Synopsis
Recent progress has demonstrated the feasibility of using
the SQUID sensor arrays in MEG helmets to record MRI data. Here we describe the
basic principles of MRI as well as the special requirements and solutions
needed to perform ultra-low-field MRI concurrently with MEG. We consider it is
feasible to build practical MEG-MRI instruments for scientific experimentation
and for clinical use. An MRI with 2 mm spatial resolution, sufficient signal-to-noise
ratio and contrast appears achievable without essentially lengthening the normal
MEG measurement time.Disclaimer
The content of this syllabus was adopted from “Ultra-Low-Field MRI and
its combination with MEG“ by Lauri Parkkonen, Risto J. Ilmoniemi, Fa-Hsuan Lin,
and Michelle Espy from “Magnetoencephalography: from signals to dynamic
cortical networks“, Ed: Selma Supek, Cheryl J. Aine, 2014, Springer press
Introduction
The availability of large arrays of highly sensitive SQUID magnetometers in modern MEG devices enables one to measure magnetic fields other than those produced by neuronal electrical activity. Perhaps
the
most promising such
possibility is to measure magnetic resonance imaging (MRI) signals. This would bring
several benefits such as improved registration of MEG and MRI, improved work flow, structural images with less distortion, and information
about the conductivity
of the brain.
In principle, combining MEG and MRI is straightforward: simply build a magnet as well as gradient and radio-frequency (RF) coils around an MEG sensor array. The problem is that MEG devices are generally
designed to measure femtotesla-level fields and have a dynamic range only up to some tens or hundreds of nanoteslas while in MRI the fields go up to several tesla, i.e., 15 orders of magnitude above the
weakest fields measured by SQUIDs. The solution is ultra-low-field MRI: the recordings are performed in a field of about 100 microteslas. If conventional MRI approaches, including tuned inductive receiver coils, would be used at these low fields, the resulting signal-to-noise
ratio (SNR) would be very low and
the
system practically unusable, because the SNR is proportional to the square of the field; the signal at
100 microteslas would be 9 orders of magnitude weaker than at 3 tesla. Fortunately, we can use three
methodologies to counteract this problem. First, unlike tuned receivers, the sensitivity of SQUID sensors
is
independent of frequency; second, we can
use
pre-polarization techniques to magnetize the sample
before the MRI data acquisition; third, we can gain from parallel data acquisition made available by the large number (up to 306 currently) of SQUID sensors in a typical MEG array.
Basic principles of ultra-low field MRI
NMR
There is a fundamental difference between ULF and high-field (HF: B > 1 T for this discussion) NMR and MRI. In
HF,
the process of polarization, magnetization reorientation,
spin evolution,
and measurement all occur within a single magnetic field provided (typically) by a large superconducting magnet. At ULF, however, these fields may all be different and produced by different magnets. Because the magnetization and thus signal is proportional to the field, one
strategy is to use pre-polarization in a higher field (Bp ~ 1 -
200 mT) followed by readout in a much lower measurement field (Bm ~ 1 - 200 μT).
There are also differences in approaches to
reorient spin between conventional HF
MRI
and ULF MRI. In HF MRI, spin reorientation is typically accomplished
by a
time-varying magnetic field applied at the Larmor frequency, in a direction orthogonal to the direction of main magnetic field. In many ULF-MRI configurations, Bp and Bm are orthogonal. Thus, one can simply begin precession by a rapid (non-adiabatic, dBp/dt >> γBm2) shut-off of Bp. In a non-adiabatic process, the magnetization cannot follow the
field change, and is left aligned orthogonal to Bm. Precession will then begin automatically, without a B1 pulse.
Once precession has begun, detection of the magnetization can begin. In HF MRI, the typical scanner strengths are 1.5 or 3 T. This translates to a proton Larmor frequency of 64 or 128 MHz. A tuned
induction coil is highly sensitive in this range. However, in ULF MRI precession occurs in the Bm field, typically on the order of 100
μT, corresponding to a Larmor frequency of 4.26 kHz. In HF MRI with a Faraday coil, the signal scales as B2. One order arising from the magnetization being proportional to B and
the other one for the induced signal being proportional to the Larmor frequency. However, this relationship no longer holds in ULF MRI. While
pre-polarization is an approach to improve the former
factor (Macovski and Conolly 1993), using an ultra-sensitive detector such as the SQUID is a way to
mitigate the latter. SQUIDs, broadband detectors with unsurpassed sensitivity of about 1 fT/
Hz
in the frequency range of ULF MRI, are almost two orders more sensitive than a Faraday coil in this regime
(Myers et al. 2006; Matlashov 2011). Because the SQUID is also the detector of choice for MEG, the combination of MEG and ULF MRI in a single device becomes obvious. We should, however, mention
that getting
a SQUID to work in the dynamic environment of MRI, even at ultra-low fields, is quite
challenging.
