Synopsis
This presentation will provide an overview of the main gradient echo based (gradient spoiled, RF spoiled and balanced steady state free precession) and conventional/fast spin echo based pulse sequences and will illustrate some methods by which their behaviour can be simulated Introduction
The behaviour
of the magnetisation in a pulse sequence is described by the Bloch equations (1) and a number of methods have been
utilised to simulate the behaviour of spins in an imaging sequence such as the matrix
formalism (2), the use of extended phase graphs
(EPG) (3, 4), pictorial representations (5) and analytical descriptions (6, 7). The examples given in this lecture
are based upon code from Brian Hargreaves' Bloch equation simulations (http://wwwmrsrl.stanford.edu/~brian/bloch/),
Matthias Weigel's EPG simulations (http://epg.matthias-weigel.net/) and a
Matlab implementation of the C program given in Klaus Scheffler's article (5).
Gradient Echo (GE)
A gradient echo sequence comprises a train of equally spaced radiofrequency (RF) excitation pulses and the necessary gradients to perform slice selection, phase encoding and frequency encoding. The echo signal is formed by the application of a negative dephasing gradient followed by the frequency encoding gradient. The minimum echo time (TE) is dependent upon the RF pulse duration, receiver bandwidth and available gradient performance. Usually the time between excitation pulses (TR) is short and the transverse magnetisation has not had time to fully dephase. Since the phase encoding gradient is the only gradient that changes between TRs the transverse magnetisation will have a position-dependent distribution of resonant offset phase angles across the slice, which will affect the signal created by the subsequent RF pulse. As a result, bands of varying signal intensity, known as "FLASH-bands" may appear degrading the image quality (Figure 1a). The three main gradient echo sequences differ in how they deal with this residual transverse magnetisation: a) the gradient spoiled sequence averages the resonant offset phase distribution across a voxel (Figure 1b), b) the radiofrequency (RF) spoiled sequence eliminates the transverse magnetisation (Figure 1c) whilst c) the balanced steady state free precession (bSSFP) sequence rephases the phase distribution. Example pulse sequence diagrams and images are shown in Figure 2.
Gradient
Spoiling
The
gradient spoiled sequence has additional gradients applied after data
acquisition on the slice select and frequency encoding axes (Figure 2a) that averages
the resonant offset phase distribution across a voxel. The phase encoding
gradient is "rewound" so that the dephasing gradients are constant
across TRs. Figure 2b shows an axial head image obtained using a gradient
spoiled sequence.
RF Spoiling
The
gradients are the same for both the gradient and RF spoiled sequences (Figure
2a), but the phase of the RF excitation pulse is incremented, usually quadratically,
on each excitation.
$$ ϕ_j = \frac{1}{2} ϕ_0 (j^2+j+1),j=0,1,2,…
$$
There
are certain “magic numbers”, e.g. $$$ϕ_0=117^0$$$, for the for the phase increment that can produce excellent spoiling across a
range of T1, T2
and excitation flip angles (8) and is equal to the steady-state transverse
magnetisation for a perfectly spoiled sequence given by the Ernst equation
$$ M_{ss}=M_0 (1-exp(-TR/T_1 ))/(1-cos(α) exp(-TR/T_1 ) ) \cdot sin(α) $$
Balanced
Steady State Free Precession(bSSFP)
The
bSSFP sequence (Figure 2d) has all gradients balanced, i.e. there is zero net
gradient area on each axis over a TR period. Any resonant offset phase errors
will therefore not be averaged. This means that the signal response is
dependent upon the phase offset. Figure 3 shows a simulation of the signal
behaviour as a function of resonant offset phase angle for two excitation
schemes. In Fig 3a and 3b there is no phase offset applied between subsequent RF
pulses. In Figure 3c and 3d the phase of the RF excitation is alternated
between 0 and 180°. In this example a large left-right shim offset of
14.7Hz/cm was applied. The resonant offset is given by $$$θ=2π\cdot Δf\cdot TR$$$ and the signal nulls or bands appear at $$$θ=±π$$$
, or when $$$Δf=±\frac{1}{2\cdot TR} $$$ . In this case the TR
was 5ms resulting in bands appearing at $$$±$$$100Hz . Figure 2e shows the
image obtained using a bSSFP sequence. Note an off resonance band can be seen
anteriorly.
