Synopsis
This lecture
will review the clinical applications of three-dimensional sequences FSE
sequences in musculoskeletal MR imaging.
Magnetic
resonance (MR) imaging is one of the most commonly utilized imaging modality
for evaluating patients with joint pain in clinical practice. Musculoskeletal MR protocols at most
institution consist of two-dimensional intermediate-weighted and T2-weighted
fast spin-echo (FSE) sequences repeated in multiple planes. These sequences have high in-plane spatial
resolution and excellent tissue contrast which can be used to evaluate all
joint structures. However,
two-dimensional FSE sequences produce images with relatively thick slices and
gaps between slices which may obscure joint pathology secondary to partial
volume averaging.
Three-dimensional
FSE sequences have also been used to evaluate patients with joint pain. Three-dimensional FSE sequences have many potential
advantages over two-dimensional FSE sequences for evaluating the
musculoskeletal system.
Three-dimensional FSE sequences produce thin continuous images with high
in-plane spatial resolution which limit the effects of partial volume averaging. In addition, multi-planar reformat images can
be created from the volumetric source data which eliminates the need to repeat
sequences with identical tissue contrast in different orientations. The use of three-dimensional FSE sequences in
clinical practice would significantly decrease MR examination times which would
improve patient comfort, reduce motion artifact, and optimize the clinical
efficiency of the MR scanner. This lecture
will review the clinical applications of three-dimensional sequences FSE
sequences in musculoskeletal MR imaging.
Three-Dimensional
Knee Joint Imaging
Cartilage Imaging
Three-dimensional FSE sequences such as fast spin-echo Cube
(FSE-Cube, GE Healthcare) and sampling perfection with application oriented
contrasts using different flip angle evolutions (SPACE, Siemens Medical Systems)
have been used to evaluate the articular cartilage of the knee joint. These sequences utilize flip angle modulation
to constrain T2 decay over an extended echo train which allows intermediate-weighted
images of the knee joint with isotropic resolution to be acquired with minimal
blurring. Three-dimensional FSE
sequences typically use frequency selective fat-saturation to suppress signal
from adipose tissue (1, 2). However, FSE-Cube has recently
been combined with IDEAL fat-water separation to provide more robust
fat-suppression in areas of magnetic field inhomogeneity (3). FSE-Cube and SPACE have
lower in-plane spatial resolution and greater image blurring when compared to
other three-dimensional cartilage imaging sequences with similar acquisition
times which may decrease the conspicuity of superficial cartilage lesions (4, 5). However, these sequences provide excellent visualization
of high contrast structures such as small cartilage fissures surrounded
by joint fluid (6).
Multiple previous studies have compared image quality of three-dimensional
FSE sequences with other cartilage imaging sequences for evaluating the knee
joint at 3.0T. Friedrich and associates
compared water excitation FLASH, water excitation true-FISP, water excitation
DESS, and fat-saturated intermediate-weighted SPACE sequences in 10
asymptomatic volunteers and found that water-excitation DESS had the highest contrast-to-noise
ratio (CNR) between cartilage and synovial fluid and the greatest overall
performance on qualitative cartilage assessment (4). In a study
comparing fat-saturated intermediate-weighted FSE-Cube, IDEAL-SPGR,
IDEAL-GRASS, multi-echo in steady-state acquisition (MENSA), coherent
oscillatory state acquisition for manipulation of image contrast (COSMIC), and
VIPR-SSFP sequences in 5 asymptomatic volunteers and 5 patients with
osteoarthritis, Chen and associates found that FSE-Cube had the highest
cartilage SNR, highest CNR between cartilage and synovial fluid, and top rank
for cartilage lesion conspicuity (5). Only one previous study has
compared three-dimensional FSE sequences with other cartilage imaging sequences
for detecting surgically confirmed cartilage lesions within the knee joint. In
this study, Kijowski and associated
found that fat-saturated intermediate-weighted FSE-Cube had similar sensitivity
and specificity as two-dimensional fat-saturated FSE sequences for detecting
cartilage lesions in 100 patients with surgical correlation at 3.0T (7). Additional studies are needed
to determine whether three-dimensional FSE sequences have superior diagnostic
performance than other two-dimensional and three-dimensional sequences for
evaluating the articular cartilage of the knee joint.