Basic principles of image acquisition
Here we will discuss that in addition to lower polarization leading to lower signal, the lower strength of imaging gradients also poses a unique new challenge for
ULF
MRI in terms of image acquisition time.
To improve the spatial resolution, one either has to increase the gradient or the acquisition (or encoding) time. The gradients in ULF MRI are typically on the order of 104 T/m. This is about 1/100 of those used in HF MRI. Thus, if we want to keep the same resolution
as
in HF MRI, with these weaker gradients we must
increase the acquisition time by a factor of 100.
Why don't we just increase the gradient strength? There are two reasons: the first being related to concomitant
fields. These are the unwanted magnetic fields that inevitably arise in directions orthogonal to the measurement field Bm. In HF MRI, these concomitant
fiels are neglected, because the main magnetic field (B0 > 1 T) is typically much
higher than the gradients (10-2 T/m) and thus frequency variations
produced by the stray fields are small. However, in ULF MRI the main magnetic
field Bm ~ 10-4 T and the
gradients G
~ 10-4 T/m, which generates a magnetic field variation of 0.2x10-4 T in a 20-cm FOV. The
concomitant fields, not in parallel with Bm, are of similar order of magnitude as G. Thus, the frequency
variations are non-negligible and the
total magnetic field experienced by the spin system is no longer in a
plane orthogonal to Bm. In general, concomitant gradients can be accounted for with some effort (e.g.,
Volegov et al. 2005; Nieminen et al. 2010; Hsu et al. 2014) but they do pose a constraint.
The other reason that gradients cannot be arbitrarily strong is related to bandwidth. Consider a simple
case: a HF MRI system with B0 = 1 T and 10-2 T/m gradients. A 20-cm object would have a central
frequency of 42.6 MHz, and a frequency spread across the object of Δω ~ 85.2 kHz. However, if Bm = 10-4 T and G = 10-4 T/m, the central frequency is 4.26 kHz
and
the frequency spread within a FOV of 20 cm
is 852 Hz. If we turn up the gradients, we will further widen the
frequency spread across the object and consequently need to measure part of the MRI signal in the challenging low-frequency regime.
Thus, our only choice to maintain spatial resolution at ULF appears to be longer acquisition times. But at
ULF, the T1 and T2 times for many interesting tissues are approximately of the length required for ta (Zotev et al. 2009), so we are also running out of signal at the same time.
A simple way to speed up imaging is provided by parallel imaging methods (Pruessmann et al. 1999). In this approach, the spatial sensitivity of an array of SQUIDs is used to replace spatial encoding steps.
Using less encoding steps speeds up the acquisition but usually results in an
aliased image. By using, for
example,
the sensitivity maps
from an array of
coils, a full image
may
still be
reconstructed.
In applications like combined MEG and MRI, where a SQUID array is available, this has been demonstrated as a viable
approach (Zotev et al. 2008a). Some of
the
complications due to inductive
coupling between
tuned
receiver coils in HF MRI parallel imaging are greatly reduced with untuned
SQUIDs. However, it should be emphasized that parallel imaging accelerates the
MRI
acquisition at the cost of SNR, which is
already scarce in ULF MRI.