The
steady state signal for a bSSFP sequence, assuming that $$$TR≪T_1,T_2$$$, is given by
$$M_{ss}=M_0 \frac {sin(α)} {(1+cos(α)+(1-cos(α) ) \cdot \frac{T_1}{T_2}}$$
The
optimal flip angle is given by
$$cos(α)=\frac{\frac{T_1}{T_2} -1}{\frac{T_1}{T_2} +1}$$
which
results in a steady state signal response of
$$M_{ss}=\frac{1}{2} M_0 \sqrt{\frac{T_2}{T_1}}$$
Hence
tissues with a high $$$\frac{T_2}{T_1}$$$ ratio, e.g. fluids, will exhibit a high
signal.
Spin Echo (SE) and Fast Spin Echo (FSE)
Spin
Echo
Spin
echo (SE) is a basic pulse sequence comprising typically of a 90° and a 180° RF
pulse. The initial 90° excitation pulse tips the longitudinal magnetisation
into the transverse plane where it naturally dephases due to T2*
relaxation processes. After a time τ
the 180° refocusing pulse is applied which rotates the transverse magnetisation
180° about the axis along which the pulse is applied. This rotation effectively
inverts the phase of the magnetisation so that it now naturally rephases,
coming fully back into phase, and forming what is known as a spin echo, after a
further time period τ. The time between
the initial 90° excitation pulse and the centre of the echo is known as the
echo time (TE) and is equal to 2τ. The
effect of the refocusing pulse is to eliminate the effect of any static
magnetic field non-uniformities from the T2* signal
therefore leaving any signal differences due to irreversible T2
interactions. The magnetisation will then recover along the z-direction due to
T1 relaxation until the application of the next 90° pulse. The time
between repeat applications of the 90° pulse is the repetition time (TR). The
signal equation is:
$$M=M_0 (1-exp(-TR/T_1 )\cdot exp(-TE/T_2)$$
Fast Spin Echo
Proton density or T2-weighted spin echo sequences are
particularly slow since they require a long TR in order for there to be
appreciable longitudinal magnetisation recovery before the next excitation
pulse. Although the acquisition of other slices can be interleaved within the
TR period the sequence is still not particularly efficient. Fast, or turbo,
spin echo (FSE/TSE) was developed as a method for reducing the overall
acquisition time (9). The basic principle is to use a Carr-Purcell-Meiboom-Gill
(CPMG) multi-echo spin echo acquisition, where the transverse magnetisation is
repeatedly refocused by the application of a train of refocusing pulses forming
multiple spin echoes each at correspondingly increasing TEs. If each echo is
individually phase encoded, then they can be used in the same image k-space resulting in a shorter
acquisition time. It is necessary to arrange the amplitudes of the phase
encoding gradients so that the echo nearest the desired TE (known as the
effective echo time (TEeff), has the lowest amplitude phase encoding
gradient, i.e. that echoes’ data will be nearest the centre of k-space. The interval between refocusing
pulses is known as the Echo Spacing (ESP) and the number of refocusing pulses
per TR is known as the Echo Train Length (ETL). Since multiple echoes are
acquired per TR there is also a filtering effect due to the T2 decay
of the echo amplitudes over the ETL. The T2 decay results in a broad
point spread function which has the effect of blurring the images.
