Barbara Cervantes1, Qinwei Zhang2, Kim van de Ven3, Hendrik Kooijman4, Ernst Rummeny1, Axel Haase5, Gustav J Strijkers2, Jan S Kirschke6, Aart J Nederveen2, and Dimitrios C Karampinos1
1Diagnostic and Interventional Radiology, Technische Universität München, Munich, Germany, 2Radiology, Academic Medical Center, Amsterdam, Netherlands, 3Philips Healthcare, Best, Netherlands, 4Philips Healthcare, Hamburg, Germany, 5Zentralinstitut für Medizintechnik, Garching, Germany, 6Neuroradiology, Technische Universität München, Munich, Germany
Synopsis
Quantitative MRI is becoming a promising tool in
the assessment of peripheral nerve pathologies and anomalies. Peripheral
neuropathy is frequently accompanied by neuropathic changes, which can be quantified
with diffusion tensor imaging (DTI). Given the small sizes and oblique
geometries of many peripheral nerves, peripheral-nerve DTI requires an acquisition
method that can provide high-resolution, distortion-free images in acceptable
clinical scanning times. The present work demonstrates isotropic- and
sub-millimeter-resolution, artifact-free DTI of the nerves in the lower
extremity using flow-compensated diffusion-prepared 3D turbo spin echo (TSE).Purpose
MR neurography is arising as a reliable method in the
diagnosis of nerve disease [1]. Quantitative MRN has been shown to be more
precise than conventional T2 imaging for detecting nerve injuries [2].
Diffusion tensor imaging (DTI) can provide precise information about the
microstructure of nerve fibers [3,4] and has been shown to have the capability
to detect disease in human nerves [5]. In high-resolution imaging, conventional
2D single-shot echo planar imaging (SS-EPI) becomes highly susceptible to
distortion artifacts from eddy currents and off-resonance effects [6]. Among
the existing alternatives to SS-EPI, diffusion-prepared multi-shot turbo spin
echo (TSE) can minimize geometric distortions and chemical shift artifacts [7].
Diffusion-prepared multi-shot TSE alone is highly sensitive to phase errors
that originate from eddy currents and motion [8].
Phase errors induce a modulation of magnitude
after the tip-up pulse [9,10,11]. Therefore, phase errors need to be considered if
using multi-shot TSE for quantitative diffusion imaging. The present work
introduces a flow-compensated diffusion-prepared 3D TSE sequence and
demonstrates high-resolution DTI of the nerves in the lower extremity.
Methods
Flow-compensated diffusion-prepared sequence: A
sequence combining a flow-compensated diffusion-preparation with a 3D TSE
readout was developed by adding dephasing and rephasing gradients during the
preparation and in the TSE readout
to reduce sensitivity to phase errors [9,10,11] (Fig.1).
SNR efficiency optimization: Refocusing angle schemes and
corresponding signal of nerve (T1=1000,T2=75ms) were simulated using extended
phase graphs (EPG) for 3D TSE for variable echo train length and TR. T1 and T2
relaxation effects were included during the refocusing pulse train and T1
relaxation was considered between shots. SNR and SNR efficiency values were
computed as: $$$\text{SNR}=s_{0}\,\Delta\text{x}\Delta\text{y}\Delta\text{z}\sqrt{\Delta
t\,N_{x}N_{y}N_{z}}$$$ and $$$\text{SNR}_{\text{eff}}=\text{SNR}\sqrt{\text{t}_\text{scan}}$$$,
where s0 is the signal at k=0, $$$\Delta\text{x}$$$, $$$\Delta\text{y}$$$
and $$$\Delta\text{z}$$$ are the acquisition voxel sizes, Nx, Ny and
Nz are the matrix dimensions, $$$\Delta\text{t}$$$ is the sampling interval and
tscan is the scan time [12].
