Effects of the acquisition window length on articular cartilage sodium MR image quality
Alireza Akbari1,2 and Michael Noseworthy1,2,3

1School of Biomedical Engineering, McMaster University, Hamilton, ON, Canada, 2Imaging Research Centre, St Joseph's Healthcare, Hamilton, ON, Canada, 3Electrical and Computer Engineering, McMaster University, Hamilton, ON, Canada

Synopsis

We assessed articular cartilage sodium MR image SNR and blurring as a function of acquisition window length. Increasing the acquisition window dramatically improved the SNR with minimal reduction in image quality as shown by minimal blurring.

Introduction

Sodium magnetic resonance imaging (MRI) is challenging because it is a signal-to-noise-ration (SNR)-limited technique. It also suffers from image blurring due to its short biexponential transverse relaxation, T2. One way to improve SNR, would be lengthening the acquisition window while keeping the overall acquisition time the same. SNR is proportional to the square root of the acquisition window, Tacq (1). Therefore, if for example, Tacq were doubled then SNR would theoretically increase by $$$\sqrt[]{2}$$$. However, the SNR gain would be at the cost of increased image blurring. One sodium MRI application is in assessment of articular cartilage as studies have shown direct correlation between osteoarthritis (OA) severity and articular cartilage sodium content (2-4). Enhancing articular cartilage sodium MR image quality would lead to more accurate sodium quantification and a better assessment of OA severity. The goal of this work was to investigate the balance between sodium image SNR gain and deleterious image blurring when lengthening Tacq.

Methods

A density-adapted 3-dimentional projection reconstruction (DA-3DPR) sequence (5) was implemented on a GE MR750 3T (General Electric Healthcare, Milwaukee WI) and home-built 12-rung 23Na split design 18cm diameter birdcage transmit/receive RF coil was used for 23Na image acquisition. Sodium DA-3DPR datasets corresponding to Tacq = 4, 8, 12, 16, 20, and 25ms were acquired with the following imaging parameters : TE/TR = 0.25/100ms, 11310 projections, isotropic resolution/FOV = 3mm/18cm, and averaging = 2. All images were reconstructed into 60 slices of 540x540 (i.e. 0.3mm in-plane resolution x 3mm thick) using a non-uniform fast Fourier transform (NUFFT) (6). To quantify the effect of readout window length on blurring, the full-width-at-half-maximum (FWHM) of the slice profile across patellar cartilage was measured for each Tacq. The SNR was calculated in the patellar cartilage according to Madelin et al.(7).

Results

In vivo axial views of the knee with various Tacq durations are shown in figure 1. The background noise was noted to markedly decrease as Tacq was increased, with only a slight increase in image blurring. The SNR measurement for patellar cartilage is shown in figure 2 for each Tacq. The SNR increased considerably with respect to the Tacq=4ms data as indicated by the numbers above each bar. The blurring caused by Tacq lengthening in the patellar cartilage is shown in Figure 3. The FWHM as a measure of blurring was observed to slightly broaden by a maximum of 1mm in the articular cartilage as Tacq was increased from 4 to 25ms.

Discussion

Our results indicate that SNR is doubled when Tacq is increased from 4 to 25 ms. This is a significant improvement in SNR. However, one would expect this increase to be greater, i.e. $$$\sqrt[]{25 ms/4 ms} = 2.5$$$, or 250% increase. One reason SNR imrovement was less than theoretical may be because articular cartilage has short T2 relaxation (10-30ms) (8), that leads to signal loss with longer acquisition window lengths. In fact, SNR is proportional to the square root of the acquisition window in the absence of T2 effects. Image blurring is expected due to biexponential T2 relaxation; however, the FWHM measurements indicate small image blurring. This is in agreement with the measured amount of blurring for a DA-3DPR acquisition scheme (5) even when the acquisition window is extended beyond the T2s.

Conclusion

This work demonstrates that the benefits of increasing Tacq in terms of SNR gain outweighs the minimal adverse effects of blurring on articular cartilage sodium MR image quality using DA-3DPR.

Acknowledgements

No acknowledgement found.

References

1. Haacke EM, Brown RW, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design. New York: A John Wiley and Sons; 1999. 2. Lesperance LM, Gray ML, Burstein D. Determination of fixed charge density in cartilage using nuclear magnetic resonance. J Orthop Res. 1992;10(30):1-13. 3. Reddy R, Insko EK, Noyszewski E a, Dandora R, Kneeland JB, Leigh JS. Sodium MRI of human articular cartilage in vivo. Magn Reson Med. 1998;39(5):697-701. 4. Shapiro EM, Borthakur A, Gougoutas A, Reddy R. 23Na MRI accurately measures fixed charge density in articular cartilage. Magn Reson Med. 2002;47(2):284-291. 5. Nagel AM, Laun FB, Weber MA, Matthies C, Semmler W, Schad LR. Sodium MRI using a density-adapted 3D radial acquisition technique. Magn Reson Med. 2009;62(6):1565-1573. 6. Fessler J a. On NUFFT-based gridding for non-Cartesian MRI. J Magn Reson. 2007;188(2):191-195. 7. Madelin G, Babb JS, Xia D, Chang G, Jerschow A, Regatte RR. Reproducibility and repeatability of quantitative sodium magnetic resonance imaging in vivo in articular cartilage at 3 T and 7 T. Magn Reson Med. 2012;68(3):841-849. 8. Madelin G, Lee J, Regatte RR, Jerschow A. Sodium MRI: methods and applications. Prog Nucl Magn Reson Spectrosc. 2014;79:14-47.

Figures

Figure 1. Sodium 23Na images of a healthy human knee obtained with various acquisition window lengths while other imaging parameters were kept constant. SNR was improved significantly as Tacq was increased, at the cost of only slight blurring.

Figure 2. SNR measurements in patellar cartilage from the images shown in figure 1. The SNR gain relative to Tacq = 4ms is indicated by the number above each bar for each acquisition window length. Maximal SNR improvement of 200% was seen with Tacq = 25ms.

Figure 3. Patellar cartilage slice profiles obtained from images shown in figure 1. Full-width-at-half-maximum (FWHM) as a measure of image blurring in patellar cartilage was calculated for each image. The FWHM broadened by only by 3 pixels (i.e. 1mm) when Tacq was increased from 4 to 25ms. The location of the slice profile is indicated by the orange line drawn in the image at the top-left corner of the figure.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
4503