A
limitation of FSE is that the large power deposition associated with long echo
trains. A common solution to this problem is to use reduced flip angle
refocusing pulses that substantially reduce the power with only a minor impact
on the signal-to-noise ratio (SNR). The
use of refocusing pulses $$$\neq$$$180° do, however, give rise to a number of
coherence pathways which lead to the creation of a so-called
pseudo-steady-state (PSS) (10). These coherence pathways can be most easily
studied using the EPG method in which the magnetisation is described as
substates which are defined as groups of spins with different Mx, My and Mz (3). Figure 4 shows a hand-drawn EPG for an
initial excitation pulse followed by three refocusing pulses . The essential feature
of the phase graph concept is that after an RF pulse has been applied the
magnetisation behaves as a composition of three parts a) dephasing transverse
magnetisation, b) rephasing transverse magnetisation and c) longitudinal
magnetisation (4). The dephased magnetisation after the excitation pulse is
labelled F1. The first refocusing
pulse will invert a component of the magnetisation, creating the F-1 substate and will also produce z-magnetisation
represented by Z1 and its complex
conjugate Z-1. Since the phase
information does not evolve with time these longitudinal substates are represented
as horizontal (red) lines. The dephasing of F1 gives rise to F2, whilst rephasing of F-1 causes the formation of a echo (F0) shown by the pink
circle. Further time evolution creates higher order substates which contribute
to the amplitude of each echo. The
evolutions of transverse substates from larger number of pulses are usually
displayed as shown in Figure 5.
Summary
The
methods outlined in this article, and presented in depth in the references, can
be used to simulate a range of MRI pulse sequences. Example applications for
these methods include designing magnetisation preparation schemes to rapidly
stabilise the signals at the start of a spoiled gradient echo (11), bSSFP (12)
or CPMG FSE (13) acquisition or developing variable flip angle schemes to
manage RF power deposition and blurring (14-16).
Acknowledgements
Addenbrooke’s Charitable Trust and the NIHR comprehensive
Biomedical Research Centre award to Cambridge University Hospitals NHS
Foundation Trust in partnership with the University of Cambridge.References
1. Bloch F. Nuclear induction. Phys Rev 1946; 70:460-474.
2. Jaynes E. Matrix treatment of nuclear induction. Phys Rev 1955; 98:1099-1105.
3. Hennig J. Multiecho imaging sequences with low refocusing flip angles. J Magn Reson 1988; 78:397-407.
4. Weigel M. Extended phase graphs: dephasing, RF pulses, and echoes - pure and simple. J Magn Reson Imaging 2015; 41:266-295.
5. Scheffler K. A pictorial description of steady-states in rapid magnetic resonance imaging. Concepts in Magn. Res. 1999; 11:291-304.
6. Sobol WT, Gauntt DM. On the stationary states in gradient echo imaging. J Magn Reson Imaging 1996; 6:384-398.
7. Hanicke W, Vogel HU. An analytical solution for the SSFP signal in MRI. Magn Reson Med 2003; 49:771-775.
8. Zur Y, Wood ML, Neuringer LJ. Spoiling of transverse magnetization in steady-state sequences. Magn Reson Med 1991; 21:251-263.
9. Hennig J, Nauerth A, Friedburg H. RARE imaging: a fast imaging method for clinical MR. Magn Reson Med 1986; 3:823-833.
10. Alsop DC. The sensitivity of low flip angle RARE imaging. Magn Reson Med 1997; 37:176-184.
11. Busse RF, Riederer SJ. Steady-state preparation for spoiled gradient echo imaging. Magn Reson Med 2001; 45:653-661.
12. Le Roux P. Simplified model and stabilization of SSFP sequences. J Magn Reson 2003; 163:23-37.
13. Hennig J, Scheffler K. Easy improvement of signal-to-noise in RARE-sequences with low refocusing flip angles. Rapid acquisition with relaxation enhancement. Magn Reson Med 2000; 44:983-985.
14. Hennig J, Weigel M, Scheffler K. Calculation of flip angles for echo trains with predefined amplitudes with the extended phase graph (EPG)-algorithm: principles and applications to hyperecho and TRAPS sequences. Magn Reson Med 2004; 51:68-80.
15. Worters PW, Hargreaves BA. Balanced SSFP transient imaging using variable flip angles for a predefined signal profile. Magn Reson Med 2010; 64:1404-1412.
16. Busse RF. Reduced RF power without blurring: correcting for modulation of refocusing flip angle in FSE sequences. Magn Reson Med 2004; 51:1031-1037.