Barbara Cervantes1, Dominik Weidlich1, Hendrik Kooijman2, Ernst Rummeny1, Axel Haase3, Jan S Kirschke4, and Dimitrios C Karampinos1
1Diagnostic and Interventional Radiology, Technische Universität München, Munich, Germany, 2Philips Healthcare, Hamburg, Germany, 3Zentralinstitut für Medizintechnik, Garching, Germany, 4Neuroradiology, Technische Universität München, Munich, Germany
Synopsis
Diffusion weighted imaging (DWI) can describe the
microstructure of nerve fibers and is therefore a powerful tool in the study of
neuropathic changes. Diffusion imaging in the body faces challenges with motion
and B0 and B1 field effects. Particularly, imaging of nerves has resolution
requirements that can limit the performance of the acquisition method. The
lumbar plexus is a particularly difficult region to image due to the complexity
of the nerve geometry there and its susceptibility to motion and field
inhomogeneity. Therefore, careful considerations need to be taken in the design
of a diffusion-prepared sequence for imaging the lumbar plexus. The present
work proposes an adiabatic diffusion preparation module in combination with a
3D TSE readout to achieve high-resolution DWI of the lumbar plexus.Purpose
Diffusion weighted imaging (DWI) can describe
the microstructure of nerve fibers [1,2] and is therefore a very powerful tool
in the study of neuropathic changes [3].
The lumbar plexus is a challenging region due to the need for high
isotropic resolution to depict the oblique geometry of the lumbosacral nerve
roots. T2-weighted imaging with 3D turbo spin echo (TSE) has been shown to meet
these resolution requirements [4] but diffusion imaging is challenged by
motion and field inhomogeneity present in this region.
Diffusion-prepared sequences have been recently applied
in a range of organs and tissues in order to achieve high-resolution
diffusion-weighted imaging (DWI) or diffusion tensor imaging (DTI), free of the
sensitivity to off-resonance effects associated with the use of single-shot EPI
[5-7].
Diffusion preparation is usually combined with
velocity compensation in order to reduce the sensitivity of the preparation to
motion effects [7].
Effects from eddy currents also need to be
considered when designing a diffusion preparation [8].
Another important consideration when applying
diffusion in the body is the sensitivity of the preparation module to transmit
B1 inhomogeneity effects. Recent works focusing on non-diffusion weighted T2
preparation have proposed the use of MLEV or hyperbolic secant pulses in order
reduce the sensitivity of the preparation to transmit B1 effects, but without completely
removing them [9,10].
An adiabatic T2 preparation using a BIR-4 pulse
has been shown to be robust to both B1 and B0 effects when applied for T2
preparation [10,11].
The present work proposes a BIR-4-based
diffusion preparation module in combination with a 3D TSE readout to achieve
high-resolution DWI of the lumbar plexus.
Methods
Sequence
development: Two alternative diffusion
preparation schemes were used: one employing the adiabatic BIR-4 RF pulse (Fig.
1a) and one employing double-refocused hyperbolic secant pulses (DRHS) (Fig. 1b).
Diffusion gradients were added around the refocusing pulses of the preparation
in order to achieve velocity compensation (first gradient moment nulling). The
diffusion preparation was combined with a 3D TSE readout. A dephasing gradient was
placed before signal restoration and later balanced in the 3D TSE readout to
reduce the effects of phase from eddy currents and motion.
In vivo measurements:
DWI of the lumbar plexus of two healthy volunteers was conducted using a
16-channel torso coil and the posterior coil on a 3T Philips system
with the developed sequence. Coronal acquisitions were performed with the
sequence parameters: FOV=380×380×80$$$\text{mm}^{3}$$$, acquisition voxel=2×2×2$$$\text{mm}^{3}$$$,
reconstruction voxel=0.7×0.7×1$$$\text{mm}^{3}$$$, TR/TE=1500/24ms, TSEfactor=60,
averages=2, duration=15m50s. DWI: b=0 and b=400 in six directions were acquired
with TEprep=58ms. For one subject, the DWI acquisition was performed with both
the DRHS and the BIR-4 diffusion preparation.
Post-processing: Iso-diffusion-weighted images (iso-DWIs) were
generated from the b=600 data.
Results
In
comparison to DRHS (Fig.1b), the BIR-4 implementation yielded higher signal in
the acquisitions without diffusion in regions with high B1 inhomogeneity
(Fig.2). In those regions, DWI with DRHS showed severe signal loss in the individual
DWI images (Fig.3) that resulted in the loss of small features especially in
the upper region of the coronally acquired images. In the same regions, DWI
with BIR-4 showed betterimage quality. High-resolution
iso-DWIs (Fig.4) showed high signal of small nerve roots and nerve branches
with an isotropic resolution of 2 mm.
Discussion & Conclusion
Diffusion
imaging in the presence of motion and eddy currents without careful
consideration in the design of the acquisition method faces many challenges.
Flow compensation is necessary to minimize phase accumulation from first-order
motion that leads to signal loss. Eddy currents inevitably occur in the
implementation of diffusion gradients and can result large spatially varying
signal losses. Even when the effects of motion and of eddy currents are
minimized, sensitivity to transmit field inhomogeneity results in the inaccurate
restoration of diffusion-prepared signal and consequently in signal loss. DWI
of the lumbar plexus is susceptible to motion and to B1 inhomogeneity particularly
in the higher regions of the FOV in the vicinity of the abdominal organs.
In vivo results demonstrate that the
proposed method is robust to motion and eddy current effects.
In vivo results also show that the
adiabatic BIR-4 pulse design improves the performance of the diffusion
preparation in regions with high B1 inhomogeneity in the cases with and without
diffusion. In the presence of diffusion, where further signal losses are
expected, the complete loss of small nerve structures results in regions with
high B1 inhomogeneity when the diffusion preparation uses hyperbolic secant
pulses. Therefore, the present work presents a diffusion preparation that can
achieve high isotropic-resolution DWI of the lumbar plexus in the presence of
motion, eddy currents and transmit B1 effects.
Acknowledgements
The present work was supported by Philips Healthcare.References
[1]
Alexander, Magn Reson Med 44:283, 2000
[2]
Mori, Neuron 51:527, 2006
[3]
[2] Mathys, Muscle Nerve 48:889, 2013
[4]
Cervantes, J Magn Reson Imaging, doi: 10.1002/jmri.25076
[5]
Xie, J Cardiovasc Magn Reson 16:67, 2014
[6]
Nguyen, Magn Reson Med 72:1257, 2014
[7]
Nguyen, Magn Reson Med, doi: 10.1002/mrm.25609
[8]
Sinha, J Magn Reson Imaging 6:657, 1996
[9]
Nezafat, Magn Reson Med 55:858, 2006
[10]
Jenista, Magn Reson Med 70:1360, 2013
[11]
Nezafat, Magn Reson Med 61:1326, 2009