Synopsis
MRI close to metallic implants is often desired to monitor
and diagnose the tissue close to the metal. However, MRI close to such implants
is difficult due to the metal’s interaction with the magnetic fields, which
causes image artifacts. In recent years, powerful methods have been developed
to overcome these artifacts. This talk will discuss how artifacts close to
metal can be reduced, both by the choice of scan parameters and also by using
advanced methods designed for artifact reduction.Highlights
- Metallic implants distort the magnetic field in
MRI, resulting in image artifacts
- The right choice of scan settings and parameters
can reduce these artifacts
- Artifacts can be corrected further by the use of
more advanced methods, such as View-Angle Tilting for in-plane distortions and
MSI methods for through-plane distortions
Target audience
Students, researchers, and healthcare
professionals interested in MRI close to metal implants.
Outcome/objectives
Audience members will have learned about the
challenges of MRI close to metal implants, the main associated artifacts and their causes,
and methods for correcting these artifacts.
Purpose
Arthroplasty procedures, such as hip and knee replacements,
are becoming more common in the United States, a trend which is projected to
continue. Often, further operations or monitoring is necessary due to
complications related to the implant. MRI close to the implant can be useful
for determining whether such revision surgery is needed, since it gives good
soft tissue contrast. However, MRI close to such devices is difficult due to
the metal’s interaction with the magnetic fields, which can cause severe
imaging artifacts. In recent years, powerful methods have been developed to
overcome these artifacts. This talk will discuss the main methods for reducing
artifacts when imaging close to metal, both in terms of choosing scan
parameters and using methods designed for artifact reduction.
Methods
Close to metal implants, the distortion in the main field
can be so great that it makes spins precess at varying rates within a single
voxel. This leads to signal loss from T2* dephasing. This signal drop-off can
be greatly reduced by applying spin-echo imaging instead of gradient echo
imaging.
Fat suppression techniques that
rely on the frequency difference between fat and water will also often fail
close to metal, since the frequency variation caused by the field inhomogeneity
can be much larger than the expected fat-water frequency difference. This can
be solved by using STIR imaging, which relies on differences in T1, rather than
frequency, between fat and water.
During signal readout, the field
distortions caused by the metal result in errors in the readout gradient. This
will lead to signals being mapped to the wrong location, resulting in image
distortion, signal loss, and pile-up. These in-plane distortions can be
resolved by using a large readout bandwidth and a View-Angle Tilting (VAT)
gradient. View-Angle Tilting involves replaying the slice-select gradient
during the readout.
Similarly, the metal will cause
errors in the slice-select gradient, resulting in an unintended shape of the
region that gets excited by the pulse. These through-plane, or slice,
distortions can lead to the excited slices being curved, thinned, thickened, or
split. The large frequency variations can also cause regions to simply not be
excited, since they are at a frequency outside the excitation bandwidth. Such errors
can be reduced by using a large excitation bandwidth and a strong slice-select
gradient. Even better correction can be achieved using Multi-Spectral Imaging
(MSI) methods, such as MAVRIC and SEMAC. These resolve the slice distortions
with 3D phase encoding, mapping each excited voxel to its correct location.
Results
Using Spin Echo imaging instead of Gradient Echo imaging greatly
reduces signal drop-off close to the metal [1]. Through-slice and in-plane
artifacts remain, resulting in distortion, signal loss, and pile-up artifacts.
STIR imaging [2] gives better fat suppression than conventional
fat-saturated imaging. However, it yields lower signal-to-noise (SNR) than
other methods, and is usually not a good option for contrast-enhanced imaging,
since it will suppress the tissue that takes up contrast, due to the shortened
T1.
The use of a VAT gradient largely removes in-plane
distortions, such as signal pile-up, due to metal [3]. Artifacts from
through-plane distortions, such as slice thinning and thickening, remain. The
use of a VAT gradient can result in blurring, which can be minimized by using a
large readout bandwidth and reducing the readout duration to less than that of
the main lobe of the RF pulse [4].
Using the MAVRIC or SEMAC MSI techniques [5,6] resolves the
through-plane artifacts, making the image useful for diagnosis close to the
metal. This comes at the cost of increased scan time from the
additional phase encoding. Some artifacts remain very close to the metal. These
are due to the very strong gradient errors, induced very close to the metal,
disrupting the Fourier encoding process [7].