Comprehensive
Joint Assessment
Three-dimensional
isotropic resolution FSE sequences have also been used to provide comprehensive
knee joint assessment. These sequences
produce images with intermediate-weighted contrast which is the most versatile
and commonly utilized tissue contrast in musculoskeletal MR imaging (8-12). Multiple studies with surgical correlation
have compared the diagnostic performance of two dimensional and
three-dimensional intermediate-weighted FSE sequences for evaluating the knee
joint at 3.0T. Kijowski and associated
found that fat-saturated FSE-Cube had similar sensitivity and specificity as
two-dimensional FSE sequences for detecting cartilage lesions, cruciate and
collateral ligament tears, meniscal tears, and bone marrow edema lesions in 100
patients with surgical correlation (7). Jung and associates also found no significant
difference between a three-dimensional
isotropic resolution FSE sequence and two-dimensional FSE sequences in
the sensitivity and specificity for
detecting surgically confirmed cruciate ligament and meniscal tears in 85
patients (13). Additional studies performed by
Notohamiprodjo and associates on 18 patients with surgical correlation (2) and by Subhas and associates
on 100 patients with surgical correlation (14) found that fat-saturated SPACE
sequences had similar diagnostic performance as two-dimensional FSE sequences
for detecting cartilage lesions, cruciate and collateral ligament tears, and
meniscal tears.
Preliminary
studies on the use of three-dimensional isotropic resolution FSE sequences for
providing comprehensive knee joint assessment are promising. However, Ristow and associates performed a
detailed subjective comparison of a fat-saturated intermediate-weighted
FSE-Cube sequence and two-dimensional FSE sequences in 50 patients at 3.0T and
described potential limitations of using FSE-Cube for evaluating the knee
joint. In this study, FSE-Cube was found
to have significantly lower image quality when compared to the two-dimensional sequences with a greater degree of blurring
and indistinctness of structural edge which may limit the detection of low
contrast structures such as meniscal tears and bone marrow edema lesions (6). In a study involving 250 patients with
surgical correlation, Kijowski and associates found that a fat-saturated
intermediate-weighted FSE-Cube sequence had similar sensitivity and specificity
for detecting medial meniscal tears and similar specificity for detecting
lateral meniscal tears at 3.0T when compared to two-dimensional FSE
sequences. However, FSE-Cube had
significantly lower sensitivity for detecting lateral meniscal tears which was
mainly attributed to its decreased ability to detect tears of the posterior
root of the lateral meniscus (15). At the current time, there is strong evidence
to suggest that three-dimensional isotropic resolution FSE sequences can be
used to provide rapid comprehensive knee joint assessment in patients with
severe pain or claustrophobia who cannot tolerate a long MR examination. However, the slight reduction in image
quality of three-dimensional FSE sequences may limit the detection of subtle
joint pathology. Additional large
clinical studies are needed before three-dimensional isotropic resolution FSE
sequences can replace two-dimensional FSE sequences in clinical practice.
Three-Dimensional
Ankle Joint Imaging
There has been
recent interest in using three-dimensional FSE sequences for providing rapid
comprehensive ankle joint assessment.
Yao and associates described the use of a three-dimensional isotropic
resolution Fourier transform FSE sequence with intermediate-weighted contrast
for evaluating the ankle joint at 3.0T (16). The authors emphasized the unique advantage
of the three-dimensional isotropic resolution sequence was its ability to
create high quality reformat images in arbitrary planes tailored to specific
anatomic regions of interest. In
addition, Stevens and associates compared intermediate-weighted FSE-Cube and
two dimensional FSE sequences for evaluating the ankle joint in 10 healthy
volunteers at 3.0T and found that FSE-Cube had significantly higher SNR
efficiency of cartilage, synovial fluid, and muscle. Image blurring was noted to be significantly
greater for FSE-Cube, but there was no significant difference between the
two-dimensional and three-dimensional sequences in the degree of artifact or
overall image quality (17). No previous study has compared
three-dimensional FSE sequences with currently used two-dimensional sequences
for evaluating the cartilage, ligaments, tendons, and osseous structures of the
ankle joint. For this reason, it is
unknown whether the thin, continuous slices of the three-dimensional sequences
and their ability to view complex anatomy in oblique and curved planes can
improve the detection of ankle joint pathology.
