Meredith Taylor1, Haonan Wang1, Antony JR Palmer2, Andrew J Carr2, Sion Glyn-Jones2, Daniel Park2, and Neal K Bangerter1
1Electrical Engineering, Brigham Young University, Provo, UT, United States, 2Nuffield Department of Orthopaedics, Rheumatology, and Musculoskeletal Sciences, University of Oxford, Oxford, United Kingdom
Synopsis
In this study,
we (1) implemented a two-acquisition 3D phase-cycled bSSFP protocol at 7 Tesla
that achieves 0.31mm isotropic resolution in under 9 minutes of scan time, (2)
implemented a 3D DESS protocol at 7 Tesla that achieves 0.36mm isotropic
resolution in just under 7 minutes, (3) performed a contrast optimization to
identify flip angles that maximize both cartilage/muscle and cartilage/synovial
fluid contrast, and (3) compared to a 3D SPACE acquisition at 7T that achieves
0.55mm isotropic resolution in a scan time of 11:37.Introduction:
7 Tesla whole body
MRI systems equipped with knee and extremity coils are seeing increasing
application in musculoskeletal MRI. The
higher magnetic polarization yields an SNR increase that can be exploited to
achieve higher resolution 3D scans than practical in clinically acceptable scan
times at 1.5 and 3 Tesla. These very
high-resolution 3D scans can potentially be used to make even more precise
measurements of cartilage morphology.
High SNR combined with high contrast between cartilage and adjacent
musculoskeletal tissues (such as synovial fluid and muscle) can also
potentially improve semi-automated and automated segmentation techniques,
helping reduce the need for painstaking (and sometimes error-prone) manual
segmentations.
Balanced SSFP
(bSSFP), DESS, and 3D SPACE sequences are all common for morphological
evaluation of cartilage at 1.5 and 3 Tesla, but they have not yet been commonly
used at 7T. Balanced SSFP further
suffers from sensitivity to off-resonance, typically resulting in bands of
signal null across the image (“banding artifact”). High field strengths compound the banding
artifact problem with bSSFP, making some kind of banding artifact removal
essential for application of bSSFP at 7T.
One of the most common techniques for banding artifact removal is the
acquisition of multiple phase-cycled bSSFP acquisitions, followed by a root
sum-of-squares (SOS) combination of the individual phase-cycled images [1].
In this study,
we (1) implemented a two-acquisition 3D phase-cycled bSSFP protocol at 7 Tesla
that achieves 0.31mm isotropic resolution in under 9 minutes of scan time, (2)
implemented a 3D DESS protocol at 7 Tesla that achieves 0.36mm isotropic
resolution in just under 7 minutes, (3) performed a contrast optimization to
identify flip angles that maximize both cartilage/muscle and cartilage/synovial
fluid contrast, and (3) compared to a 3D SPACE acquisition at 7T that achieves
0.55mm isotropic resolution in a scan time of 11:37.
Methods:
Phase-Cycled bSSFP Protocol:
Two phase-cycled acquisitions were performed followed by a root
sum-of-squares combination with the following parameters: TR/TE = 4/35/1.88ms, 25 degree flip angle
(max due to SAR constraints), 0.31mm isotropic resolution over a 15.8cm x
15.8cm x 12cm FOV, sagittal acquisition slab, total scan time = 8:50.
DESS Protocol:
The 3D DESS protocol utilized the following parameters: TR/TE = 8.19/2.26ms, 25 degree flip angle
(max due to SAR constraints), 0.36mm isotropic resolution over a 15.8cm x
15.8cm x 12cm FOV, sagittal acquisition slab, total scan time = 6:54.
Contrast Optimization:
Signal equations for bSSFP and DESS were used to perform a numerical
optimization of cartilage/muscle contrast-to-noise ratio (CNR) and
cartilage/synovial fluid CNR as a function of flip angle. Actual DESS scans were also performed across
a range of flip angles, and the CNRs measured from the scans for both to
compare with the theoretical predictions.
All scans were performed on
a 7 Tesla Siemens whole body scanner equipped with a 32-channel knee coil. The phase-cycled bSSFP and DESS scans were
then compared with a 3D SPACE protocol that achieved 0.55mm isotropic
resolution over a similar FOV in a total scan time of 11:37.
Results:
Results from the
numerical contrast optimization are shown in Figure 1. The measured DESS CNR values are in excellent
agreement with those predicted by the numerical simulation up to a flip angle
of 25 degrees. (Flip angles above 25
degrees could not be validated due to SAR constraints.) Cartilage/fluid CNR is optimized for DESS at
~25 degrees and for bSSFP at ~35 degrees, so a 25 degree flip angle (the
maximum achievable due to SAR constraints) was used achieve the highest
cartilage/fluid CNR possible.
The phase-cycled
bSSFP protocol at 0.31mm resolution yielded exquisite results. Native sagittal images along with axial and
coronal reformats are shown in Figure 2.
A comparison between the 3D phase-cycled bSSFP images, the 3D DESS
images, and the 3D SPACE images is shown in Figure 3. All of the sequences achieved beautiful high
resolution images, although the bSSFP images provided the best combination of
high resolution (0.31mm isotropic) and excellent cartilage/synovial fluid
contrast at a scan time of under 9 minutes.
Finally, both the SNR
(normalized by voxel volume) and SNR efficiency (SNR normalized by square root
of scan time) were measured in cartilage in the patella, the anterior femur,
and the posterior femur for both the bSSFP and DESS scans. These are reported in Table 1.
Discussion:
All
three of the tested protocols provide excellent delineation of cartilage
morphology, high resolutions and SNR, and short scan times. While the 3D DESS scan provides superior
cartilage SNR than bSSFP at the flip angle achievable, bSSFP provides superior
cartilage/fluid contrast while achieving higher resolution at acceptable SNR.
Acknowledgements
No acknowledgement found.References
[1] N. K.
Bangerter, et al., MRM, vol.
51, no. 5, pp. 1038–1047, 2004.
[2] Hanicke, W., and Vogel, H. U. 2003. An
analytical solution for the SSFP signal in MRI. Magn. Reson. Med. Phys. 17:1004-1010.