_{up}}{N_{down}} = e^{- \frac{\triangle E}{kY}}$$
It is the difference in ‘up’ and ‘down’ spins (approximately 1
in 106 at room temperature) which is responsible for measured NMR signal.
A Classical Approach
Following Farrah & Becker [3], "Our approach is to treat M as a measurable and easily pictured quantity and to avoid, in so far as possible, detailed discussion of quantization conditions".
If we place tissue (or anything else containing suitable spin half nuclei) in a magnetic field it becomes magnetised, with an overall magnetic moment, M, aligned along the main magnetic field, B0. Individual “spinning” nuclei behave like gyroscopes, and if disturbed, they react by moving in a direction d at right angles to both the applied (magnetic) force F and to their spin (along M) (Figure1). The spins precess around B0 at the Larmor frequency, given (in radians s-1) by:
$$\omega_{0} = 2 \pi \nu_{0} = \gamma B_{0}$$
We generally need to consider only the net magnetisation (i.e. the vector sum of the individual nuclear moments), which acts like a bar magnet (although we will sometimes also need to consider individual isochromats – i.e. subsets of the spins that are behaving identically in terms of their orientation and precession). The net magnetisation has magnetic moment given by $$$M_{0} = \chi B_{0}$$$, where χ = susceptibility and B0 is the applied magnetic field. An additional (small) internal field, caused by magnetisation itself and equal in value to it, means total field is: $$$B_{int} = B_{0} (1 + \chi)$$$. (This susceptibility effect can often be ignored, but is important when discussing gradient echoes, echo-planar imaging and spectroscopy).
The Rotating Frame, Nutation Pulses and the Free Induction Decay
The Rotating Frame
So far we have viewed M in a static (laboratory) frame of reference. Any
reference frame is equally valid, but it may be much easier to explain certain
effects in one frame than another. (Describing the trajectory of a ball thrown
inside a moving train is much easier in a frame of reference moving at the same
speed as than train than it it is from the viewpoint of an observer stood on the track
side, for example). For NMR/MRI, the most convenient frame to use is one rotating with respect to laboratory frame. If this frame is rotating in same direction as, but slightly
slower than, the precessing magnetization, then in this frame the magnetization
will appear to be precessing more slowly than in laboratory frame; if we
gradually increase speed of the rotating
frame, the magnetization will eventually appear stationary, making calculation of any other motion of the net magnetisation much easier. Since we have defined the precession
frequency in terms of magnetic field, however, (through the Larmor Equation, $$$\omega_{0} = \gamma B_{0}$$$), then if ω = 0 then
B (in this frame) must be zero (!).
Mathematically, we allow for this by considering the to magnetization to see an effective field Beff, equal to the field in laboratory frame (i.e. B0) plus or minus an
extra 'fictitious' field ω
/γ (with the + or - depending on direction of rotation):
$$B_{eff} = B_{0} -\frac{\omega}{\gamma}$$
When ω = ω0 (i.e. when the rotation speed of the
frame of reference is identical to the rate of precession of the spins), Beff
= 0, making our mathematical description consistent with the fact that the
spins are no longer moving (relative to the rotating frame).
Nutation pulses
If we now (briefly) apply a field B1, perpendicular to B0
and rotating at ω0, then in the rotating frame (also rotating at ω0), B1 will be stationary. B1 is the only effective field in this frame, and M will therefore precesses
about it, rotating away from z-axis towards xy-plane at an angular
rate of $$$\omega_{1} = \gamma B_{1}$$$ (the Larmor
Equation again). When we turn off B1, the magnetisation will have been rotated by a tip angle (or flip
angle) given by
$$$\theta = \omega_{1}\tau = \gamma B_{1}\tau$$$
where τ
= duration of pulse. A 90o pulse has a duration
of $$$\frac{\pi}{2\gamma B_1}$$$, while a 180o
pulse has a duration of $$$\frac{\pi}{\gamma B_1}$$$, etc, etc…
Resonance
Resonance is “The condition of a body when it is subjected
to periodic disturbance of the same frequency as the natural frequency of the
body or system” (Collins English Dictionary) (think of a wine glass 'ringing') or “The condition of a system in which
there is a sharp maximum probability for the absorption of electromagnetic
radiation” (Ibid).
