Synopsis
The fundamental signal generation in magnetic resonance imaging (MRI)
sequences is based on the principle of either spin echoes or gradient echoes
or a combination of
the two. This course elucidates concepts and basic properties of gradient echo
methods with a special focus on fast gradient echo sequences.Introduction
The fundamental signal generation in magnetic resonance imaging (MRI)
sequences is based on the principle of either spin echoes (1,2) or gradient echoes
(3–7) or a combination of
the two. This course elucidates concepts and basic properties of gradient echo
methods with a special focus on fast gradient echo sequences.
Gradient echoes ‑ also called gradient-recalled echoes (GRE),
gradient-refocused echoes or field echoes – are the basis of many applications on
modern MRI systems (5–7). Figure 1 shows the basic principle of such a GRE, where the free
induction decay (FID) after radio-frequency (RF) excitation is measured.
The most striking differences in Fig. 1 compared to a conventional
spin echo sequence are, generally, the lacking refocusing pulse, the typically
low excitation flip angle of α<90°, and the read gradient reversal.
A GRE sequence as outlined in Fig. 1
represents a quite “simple” form, since the sketch shall imply a repetition
time (TR) that is much longer than $$$T_1$$$ and $$$T_2$$$. Therefore,
the transverse magnetization ($$$M_{xy}$$$) decays completely and the longitudinal
magnetization ($$$M_z$$$) fully recovers until the next excitation pulse. Then,
$$$M_{xy}$$$ and, hence, the signal behavior and image contrast is dominated
by the frequently introduced transverse decay time $$$T_2^*$$$:
$$\begin{align} \frac{1}{T_2^*}=\frac{1}{T_2}+\frac{1}{T_2^{'}}. && \tag{1}\end{align}$$
As defined in Eq. [1], $$$T_2^*$$$ is a combination of the
tissue characteristic transverse relaxation time ($$$T_2$$$) and a relaxation
term, $$$T_2^{'}$$$, which reflects signal decay due to static field
inhomogeneity and susceptibility effects. Note that it is always $$$T_2^*$$$<$$$T_2$$$.
The exhibited $$$T_2^*$$$ contrast of such a simple (and slow) GRE
sequence is dependent on the echo time TE as follows:
$$\begin{align} SI(TE) = SI_0 \cdot exp(-TE/T_2^*). && \tag{2}\end{align}$$
It should be noted here that the introduction of $$$T_2^{'}$$$ and,
thus, $$$T_2^*$$$ in Eq. [1] is an approximation which performs often
well in properly describing the observed signal behavior depicted by
Eq. [2].
It has always been a necessity in MRI to actually shorten the
acquisition time (TA) that is directly proportional to TR, which leads under
certain circumstances to a dynamic
magnetization equilibrium or steady
state signal due to the incomplete $$$T_1$$$ and $$$T_2$$$
relaxation until the next RF excitation pulse.
Steady State Free Precession Sequences
As originally introduced in 1958 by Carr (8) for NMR spectroscopy, a
dynamic equilibrium or steady-state in the magnetization can be established by
a train of RF excitation pulses interleaved by periods of ‘free precession’, commonly
referred to as steady state free
precession (SSFP) (see Fig. 2).
The quick succession of RF pulses prevents the magnetization from
returning to thermal equilibrium and each RF pulse therefore acts on both remaining
transversal and longitudinal magnetization generating rather complex patterns of spatial magnetization
distribution even after a few RF pulses. Nevertheless, the magnetization between
consecutive excitation pulses can reach a dynamic equilibrium, i.e. a steady
state, if the following conditions are fulfilled, see (9,10):
$$$\qquad$$$ · The dephasing from gradients (G) within TR, TR itself and
the flip angle (α) must be constant.
$$$\qquad$$$ · The phases (Ф) of the RF pulse must satisfy
the equation: Ф$$$_n = a + b×n + c×n^2.$$$
Transition to
steady state from thermal equilibrium (or any other magnetization prepared
state) is completed after 5x$$$T_1$$$, however, this is frequently not an
acceptable waiting time for a fast imaging method. As a result, several
preparation methods have been proposed to facilitate, enhance or smoothen this
transition. In the following, we will assume that a steady state could have
been established after sufficient RF pulses.
The class of fast GRE sequences that are based on the dynamic equilibrium effect of SSFP are commonly called SSFP
sequences. Generally, the measured signal from such SSFP sequences will depend on
relaxation ($$$T_{1,2}$$$) and on diffusion or flow effects, but also becomes
a function of the repetition time (TR), the echo time (TE), the flip angle (α) and the RF pulse
phase increment ($$$\phi_n - \phi_{n-1}$$$), and of the
gradient switching pattern.
