Synopsis
·
EPI favors high bandwidth acquisitions to reduce
susceptibility artifacts.
·
fMRI acquisition methods critically depend on the
targeted spatiotemporal resolution.
·
The spatiotemporal resolution of fMRI can be
optimized by a combination of k-space trajectory design, receiver coil array,
and reconstruction algorithm.
·
Sequences using spin-echo or gradient-echo, the echo
time, and the flip angle can tune the sensitivity of fMRI acquisitions.
·
Physiological noise is a dominant noise source in
high-field fMRI experiments.
· Care must be taken to get the best shimming and to minimize motion as well as acoustic noise/vibration.
Introduction
Functional MRI (fMRI) in humans1 using the
blood-oxygen level dependent (BOLD) contrast2,3 allows
non-invasive detection of hemodynamic responses associated with neural
activity. Neuronal activity results in a complex series of hemodynamic changes
in blood flow, volume, and oxygenation, whose net effect results is BOLD signal
increase or decrease4.
Single-shot echo-planar imaging (EPI), which has been the principal technology
for fMRI, has a sampling rate of 1–3 s and spatial resolution of 3–5 mm for 3D
brain imaging. Note that spiral imaging (for review, see 5) is also
a widely used method in fMRI. Since EPI and spiral imaging share a lot of
features, here we only discuss EPI.
The readout of EPI typically
lasts for around 50–100 ms, which makes it susceptible to any disturbance
during data acquisition. Notably, susceptibility can disturb the resonance
frequency distribution and cause image distortion. Meanwhile, EPI can also
suffer from signal loss due to intra-voxel dephasing. Taken together,
acquisition methods with (effectively) high bandwidth and short readout without
compromising the FOV and spatial resolution are desired.
High
spatiotemporal resolution fMRI acquisition
The desired spatiotemporal resolution of EPI in an fMRI experiment
depends on the hypothesis to be tested. The canonical hemodynamic response, the
impulse response of the BOLD signal after a brief neuronal activity, has been
commonly considered temporally smooth with most of the energy below 0.1 Hz.
This spectral property supports the protocol of setting EPI repetition time (TR,
the time between two consecutive volume acquisition) to about 2 s. While this
TR is used in most fMRI experiments, faster EPI with a shorter TR may be
favorable in experiments attempting to critically monitor physiological noise
and interested in the fine temporal features of the BOLD signal.
Partial Fourier acquisition
is a method of reducing the data acquisition time. It is based on the
mathematical assumption that the k-space
data points are partially redundant. In practice, one can take 6/8 of the k-space by leaving out the ¼ of high
spatial harmonics k-space to save 25%
of the data acquisition time at the cost of reduced signal-to-noise ratio.
Multi-shot sequences separate the k-space traversal into multiple acquisitions.
Effectively this method can increase the bandwidth and reduce the echo spacing
(thus reduce image distortion caused by susceptibility). However, multi-shot
sequences can also be susceptible to shot-to-shot instabilities caused by subject
motion and/or physiological noise6.
Importantly, the sampling rate of a multi-shot sequence is much lower than that
of a single-shot sequence.
Sharing part of the k-space data during dynamic scanning can also improve the sampling
rate at the cost of losing some dynamic information. Key-hole
imaging7,8 is a
method that only updates the central part of the k-space in a dynamic scan.
Thus the sampling rate can be improved without reducing the spatial resolution.
However, the contrast can be reduced since the high spatial harmonic
information is repetitively used.
Parallel MRI is a method of reconstructing images
using spatial information derived from multichannel RF receiver coil arrays.
Parallel MRI can dramatically improve the sampling rate of dynamic MRI because
the spatial encoding no longer completely depends on gradient, but rather the
combination gradient and RF coil sensitivity9-11.
Accelerated multi-slice EPI acquisitions based on simultaneous excitation,
simultaneous echo refocusing, and signal separation using coil sensitivity
profiles have been demonstrated at both 3T and 7T, offering maximal full-brain
sampling resolutions of about 0.4 seconds12-14. There
are also methods of single-shot highly accelerated fMRI affording a sampling
rate up to 10 Hz (TR = 0.1 s) with whole-brain coverage15-17. Preliminary
results suggest that the BOLD signal can carry physiologically meaningful
information at the time scale of hundreds of milliseconds18,19. However,
these methods have to lower the spatial resolution in order to achieve a high
sampling rate.
Aside from
improving the temporal resolution of fMRI acquisition, there have also been
efforts in pushing the limits of spatial resolution, which can be implemented
by using specialized gradient coils and/or parallel MRI methods. Together with
an advanced structural MRI reconstruction, EPI of 1 mm3 isotropic
resolution has been demonstrated to analyze the specificity of cortical laminar
layers20. The methods of high spatial resolution fMRI
have brought remarkable results. For example, in characterizing the human visual
system, there have been reports on ocular dominance columns21 and
orientation dominance columns22 mapping.
