Matthew Robert Smith1, Jin Jung Kweon2, Eun Sang Choi2, Curtis Wiens1, Nathan Artz3, and Scott B Reeder1,4
1Radiology, University of Wisconsin, Madison, WI, United States, 2Florida State University, Tallahassee, FL, United States, 3St. Jude's Children's Hospital, Memphis, TN, United States, 4Medical Physics, University of Wisconsin, Madison, WI, United States
Synopsis
Despite important recent
development, further progress of MR imaging around metallic prostheses is
dependent on the ability to model the field perturbations surrounding the
prostheses. These calculations require knowledge of the magnetic susceptibility
of the metal which is not reported by the manufacturers. The purpose of this
work was to estimate the magnetic susceptibility of commonly implanted metal
alloys by measuring the magnetic moment across a range of clinical field
strengths (0-5 Tesla) using a SQUID (Superconducting QUantum Interference
Device) magnetometer. Linearity of the susceptibility across field strengths was
also assessed.Purpose
There has been
significant advancement in the ability to perform MR imaging around metallic
prostheses
1-3. Accurate modeling of the large field perturbations
induced by the metal is essential for the design and optimization of MRI acquisition
and reconstruction strategies near metallic implants
4. These field
perturbations are affected by composition (magnetic susceptibility of implant),
geometry, orientation relative to the main magnetic field, and field strength. While
the magnetic susceptibility values of pure metals are well understood
5,
manufacturers do not provide susceptibility values for specific commercial implantable
alloys. The purpose of this work was to estimate the magnetic susceptibility
of commonly implanted metal alloys by measuring the magnetic moment across a
range of clinical field strengths (0-5 Tesla) using a SQUID (Superconducting
QUantum Interference Device) magnetometer.
Methods
Representative samples of popular total hip and
knee metallic prostheses (Figure 1, Table 1) were obtained from an orthopedic
manufacturer (Zimmer Biomet, Warsaw, ID) and sent to the National High Magnetic
Field Laboratory (Tallahassee, FL). Sample preparation included cutting to a
desired shape (~6x6x6 mm3) by
electrical discharge machining (samples 1-5) or by milling (sample 6) and washing with dilute acid to remove any ferromagnetic contamination
introduced during the cutting process. Samples 1-5 were treated in a 5% HCl
solution for 1 hour to remove any residue. Sample 5 was additionally treated
for 24 hours. Sample 6 was washed with ethanol. The rough and porous surface of
Samples 2 and 3 required multiple washing and drying cycles to achieve a steady
mass over time (washing: deionized water, 2 hours; drying: dry box, 1 week).
Sample masses were recorded before and after the washing
process and no significant changes were observed.
The measurements of the magnetic moments were
conducted using a Quantum Design, MPMS-5 SQUID magnetometer. The magnetic
moment of each sample was measured at 310 K (normal body temperature) as the
field strength swept from 0 to 5 Tesla and back down to -0.2 Tesla. Measurements
were then normalized by the respective masses. Mass susceptibility was taken as
the slope of the averages of the linear fit done for both up and down sweeps. The
following equation allowed conversion of mass susceptibility (χg) to
volume
ppm susceptibility: χppm = χg * ρ * 4π * 106
(multiplied by: ρ (density) for volume
susceptibility, 4π for SI quantity, 106 for ppm).
Results
The magnetic moments of the measured alloys and
plastic demonstrated a linear dependence up to 5 T (Figure 2). The
magnetization of the Co-Cr-Mo alloys compared to the Ti-6Al-4V alloys were
approximately double. Sample 1, 2, 3, 4, and 5 demonstrated paramagnetism
(positive linear) while Sample 6 (UHMWPE) demonstrated diamagnetism (negative
linear). The magnetic susceptibility values for the samples are shown in Table
1.
Discussion and Conclusion
This work estimated the
magnetic susceptibility of several commonly implanted metallic alloys
across a range of field strengths using a SQUID magnetometer. The reported
susceptibility values were noted to differ with values used previously in the
literature that were obtained by estimation with MRI
6-7. Furthermore,
a commonly made assumption regarding the linearity of magnetization across
clinically relevant field strengths was experimentally validated.
Acknowledgements
The authors are grateful to the National High
Magnetic Field Laboratory for their assistance in susceptibility measurements
and Zimmer Biomet for providing the samples. The authors also wish to
acknowledge support from the NIH (UL1TR00427), as well as GE Healthcare for
their support.References
1. Koch KM, Lorbiecki JE,
Hinks RS, King KF. A multispectral three-dimensional acquisition technique for
imaging near metal implants. Magn Reson Med 61, 381-390, doi:10.1002/mrm.21856
(2009).
2. Lu W, Pauly KB, Gold
GE, Pauly JM, Hargreaves BA. SEMAC: Slice Encoding for Metal Artifact
Correction in MRI. Magn Reson Med 62, 66-76, doi:10.1002/mrm.21967 (2009).
3. Artz NS, Hernando D, Taviani V, Samsonov A,
Brittain JH, Reeder SB. Spectrally resolved fully phase-encoded
three-dimensional fast spin-echo imaging. Magn Reson Med, doi:10.1002/mrm.24704
(2013).
4. Smith MR, Artz NS, Wiens C, Hernando D,
Reeder SB. Characterizing the limits of MRI near metallic prostheses. Magn
Reson Med, doi:10.1002/mrm.25540 (2014).
5. Schenck JF. The role of magnetic
susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the
first and second kinds. Med Phys 23, 815-850 (1996).
6. Koch KM, Hinks RS. in
Proceedings of the 16th Annual Meeting of the ISMRM 1180 (Toronto, Canada,
2008).
7. Bartusek K, Dokoupil Z,
Gescheidtova E. Magnetic field mapping around metal implants using an
asymmetric spin-echo MRI sequence. Measurement Science and Technology 17, 3293
(2006).