SSFP Classification
Overall, GRE sequences can be broadly classified
as incoherent or coherent depending on whether or not any remnant transverse
magnetization is spoiled prior to the next RF pulse (see Table 1). Since its
introduction more than half a century ago, the use of the SSFP signal has
become increasingly popular for imaging and a large
number of SSFP imaging methods have been described so far under a
sometimes confusing array of acronyms, e.g., see Refs. (5–7,11). For clearness, we
will stick to the following generic SSFP nomenclature:
$$$\qquad$$$ ·
The term ‘SSFP’
embeds all steady state sequences and variants thereof since it just indicates
the most basic SSFP principle as introduced by Carr.
$$$\qquad$$$ ·
The term ‘balanced’
is used to indicate that all gradient moments are fully rephased prior to the
next excitation pulse.
$$$\qquad$$$ ·
The term ‘nonbalanced’
is used to indicate the presence of some crusher gradients, i.e. dephasing
gradient moments, prior to the next excitation pulse.
$$$\qquad$$$ ·
The term ‘spoiled’
is used to indicate that the transverse magnetization components can be assumed
to be zero before the next excitation pulse, i.e., there will be no
contribution of transverse magnetization components to the signal.
$$$\qquad$$$ ·
The term ‘FID’
or ‘Echo’ in combination with SSFP is
used to indicate whether the signal refers to the transverse magnetization just
after or before the RF pulse (see also Fig. 2).
For common sequence acronyms
expressed in generic terms, see Table 1.
SSFP Sequences and Contrast
For incoherent SSFP imaging, the contribution
from any residual transverse magnetization prior to the next excitation pulse is
assumed to be zero, or spoiled. As a
result, spoiled sequences show a pure $$$T_1$$$ contrast, and the steady
state signal immediately after the excitation pulse is given by
$$\begin{align} M_{xy} = M_0 \frac{1-E_1}{1 - E_1 cos(\alpha)} sin(\alpha), && \tag{3}\end{align}$$
where $$$E_1
:= exp(–TR/T_1)$$$. Equation [3] is also known as the
“Ernst equation” (5–7,11). However, the tricky part is to actually get rid of all
the transverse magnetization components prior to the next excitation pulse or
to find a clever method to avoid any significant contribution in subsequent
repetition periods.
The simplest way of efficient
spoiling is a long enough waiting time. For repetition times of about three
times $$$T_2$$$, the transverse magnetization has decayed sufficiently;
however, in the light of a minimization of both TR and TA (c.f. beginning) this
approach is not very practical and a different spoiling strategy has to be
used.
An elegant and rather efficient
approach for spoiling transverse magnetization is to adapt the phase of the RF
excitation pulse Ф in every TR
interval (c.f. Fig. 3a) to have a linear phase increment according to the
formula
$$\begin{align} \phi_n - \phi_{n-1} = \psi_0 + n \cdot \psi. && \tag{4}\end{align}$$
This method ‑ commonly referred to
as RF spoiling (10,12,13) ‑ generates quadratically increasing phase offsets
in the residual transverse magnetization. A proper choice of ψ (e.g., ψ = 50° or ψ = 117°) leads to RF phase cycling conditions
that lead to a pure $$$T_1$$$-weighted signal, i.e., to a near complete
destructive interference of all residual transverse magnetization components.
For SSFP, the dephasing from
gradients must be constant and phase encoding gradients need to be rewinded ‑ a
possible sequence diagram for the (RF) spoiled SSFP is shown in Fig. 3a. Asides,
RF spoiling always comes with crusher gradients ‑ sometimes also (improperly) called
spoiler gradients.
Figure 3 provides a general
overview of four major types of SSFP sequences. In fact, all four major types (RF
spoiled SSFP, SSFP-FID, SSFP-Echo, balanced SSFP) can be regarded as a variant
from a generic SSFP sequence. Their sole difference is whether the particular
type of SSFP employs RF spoiling and crusher gradients in each case (c.f.
Table 1); however, the resulting signal behavior and image contrast is
different.
Figure 3a displays the already
discussed RF spoiled SSFP sequence (including a crusher gradient in the readout
direction) that facilitates a dominant $$$T_1$$$ contrast. Omitting RF
spoiling but keeping the application of crusher gradients prior to the next
excitation pulse, leads to the gradient-spoiled (i.e., also nonbalanced)
SSFP-FID sequence (Fig. 3b).
