Active & Passive Devices
Kevan Anderson

Synopsis

Specialty Area:

Interventional MRI - Devices & Cardiovascular Applications

Speaker Name:

Kevan Anderson (kanders@sri.utoronto.ca)

Target Audience:

Physicists and clinicians interested in interventional MRI

Outcomes / Objectives:

1. Describe imaging strategies for passive and active device visualization

2. Recognize safety considerations for the use of cardiovascular devices in the MRI environment

Introduction:

Devices for cardiovascular applications consist mainly of guidewires and catheters. Guidewires are long thin wires that are inserted into an artery or vein and manipulated to form a path to a particular target through the vasculature. Guidewires are typically made of metals such as nitinol or stainless steel. Catheters for vascular applications are tubes with either a single of multiple lumens made of braided thermoplastic and are used to provide mechanical support for guidewires. For many non-vascular applications, such as electrophysiology procedures, catheters may be more sophisticated and could include additional features such as needles, ablation electrodes, and thermocouples.

The use of cardiovascular devices can be very challenging in the MRI environment. One must ensure that the materials used in their construction are MR-compatible to the extent that they will not experience any significant forces through interaction with the static magnetic field and that they will not significantly degrade the ability to image the anatomy of interest. Moreover, the use of long conductive structures, such as guidewires and conductors in catheters, can give rise to significant tissue heating. Techniques to visualize devices are also important, as one needs to know the position of the device with respect to the anatomy.

Passive Devices:

Broadly, devices developed for use in the MR scanner can be separated into two major groups: passive devices and active devices. Passive devices are broadly defined as devices that are not electrically connected to any other equipment. The use of passive devices is largely motivated by the greater simplicity and perceived safety compared to the collection of active devices which are described below. The visualization of passive devices is reliant strictly on image contrast. Several techniques have been proposed in order to improve the ability to visualize a passive device on an MR image. The use of susceptibility markers enables the visualization of devices through the presence of signal voids induced by magnetic susceptibility artifacts (1-4). Another mechanism involves filling a region of the catheter with a medium that creates either a MR signal void or an area of strong MR signal relative to that of the surrounding anatomy. Several other sources of contrast methods have been used. These include carbon dioxide gas (5) and gadolinium (6,7).

A significant limitation of passive visualization techniques is that the device must be manipulated within the scan plane (typically 5-10 mm thick) in order for it to be visualized. This can be difficult, especially during procedures in tortuous vessels and in the cardiac chambers. Moreover, if the device moves outside of the imaging slice, increasing the slice thickness will degrade contrast as more background signal from tissues will be present. Also, automatic scan plane prescription is not possible because catheter localization is performed visually and no quantitative information about catheter position is generated for feedback to the scanner. The time it takes to visualize the catheter is limited by the time required for image acquisition, image reconstruction, and image display. Slow reconstruction rates and display times can result in a decreased temporal resolution.

Active Devices:

Unlike passive devices, active devices have an electrical connection to the MR scanner. Active devices are typically used by acquiring MR signal from small imaging coils located along their length or at their tip. One common method for visualizing active devices is to superimpose an image reconstructed from receive coils located on the device onto an anatomical image acquired from a larger surface coil (8-10). Using this method, the small sensitivity regions of the catheter-based receive coils indicate where the catheter is located. The images obtained from the device-based coils can be colour-coded to improve visualization. Due to the fact that imaging coils are able to receive MR signal from a significant volume around them, the device can appear blurred when it is visualized using this method. As a result, it is difficult to ascertain its exact position. Another limitation with this technique is that, like passive techniques, the device must be manipulated within the scan plane.

One popular catheter tracking technique involves projecting the MR signal obtained from a solenoid receive coil located on the distal tip of the catheter onto three orthogonal planes (11,12). This is accomplished by exciting the entire volume and generating gradient recalled echoes in three orthogonal directions while frequency encoding the position of the microcoil. The position of the coil can then be determined by locating the signal peak in each of the three projections. The calculated position of the catheter can then be displayed on a real-time anatomical image to visualize its position in an anatomical context.

