Synopsis
Specialty Area:
Interventional MRI -
Devices & Cardiovascular Applications
Speaker Name:
Kevan Anderson (kanders@sri.utoronto.ca)
Target Audience:
Physicists and clinicians interested in
interventional MRI
Outcomes / Objectives:
1.
Describe imaging strategies for passive and active
device visualization
2. Recognize safety considerations
for the use of cardiovascular devices in the MRI environment
Introduction:
Devices for cardiovascular applications consist mainly of guidewires
and catheters. Guidewires are long thin
wires that are inserted into an artery or vein and manipulated to form a path
to a particular target through the vasculature.
Guidewires are typically made of metals such as nitinol or stainless
steel. Catheters for vascular applications are tubes with either a single of
multiple lumens made of braided thermoplastic and are used to provide
mechanical support for guidewires. For many
non-vascular applications, such as electrophysiology procedures, catheters may
be more sophisticated and could include additional features such as needles,
ablation electrodes, and thermocouples.
The use of cardiovascular devices can be very challenging in the MRI
environment. One must ensure that the
materials used in their construction are MR-compatible to the extent that they
will not experience any significant forces through interaction with the static
magnetic field and that they will not significantly degrade the ability to image
the anatomy of interest. Moreover, the use of long conductive structures, such
as guidewires and conductors in catheters, can give rise to significant tissue
heating. Techniques to visualize devices are also important, as one needs to know
the position of the device with respect to the anatomy.
Passive Devices:
Broadly, devices developed for use in the MR scanner can be
separated into two major groups: passive devices and active devices. Passive devices are broadly defined as
devices that are not electrically connected to any other equipment. The use of
passive devices is largely motivated by the greater simplicity and perceived
safety compared to the collection of active devices which are described below.
The visualization of passive devices is reliant strictly on image contrast.
Several techniques have been proposed in order to improve the ability to
visualize a passive device on an MR image.
The use of susceptibility markers enables the visualization of devices
through the presence of signal voids induced by magnetic susceptibility
artifacts (1-4).
Another mechanism involves filling a region of the catheter with a medium that
creates either a MR signal void or an area of strong MR signal relative to that
of the surrounding anatomy. Several other sources of contrast methods have been
used. These include carbon dioxide gas (5) and
gadolinium (6,7).
A significant limitation of passive visualization techniques is that
the device must be manipulated within the scan plane (typically 5-10 mm thick)
in order for it to be visualized. This can be difficult, especially during
procedures in tortuous vessels and in the cardiac chambers. Moreover, if the
device moves outside of the imaging slice, increasing the slice thickness will
degrade contrast as more background signal from tissues will be present. Also,
automatic scan plane prescription is not possible because catheter localization
is performed visually and no quantitative information about catheter position
is generated for feedback to the scanner. The time it takes to visualize the
catheter is limited by the time required for image acquisition, image
reconstruction, and image display. Slow reconstruction rates and display times
can result in a decreased temporal resolution.
Active Devices:
Unlike passive devices, active devices have an electrical connection
to the MR scanner. Active devices are typically used by acquiring MR signal
from small imaging coils located along their length or at their tip. One common
method for visualizing active devices is to superimpose an image reconstructed
from receive coils located on the device onto an anatomical image acquired from
a larger surface coil (8-10).
Using this method, the small sensitivity regions of the catheter-based receive
coils indicate where the catheter is located. The images obtained from the
device-based coils can be colour-coded to improve visualization. Due to the fact that imaging coils are able
to receive MR signal from a significant volume around them, the device can
appear blurred when it is visualized using this method. As a result, it is
difficult to ascertain its exact position. Another limitation with this
technique is that, like passive techniques, the device must be manipulated
within the scan plane.
One popular catheter tracking technique involves projecting the MR
signal obtained from a solenoid receive coil located on the distal tip of the
catheter onto three orthogonal planes (11,12). This
is accomplished by exciting the entire volume and generating gradient recalled
echoes in three orthogonal directions while frequency encoding the position of
the microcoil. The position of the coil can then be determined by locating the
signal peak in each of the three projections. The calculated position of the
catheter can then be displayed on a real-time anatomical image to visualize its
position in an anatomical context.
Implementation of the projection technique into a rapid-acquisition
sequence is done by intermittently acquiring projection data sets from the
microcoil between rapid image acquisitions from a larger coil. Common types of
acquisitions in which to intersperse the tracking sequence are spiral
acquisitions and Cartesian steady state free procession (SSFP) acquisitions (13,14). Despite being a very rapid method for
localizing a device, there are some limitations with the technique. The peaks
of the projections that are acquired correspond to the location of maximum MR
signal received by the coil. Depending on the structure of the surrounding
medium, this region of maximum signal may not be located at the centre of the
coil and the coil’s orientation within the main magnetic field can have the
effect of creating dual-peaked signal projections. Another disadvantage with the projection
method is that it is not able to provide any information about device
orientation without the use of more than one device-based coil. Several types
of active devices have been developed to address the problem of directionality.