Quantum mechanically, transitions between energy levels
occur only when a photon of the correct energy is applied, while in the classical
formalism magnetic moments (and therefore the net magnetisation) are only
affected when a radio wave of a particular frequency is applied; in the static
(laboratory) frame, nutation pulses correspond to magnetic fields
varying at the resonant frequency, i.e. to radio frequency (RF)
pulses.
Free Induction Decay
Faraday's Law tells us that a moving magnet will induce a current
in a surrounding/nearby electrical conductor, and following a 90o nutation
pulse, the transverse magnetization (Mxy) constitutes such a moving
magnet. A nearby coil can therefore detect a Free Induction
Decay (FID) signal. The initial amplitude of the FID is proportional to the proton density (i.e. the concentration
of the nuclei contributing to the signal), and the signal decays to zero with
time constant T2* (see later). (NB - Mz does not give a signal, as it is stationary with respect to the laboratory frame, and therefore the coil).
Spatial Localisation,Echoes and Imaging
Field Gradients
By convention, the main magnetic field, B0, is
aligned with z (usually along the bore of the magnet); a field gradient along x, for example, means that the
strength of the magnetic field (still itself aligned in z direction) varies with
position along x (say).
$$B = B_0 + G_x x$$
The scanner hardware can create linear gradients in x, y,
and z and, by applying gradients simultaneously on multiple axes, in any
arbitrary direction. To understand why this is useful, consider what happens
when magnetic field, B, is not uniform over the object:
If the field increases linearly with position, then resonant
frequency must also increases linearly with position, which implies that we can
use frequency to spatially encode
the signal; by acquiring signals while gradients are applied in a number of
direction, we can build up projections from which a complete
image can be reconstructed. Early scanners used such a projection-reconstruction
approach (which is similar to how a CT scanner works), but it turns out to be difficult
to reconstruct artefact free images from projections, which has led to an
alternative approach called "spin warp" imaging (which uses a phase
encoding technique) becoming much more common. It is also technically difficult to measure a
signal immediately after a 90o pulse (and doing so doesn’t allow us
to control the degree of signal decay due to T2 or T2* (which we discuss
below), so images are usually acquired using an echo based technique
(also discussed below).
Spin Warp Imaging
Creating an image requires encoding of spatial positions in
3 dimensions. In 2D imaging, we first select a slice, and then perform spatial
encoding in 2 directions, usually repeating the process for a number of slices,
in order to cover the whole object. For 3D imaging, we may choose to select
a “slab” (i.e. a single thick slice) from which we will collect signal, or may opt
to collect signal from the whole object being imaged, but in either case will
need to spatial encode the resulting signal in 3 directions.
A full description of spatial encoding techniques commonly
used is beyond the scope of this talk, but very briefly:
- Slice selection: Slice (or slab) selection is achieved by applying a slice select gradient simultaneously with a shaped RF pulse for the RF pulse(s) needed to excite a signal. By changing the shape and
timing of the RF pulse, and the amplitude of the gradient, it can be arranged
that only a thin section of the object being imaged has a resonance frequency
within the bandwidth of the pulse, and thus only this section (slice) will be
excited. (Figure 2a).
- Frequency Encoding: If we apply a gradient during data
acquisition (often called read
or readout gradient),
then spins at different positions experience slightly different local fields
and therefore precess at different rates. The resulting signals thus have
different frequencies (which can be distinguished using a Fourier Transform).
(Figure 2b).
- Phase Encoding: If we apply a gradient before data
acquisition (usually called the phase
encoding gradient) then, as before,
spins at different positions experience slightly different local fields and
precess at different rates. By end of gradient they have therefore precessed
through different angles, giving signals with spatially dependent phase. Because
we can only measure phase to within 360o, however, phase encoding
must be repeated many times (e.g. 128 or 256); while the signal phase itself is not unique, the way in which phase changes with gradient strength is. (Figure 2c).