Implementation of the projection technique into a rapid-acquisition sequence is done by intermittently acquiring projection data sets from the microcoil between rapid image acquisitions from a larger coil. Common types of acquisitions in which to intersperse the tracking sequence are spiral acquisitions and Cartesian steady state free procession (SSFP) acquisitions (13,14). Despite being a very rapid method for localizing a device, there are some limitations with the technique. The peaks of the projections that are acquired correspond to the location of maximum MR signal received by the coil. Depending on the structure of the surrounding medium, this region of maximum signal may not be located at the centre of the coil and the coil’s orientation within the main magnetic field can have the effect of creating dual-peaked signal projections. Another disadvantage with the projection method is that it is not able to provide any information about device orientation without the use of more than one device-based coil. Several types of active devices have been developed to address the problem of directionality. One utilizes more than one device-based microcoil placed in close proximity to each other (15,16). By determining the position of both micocoils using the projection method, and by assuming that the device does not bend significantly between the two coils, it is possible to determine the direction of the device. Another utilizes phase information in the MR signal to determine the orientation of the microcoil with respect to the static field (17).

Another class of techniques utilizes external coupling devices that inductively couple to conductive devices that pass through them (18-21). This makes it possible to receive MR signal from around the device without having a wired electrical connection between the intravascular device and the scanner. Using such a technique, one can actively visualize a conventional guidewire that could otherwise only be visualized using passive techniques.

Intravascular Imaging with Active Devices:

The use of the MR signal received by active devices is not limited solely to device visualization and tracking purposes but can also be used for anatomical imaging. Using small device-based receive coils for imaging can be beneficial as high-resolution images with a small field-of-view can be produced.

The majority of intravascular imaging applications have focused on the imaging of the vessel wall and atherosclerotic plaques. Towards this goal, the majority of active devices developed for intravascular imaging have been designed to be ‘side-looking’. One example is the opposed-solenoid intravascular imaging catheter. This device consists of two solenoidal coils that are separated by 1-2cm and are concentric with the axis of the catheter (16,22-24). In addition to being able to image only beside the catheter, the opposed-solenoid design suffers from the limitation that the catheter must be positioned such that the anatomy of interest is centred between the two coils.

A novel active imaging device that overcomes this limitation is the loopless antenna, also commonly referred to as an imaging guidewire. Unlike the opposed-solenoid, the device does not contain any solenoidal coils and instead consists of a folded dipole antenna that is located on the tip of a coaxial cable (25). The sensitivity of the device to MR signal is much more distributed, and although the area of high sensitivity is still located to the side of the device, it is present for a greater distance over its length. This makes imaging much less sensitive to precise device positioning. Moreover, the imaging sensitivity of the device falls off much less rapidly with distance when compared to solenoidal designs. The loopless antenna can also be fabricated to be much thinner and more flexible than imaging devices that contain solenoidal coils.

Tissue Heating Associated with Conductive Devices:

One of the major limitations of active devices and guidewires is that significant safety concerns are associated with the use of long conducting structures in an MR scanner as tissue heating can occur due to coupling between the conductive structure and the electric component of radio-frequency (RF) excitation field (26-28). Tissue heating occurs because currents on the conductor can create large electric fields at the device tip thereby resulting in tissue heating.

Several solutions have been proposed to mitigate the heating of tissues around conductive devices and many have focused specifically on methods for reducing currents on the transmission lines that connect active devices to the MR scanner. Ladd et al. demonstrated that the amount of tissue heating at the tip of a transmission line could be reduced by incorporating quarter-wavelength coaxial current chokes onto the transmission line (29). Weiss et al. demonstrated that heating could be reduced through the use of transformers positioned at discrete intervals along the length of the transmission line (30,31). Both of these techniques, although shown to mitigate the heating effect, are restricted to use in thicker devices such as catheters as they require additional space in order to be implemented. For instance, the current choke as proposed by Ladd et al. requires the use of a tri-axial cable (which is thicker than ordinary coaxial cable) and transformers need to be of sufficient size to enable efficient signal propagation. As a result, they cannot be applied to thinner devices such as active and conventional guidewires.

For thinner devices, several strategies to mitigate heating have focused on the development of techniques to identify unsafe RF power levels used to excite spins in an MR scanner. Characterization studies have attempted to identify the power limits that can safely be used in the presence of interventional devices (32-34). Attempts have also focused on techniques to detect the presence of currents on devices before the tissue experiences a significant temperature rise. One such technique uses a low-power reverse-quadrature excitation (35). Another detects currents through the use of an inductively-coupled pick-up coil positioned over top of the wire (36). Current research is also focusing on excitation techniques using surface coil arrays to minimize the coupling with conductive devices (37,38).

Acknowledgements

No acknowledgement found.

References

(1) Bakker CJ, Hoogeveen RM, Weber J, van Vaals JJ, Viergever MA, Mali WP. Visualization of Dedicated Catheters Using Fast Scanning Techniques with Potential for MR-Guided Vascular Interventions. Magn Reson Med 1996;36:816–820.