One utilizes more than one device-based microcoil placed in close proximity to
each other (15,16). By
determining the position of both micocoils using the projection method, and by
assuming that the device does not bend significantly between the two coils, it
is possible to determine the direction of the device. Another utilizes phase
information in the MR signal to determine the orientation of the microcoil with
respect to the static field (17).
Another class of techniques utilizes external coupling devices that inductively
couple to conductive devices that pass through them (18-21). This
makes it possible to receive MR signal from around the device without having a
wired electrical connection between the intravascular device and the
scanner. Using such a technique, one can
actively visualize a conventional guidewire that could otherwise only be
visualized using passive techniques.
Intravascular Imaging with
Active Devices:
The use of the MR signal received by active devices is not limited
solely to device visualization and tracking purposes but can also be used for
anatomical imaging. Using small device-based receive coils for imaging can be
beneficial as high-resolution images with a small field-of-view can be
produced.
The majority of intravascular imaging applications have focused on
the imaging of the vessel wall and atherosclerotic plaques. Towards this goal,
the majority of active devices developed for intravascular imaging have been
designed to be ‘side-looking’. One example is the opposed-solenoid
intravascular imaging catheter. This device consists of two solenoidal coils
that are separated by 1-2cm and are concentric with the axis of the catheter (16,22-24). In
addition to being able to image only beside the catheter, the opposed-solenoid
design suffers from the limitation that the catheter must be positioned such
that the anatomy of interest is centred between the two coils.
A novel active imaging device that overcomes this limitation is the
loopless antenna, also commonly referred to as an imaging guidewire. Unlike the
opposed-solenoid, the device does not contain any solenoidal coils and instead
consists of a folded dipole antenna that is located on the tip of a coaxial
cable (25). The
sensitivity of the device to MR signal is much more distributed, and although
the area of high sensitivity is still located to the side of the device, it is
present for a greater distance over its length. This makes imaging much less
sensitive to precise device positioning. Moreover, the imaging sensitivity of
the device falls off much less rapidly with distance when compared to
solenoidal designs. The loopless antenna can also be fabricated to be much
thinner and more flexible than imaging devices that contain solenoidal coils.
Tissue Heating Associated
with Conductive Devices:
One of the major limitations of active devices and guidewires is
that significant safety concerns are associated with the use of long conducting
structures in an MR scanner as tissue heating can occur due to coupling between
the conductive structure and the electric component of radio-frequency (RF)
excitation field (26-28). Tissue heating occurs because currents on the
conductor can create large electric fields at the device tip thereby resulting
in tissue heating.
Several solutions have been proposed to mitigate the heating of
tissues around conductive devices and many have focused specifically on methods
for reducing currents on the transmission lines that connect active devices to
the MR scanner. Ladd et al. demonstrated that the amount of tissue heating at
the tip of a transmission line could be reduced by incorporating
quarter-wavelength coaxial current chokes onto the transmission line (29).
Weiss et al. demonstrated that heating could be reduced through the use of
transformers positioned at discrete intervals along the length of the transmission
line (30,31). Both
of these techniques, although shown to mitigate the heating effect, are
restricted to use in thicker devices such as catheters as they require additional
space in order to be implemented. For instance, the current choke as proposed
by Ladd et al. requires the use of a tri-axial cable (which is thicker than
ordinary coaxial cable) and transformers need to be of sufficient size to
enable efficient signal propagation. As a result, they cannot be applied to
thinner devices such as active and conventional guidewires.
For thinner devices, several strategies to
mitigate heating have focused on the development of techniques to identify
unsafe RF power levels used to excite spins in an MR scanner. Characterization
studies have attempted to identify the power limits that can safely be used in
the presence of interventional devices (32-34). Attempts have also focused on techniques to
detect the presence of currents on devices before the tissue experiences a
significant temperature rise. One such technique uses a low-power
reverse-quadrature excitation (35). Another detects currents through the use of an
inductively-coupled pick-up coil positioned over top of the wire (36). Current research is also focusing on excitation
techniques using surface coil arrays to minimize the coupling with conductive
devices (37,38).
Acknowledgements
No acknowledgement found.References
(1) Bakker
CJ, Hoogeveen RM, Weber J, van Vaals JJ, Viergever MA, Mali WP. Visualization of
Dedicated Catheters Using Fast Scanning Techniques with Potential for MR-Guided
Vascular Interventions. Magn Reson Med 1996;36:816–820.