Optimizing
the sensitivity of fMRI acquisitions
Draining veins can cause strong BOLD signal in typical gradient-echo-type
sequences. This can strongly bias the estimation of true site of neuronal
activity. Spin-echo-type sequences have been suggested to have higher
specificity of the extra-vascular BOLD signal than gradient-echo-type sequences23,24. However,
one of the challenges in spin-echo-type sequences is the reduced sensitivity25,26. At high
fields, spin-echo magnetization preparation raises further concerns on
specific-absorption rate and the sensitivity to inaccurate spin-echo due to a
shorter wavelength.
The other imaging parameter
to be optimized in fMRI experiment is the flip angle. It has been well-known
that the flip angle can be set to the Ernst angle (arccos(exp(-TR/T1)) to
maximize the signal strength. A recent study shows that, the flip angle can be
set below this Ernst angle without detrimental effects when physiological noise
is taken into consideration27.
Similarly, the echo time
(TE) should be also optimized in fMRI. Since BOLD is a T2*-weighted contrast,
it can be derived that setting TE to T2* can obtain the maximal sensitivity.
The T2* value depends on the magnetic field strength. However, TE = 30 ms and
20 ms has been quite commonly used in 3T and 7T studies, respectively. Note
that T2* of the gray matter can also change among brain areas. Thus TE may be
further tuned to improve the BOLD signal sensitivity if local T2* is known.
Physiological
noise
The noise sources confounding the BOLD-contrast fMRI data processing can be
categorized into two types: system noise and sample noise. System noise can
arise from suboptimal instrumental performance. This includes, but is not
limited to, thermal noise in the radio-frequency coils, preamplifiers, and
other electronic components in the receiver processing chain. Sample noise is related to the properties of
the object to be imaged. For example, resistive and dielectric losses due to
the presence of the sample inside the RF coil contribute to sample noise. In
fMRI experiments, motion during data acquisition is another significant source
of noise28. Motion
effects can be effectively reduced by either restricting head movement of the
participant inside the RF coil or using image volume alignment to reduce
image-to-image signal variation under the assumption of rigid body motion
between acquisitions29,30. Other
sources of sample noise can result from intrinsic physiological processes. In
fMRI experiments, physiological noise can be further separated into echo-time
and non-echo-time dependent components31, with the
latter component closely related to periodic cardiac and respiratory activity.
Comparing system and sample noise in terms of improving contrast-to-noise ratio
(CNR) in high field fMRI experiments, the latter constitutes the major
limitation. Physiological noise is generally proportional to the signal and
going to higher field strength increases its contribution to overall variance31. In
addition, at a given field strength (e.g.
3T), improvements to receiver hardware and signal reception32 can
result in physiological noise dominating the variance in fMRI time-course data.
Physiological noise in fMRI
data can be reduced by a few approaches. First, it is common to use the pulse
oximeter and respiration belt to monitor the cardiac and respiratory cycles
synchronously with EPI acquisitions. Post-processing methods (DRIFTER33 and RETROICOR34, for
example) can computationally remove these two major fluctuations from EPI time
series. Alternatively, it has been suggested using high spatial resolution
protocol to reduce physiological noise. This is because the physiological noise
scales with the voxel size. Averaging imaging voxels of high spatial resolution
can thus reduce the noise level without compromising the signal strength.
Lastly, acquiring fMRI at the rate higher than the Nyquist frequency, e.g. 3 Hz, allows straightforward
filtering of both cardiac and respiratory fluctuations.
Other
concerns in fMRI data acquisitions
Since EPI uses a fast switching gradient to complete the k-space traversal in a fraction of a
second, the Lorentz force generated by the gradient coil generates strong
acoustic noise, which can even elicit complex neuronal responses35. One way
to reduce the acoustic noise is tuning the echo-spacing to avoid acquiring data
at peaks of acoustic resonance frequencies. However, cautions must be taken
because echo-spacing also directly affects the image distortion in EPI.
Like other MRI experiments,
fMRI also favors highly homogeneous magnetic fields over the course of the
experiment. This suggests the importance of high quality shimming. Note that
strong off-resonance occurs at air-tissue interface. Thus brain around
bi-hemispheric ear canals (inferior temporal lobes) and nasal cavity
(orbitofrontal lobe) usually shows serious EPI distortion and voids. Such
artifacts may be partially alleviated by tuning high-order shim, adjusting
imaging parameters, and using localized shimming hardware.
Motion artifacts needs to be
carefully controlled in fMRI measurements in order to ensure that the signal
dynamics faithfully represent neuronal/vascular responses over time. Before the
measurement, it is helpful to ensure that the subject’s head is properly
positioned and stabilized using pillows/padding materials (or even a bite-bar).
