Xinzeng Wang1, Crystal E. Harrison1, Yogesh K. Mariappan2, Karthik Gopalakrishnan2, Avneesh Chhabra1,3, Robert E. Lenkinski1,3, and Ananth J. Madhuranthakam1,3
1Radiology, UT Southwestern Medical Center, Dallas, TX, United States, 2Philips Innovation Campus, Philips Healthcare, Bangalore, India, 3Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, TX, United States
Synopsis
Volumetric Brachial Plexus
imaging at 3T often suffers from incomplete fat suppression and reduced SNR with
standard STIR and SPAIR due to increased B1 and B0 inhomogeneities. Dual-echo
Dixon TSE has been shown to achieve uniform fat suppression without increasing
total scan time or SNR penalty by acquiring two echoes in the same repetition.
In this work, we compared 3D dual-echo Dixon TSE against current standard of
care 3D STIR and 3D SPAIR for brachial plexus imaging with respect to fat
suppression, blood suppression, nerve visualization and SNR at 3T. Overall, the
3D dual-echo Dixon TSE showed significantly improved performance.Introduction
3D
magnetic resonance neurography (MRN) enables multiplanar coverage and
characterization of the peripheral nerves. 3D Short tau inversion recovery
(STIR) or spectral adiabatic inversion recovery (SPAIR) are typically used to
suppress fat and improve the conspicuity of the nerves. However, considerable B1
and B0 inhomogeneities in difficult regions such as brachial plexus limit the
performance of these techniques, resulting in inhomogeneous fat suppression and
reduced SNR, particularly on 3T scanners. Recently introduced Dixon method with
3D TSE achieves homogenous fat suppression with high SNR, often at the expense
of scan time
[1,2]. However, the scan time can be made equivalent to 3D STIR or
SPAIR by acquiring both echoes in the same repetition using dual-echo Dixon TSE
[3].
The
purpose of this work was to compare the performance of 3D dual-echo Dixon TSE against
current standard of care 3D STIR and SPAIR in brachial plexus of volunteers and
patients at 3T using similar acquisition times.
Methods
All experiments were performed on a 3T Ingenia
scanner (Philips Healthcare, The Netherlands). Dual-echo Dixon TSE was designed
to acquire both in-phase and out-of-phase images in the same repetition. 3D
STIR and 3D dual-echo Dixon TSE were performed on the brachial plexus of 3 healthy
volunteers and 4 patients with IRB approval and written informed consent. 3D
SPAIR TSE was performed on 4 normal volunteers, but was not performed subsequently
on the patients due to incomplete fat suppression observed in the preliminary
studies. The dual-echo Dixon images (Δt=1.1 ms) were acquired using a modulated
refocusing flip angle 3D TSE with the following parameters: α
min=20
o, α
center=100
o and α
max=120
o; TR = 2000ms; TE = 107ms; FOV = 320×321×123mm;
resolution = 1.4×1.4×1.4mm (reconstructed to isotropic 0.7 mm); SENSE
= 2.4 (phase encoding direction) and total scan time = 6 minutes (coronal
orientation). 3D STIR and 3D SPAIR were acquired using the same parameters
except for a TR of 3000ms to increase the SNR; TE of 68ms; and TI of 240ms. All images were qualitatively
evaluated by a musculoskeletal radiologist (AC) with 19 years of experience.
The images were scored on a three-point grading scale. Visualization of the
nerves, fat suppression: 0 – poor and non-diagnostic; 1 – moderate and
non-uniform; and 2 – excellent and uniform. Blood suppression: 0 – unsuppressed
and non-diagnostic; 1 – unsuppressed but diagnostic; and 2 – suppressed and
diagnostic. These scores were analyzed with a Wilcoxon signed-rank test. Additionally,
quantitative signal intensities were measured on the dorsal nerve root ganglion
and C6 nerve. The SNR was calculated as the ratio of the mean to the standard
deviation and the measurements were compared using a paired Student’s t-test.
Results
Figure 1 shows targeted maximum intensity
projections (MIP) of a representative volunteer comparing 3D STIR, 3D SPAIR and
water-only images from 3D dual-echo Dixon TSE. The conspicuity of the nerves is
relatively poor on 3D STIR (fig.1a) due to reduced SNR combined with
non-uniform fat suppression due to its sensitivity to both B1 and B0
inhomogeneities. The conspicuity of the nerves is improved in 3D SPAIR (fig.1b), but still suffers from incomplete fat suppression in the areas of
increased B0 inhomogeneity (dashed arrow in fig.1b). The 3D dual-echo Dixon
TSE (fig.1c) achieves uniform fat suppression increasing the conspicuity of
the nerves throughout the volume. The brachial plexus images of a patient with stretch
injury of the left upper trunk, axillary and suprascapular nerve are shown in
figure 2, where the B1 inhomogeneity caused significant shading artifacts on
the right side in 3D STIR (fig.2a). The
uniform fat suppression achieved with 3D dual-echo Dixon TSE (fig.2b) provided
enhanced visualization of the entire brachial plexus as well as the lesion. The fat suppression, arterial and venous suppression
and nerve visualization scores across all subjects (3 normal volunteers and 4
patients) are summarized in figure 3a. The fat suppression and nerve
visualization with 3D dual-echo Dixon TSE were significantly higher than 3D
STIR (P<0.05), while the arterial and venous suppression were better, but
not statistically significant. Figure 3b
shows statistically significant SNR increase with 3D dual-echo Dixon TSE
compared to 3D STIR (P<0.001).
Discussion
Compared to 3D STIR and SPAIR, 3D dual-echo
Dixon TSE sequence provided robust fat suppression with superior SNR and better
nerve visualization in similar acquisition times. The modulated refocusing flip
angle 3D TSE sequence with low refocusing flip angles has the intrinsic ability
to suppress vessels increasing the conspicuity of the nerve visualization.
Conclusion
3D dual-echo Dixon TSE sequence can be used for
optimal brachial plexus MR Neurography for selective nerve visualization at 3T.
Acknowledgements
No acknowledgement found.References
1) Reeder SB et. al. JMRI 24 (2006). 2)
Shankaranarayanan A et. al. ISMRM, 19 (2011); 3) Wang XZ et. al. ISMRM 23 (2015).