A time-reversal of the readout
gradient as displayed in Fig. 3c leads to the (gradient-spoiled) SSFP-Echo
sequence.
Finally, balanced SSFP refers to an
acquisition scheme as given in Fig. 3d, where all gradients have a zero net
area within any TR (= balanced); and any residual phase accruals within any TR
are therefore closely related to field inhomogeneities. As a result, balanced
SSFP is prone to off-resonances that can lead to prominent signal voids or
banding artifacts in regions of strong susceptibility variations and with poor
shimming.
It is evident that except for the
spoiled SSFP acquisition all other SSFP sequences show besides a $$$T_1$$$
contrast also some $$$T_2$$$ contrast due to the contribution from
unspoiled transverse magnetization components. To leading order, the
nonbalanced and balanced SSFP signal depends on $$$T_1/T_2$$$,
which generally leads to a prominent contrast between tissues and fluids (Fig.
4).
Magnetization Preparation & Applications
GRE sequences are often combined with magnetization preparation to
obtain dedicated contrasts and to increase the variety of MRI applications.
Furthermore, an important reason for using GRE (SSFP) based MRI sequences is
the ability of acquiring 2D and 3D volumes in short experiment times of a few
minutes down to less than a second. In the following, some important
applications are selected.
Dynamic Imaging & CINE Imaging
Modern rapid GRE approaches allow time resolved measurements of dynamic
processes. One example is a possible bolus tracking during contrast enhanced
MRI based angiography (see also ‘angiography’ below).
Although MRI acquisitions are generally too slow to generate a “direct
movie” of the beating human heart,
e.g., a balanced SSFP sequence can be synchronized with the cardiac cycle via ECG
triggering. Then, only segments or parts 1/N of k-space data are acquired at
defined points of time within the cardiac cycle. After N cardiac cycles the
data is measured completely and a sequence of images along the cardiac cycle
can be reconstructed. The image data set can be viewed as a CINE loop (“movie”)
to reveal the dynamics during an averaged cardiac cycle. This approach is
referred to as CINE imaging (14,15).
Inversion Recovery Preparation
For very fast SSFP approaches with short TR and low flip angles, the
resulting signal intensity diminishes more and more and the inherent contrast
becomes rather poor. A favorite approach to reinstate again stronger $$$T_1$$$
weighting in RF spoiled SSFP sequences is to use an initial 180° inversion
pulse (inversion recovery preparation). Furthermore, in dependence of the
inversion time TI, the signal intensity of certain tissues with a specific T1
can be enhanced, suppressed, or even nulled. Particularly for $$$T_1$$$
weighted 3D imaging of the entire head such a magnetization prepared rapid
gradient echo (MPRAGE) approach (16) has become popular in recent years.
Magnetization Transfer Preparation
Neglecting fat tissue for the moment, conventional clinical MRI sequences
are only able to acquire signal of the “free water”; i.e., only from 1H
nuclei (protons) that are bound to water molecules that are free to move in a
given tissue (17). Although they are excited, protons that are bound to macromolecules
cannot be observed, because their $$$T_2$$$ is too short (c.f. NMR
spectroscopy / MRS !). However, these protons are much more sensitive to off-resonant
excitation, i.e., RF excitation pulses which are off the Larmor frequency. Excited
macromolecular protons transfer their excitation to the “free protons”. This
effect is called magnetization transfer
or saturation transfer. Hence, adding
an off-resonant excitation pulse as a magnetization preparation to a GRE
sequence will pre-excite tissues with a high macromolecular content. As a
result, these tissues will display reduced signal intensity (-> MT effect).
MT preparations are usually combined with RF spoiled SSFP sequences.
Inflow Enhancement & Angiography
In GRE imaging the signal intensity of static tissue will be partially
saturated in accordance with the steady state conditions. In contrast, blood
that flows into the imaging slice or volume still carries “fresh”, i.e.,
unsaturated magnetization – a signal enhancement occurs. Thus, the signal of
blood is bright compared to the surrounding static tissue. This effect is
exploited in time-of-flight (TOF) imaging (11,18), a non-invasive MR
angiography (MRA) method without the administration of any external contrast
agent. TOF imaging is usually based on fast RF spoiled SSFP sequences, since
compared to SSFP-FID and balanced SSFP sequences they also suppress the bright
signal from cerebrospinal and other fluids due to strong $$$T_1$$$
weighting.