Three-Dimensional
Shoulder Joint Imaging
There has been
much recent interest in using three-dimensional FSE sequences to decrease the
time required to evaluate the shoulder joint during MR arthrography. Multiple studies with surgical correlation
have compared three-dimensional isotropic resolution FSE sequences with
currently used multi-planar two-dimensional fat-saturated T1-weighted FSE
sequences for detecting shoulder joint pathology during MR arthrography. Rybak and associates compared a fat-saturated
intermediate-weighted SPACE sequence with two-dimensional FSE sequences for
evaluating the shoulder joint during MR arthrogaphy in 28 patients at 1.5T and
found that the three-dimensional sequence provided significantly better
visualization of the biceps tendon anchor, subscapularis tendon insertion, and
articular cartilage and equivalent visualization of other joint
structures. SPACE was also found to have
similar sensitivity but significantly lower specificity than the
two-dimensional FSE-sequences for detecting cartilage lesions, rotator cuff
tears, and labral tears in the 11 patients with surgical correlation (18). In
addition, Jung and associates found that a fat-saturated intermediate-weighted
SPACE sequence had similar diagnostic performance as two-dimensional FSE
sequences during MR arthrography at 3.0T for detecting surgically confirmed
rotator cuff tears in 87 patients.
However, SPACE was found to have significantly lower accuracy than the
two-dimensional FSE sequences for predicting the exact depth and location of
the rotator cuff tears (19).
Three-dimensional
FSE sequences have also been used to evaluate the shoulder joint during
non-contrast MR imaging. However, these
sequences have been used much less extensively during non-contrast MR imaging
than during MR arthrography with only one study reported thus far in the
literature. Hill and associates compared
a fat-saturated T2-weighted FSE-Cube sequence and multi-planar two-dimensional
fat-saturated T2-weighted FSE sequences for evaluating the shoulder joint in 19
patients at 3.0T and found that the sequences had fair to good agreement and
similar confidence level for grading the rotator cuff tendon. The authors also described the use of a
novel reformat plane called the radial oblique coronal plane in which thin
continuous images perpendicular to each portion of the rotator cuff tendon
could be obtained from the FSE-Cube
source data . The authors
postulated that the curved oblique reformat images could reduce the effects of
partial volume averaging which would be
particularly helpful for identifying tears within the far anterior portion of
the rotator cuff tendon (20). There
are many potential benefits of using three-dimensional isotropic resolution
sequences for evaluating the shoulder joint including their ability to produce
thin continuous images in any orientation following a single acquisition. However, no previous studies with surgical
correlation have compared three-dimensional isotropic resolution sequences and
currently used two-dimensional FSE sequences for detecting shoulder joint
pathology during non-contrast MR imaging.
Three-Dimensional
Elbow Joint Imaging
A recent study has
investigated the use of three-dimensional FSE sequences for evaluating the elbow
joint. Shapiro and associates compared fat-saturated
intermediate-weighted FSE-Cube and two-dimensional FSE sequences for evaluating
the elbow joint in 10 asymptomatic volunteers at 3.0T. FSE-Cube was found to have significantly
higher synovial fluid SNR and significantly higher CNR between cartilage and synovial
fluid than the two-dimensional sequence but had slightly worse overall image
quality and greater blurring on subjective analysis. (21). Three-dimensional FSE sequences produce high
quality multi-planar reformat images which allow the complex anatomy of the
elbow joint to be evaluated in any orientation following a single
acquisition. However, additional studies
are needed to determine whether the improved visualization of elbow joint anatomy
can lead to better detection of elbow joint pathology.
Three-Dimensional
Wrist Joint Imaging
Three-dimensional
sequences produce images with thinner slices and higher in-plane resolution than
two-dimensional sequences and thus appear better suited for assessing the thin
articular cartilage and small ligaments and fibrocartilage structures of the
wrist joint. However, few previous
studies have directly compared various two-dimensional and three-dimensional sequences
for detecting wrist joint pathology.
