S Sivaram Kaushik1, Abhishiek Sharma2, Rajeev Mannem1, Cathy Marszalkowski1, Scott Rand1, Dennis Maiman2, and Kevin M Koch1
1Radiology, Medical College of Wisconsin, Milwaukee, WI, United States, 2Neurosurgery, Medical College of Wisconsin, Milwaukee, WI, United States
Synopsis
Failed back surgery syndrome (FBSS) is a commonly encountered clinical condition for which the root cause often remains elusive after lengthy clinical assessment. While MRI provides great potential utility for uncovering the causes of failed back surgery syndrome, it also is confounded by metal-induced artifacts. This study explores the benefits of a prospectively calibrated 3D-MSI metal artifact reduction technique in assessing instrumented FBSS. Residual artifacts and clinical diagnostic impact were assessed on a cohort of symptomatic FBSS subjects. The results of the study indicate diagnostic advantages using calibrated 3D-MSI, even in the presence of lower-susceptibility titanium spinal hardware.
Spinal fusion is the most
commonly employed surgical procedure for managing conditions of the spine.
Unfortunately, these procedures often do not produce expected results. The
failure of surgical intervention in the spine to relieve symptoms often is
described by the generalized term ‚”Failed Back Surgery Syndrome” (FBSS)
[1].
Fusion procedures often
implant metallic hardware, which generate MRI artifacts. In practice, spine
surgeons often install titanium alloyed hardware components. While titanium
implants induce far less artifact than other metallic alloys, easily
discernible artifacts remain in conventional MRI. To elucidate the nature of these subtle
artifacts, both conventional metal artifact reduction sequences (MARS) and
advanced 3D multi-spectral imaging (3D-MSI) techniques were applied and
analyzed on a cohort of symptomatic FBSS subjects.
Commercially available
3D-MSI sequences are not yet equipped with prospective capabilities to adjust
to different implant scenarios. The
lower susceptibility titanium alloys utilized in spinal fusion devices are
often imaged with 3D-MSI parameters that are‚ ”overkill”‚ for the problem at
hand. Therefore, in this study, a
prospective calibration scan (~1 min) was utilized to calibrate each 3D-MSI
sequence to the implanted fusion device [2,3].
This spectral calibration provides crucial efficiency improvements to
3D-MSI, which further allows for higher resolution and higher SNR spinal
imaging, often with shorter acquisition times than conventional MARS
techniques.
Methods
MRI were collected from a
GE Healthcare Optima 450w GEM 1.5T system. GEM variable receiver array
configurations were utilized for all spine acquisitions. A cohort of
symptomatic FBSS subjects (N=21) provided written consent for an imaging
research protocol approved by the MCW Institutional Review Board. Subjects were
imaged with a routine MARS acquisition protocol as well as a series of T1w and
T2w prospectively calibrated MAVRIC SL 3D-MSI images. The most commonly fused segments were
cervical and lumbar, with 10 subjects each. One subject underwent evaluation
after thoracic.
To quantify artifact reduction, a subset of subjects
was identified from the study cohort (N=6, 3 L-Spine, 3 C-Spine) who previously
had CT imaging of their spinal fusion.
Bone CT images of titanium fusion devices have limited beam scattering
artifacts and allow for easy identification of implant interfaces. After affine co-registration of
CT/MARS/3D-MSI image volumes, implant or artifact volumes were identified using
a combination of manual segmentation (MRI) and seeded threshold-based region
growing (CT). The difference of the
implant/artifact volumes between the MRI and the CT provided an estimate of
residual artifact volume. Multiple-observer
measurements were used to estimate the error bars in these volume-estimates. To asses the diagnostic impact of calibrated
3D-MSI, the MARS and 3D-MSI image sets were reviewed and compared with clinical
records to independently identify complications in each image set.
Results
Figure 1 provides
volumetric renderings of a segmented lumbar spine fusion device. Quantitatively, the subtraction of measured
and rendered volumes in this case yielded 3.9±1.6 cm3 of residual artifact in
the MAVRIC SL 3D-MSI images and 58.0±12.2cm3 of residual artifact in the MARS
images. It is important to note that the dominant volume displacement is occurring
in the vertical dimension, which is the slice-selective dimension in this axial
MARS acquisition.
Figure
2 shows that for all subjects studied, the measured artifact volume in the MARS
image was significantly higher than in the MAVRIC images. The 3D-MSI residual
artifact was subtle and within the measurement error. For the subject shown in Figure 1, Figure 3
shows a 3D rendering of the segmented nerve-roots from the MAVRIC and MARS
images. It is clear where the expanded artifact volume of the MARS image
interferes with the nerve root and can impact clinical assessments.
Clinically, 10 of the
subjects exhibited normal post-surgical findings on plain-film radiography, CT,
and MRI. Imaging findings identified adjacent segment disease in eight
subjects; in which it was at least partially identified on both MARS and
3D-MSI. Of these cases, 3D-MSI uniquely identified foraminal pathology in three
subjects, nerve root edema in one subject, epidural fibrosis in one subject,
and a malpositioned screw in one subject.
Figure 4 provides sample
images from this MARS preliminary clinical analysis. Foraminal pathology
adjacent to a pedicle screw is highlighted in the indicated zoomed region. In
the MAVRIC SL image, a compression of the nerve root is identified (dashed
white arrow). This compression is obscured by artifact (dashed white arrow) in
the corresponding MARS image.
Conclusion
A prospectively calibrated 3D-MSI technique
provided clear visible and quantifiable improvements and multiple diagnostic
benefits on a cohort of symptomatic titanium alloy instrumented FBSS subjects.
Acknowledgements
Advancing a Healthier
Wisconsin Research and Education Fund, #5520357
GE Healthcare Technical
Development Grant Support
References
[1]
F. Shafaie, C. Bundschuh, and J. Jinkins. The post-therapeutic lumbosacral spine. In J. Jinkins, editor,
Post-therapeutic neurodiagnostic imaging, pages 223–243. Lippincott-Raven, 1997.
[2] K.M Koch,
Metal Implant-Induced Spectral Range Optimization using Rapid 3D-MSI Calibration Scans, In: Proc. ISMRM, 2015, 2511
[3] S. Kaushik, C Marszalkowski, K.M Koch, External Calibration of the Spectral Coverage for 3D Multi-Spectral Magnetic Resonance Imaging, Magn. Reson. Med (in press)