(2) Bakker
CJ, Bos C, Weinmann HJ. Passive Tracking of Catheters and Guidewires by
Contrast-Enhanced MR Fluoroscopy. Magn Reson Med 2001;45:17–23.
(3) Mekle
R, Hofmann E, Scheffler K, Bilecen D. A Polymer-Based MR-Compatible Guidewire:
a Study to Explore New Prospects for Interventional Peripheral Magnetic
Resonance Angiography (ipMRA). J. Magn. Reson. Imaging 2006;23:145–155.
(4) Krueger
S, Schmitz S, Weiss S, Wirtz D, Linssen M, Schade H, Kraemer N, Spuentrup E,
Krombach G, Buecker A. An MR Guidewire Based on Micropultruded Fiber-Reinforced
Material. Magn Reson Med 2008;60:1190–1196.
(5) Miquel
ME, Hegde S, Muthurangu V, Corcoran BJ, Keevil SF, Hill DLG, Razavi RS.
Visualization and Tracking of an Inflatable Balloon Catheter Using SSFP in a
Flow Phantom and in the Heart and Great Vessels of Patients. Magn Reson Med
2004;51:988–995.
(6) Omary
RA, Unal O, Koscielski DS, Frayne R, Korosec FR, Mistretta CA, Strother CM,
Grist TM. Real-Time MR Imaging-Guided Passive Catheter Tracking with Use of
Gadolinium-Filled Catheters. Journal of vascular and interventional radiology :
JVIR 2000;11:1079–1085.
(7) Unal
O, Korosec FR, Frayne R, Strother CM, Mistretta CA. A Rapid 2D Time-Resolved
Variable-Rate K-Space Sampling MR Technique for Passive Catheter Tracking
During Endovascular Procedures. Magn Reson Med 1998;40:356–362.
(8) Serfaty
JM, Yang X, Aksit P, Quick HH, Solaiyappan M, Atalar E. Toward MRI-Guided
Coronary Catheterization: Visualization of Guiding Catheters, Guidewires, and
Anatomy in Real Time. J. Magn. Reson. Imaging 2000;12:590–594.
(9) Lederman
RJ, Guttman MA, Peters DC, Thompson RB, Sorger JM, Dick AJ, Raman VK, McVeigh ER.
Catheter-Based Endomyocardial Injection with Real-Time Magnetic Resonance
Imaging. Circulation 2002;105:1282–1284.
(10) Karmarkar
PV, Kraitchman DL, Izbudak I, Hofmann LV, Amado LC, Fritzges D, Young R,
Pittenger M, Bulte JWM, Atalar E. MR-Trackable Intramyocardial Injection
Catheter. Magn Reson Med 2004;51:1163–1172.
(11) Dumoulin
CL, Souza SP, Darrow RD. Real-Time Position Monitoring of Invasive Devices
Using Magnetic Resonance. Magn Reson Med 1993;29:411–415.
(12) Dumoulin
CL, Mallozzi RP, Darrow RD, Schmidt EJ. Phase-Field Dithering for Active
Catheter Tracking. Magn Reson Med 2010;63:1398–1403.
(13) Elgort
DR, Wong EY, Hillenbrand CM, Wacker FK, Lewin JS, Duerk JL. Real-Time Catheter
Tracking and Adaptive Imaging. J. Magn. Reson. Imaging 2003;18:621–626.
(14) Bock
M, Volz S, Zühlsdorff S, Umathum R, Fink C, Hallscheidt P, Semmler W. MR-Guided
Intravascular Procedures: Real-Time Parameter Control and Automated Slice
Positioning with Active Tracking Coils. J. Magn. Reson. Imaging
2004;19:580–589.
(15) Flask
C, Elgort D, Wong E, Shankaranarayanan A, Lewin J, Wendt M, Duerk JL. A Method
for Fast 3D Tracking Using Tuned Fiducial Markers and a Limited Projection
Reconstruction FISP (LPR-FISP) Sequence. J. Magn. Reson. Imaging
2001;14:617–627.
(16) Hillenbrand
CM, Elgort DR, Wong EY, Reykowski A, Wacker FK, Lewin JS, Duerk JL. Active
Device Tracking and High-Resolution Intravascular MRI Using a Novel
Catheter-Based, Opposed-Solenoid Phased Array Coil. Magn Reson Med
2004;51:668–675.
(17) Anderson
KJT, Scott GC, Wright GA. Catheter Tracking with Phase Information in a
Magnetic Resonance Scanner. Medical Imaging, IEEE Transactions on
2012;31:1173–1180.