Meier and associates compared two-dimensional fat-saturated
intermediate-weighted and T2-weighted FSE sequences, a fat-saturated
intermediate-weighted FSE-Cube sequence, a three-dimensional fat-saturated
ultra-fast gradient-echo (FGRE) sequence, and a three-dimensional fat-saturated
multiple echo recombined gradient-echo (MERGE) sequence for evaluating
articular cartilage in 5 cadaveric wrist joints at both 1.5T and 3.0T. The FGRE sequence acquired at 3.0T was found
to be best imaging technique and had the highest cartilage SNR, highest CNR
between cartilage and bone, and highest diagnostic performance for detecting
surgically confirmed cartilage lesions.
The anisotropic resolution FGRE sequence had higher in-plane spatial
resolution and decreased image blurring when compare to the isotropic resolution
FSE-Cube sequence which provided better detection of subtle cartilage lesions
within the wrist joint (22).
Only one
study has investigated the use of three-dimensional FSE sequences for
evaluating the wrist joint. Stevens and
associates compared fat-saturated intermediate-weighted FSE-Cube and
two-dimensional FSE sequences for evaluating the wrist joint during
non-contrast MR imaging in 10 asymptomatic volunteers at 1.5T. FSE-Cube was found to have similar cartilage
and synovial fluid SNR and similar CNR between cartilage and adjacent joint
structures as the two-dimensional sequences.
While the coronal FSE-Cube source images had significantly better
overall image quality on subjective analysis than the two-dimensional FSE
images, the axial FSE-Cube reformat images were degraded by blurring due to T2
decay over the extended echo train (23). The use of three-dimensional FSE sequences could
decrease the time required to evaluate the wrist joint by eliminating the need
to repeat sequences with identical tissue contrast in multiple planes. An additional advantage of these sequences is
their ability to create high quality oblique reformat images through the carpal
ligaments and triangular fibrocartilage complex which may improve the detection
of subtle tears. Additional studies are
needed to document the clinical benefits of using three-dimensional isotropic
resolution sequences for evaluating the wrist joint.
Acknowledgements
No acknowledgement found.References
1. Gold GE, Busse RF, Beehler C, et al.
Isotropic MRI of the knee with 3D fast spin-echo extended echo-train
acquisition (XETA): initial experience. AJR American journal of roentgenology.
2007;188(5):1287-93.
2. Notohamiprodjo M, Horng A,
Pietschmann MF, et al. MRI of the knee at 3T: first clinical results with an
isotropic PDfs-weighted 3D-TSE-sequence. Invest Radiol. 2009;44(9):585-97.
3. Madhuranthakam AJ, Yu H, Shimakawa
A, et al. T(2)-weighted 3D fast spin echo imaging with water-fat separation in
a single acquisition. J Magn Reson Imaging. 2010;32(3):745-51.
4. Friedrich KM, Reiter G, Kaiser B, et
al. High-resolution cartilage imaging of the knee at 3T: Basic evaluation of
modern isotropic 3D MR-sequences. Eur J Radiol. 2010.
5. Chen CA, Kijowski R, Shapiro LM, et
al. Cartilage morphology at 3.0T: assessment of three-dimensional magnetic
resonance imaging techniques. Journal of magnetic resonance imaging : JMRI.
2010;32(1):173-83.
6. Ristow O, Steinbach L, Sabo G, et
al. Isotropic 3D fast spin-echo imaging versus standard 2D imaging at 3.0 T of
the knee--image quality and diagnostic performance. Eur Radiol.
2009;19(5):1263-72.
7. Kijowski R, Davis KW, Woods MA, et
al. Knee joint: comprehensive assessment with 3D isotropic resolution fast
spin-echo MR imaging--diagnostic performance compared with that of conventional
MR imaging at 3.0 T. Radiology. 2009;252(2):486-95.
8. Lal NR, Jamadar DA, Doi K, et al.
Evaluation of bone contusions with fat-saturated fast spin-echo proton-density
magnetic resonance imaging. Can Assoc Radiol J. 2000;51(3):182-5.