Echoes
Spin Echo
imaging is the basis for many structural MRI (sMRI) sequences, and uses an
extra (180o) pulse to refocus effects of dephasing. As shown in Figure 3a, an initial 90o
pulse (along x') produces transverse magnetisation along y', which dephases due
to variations (inhomogeneities) in the local magnetic field. Applying a 180o
pulse (along x') flips spins to a different (mirrored) position in the
transverse plane, after which they continue to precess in same direction and at
same speed as before (as they are still experiencing the same local magnetic
field). The transverse magnetisation therefore
rephases along the –y’ axis, producing an
echo signal that reaches a maximum at the echo time, TE, before decaying away
again.
Gradient Echo
imaging, which creates an echo without using a 180o pulse, is
used for both structural and functional MRI (fMRI). As shown in Figure 3b, an initial excitation
pulse (show here as a 90o pulse (along x'), although lower flip
angles can also be used) produces transverse magnetisation along y', in the
same way as for a spin echo. If a
magnetic field gradient is now applied, the transverse magnetisation will dephase
due to both time-invariant local field inhomogeneities (as in the spin echo
case) and the applied gradient. If the gradient
is then reversed, spins will precess in opposite direction and at (almost) the same
speed as before, so that the transverse magnetisation rephases, again
producing an echo signal that reaches a maximum at the echo time, TE, before
decaying away again. Since for the
effect of local inhomogeneities remains the same throughout the whole TE
period, however (in contrast to the applied gradient, which is positive during one
half of TE and negative during the other), the effect of local in homogeneities
is NOT refocused.
Pulse Sequence Diagrams
A Pulse Sequence
Diagram (e.g. Figure 4,
which shows a spin echo sequence) can be used to diagrammatically show all
gradients & RF pulses that are applied in order to form an image. It shows
the relative timing and amplitudes of each pulse (although it may not always be drawn exactly
to scale). Shading is often used to indicate gradient areas which should be equal
(or equal and opposite), with arrows indicating exact timing of events (in
particular TE, the Echo Time and TR, the Repetition Time (i.e. the length of time between applying the first excitation pulse
in a sequence, and repeating this (for signal averaging, or to collect the next
phase encode step)).
Relaxation and Image Contrast
Relaxation Times, TE & TR
After we disturb the magnetisation in any way, it relaxes back to its original
configuration. This effect is described
by two main time constants, T1 and T2. (We’ll also
mention T2*). These relaxation times are fundamental MR
parameters reflecting the ‘tissue environment’ (the binding and mobility of
water within tissues).
Along with PD (the proton density), T1 and T2
are responsible for (most) of the contrast in MR images. Different combinations of TE
and TR can be used to produce different degrees of contrast (intensity
difference) between tissues: changes of TR lead to changes in contrast
between tissues with different T1 relaxation times, while changes in TE affect contrast between tissues with different T2s.
Mathematically, for a spin echo sequence:
$$I_{SE} = g {PD} \exp(-\frac{TE}{T_2}) [1 - \exp(-\frac{TR}{T_1})]$$
(A similar equations holds for some gradient echo sequences,
with T2 being replaced by T2* to reflect the difference
in signal loss due to dephasing in local field inhomogenities).
Relaxation Times
T1 is known as the Longitudinal or Spin-lattice relaxation time, and
describes the return of the net magnetisation towards its equilibrium position
(parallel to B0) after it has
been disturbed (e.g. by a 90o pulse flipping it into the transverse
plane). T1 is time taken for
z-magnetisation to recover to 67% of its initial value following a 90o
pulse. It varies between tissues, and T1 contrast often gives good
differentiation of anatomical structures. (Figure 5a).