ULF-MRI Instrumentation
In the discussion below we will generally assume that the
application is for the
combination of ULF MRI and MEG. Thus, we are assuming that ULF MRI is being done in the
presence of a magnetically shielded room (MSR). We will also assume that we have an array of SQUIDs. We note that for applications of
ULF
MRI that do not include
MEG, an MSR may not be
required. For example, the
Clarke group at UC
Berkeley operates in an aluminum eddy current shield only. We also note that there has been progress
using sensors other than SQUIDs (namely the atomic magnetometer) both for MEG (Xia et al. 2006) and
ULF MRI (Savukov et al. 2009), which we
will not discuss here. In ULF MRI, one aims at the highest pre-polarization that can be tolerated (while maintaining the
benefits of the
ULF
regime) and at the most sensitive
sensor. The
heart of any ULF MRI instrument is the SQUID sensor array. Large (hundreds of sensors) SQUID arrays have been used for decades for MEG
(Hämäläinen et al. 1993). The noise level of a SQUID can be as low as 10-15 T, enabling it to detect the very weak (10-12 to 10-15 T) magnetic
fields from brain activity from outside the head. A challenge for ULF MRI is that the changing magnetic fields are many orders of magnitude larger than
the dynamic range of SQUIDs. Several strategies have been implemented to deal with this. One approach is to encapsulate
the
SQUID chips in sealed Pb boxes that have a critical field of about 80 mT. The SQUID can also be locally
shielded by Nb plates on the chip (Luomahaara et al. 2011). In either case, the pick-up coil extends
outside the shield to detect fields of interest. Thus, the input coil circuit also needs current limiters. One
can
use
externally controlled superconducting cryo-switches (Zotev et al. 2007), which become resistive when heated above the critical temperature and thereby make
the pick-up coil much less sensitive to magnetic fields. Other groups have used arrays of Josephson junctions (Hilbert et al. 1985) that become resistive above the Josephson critical current and thereby limit the current.
To achieve
a high
pre-polarization field, a variety of magnet designs have been proposed. In
some sense,
generating Bp is
much simpler than at HF because the fields are so much lower and the field homogeneity requirement is less stringent (McDermott et al. 2002; Burghoff et al. 2005). Bp can
be relatively (a few
percent) inhomogeneous compared to the parts-per-million or better requirement for B0 in
HF MRI.
However, there are considerations to producing (and removing) Bp. In fact, some of the advantages
(reduced homogeneity requirement, shorter T1 times) can be disadvantages if not accounted for. The first
consideration is that it is not trivial to make a pulsed field at > 50 mT. The coil will heat up, dissipated energy must be removed, and the proximity of a large amount
of conductor near the SQUIDs can introduce Johnson noise. The Bp coil should be physically disconnected during the measurement (via a
relay) to reduce the antenna effect.
There have been several approaches to producing a pulsed Bp, including water-cooled (Myers 2006), coolant (Fluorinert) and liquid nitrogen (LN) cooled coils (Sims et al. 2010). The
LN coil has the benefit of 7× lower resistance, but requires an additional cryostat. Recently, a self-shielded (Nieminen et al. 2011) pulsed superconducting coil (Vesanen et al. 2012a) has been demonstrated for ULF MRI; the Bp coil was integrated into the cryostat with the SQUIDs. The choice of materials for Bp can
be important.
For example, we have found that multi-stranded Litz wire performs much better than solid wire in terms of noise. According to Vesanen et al. (2012a), the superconducting wire magnetized if too high of a current (> 12 A) was applied, producing spurious gradients that influenced the image quality, limiting Bp in that work to < 24 mT.
Perhaps the most challenging aspect of a pulsed Bp is
how to remove it appropriately. When the Bp field changes transient eddy currents will be induced in nearby conductors, which can impose a long dead-time if the magnetic
fields from the transients exceed the
dynamic range of the SQUIDs.
There
are two approaches to switching off Bp: adiabatic and non-adiabatic. In a non-adiabatic
ramp-down, dBp/dt >> γBm2
such that the magnetization is left aligned with the original direction of Bp . If Bm is orthogonal to Bp, precession will begin automatically. In principle, this approach can minimize the time
between beginning precession and measurement. In reality, however, the
faster the
ramp-down of Bp, the
larger are the transients that are induced in nearby conductors. When measurements are made inside a
conductive MSR, these transients can become a serious confound as they may have components that
persist for hundreds of milliseconds (Vesanen et al. 2012b) and are
inherently low frequency and hard to be
removed from the
MEG.
Even in the absence of an MSR, anything conducting nearby will also support transients that may impact the image and/or impose long wait times. One added consideration in the non-adiabatic field removal is the non-uniformity of Bp. Not requiring a uniform Bp greatly simplifies the
magnet design, but signal is lost because of this non-uniformity; when
precession starts, the spins are not
all in phase. In addition, there are technical problems associated with the requirement to dissipate the energy stored in the Bp coil.