(2) Bakker CJ, Bos C, Weinmann HJ. Passive Tracking of Catheters and Guidewires by Contrast-Enhanced MR Fluoroscopy. Magn Reson Med 2001;45:17–23.

(3) Mekle R, Hofmann E, Scheffler K, Bilecen D. A Polymer-Based MR-Compatible Guidewire: a Study to Explore New Prospects for Interventional Peripheral Magnetic Resonance Angiography (ipMRA). J. Magn. Reson. Imaging 2006;23:145–155.

(4) Krueger S, Schmitz S, Weiss S, Wirtz D, Linssen M, Schade H, Kraemer N, Spuentrup E, Krombach G, Buecker A. An MR Guidewire Based on Micropultruded Fiber-Reinforced Material. Magn Reson Med 2008;60:1190–1196.

(5) Miquel ME, Hegde S, Muthurangu V, Corcoran BJ, Keevil SF, Hill DLG, Razavi RS. Visualization and Tracking of an Inflatable Balloon Catheter Using SSFP in a Flow Phantom and in the Heart and Great Vessels of Patients. Magn Reson Med 2004;51:988–995.

(6) Omary RA, Unal O, Koscielski DS, Frayne R, Korosec FR, Mistretta CA, Strother CM, Grist TM. Real-Time MR Imaging-Guided Passive Catheter Tracking with Use of Gadolinium-Filled Catheters. Journal of vascular and interventional radiology : JVIR 2000;11:1079–1085.

(7) Unal O, Korosec FR, Frayne R, Strother CM, Mistretta CA. A Rapid 2D Time-Resolved Variable-Rate K-Space Sampling MR Technique for Passive Catheter Tracking During Endovascular Procedures. Magn Reson Med 1998;40:356–362.

(8) Serfaty JM, Yang X, Aksit P, Quick HH, Solaiyappan M, Atalar E. Toward MRI-Guided Coronary Catheterization: Visualization of Guiding Catheters, Guidewires, and Anatomy in Real Time. J. Magn. Reson. Imaging 2000;12:590–594.

(9) Lederman RJ, Guttman MA, Peters DC, Thompson RB, Sorger JM, Dick AJ, Raman VK, McVeigh ER. Catheter-Based Endomyocardial Injection with Real-Time Magnetic Resonance Imaging. Circulation 2002;105:1282–1284.

(10) Karmarkar PV, Kraitchman DL, Izbudak I, Hofmann LV, Amado LC, Fritzges D, Young R, Pittenger M, Bulte JWM, Atalar E. MR-Trackable Intramyocardial Injection Catheter. Magn Reson Med 2004;51:1163–1172.

(11) Dumoulin CL, Souza SP, Darrow RD. Real-Time Position Monitoring of Invasive Devices Using Magnetic Resonance. Magn Reson Med 1993;29:411–415. (12) Dumoulin CL, Mallozzi RP, Darrow RD, Schmidt EJ. Phase-Field Dithering for Active Catheter Tracking. Magn Reson Med 2010;63:1398–1403.

(13) Elgort DR, Wong EY, Hillenbrand CM, Wacker FK, Lewin JS, Duerk JL. Real-Time Catheter Tracking and Adaptive Imaging. J. Magn. Reson. Imaging 2003;18:621–626.

(14) Bock M, Volz S, Zühlsdorff S, Umathum R, Fink C, Hallscheidt P, Semmler W. MR-Guided Intravascular Procedures: Real-Time Parameter Control and Automated Slice Positioning with Active Tracking Coils. J. Magn. Reson. Imaging 2004;19:580–589.

(15) Flask C, Elgort D, Wong E, Shankaranarayanan A, Lewin J, Wendt M, Duerk JL. A Method for Fast 3D Tracking Using Tuned Fiducial Markers and a Limited Projection Reconstruction FISP (LPR-FISP) Sequence. J. Magn. Reson. Imaging 2001;14:617–627.

(16) Hillenbrand CM, Elgort DR, Wong EY, Reykowski A, Wacker FK, Lewin JS, Duerk JL. Active Device Tracking and High-Resolution Intravascular MRI Using a Novel Catheter-Based, Opposed-Solenoid Phased Array Coil. Magn Reson Med 2004;51:668–675.

(17) Anderson KJT, Scott GC, Wright GA. Catheter Tracking with Phase Information in a Magnetic Resonance Scanner. Medical Imaging, IEEE Transactions on 2012;31:1173–1180.