T2 (the Transverse or Spin-spin relaxation
time) also varies between tissues, and is often (non-specifically) increased in
pathological regions. (See Figure 5b). After
excitation, any differences in the local magnetic field the experience will
lead to spins precessing at different frequencies; the overall net transverse
magnetisation will therefore dephase,
and the measured signal will decrease.
T2 is time taken for signal to decay, in a homogenous static
magnetic field, to 37% of its initial
value, while T2* is the equivalent time in an inhomogeneous
field.
The Bloch Equations
The Bloch
equations [4] give a complete classical description of the behaviour of
magnetisation under influence of an applied field, and of relaxation. The applied
field can include:
- any Bz present in rotating frame (i.e.
off-resonance effects)
- field gradients (which may vary with time)
- the RF field B1 (which may be applied
x' or y' axis, or a combination, and may vary with time)
Precession of M around B is described by:
$$\frac{\text{d}M}{\text{d}t} = \gamma M\times B$$
... which expands to:
$$\frac{\text{d}M_x}{\text{d}t} = \gamma (M_yB_z - M_zB_y)$$
$$\frac{\text{d}M_y}{\text{d}t} = \gamma (M_zB_z - M_zB_y)$$
$$\frac{\text{d}M_z}{\text{d}t} = \gamma (M_yB_z - M_zB_y)$$
Relaxation is dealt with by additional (phenomenological)
terms:
$$\frac{\text{d}M_x}{\text{d}t} = \gamma (M_yB_z - M_zB_y) - \frac{M_x}{T_2}$$
$$\frac{\text{d}M_y}{\text{d}t} = \gamma (M_zB_z - M_zB_y) - \frac{M_y}{T_2}$$
$$\frac{\text{d}M_z}{\text{d}t} = \gamma (M_yB_z - M_zB_y) - \frac{M_0 - M_z}{T_1}$$
... and additional terms can also be added (e.g. to allow for
diffusion or other weightings).
Summary
- Quantum Mechanics says that some nuclei will have
angular momentum, or “spin”; this includes
protons (hydrogen nuclei).
- Spinning,
charged, particles create a (tiny) magnetic field, so hydrogen nuclei - in water, and organic molecules - have a magnetic moment (much
like a bar magnet).
- In
a magnetic field, such NMR visible nuclei align themselves with direction of the magnetic field (B0),
some “up” and some “down”, and the overall
effect of all up and down spins can be described in terms of a single net magnetisation, M.
- M
behaves like a gyroscope and precesses around
magnetic field (B0) at Larmor frequency, $$$\omega_{0} = \gamma B_{0}$$$.
- If
‘flipped’ into transverse plane, M induces a
detectable voltage in surrounding conductors.
- Gradients for allow us to spatially localise this signal, forming an image.
- Differences
in relaxation times determine
how much signal we measure, allowing us to manipulate image contrast.
- The Bloch equations allow the overall magnetisation
at any time to be calculated, allowing for effects of RF pulses, relaxation, etc.
Conclusions
The classical model of the NMR phenomenon provides a simple (and importantly relatively intuitive) frame work within which to visualise and describe many NMR/MRI phenomena. While quantum mechanical explanations may be necessary in more complex situations (for example when describing (homo- or hetero-nuclear) interactions between different spin systems) for simple systems the Bloch equations give complete classical description of the behaviour of magnetisation under influence of an applied field, and of relaxation.
Acknowledgements
Thanks to all who have, over the years, contributed to the material in this lecture.References
[1] E. M.
Purcell, H. C. Torrey, and R. V. Pound, Resonance Absorption by Nuclear
Magnetic Moments in a Solid.
Phys. Rev. 69, 37 (1946)
[2] F.
Bloch, W. W. Hansen, and M. Packard. The Nuclear Induction Experiment. Phys.
Rev. 70, 474 (1946)
[3] T. Farrar and E. Becker, Pulse and Fourier
Transform NMR. Introduction to Theory and Method, Elsevier, 1971
[4] F. Bloch, Nuclear Induction, Phys. Rev. 70, 460 (1946)