If,
instead, an adiabatic ramp (dBp/dt << γBm2)
is used, the final magnetization will be aligned with low Bm field, which is easy to make uniform. Further, phase coherence is typically improved due
to lack of
transients. A spin-flip pulse is then required
to start precession. In an
adiabatic ramp, dB/dt is lower and
thus there
is
less danger of heating metallic implants or affecting therapeutic electronic devices the
subject may be
carrying This is a special consideration when imaging near metal. In an ULF MRI system, Bp is typically 10
– 200
mT
and it is removed within 10 – 100 ms (depending on the approach). Thus, the
field change dB/dt may
range from 0.1 – 20 T/s. Even with relatively high
pre-polarization, the
dB/dt in a ULF MRI system is typically lower than
that in HF
MRI
systems. However, the adiabatic ramp-down takes longer, and signal is lost due to T1 relaxation
during that time.
The generation of the other magnetic fields required for imaging (Bm and gradients) is relatively simple
given their low field strengths. Typically, simple wire-wound coils can be utilized. An interesting and important aspect regarding the measurement and gradient coils is how we
power them. Because
the
noise level should be as low as possible (~ 1 fT/Hz is a typical goal) at the frequencies of interest, the current noise
of
the power supplies must also be as low as possible. While
this sort of noise is
not
an issue
at the
frequencies of HF MRI, at ULF it can be a problem. Typically this has been dealt with by the use of batteries and heavily filtered circuits. Unfortunately, these solutions limit the sorts of pulse
sequences that can be used. Thus, it is likely that one critical advance
to ULF MRI instrumentation will be developing low-noise electronics that enable flexible pulse sequences. For example, the Bm field could also be oriented in any direction on the fly (enabling a totally new kind of projection imaging that would provide maximum sensitivity in a helmet-like configuration of sensors) but this approach also requires three Bm field coil sets to be
on
simultaneously.
Potential applications of ULF MRI
Medical imaging
MRI at ultra low fields shows improved T1 contrast compared to high fields,
which may translate in to a unique capability of ULF MRI to help delineate certain tissue types better than
high-field MRI. It has already been shown that, for example, biopsies of prostate
cancer tissue exhibit a significantly shorter T1 time at Bm = 132 µT than healthy tissue (Clarke et al. 2007) and cancerous vs.
normal rat liver shows similarly high contrast at Bm = 100 µT (Liao et al. 2010). The complex behavior of the proton relaxation dispersion of water when approaching zero field (Hartwig et al. 2011) may be the physical background of the enhancement of T1 contrast between healthy tissue and tumors at low fields.
The grey
white matter border has a relatively high T1 contrast. This border is blurred at focal cortical
dysplasias (FCDs), malformations generated during cortical development. FCDs are
highly epileptogenic and frequently cause intractable epilepsy. Unfortunately, a considerable
fraction of FCDs can not be discerned in high-field MRI and are only detected in tissue microscopy after removal of the cortical region. Thanks to its higher T1 contrast, ULF MRI may be able to visualize FCDs better than high-field
MRI.
High-field MRI cannot be applied to patients with
pacemakers, stimulators or metal in the body. However, there are no such restrictions with ULF MRI, which is inherently safe.
Temperature mapping
Vesanen et al. (2012) utilized the dependence of T1 relaxation time of agarose gel to demonstrate the ability of ULF MRI to measure temperature. Although this method can prove useful in special cases, one must bear in mind that in human tissue, T1 is influenced much more by the detailed structure
of
the tissue
than by temperature. Therefore, this method can be used only for indicative purposes and to monitor
possible changes of temperature when other factors affecting T1 can be assumed fixed.
Conductivity imaging
If electric current is applied to a conducting object using two electrodes, the current will distribute itself between different paths in proportion to the conductivities of the
paths. It was shown by Nieminen et al. (2012) and Vesanen et al. (2013) that the direction and
amplitude of the current distribution can be
measured with low-field MRI, provided, of course, that the signal-to-noise ratio is sufficient. This is not
possible with a high-field MRI device without the
need to turn the head between two measurements. So far, only simulation studies have been done. These indicate excellent reconstruction accuracy but also the
fact that the
signal-to-noise
ratio of MRI measurements must be
significantly improved before electrical
impedance
tomography can be done in humans with safe
current strengths (on the order of 2 mA as those
in
transcranial direct current stimulation or tDCS).