In vivo measurements: DTI of the knee of two healthy
volunteers was conducted using a 16-channel knee coil on a 3T Philips system with
the developed sequence. 1) A sagittal acquisition was performed on one subject
and 2) an axial acquisition on the second subject. Readout parameters: 1) FOV=160×127×100$$$\text{mm}^{3}$$$,
acquisition voxel=1.7×1.7×1.7$$$\text{mm}^{3}$$$, reconstruction voxel=0.5×0.5×0.85$$$\text{mm}^{3}$$$,
TR/TE=1700/19ms, TSEfactor=60, averages=2, duration=15m33s. 2) FOV=140×140×52$$$\text{mm}^{3}$$$,
acquisition voxel=0.7×0.7×8 $$$\text{mm}^{3}$$$, reconstruction voxel=0.49×0.49×0.4
$$$\text{mm}^{3}$$$, TR/TE=1650/21ms, TSEfactor=40, averages=2, duration=5m28s.
DTI: b=0 and b=600 in six directions were acquired with TEprep=60ms. An axial T2-weighted
mDIXON TSE scan with an acquisition voxel=0.3×0.38×4$$$\text{mm}^{3}$$$ was
acquired as an anatomical reference.
Post-processing: A coronal reformat was obtained
from the sagittal dataset. Iso-diffusion-weighted images (iso-DWIs) were
generated. Diffusion tensors were computed using linear fitting and DTI
parameters were obtained from the derived eigenvalues. Mean values of mean
diffusivity (MD) and fractional anisotropy (FA) were measured. Projections of
the primary eigenvectors were generated for visualization purposes.
Results
Optimization of SNR
eff yielded a TSE factor of 40 and a TR
of 1650ms for the axial protocol (Fig.2) and a TSE factor of 60 and TR of 1700ms
for the sagittal protocol (not shown). Iso-DWIs show sagittal and coronal views
of the long axis of the tibial nerve (Fig.3, left). Mean values for the tibial
nerve in a region above the knee were found to be MD=$$$(1.1\pm0.1)\times10^{-9}$$$,FA=$$$0.57\pm0.06$$$. MD and FA maps of the nerve show
little spatial variation along the course of the nerve (Fig.3, center and
right). The primary eigenvector projections in Fig.4 follow the course of the
nerve along its long axis. Axial iso-DWIs and MD maps with sub-millimeter
in-plane resolution (Fig.5) show the sub-fascicular structure within the tibial
and peroneal nerves.
Discussion & Conclusion
Simulation results show that a compromise between SNR and
scan time can be found with considerations of relaxation and refocusing angle
modulation in 3D TSE. SNR efficiency becomes important in quantitative scans
with intrinsically low SNR and long acquisition times, since high noise levels
can result in the incorrect estimation of quantitative values and long scan
times can exceed the clinical limits. In
vivo results demonstrate that the developed method can generate high-SNR,
high-resolution iso-DWIs in the extremities free of eddy-current, off-resonance
and chemical-shift artifacts in acceptable scanning times. Measured MD and FA and
eigenvectors for the tibial nerve are in agreement with reported values [12]
and therefore indicate that the proposed method minimizes phase errors. Axial
acquisitions with sub-millimeter in-plane resolution show artifact-free
iso-DWIs and MD maps with resolved nerve fascicles. The present work therefore
introduces a multi-shot method that is capable of accurate, high-resolution
quantitative diffusion imaging of peripheral nerves and that is robust to many
of the problems commonly encountered in other DWI sequences.
Acknowledgements
The present work was supported by Philips Healthcare.References
[1] Mathys, Muscle Nerve 48:889, 2013
[2] Bendszus,
Nat Clin Pract Neurol 1:45, 2005
[3] Alexander, Magn Reson Med 44:283, 2000
[4]
Mori, Neuron 51:527, 2006
[5] Kakuda Neurorad 53:955, 2011
[6] Karampinos,
NMR Biomed 26:630, 2013
[7] Xie, J Cardiovasc Magn Reson 16:67, 2014
[8]
Cervantes, Proc. ISMRM 2015, p. 101
[9] Alsop, Magn Reson Med 38:527, 1997
[10]
Sinha, J Magn Reson Imaging 6:657, 1996
[11] Jeong, Magn Reson Med 56:1173,
2006
[12] Simon, J Magn Reson Imaging doi:10.1002/jmri.25056, 2015