RF spoiled SSFP sequences are also used for
another family of MRA approaches: contrast enhanced MRA (CE-MRA, (11,19)).
However, the mechanism of signal enhancement in the vessels differs. CE-MRA
methods rely on externally administered contrast agents that
reduce the $$$T_1$$$ relaxation time in
their vicinity. Thus, the RF spoiled SSFP sequences provide a means for fast $$$T_1$$$
weighted acquisitions.
Fat-Water-Imaging
An important consequence of the missing refocusing pulse in a GRE
sequence is the chemical shift induced dephasing of magnetization that is not
refocused anymore. Due to the natural Larmor frequency difference between water
and fat of approx. 220Hz at 1.5T or 440Hz at 3T, their corresponding
magnetization vectors experience a time-dependent phase difference. Particularly,
at echo times TE = 4.4ms, 8.8ms, … water and fat signal are in-phase
(1.5T), the GRE sequence acquires a so-called in-phase image. At echo times TE = 2.2ms, 6.6ms, … water and
fat signal are opposed-phase (1.5T), the GRE sequence acquires an out-of-phase or opposed phase image. The latter images display dark rims due to
signal voids in voxels with a similar content of fat and water.
Dedicated in-phase and out-of-phase images can be processed to obtain
images that only display signal either from aqueous tissue (“water image”) or
from fatty tissue (“fat image”). This method is also known as “Dixon Imaging” (11,20).
Acknowledgements
No acknowledgement found.References
1. Hahn
EL. Spin Echoes. Phys Rev 1950;80:580–594.
2. Jung
BA, Weigel M. Spin echo magnetic resonance imaging. J Magn Reson Imaging
2013;37:805–817.
3.
Edelstein WA, Hutchison JMS, Johnson G, Redpath T. Spin warp NMR imaging and
applications to human whole-body imaging. Phys Med Biol 1980;25:751.
4. Haase
A, Frahm J, Matthaei F, Hänicke W, Merboldt KD. FLASH imaging. Rapid NMR
imaging using low flip-angle pulses. J Magn Reson 1986;67:258–266.
5. Markl
M, Leupold J. Gradient echo imaging. J Magn Reson Imaging 2012;35:1274–1289.
6. Bieri
O, Scheffler K. Fundamentals of balanced steady state free precession MRI. J
Magn Reson Imaging 2013;38:2–11.
7.
Hargreaves B. Rapid gradient-echo imaging. J Magn Reson Imaging
2012;36:1300–1313.
8. Carr
HY. Steady-State Free Precession in Nuclear Magnetic Resonance. Phys Rev
1958;112:1693.
9. Sobol
WT, Gauntt DM. On the stationary states in gradient echo imaging. J Magn Reson
Imaging 1996;6:384–98.
10. Zur Y,
Wood ML, Neuringer LJ. Spoiling of transverse magnetization in steady-state
sequences. Magn Reson Med 1991;21:251–263.
11.
Bernstein MA, King KF, Zhou XJ. Handbook of MRI Pulse Sequences. New York:
Elsevier Academic Press; 2004.
12.
Crawley AP, Wood ML, Henkelman RM. Elimination of transverse coherences in
FLASH MRI. Magn Reson Med 1988;8:248–260.
13. Duyn
JH. Steady state effects in fast gradient echo magnetic resonance imaging. Magn
Reson Med 1997;37:559–568.
14.
Atkinson DJ, Edelman RR. Cineangiography of the heart in a single breath hold
with a segmented turboFLASH sequence. Radiology 1991;178:357–360.
15.
Hernandez RJ, Aisen AM, Foo TK, Beekman RH. Thoracic cardiovascular anomalies
in children: evaluation with a fast gradient-recalled-echo sequence with
cardiac-triggered segmented acquisition. Radiology 1993;188:775–780.
16. Mugler
JP, Brookeman JR. Three-dimensional magnetization-prepared rapid gradient-echo
imaging (3D MP RAGE). Magn Reson Med 1990;15:152–157.
17.
Henkelman RM, Stanisz GJ, Graham SJ. Magnetization transfer in MRI: a review.
NMR Biomed 2001;14:57–64.
18. Wehrli
FW. Time-of-flight effects in MR imaging of flow. Magn Reson Med
1990;14:187–193.
19. Koenig
SH. From the relaxivity of Gd(DTPA) − to everything else. Magn Reson Med
1991;22:183–190.
20. Dixon
WT. Simple proton spectroscopic imaging. Radiology 1984;153:189–194.