Future directions
Improvements in instrumentation
The present state of the art is not yet sufficient for scientific and clincal applications of combined MEG and ULF MRI. However, we can predict that ULF MRI can be improved to a level that will provide acceptable image quality and allow accurate registration of MEG and MRI coordinate
systems. A great
improvement in signal-to-noise
ratio can be obtained by increasing the
pre-polarization field strength and by reducing SQUID and dewar noise further; Bp > 100 mT and sensor noise level of 0.5 fT/Hz1/2 seem possible even in a large array. The improvement in signal-to-noise ratio may enable us to measure the
conductivity
structure of the head
as well. However, problems arising from the highly
sensitive SQUIDs in strong pulsed magnet fields will become more severe. Thus, the task of building a
practical MEG-MRI system is far from trivial. Elaborate methods will be needed to handle problems
caused by eddy currents in the system and in nearby structures as well as the magnetization of materials. Optimized sequences and signal processing will be needed to maximally utilize the recorded data. Next, we
will give a glimpse of one approach in developing signal processing.
Advances in signal processing
Conceivably, ULF MRI with
up to hundreds of sensors for whole-brain imaging will be developed in
the
next few years. In high-field MRI, such highly parallel signal detection has been used to improve the spatiotemporal resolution at the cost of SNR (Pruessmann et al. 1999; Sodickson and Manning 1997).
The feasibility of 3-fold
acceleration has been reported
in
an ULF MRI study with
a seven-channel SQUID
system (Zotev et al. 2008a). Although currently SNR of ULF MRI is too low to be compromized, we expect that with future SNR improvement offered by higher pre-polarization field and more sensitive
SQUID detectors, it might be possible
to trade-off SNR for a shorter acquisition time. The SNR loss in
parallel MRI (pMRI) is the consequence of the loss of data samples and
the
noise amplification in image reconstruction (Pruessmann et al. 1999). While the former loss is inevitable
in acceleration, the latter loss
can
be
compensated for by regularized reconstruction methods (Lin et al. 2004; 2005) and by increasing
the
number of parallel detectors.
The SNR penalty of accelerated ULF MRI was recently investigated by simulating a helmet-shaped sensor array
with up to 306 SQUID sensors (204 gradiometers and 102 magnetometers; VectorViewTM, Elekta Oy, Helsinki, Finland) (Lin et al. 2013). It was found that,
in all SQUID sensor geometries, image locations away from pick-up coils show a larger g-factor in general. These numbers suggest that, in
a 3D
ULF-MRI acquisition with
two phase-encoding directions, the
highest acceptable
acceleration rate may be from 9- to
16-fold (based
on
an arbitrary threshold of average g = 1.4). Consider a 3D ULF-MRI acquisition with 64 x 64 x 64 voxels. It needs 4,096 independent read-outs with 64 x 64
phase encoding
steps. With repetition
time TR = 1 s, this amounts to
more than an hour of acquisition time. Using an array of 102 sensors
and 9-fold acceleration and assuming that the SNR loss due to
reduced
samples is tolerable, the data acquisition can be completed in approximately 7 minutes. However, as suggested by these results, spatially varying noise amplifications
(i.e., g-factor) could be significant at 9- and
16-fold acceleration, resulting in inhomogeneous image quality deterioration.
In addition to or instead of aiming at achieving a higher resolution or shorter measurement time, parallel
MRI
can be applied to ULF MRI to improve
the SNR by exploiting the
redundancy among the receiver channels by enforcing k-space data consistency among them and by
adding a priori image sparsity information to further suppress noise (Lin et al. 2013).
Conclusions
By performing MRI measurements with the large arrays of SQUID sensors available in MEG helmets, one can realize
combined MEG and MRI, which offers unprecedented possibilities to obtain new kinds of information about the human brain. MEG-MRI systems will be quiet, open, and safe. They will enable highly accurate registration of MEG and MRI coordinate
systems and, if imaging of injected current
density proves practical, the determination of the three-dimensional conductivity distribution. This, in
turn, would enable us to solve
the
inverse problem of MEG (and EEG) using reliable knowledge of both measurement and conductivity geometry. However, we are still far from constructing practical devices. It will be necessary to improve the signal-to-noise ratio considerably to attain MR image quality that is sufficient
for clinical and scientific
applications.