(18) Hillenbrand
CM, Elgort DR, Wong EY, Reykowski A, Wacker FK, Lewin JS, Duerk JL. The bazooka coil: A novel dual-purpose device
for active visualization and reduction of cable currents in electrically
conductive endovascular instruments. In
Proceedings of the 13th Annual Meeting
of ISMRM,Miami,2005. 179.
(19) Etezadi-Amoli
M, Stang P, Kerr A, Pauly J, Scott G. Interventional Device Visualization with
Toroidal Transceiver and Optically Coupled Current Sensor for Radiofrequency
Safety Monitoring. Magn Reson Med 2015;73:1315-1327.
(20) Anderson
KJT, Yak N, Biswas L, Barry J, Wright GA.
In Vivo Active Visualization of
an Ablation Guidewire for the Revascularization of Occlusive Arterial Disease. In Proceedings of the 21st Annual Meeting of ISMRM,Salt
Lake City,2013. 474.
(21) Lockwood
J, Griffin GH, Wright GA, Anderson KJT. Determining
the Location of the Tip of an Active Transceive Guidewire. In Proceedings of the 10th Interventional MRI Symposium,Leipzig:,2014. P-24.
(22) Martin
AJ, Plewes DB, Henkelman RM. MR Imaging of Blood Vessels with an Intravascular
Coil. J. Magn. Reson. Imaging 1992;2:421–429.
(23) Martin
AJ, Henkelman RM. Intravascular MR Imaging in a Porcine Animal Model. Magn
Reson Med 1994;32:224–229.
(24) Hurst
GC, Hua J, Duerk JL, Cohen AM. Intravascular (Catheter) NMR Receiver Probe:
Preliminary Design Analysis and Application to Canine Iliofemoral Imaging. Magn
Reson Med 1992;24:343–357.
(25) Ocali
O, Atalar E. Intravascular Magnetic Resonance Imaging Using a Loopless Catheter
Antenna. Magn Reson Med 1997;37:112–118.
(26) Konings
MK, Bartels LW, Smits HF, Bakker CJ. Heating Around Intravascular Guidewires by
Resonating RF Waves. J. Magn. Reson. Imaging 2000;12:79–85.
(27) Liu
CY, Farahani K, Lu DS, Duckwiler G, Oppelt A. Safety of MRI-Guided Endovascular
Guidewire Applications. J. Magn. Reson. Imaging 2000;12:75–78.
(28) Nitz
WR, Oppelt A, Renz W, Manke C, Lenhart M, Link J. On the Heating of Linear
Conductive Structures as Guide Wires and Catheters in Interventional MRI. J.
Magn. Reson. Imaging 2001;13:105–114.
(29) Ladd
ME, Quick HH. Reduction of Resonant RF Heating in Intravascular Catheters Using
Coaxial Chokes. Magn Reson Med 2000;43:615–619.
(30) Weiss
S, Vernickel P, Schaeffter T, Schulz V. Transmission Line for Improved RF
Safety of Interventional Devices. Magn Reson Med 2005;54:182-189.
(31) Vernickel
P, Schulz V, Weiss S, Gleich B. A Safe Transmission Line for MRI. IEEE Trans
Biomed Eng 2005;52:1094-1102.
(32) Yeung
CJ, Atalar E. RF Transmit Power Limit for the Barewire Loopless Catheter
Antenna. J. Magn. Reson. Imaging 2000;12:86–91.
(33) Yeung
CJ, Susil RC, Atalar E. RF Safety of Wires in Interventional MRI: Using a
Safety Index. Magn Reson Med 2002;47:187–193.
(34) Yeung
CJ, Karmarkar P, McVeigh ER. Minimizing RF Heating of Conducting Wires in MRI.
Magn Reson Med 2007;58:1028–1034.
(35) Overall
WR, Pauly JM, Stang PP, Scott GC. Ensuring Safety of Implanted Devices Under
MRI Using Reversed RF Polarization. Magn Reson Med 2010;64:823–833.
(36) Zanchi
MG, Venook R, Pauly JM, Scott GC. An Optically Coupled System for Quantitative
Monitoring of MRI-Induced RF Currents Into Long Conductors. Ieee T Med Imaging 2010;29:169–178.
(37) Gudino
N, Sonmez M, Yao Z, Baig T, Nielles-Vallespin S, Faranesh AZ, Lederman RJ,
Martens M, Balaban RS, Hansen MS, Griswold MA. Parallel Transmit Excitation at
1.5 T Based on the Minimization of a Driving Function for Device Heating.
Medical physics 2015;42:359–371.
(38) Etezadi-Amoli
M, Stang P, Kerr A, Pauly J, Scott G. Controlling Radiofrequency-Induced
Currents in Guidewires Using Parallel Transmit. Magn Reson Med
2015;74:1790–1802.