9. Ha TP, Li KC, Beaulieu CF, et al.
Anterior cruciate ligament injury: fast spin-echo MR imaging with arthroscopic
correlation in 217 examinations. AJR Am J Roentgenol. 1998;170(5):1215-9.
10. Schaefer FK, Schaefer PJ, Brossmann J,
et al. Value of fat-suppressed PD-weighted TSE-sequences for detection of
anterior and posterior cruciate ligament lesions--comparison to arthroscopy.
Eur J Radiol. 2006;58(3):411-5.
11. Schweitzer ME, Tran D, Deely DM, Hume
EL. Medial collateral ligament injuries: evaluation of multiple signs,
prevalence and location of associated bone bruises, and assessment with MR
imaging. Radiology. 1995;194(3):825-9.
12. Mohr A, Roemer FW, Genant HK, Liess C.
Using fat-saturated proton density-weighted MR imaging to evaluate articular
cartilage. AJR Am J Roentgenol. 2003;181(1):280-1; author reply 1-2.
13. Jung JY, Yoon YC, Kwon JW, Ahn JH, Choe
BK. Diagnosis of internal derangement of the knee at 3.0-T MR imaging: 3D
isotropic intermediate-weighted versus 2D sequences. Radiology.
2009;253(3):780-7.
14. Subhas N, Kao A, Freire M, Polster J,
Obuchowski N, Winalski C. Comparison of 3D isotropic resolution fast spin-echo
MR imaging with conventional 2D MR imaging of knee ligaments and menisci at
3.0T. Proceedings of the Society of
Skeletal Radiology Annual Meeting. Las Vegas, 2010.
15. Kijowski R, Davis KW, Blankenbaker DG,
Woods MA, Del Rio AM, De Smet AA. Evaluation of the menisci of the knee joint
using three-dimensional isotropic resolution fast spin-echo imaging: diagnostic
performance in 250 patients with surgical correlation. Skeletal Radiol. 2011.
16. Yao L, Pitts JT, Thomasson D. Isotropic
3D fast spin-echo with proton-density-like contrast: a comprehensive approach
to musculoskeletal MRI. AJR Am J Roentgenol. 2007;188(2):W199-201.
17. Stevens KJ, Busse RF, Han E, et al.
Ankle: isotropic MR imaging with 3D-FSE-cube--initial experience in healthy
volunteers. Radiology. 2008;249(3):1026-33.
18. Rybak L, Vieira R, Shepard T, Wiggins
G, Recht M, Glaser C. 3D isotropic intermediate-weighted TSE: a new approach for shoulder arthrography at
1.5T? Proceedings of the Radiological
Society of North America Annual Scientific Meeting. Chicago, 2010.
19. Jung J, Jee W, Park M, Lee S, Kim Y.
Rotator cuff tears at 3.0T shoulder MR arthrography: diagnosis with 3D isotropic turbo spin-echo
SPACE sequence versus 2D conventional sequences. Proceedings of the Radiological Society of
North American Annual Scientific Meeting. Chicago, 2010.
20. Hill DM, Tuite MJ, Lum DP, Busse RF.
Isotropic 3D T2-weighted MR imaging of the rotator cuff tendon at 3.0T Proceedings of the Society of Skeletal
Radiology Annual Meeting. Isle of Palms, 2009. .
21. Shapiro LM, Jenkins AM, Stevens KJ, et
al. Isotropic MRI of the upper extremity with 3D-FSE-Cube. Preceedings of the 18th Annual Scientific
Meeting of the International Society of Magnetic Resonance in Medicine
Stockholm, 2010.
22. Meier
A, Rosas H, Tueting J, Kijowski R. Evaluation of the articular cartilage of the
wrist joint using two-dimensional and three-dimensional sequences at 1.5T and
3.0T. Proceedings of the 19th Annual
Scientific Session of the International Society of Magnetic Resonance in
Medicine
Montreal,
2011.
23. Stevens
KJ, Wallace CG, Chen W, Rosenberg JK, Gold GE. Imaging of the wrist at 1.5
Tesla using isotropic three-dimensional fast spin echo cube. J Magn Reson
Imaging. 2011;33(4):908-15.