During the measurement, prospective motion correction can be used to reduce the
signal fluctuations due to bulk motion. After the measurement, analyzing head motion over time can give good
hints about the quality of the fMRI time series.
Finally, scanning experiment
participants with care and necessary interactions (via microphone, for example)
can always be valuable to ensure the quality of fMRI data. It is also helpful
to monitor and check fMRI images on the scanner console during acquisition to
ensure both the participants are following your instructions and the MRI
scanner is stable. Your immediate attention may bring both safety to your
participants and high quality images to you.
Acknowledgements
I thank the support from the Ministry of Science and Technology,
Taiwan (MOST 104-2314-B-002-238, MOST 103-2628-B-002-002-MY3).References
1 Belliveau J., Kennedy D.,
McKinstry R. et al.Science.1991; 254:716-719.
2 Kwong K. K., Belliveau J.
W., Chesler D. A. et al.Proc Natl
Acad Sci U S A.1992; 89:5675-5679.
3 Ogawa S., Lee T. M., Kay
A. R. et al.Proc Natl Acad Sci U S
A.1990; 87:9868-9872.
4 Logothetis N. K., Pauls
J., Augath M. et al.Nature.2001; 412:150-157.
5 Glover G.
H.Neuroimage.2012; 62:706-712.
6 Glover G. H. & Lee A.
T.Magn Reson Med.1995; 33:624-635.
7 van Vaals J. J., Brummer
M. E., Dixon W. T. et al.J Magn Reson
Imaging.1993; 3:671-675.
8 Jones R. A., Haraldseth
O., Muller T. B. et al.Magn Reson
Med.1993; 29:830-834.
9 Griswold M. A., Jakob P.
M., Heidemann R. M. et al.Magn Reson
Med.2002; 47:1202-1210.
10 Pruessmann K. P., Weiger M.,
Scheidegger M. B. et al.Magn Reson
Med.1999; 42:952-962.
11 Sodickson D. K. &
Manning W. J.Magn Reson Med.1997; 38:591-603.
12 Feinberg D. A., Moeller S.,
Smith S. M. et al.PLoS ONE.2010; 5:e15710.
13 Larkman D. J., Hajnal J. V.,
Herlihy A. H. et al.J Magn Reson
Imaging.2001; 13:313-317.
14 Setsompop K., Gagoski B. A.,
Polimeni J. R. et al.Magn Reson
Med.2011;
15 Hennig J., Zhong K. &
Speck O.Neuroimage.2007; 34:212-219.
16 Lin F.-H., Wald L. L.,
Ahlfors S. P. et al.Magn Reson
Med.2006; 56:787-802.
17 Lin F. H., Witzel T.,
Mandeville J. B. et al.Neuroimage.2008; 42:230-247.
18 Lin F. H., Ahveninen J.,
Raij T. et al.PLoS One.2014; 9:e100319.
19 Lin F. H., Witzel T., Raij
T. et al.Neuroimage.2013; 78:372-384.
20 Polimeni J. R., Fischl B.,
Greve D. N. et al.Neuroimage.2010; 52:1334-1346.
21 Cheng K., Waggoner R. A.
& Tanaka K.Neuron.2001; 32:359-374.
22 Yacoub E., Harel N. &
Ugurbil K.Proc Natl Acad Sci U S A.2008;
105:10607-10612.
23 Duong T. Q., Yacoub E.,
Adriany G. et al.Magn Reson Med.2003; 49:1019-1027.
24 Yacoub E., Shmuel A., Logothetis
N. et al.Neuroimage.2007; 37:1161-1177.
25 Boxerman J. L., Hamberg L.
M., Rosen B. R. et al.Magn Reson
Med.1995; 34:555-566.
26 Ogawa S., Menon R. S., Tank
D. W. et al.Biophys J.1993; 64:803-812.
27 Gonzalez-Castillo J.,
Roopchansingh V., Bandettini P. A. et al.Neuroimage.2011; 54:2764-2778.
28 Hajnal J. V., Myers R.,
Oatridge A. et al.Magn Reson
Med.1994; 31:283-291.
29 Cox R. W. & Jesmanowicz
A.Magn Reson Med.1999; 42:1014-1018.
30 Woods R. P., Grafton S. T.,
Holmes C. J. et al.J Comput Assist Tomogr.1998; 22:139-152.
31 Kruger G. & Glover G.
H.Magn Reson Med.2001; 46:631-637.
32 Bodurka J., Ye F., Petridou
N. et al.Neuroimage.2007; 34:542-549.
33 Sarkka S., Solin A.,
Nummenmaa A. et al.Neuroimage.2012; 60:1517-1527.
34 Glover G. H., Li T. Q. &
Ress D.Magn Reson Med.2000; 44:162-167.
35 Zhang N., Zhu X. H. &
Chen W.Magn Reson Med.2005; 54:258-263.