(18) Hillenbrand CM, Elgort DR, Wong EY, Reykowski A, Wacker FK, Lewin JS, Duerk JL. The bazooka coil: A novel dual-purpose device for active visualization and reduction of cable currents in electrically conductive endovascular instruments. In Proceedings of the 13th Annual Meeting of ISMRM,Miami,2005. 179.

(19) Etezadi-Amoli M, Stang P, Kerr A, Pauly J, Scott G. Interventional Device Visualization with Toroidal Transceiver and Optically Coupled Current Sensor for Radiofrequency Safety Monitoring. Magn Reson Med 2015;73:1315-1327.

(20) Anderson KJT, Yak N, Biswas L, Barry J, Wright GA. In Vivo Active Visualization of an Ablation Guidewire for the Revascularization of Occlusive Arterial Disease. In Proceedings of the 21st Annual Meeting of ISMRM,Salt Lake City,2013. 474.

(21) Lockwood J, Griffin GH, Wright GA, Anderson KJT. Determining the Location of the Tip of an Active Transceive Guidewire. In Proceedings of the 10th Interventional MRI Symposium,Leipzig:,2014. P-24.

(22) Martin AJ, Plewes DB, Henkelman RM. MR Imaging of Blood Vessels with an Intravascular Coil. J. Magn. Reson. Imaging 1992;2:421–429.

(23) Martin AJ, Henkelman RM. Intravascular MR Imaging in a Porcine Animal Model. Magn Reson Med 1994;32:224–229.

(24) Hurst GC, Hua J, Duerk JL, Cohen AM. Intravascular (Catheter) NMR Receiver Probe: Preliminary Design Analysis and Application to Canine Iliofemoral Imaging. Magn Reson Med 1992;24:343–357.

(25) Ocali O, Atalar E. Intravascular Magnetic Resonance Imaging Using a Loopless Catheter Antenna. Magn Reson Med 1997;37:112–118.

(26) Konings MK, Bartels LW, Smits HF, Bakker CJ. Heating Around Intravascular Guidewires by Resonating RF Waves. J. Magn. Reson. Imaging 2000;12:79–85.

(27) Liu CY, Farahani K, Lu DS, Duckwiler G, Oppelt A. Safety of MRI-Guided Endovascular Guidewire Applications. J. Magn. Reson. Imaging 2000;12:75–78.

(28) Nitz WR, Oppelt A, Renz W, Manke C, Lenhart M, Link J. On the Heating of Linear Conductive Structures as Guide Wires and Catheters in Interventional MRI. J. Magn. Reson. Imaging 2001;13:105–114.

(29) Ladd ME, Quick HH. Reduction of Resonant RF Heating in Intravascular Catheters Using Coaxial Chokes. Magn Reson Med 2000;43:615–619.

(30) Weiss S, Vernickel P, Schaeffter T, Schulz V. Transmission Line for Improved RF Safety of Interventional Devices. Magn Reson Med 2005;54:182-189.

(31) Vernickel P, Schulz V, Weiss S, Gleich B. A Safe Transmission Line for MRI. IEEE Trans Biomed Eng 2005;52:1094-1102. (32) Yeung CJ, Atalar E. RF Transmit Power Limit for the Barewire Loopless Catheter Antenna. J. Magn. Reson. Imaging 2000;12:86–91.

(33) Yeung CJ, Susil RC, Atalar E. RF Safety of Wires in Interventional MRI: Using a Safety Index. Magn Reson Med 2002;47:187–193.

(34) Yeung CJ, Karmarkar P, McVeigh ER. Minimizing RF Heating of Conducting Wires in MRI. Magn Reson Med 2007;58:1028–1034.

(35) Overall WR, Pauly JM, Stang PP, Scott GC. Ensuring Safety of Implanted Devices Under MRI Using Reversed RF Polarization. Magn Reson Med 2010;64:823–833.

(36) Zanchi MG, Venook R, Pauly JM, Scott GC. An Optically Coupled System for Quantitative Monitoring of MRI-Induced RF Currents Into Long Conductors. Ieee T Med Imaging 2010;29:169–178.

(37) Gudino N, Sonmez M, Yao Z, Baig T, Nielles-Vallespin S, Faranesh AZ, Lederman RJ, Martens M, Balaban RS, Hansen MS, Griswold MA. Parallel Transmit Excitation at 1.5 T Based on the Minimization of a Driving Function for Device Heating. Medical physics 2015;42:359–371.

(38) Etezadi-Amoli M, Stang P, Kerr A, Pauly J, Scott G. Controlling Radiofrequency-Induced Currents in Guidewires Using Parallel Transmit. Magn Reson Med 2015;74:1790–1802.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)