Acknowledgements
Lauri Parkkonen, Risto Imoniemi and Fa-Hsuan Lin acknowledge the financial support from the Academy of Finland, and the European 7th Framework Programme (FP7/2007 - 2013); Grant number: 200859.
Fa-Hsuan Lin thanks the support from the Ministry of Science and Technology,
Taiwan (MOST 104-2314-B-002-238, MOST 103-2628-B-002-002-MY3). Michelle Espy would like to
acknowledge
the financial support of LDRD, DHS, as well as thank her colleagues and especially Andrei Matlashov for his assistance in writing.References
Burghoff M, Hartwig S, Trahms L (2005) Nuclear magnetic resonance in the nanotesla range. Applied
Physics
Letters 87:054103.
Clarke J, Hatridge M, Mössle M (2007) SQUID-detected magnetic resonance imaging in microtesla fields. Annual Review of Biomedical Engineering 9:389–413.
Hämäläinen M, Hari R, Ilmoniemi RJ, Knuutila J, Lounasmaa OV (1993) Magnetoencephalography
–theory, instrumentation, and applications to noninvasive studies of the working human brain.
Reviews of Modern Physics 65:413–497.
Hartwig S, Voigt J, Scheer H-J, Albrecht H-H,
Burghoff M, Trahms L (2011)
Nuclear magnetic relaxation in water revisited. The Journal of Chemical Physics 135:054201.
Hilbert C, Clarke J , Sleator T, Hahn EL (1985) Nuclear quadrupole
resonance detected at 30 MHz with a dc
superconducting quantum interference
device. Applied Physics Letters 47:637–639.
Hsu
YC, Vesanen PT, Nieminen JO, Zevenhoven KCJ, Dabek J, Parkkonen L, Chern IL, Ilmoniemi
RJ, Lin FH (2014) Efficient concomitant and remanence field artifact reduction in
ultra-low-field MRI using a frequency-space formulation. Magnetic Resonance in Medicine,
71:955–965.
Liao
SH,
Huang
KW,
Yang
HC,
Yen
CT,
Chen
MJ, Chen HH,
Horng HE,
Yang
SY (2010)
Characterization of tumors using high-Tc superconducting quantum interference device-detected
nuclear magnetic resonance and imaging. Applied Physics Letters 97:263701–263703.
Lin FH, Huang TY, Chen NK, Wang FN, Stufflebeam SM, Belliveau JW, Wald LL, Kwong KK (2005) Functional MRI using regularized parallel imaging acquisition. Magnetic Resonance in Medicine
54:343–353.
Lin FH, Kwong KK, Belliveau JW, Wald LL (2004) Parallel ima ging reconstruction using automatic regularization. Magnetic Resonance
in
Medicine 51:559–567.
Lin FH, Vesanen PT, Nieminen JO, Hsu YC, Zevenhoven
KC, Dabek
J, Parkkonen LT, Zhdanov A, Ilmoniemi RJ (2013) Noise amplification in parallel whole-head ultra-low-field magnetic resonance imaging
using
306
detectors.
Magnetic Resonance in Medicine, 70: 595–600
Lin FH, Vesanen PT, Hsu YC, Nieminen JO, Zevenhoven KCJ, Dabek J, Parkkonen LT, Simola J,
Ahonen AI, Ilmoniemi RJ (2013) Suppressing multi-channel ultra-low-field MRI measurement
noise using data consistency and image sparsity. PLoS ONE 8:e61652.
Luomahaara J, Vesanen PT, Penttilä J, Nieminen JO, Dabek J, Simola J, Kiviranta M, Grönberg L,
Zevenhoven CJ,
Ilmoniemi
RJ, Hassel J (2011)
All-planar
SQUIDs and pickup coils for combined MEG and MRI.
Superconductor Science and Technology 24:075020.
Macovski AS, Conolly S (1993) Novel approaches to low cost MRI. Magnetic Resonance in Medicine
30:221–230.
Matlashov AN, Schultz LJ, Espy MA, Kraus RH, Savukov
IM,
Volegov PL, Wurden
CJ
(2011) SQUIDs
vs.
induction coils for ultra-low field nuclear magnetic resonance: experimental and simulation
comparison. IEEE Transactions on Applied Superconductivity 21:465–468.