Discussion
The methods presented can reduce signal drop-off, improve
fat suppression, and correct in-plane and through-plane artifacts close to
metal implants. Other methods for reducing artifacts near metal have been
proposed, such as field mapping techniques for estimating the field errors [8]
and ultrashort TE techniques to not allow for much dephasing before the signal
is acquired [9]. MSI methods have proven very useful for correcting artifacts
close to metal but do come at the cost of increased scan time. Several methods
have been proposed to reduce MSI scan time. These include both conventional
acceleration methods such as partial Fourier and parallel imaging [10,11] as
well as using compressed sensing [12] or making use of the shape of the
distorted slices to use a more efficient Field of View (FOV), saving scan time
[13]. Since the phase encoding direction does not experience distortion, recent
work has explored phase encoding along all directions within feasible scan
times, with promising results [14]. Recently, a 2D MSI method has been
developed [15], which can save scan time and has been used for thermal
monitoring close to metallic implants [16]. Methods have also been developed to
resolve signal voids, due to strong effects on local gradients close to the
metal, in order to visualize the implant geometry [17].
Conclusion
Metallic implants can produce severe artifacts in MRI,
complicating the monitoring of tissue close to the implant. Careful selection
of conventional imaging techniques and scan parameters can reduce these
artifacts to some extent. Furthermore, using more advanced techniques, such as
View-Angle Tilting or MSI methods can reduce these artifacts even further.
Acknowledgements
NIH, GE HealthcareReferences
1. Hargreaves BA, Worters PW, Pauly KB, Pauly JM, Koch KM,
Gold GE. Metal-Induced Artifacts in MRI. AJR 2011; 197:547-555.
2. Bydder GM, Pennock JM, Steiner RE, Khenia S, Payne JA,
Young IR. The short TI inversion recovery sequence: an approach to MR imaging
of the abdomen. Magn Reson Imaging 1985; 3:251-254.
3. Cho ZH, Kim DJ, Kim YK. Total inhomogeneity correction
including chemical shifts and susceptibility by view angle tilting. Med. Phys.
1988; 15(1):7-11.
4. Butts, K, Pauly J M, Gold GE. Reduction of blurring in
view angle tilting MRI. Magn REson Med 2005; 53:418-424.
5. Koch KM, Lorbiecki JE, Hinks RS, King KF. A multispectral
three-dimensional acquisition technique for imaging near metal implants. Magn
Reson Med 2009; 61:381-390.
6. Lu W, Pauly KB, Gold GE, Pauly JM, Hargreaves BA. SEMAC:
slice encoding for metal artifact correction in MRI. Magn Reson Med 2009;62:66-76.
7. Koch KM, Hargreaves BA, Pauly KB, Chen W, Gold GE, King
KF. Magnetic Resonance Imaging Near Metal Implants. J Magn Reson Imag
2010;32:773-787.
8. Pauly KB, Gold GE. Correction of slice profile distortion
from metallic devices. ISMRM 2006, p. 2380.
9. Gold GE, Thedens DR, Pauly JM, Fechner KP, Bergman G,
Beaulieu CF, Macovski A. MR imaging of articular cartilage of the knee: new
methods using ultrashort TEs. AJR 1998;170:1223-1226.
10. Noll DC, Nishimura DG, Macovski A. Homodyne detection in
magnetic resonance imaging. IEEE Trans Med Imaging 1991;10:154–163.
11. Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V,
Wang J, Kiefer B, Haase A. Generalized autocalibrating partially parallel
acquisitions (GRAPPA). Magn Reson Med 2002;47:1202–1210.
12. Worters PW, Sung K, Stevens KJ, Koch KM, Hargreaves BA.
Compressed-
sensing multispectral imaging of the postoperative spine. J
Magn Reson Imaging 2013;37:243–248.
13. Sveinsson B, Worters PW, Gold GE, Hargreaves BA.
Hexagonal undersampling for faster MRI near metallic implants. 2015;73:662-668.
14. Artz NS, Hernando D, Taviani V, Samsonov A, Brittain JH,
Reeder SB. Spectrally resolved fully phase-encoded three-dimensional fast
spin-echo imaging. Magn Reson Med 2014;71:681–690.
15. Hargreaves BA, Taviani V, Yoon D. Fast 2D imaging for
distortion correction near metal implants. ISMRM 2014, p. 615.
16. Weber H, Yoon D, Taviani V, Pauly KB, Hargreaves BA.
T1-based MR thermometry close to metal. ISMRM 2015, p. 4056.
17. Shi X, Yoon D, Koch D, Hargreaves B.
Regularized inversion of metallic implant Susceptibility from B0 field maps.
ISMRM 2015, p. 3734.