Synopsis
An isotropic three-dimensional
T2 mapping technique with adiabatic T2 magnetization preparation for T2 relaxation
time quantification was implemented and validated at 3T in phantoms and 6 healthy volunteers.Introduction
Osteoarthritis (OA)
is a degenerative disease of the joint, with an estimated prevalence of 18% in
the United States [1], of which progressive cartilage disease is considered to
be the most important characteristic. Magnetic resonance imaging (MRI) is
considered the gold-standard for the non-invasive assessment of cartilage.
Since cartilage degeneration is not reversible, a prompt diagnosis of local damage
that precedes cartilage tissue loss is of crucial importance to guide early-stage treatment.
The cartilage T
2 relaxation time is considered as a biomarker for early
cartilage damage and degeneration, as it has been shown to be correlated with changes
in several attributes of the extracellular matrix, such as its anisotropy, water
content and collagen concentration [2, 3]. Most studies have so far been based
on two-dimensional (2D)
T
2 mapping sequences. However, due to the complex three-dimensional
(3D) structure of cartilage, an isotropic 3D acquisition is
required I) to perform a high-resolution quantitative analysis of cartilage T
2
value in 3D that can be reformatted in an arbitrary direction, and II) to
obtain a higher signal-to-noise (SNR) ratio compared to multiple 2D acquisitions
and thus a higher T
2 precision in T
2 determination. In this study, we therefore
propose an isotropic 3D T
2 mapping technique that consists of
separate sequential T
2-prepared 3D gradient recalled echo (GRE)
images, which allows for co-registration of these separate images to account
for motion. This protocol was optimized through Bloch equation simulations and a
phantom study, and validated in healthy volunteers.
Methods
The pulse sequence consists of
a segmented centric GRE acquisition with repetition time TR=4.2ms, echo time
TE=1.76ms, 100 k-space lines per segment, segmental acquisition time=700ms and
incremental adiabatic T2 preparation (T2prep) durations
of 0, 23, 38 and 53ms. Numerical simulations of the Bloch equations were
performed using Matlab (The Mathworks,
Natick, MA). The goals of these simulations were:
I) to find the optimal sequence parameters (RF excitation angle, number of excitations per segment) to maximize
the signal per unit of time and II) to determine an empirical T2-fitting
offset to correct for T1 relaxation [4] in the following equation:
$$S(TE_{T2prep})=S_{0}\cdot e^{\frac{TE_{T2prep}}{T_{2}}}+\delta,$$
where S0 is
the initial signal when T2prep= 0ms and δ is the empirical offset
determined through Bloch equation simulations. To calculate the T2
map, the signal S was fitted with the above-mentioned
empirical equation in each pixel. To assess the accuracy of the fixed empirical offset, the
T2 map of a custom-built agar-NiCl2 phantom with 3
compartments was acquired with the following additional parameters: adiabatic T2prep, voxel volume= 0.82×0.82×0.82mm3,
matrix size= 208×216×112, total scan time=9min20s, RF excitation angle 15°, 2×GRAPPA
acceleration. Spin-echo (SE) relaxation time mapping was then used as a reference
standard to determine the T1 (1072ms) and T2 (33.2ms)
values of the phantom at 3T (Prisma, Siemens Healthcare). Finally, a
T2 map of the knee was acquired in 6 healthy adult volunteers (4 right and 2 left, 3
male and 3 female with an average age of 29.5±4.8 years) with the same
parameters as in the phantom experiments. Rigid 3D image registration followed
by pixel-wise T2-mapping was performed in Matlab and the average T2
in regions of interest was calculated. T2 values were reported as
average ± standard deviation over all volunteers.
Results
and Discussion
Bloch equation
simulations resulted in an empirical offset of 0.003, which was confirmed in
the phantom study (Fig. 1). The phantom study furthermore demonstrated a homogenous T
2
distribution (T2=34.3±1.4ms) in the compartment with the relaxation times that
approximated cartilage (Fig. 1, outer ring), which agreed with the spin-echo
reference T
2 value (33.2±0.4ms). Similarly, the T
2 maps
in the volunteers resulted in T
2 values of 28.7±4.3ms and 28±2.5ms
for tibial and femoral cartilage, respectively (sagittal plane, Fig. 2a, standard deviation within the region of interest less
than 15%), which agrees well with the values reported in literature [5]. Moreover, the isotropic 3D T
2 mapping
technique allowed the quantification of the patellar cartilage T
2
value in the axial plane (37.5±2.6ms, Fig. 2b, standard deviation within the region of interest less
than 15%). The tibial and femoral cartilage layers could consistently be visualized
and T
2 value could also be quantified in the coronal plane in all
volunteers (Fig. 3, 4).
Conclusions
This study successfully demonstrated that an isotropic 3D T
2
mapping is feasible for knee cartilage characterization and results in precise
T
2 values.
Acknowledgements
No acknowledgement found.References
[1] Lawrence et al., Arthritis Rheum
41:778-799, 1998
[2] Mosher et al.,
Semin Muscoloskelet Radiol 8(4):355-368,
2004
[3] Kijowski et al., Radiology 267(2):503–513,
2013
[4] van Heeswijk et al., Circ Cardiovasc Imaging 5:1231-1239, 2012
[5] Stahl et al., Eur
Radiol 19:132-143, 2009.