McDermott R, Trabesinger AH, Mück M, Hahn EL, Pines A, Clarke J (2002) Liquid-state NMR and scalar couplings in microtesla magnetic
fields. Science 295:2247–2249.
Myers W, Slichter D, Hatridge M, Busch S, Mößle M, McDermott R, Trabesinger A, Clark J (2006a)
Calculated signal-to-noise ratio of MRI detected with SQUIDs and Faraday detectors in fields
from 10 µT to 1.5 T.
Journal of Magnetic Resonance 186:182–192.
Myers WR (2006b) Potential Applications of Microtesla Magnetic Resonance
Imaging Detected Using a Superconducting Quantum Interference Device. PhD thesis, UC Berkeley. http://escholarship.org/uc/item/19x7q88x.
Nieminen J (2012) Ultra-low-field MRI: techniques and instrumentation for hybrid MEG-MRI. PhD
thesis, Aalto University publication
series.
lib.tkk.fi/Diss/2012/isbn9789526046457/isbn9789526046457.pdf.
Nieminen JO, Ilmoniemi RJ (2010) Solving the problem of concomitant gradients in ultra-low-field MRI. Journal of Magnetic Resonance 207:213–219.
Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P (1999) SENSE: Sensitivity encoding for fast
MRI.
Magnetic Resonance in Medicine 42:952–962.
Savukov IM, Zotev VS, Volegov PL, Espy MA, Matlashov AN, Gomez JJ, Kraus RH Jr. (2009) MRI
with an atomic magnetometer suitable
for practical imaging applications. Journal of Magnetic Resonance 199:188–191.
Sims JR, Schillig JB, Swenson CA, Gardner DL, Matlashov AN, Ammerman CN (2010) Low-noise pulsed pre-polarization magnet systems for ultra-low field NMR. IEEE Transactions on Applied
Superconductivity 20:752–755.
Sodickson DK, Manning WJ (1997) Simultaneous
acquisition
of spatial harmonics
(SMASH):
Fast imaging with radiofrequency coil arrays.
Magnetic Resonance
in
Medicine 38:591–603.
Vesanen PT, Nieminen JO, Zevenhoven KCJ, Dabek J, Parkkonen LT, Zhdanov AV, Luomahaara J, Hassel J, Penttilä J, Simola J, Ahonen AI, Mäkelä JP, Ilmoniemi RJ (2012a) Hybrid ultra-low- field MRI and magnetoencephalography system based on a commercial whole-head neuromagnetometer. Magnetic Resonance in Medicine 69:1795–1804.
Vesanen PT, Nieminen JO, Zevenhoven KCJ, Dabek J, Simola J, Sarvas J, Ilmoniemi RJ (2012b) The
spatial and temporal distortion of magnetic fields applied inside a magnetically shielded room. IEEE Transactions on Magnetics 48:53–61.
Vesanen PT (2013) Combined ultra-low-field MRI and MEG: Instrumentation and applications, PhD thesis, Aalto University publication series. http://lib.tkk.fi/Diss/2013/isbn9789526051789/isbn9789526051789.pdf.
Volegov PL, Mosher JC, Espy MA, Kraus RH Jr. (2005) On concomitant gradients in ultra -low field MRI. Journal of Magnetic Resonance 175:103–113.
Xia
H, Ben-Amar Baranga BA, Hoffman D, Romalis MV (2006) Magnetoencephalography with an atomic magnetometer. Applied Physics Letters 89:211104.
Zotev VS, Matlashov AN, Volegov PL, Urbaitis AV, Espy MA, Kraus RH Jr. (2007)
SQUID-based instrumentation for ultra-low-field MRI. Superconductor Science and Technology 20:S367–S373.
Zotev VS, Volegov PL, Matlashov AN, Espy MA, Mosher JC, Kraus RH Jr. (2008a) Parallel MRI at
microtesla fields. Journal of Magnetic Resonance 192:197–208.
Zotev VS, Matlashov AN, Savukov IM, Owens T, Volegov PL, Gomez JJ, Espy MA (2009) SQUID- based microtesla MRI for in vivo relaxometry of the human brain. IEEE Transactions on Applied
Superconductivity 19:823–826.