Improved gradient warping correction for large field-of-view imaging and application to radiation therapy planning
Paul T. Weavers1, Shengzhen Tao1, Kiaran McGee1, Joshua Trzasko1, Yunhong Shu1, Erik Tryggestad2, Ken-Pin Hwang3, Seung-Kyun Lee4, Thomas KF Foo4, and Matt Bernstein1

1Mayo Clinic, Rochester, MN, United States, 2Radiation Oncology, Mayo Clinic, Rochester, MN, United States, 3MD Anderson, Houston, TX, United States, 4GE Global Research, Niskayuna, NY, United States

Synopsis

Radiation therapy, especially proton beam therapy requires exacting spatial accuracy to deliver a sterilizing dose of ionizing radiation to the target volume with confidence. The superior soft tissue contrast afforded by MRI vs. CT has increased interest in using MRI for treatment planning. However, gradient non-linearities reduce the spatial accuracy of MRI. We have developed a fiducial phantom based calibration procedure to map these gradient nonlinearities on a system-specific basis and generate up to 9th order spherical harmonic coefficients for correction. These coefficients show improved spatial accuracy vs. standard 5th order, especially at distances >400mm from magnet and gradient isocenter.

Background and Significance

Magnetic resonance imaging (MRI) based radiation treatment planning has gained wide clinical acceptance due to the superior soft tissue contrast when compared to x-ray computed tomography (CT). MRI is particularly attractive for proton beam therapy because of the need to clearly demarcate the target from surrounding radiosensitive organs at risk. There is also interest in using MR information only for both photon and proton radiation therapy planning [1], as well as for attenuation correction for PET/MR systems[2,3]. The main technical challenge in utilizing this information lies in determining the accurate and precise geometric precision of the tissues that are to be targeted or avoided during radiation therapy.

A potential source of error that can affect the geometric precision of an MR image is spatial distortion induced by spatial gradient nonlinearity (GNL). Imaging gradients are designed to vary linearly across the field-of-view (FOV), but the Maxwell equations and engineering limitations dictate that there will be some residual GNL, usually manifesting as a “pincushion” type of effect. This effect is more pronounced at the edges of the FOV when a large FOV is prescribed, with position errors of up to 5 mm from truth potentially impairing the accuracy of treatment planning. Previous work targeting a small FOV imaging device utilized up to 10th order GNL correction with spherical harmonic coefficients obtained via an iterative calibration procedure and use of the ADNI [4]. This work extends that idea to whole body systems with large FOV imaging.

Methods

This work utilizes a large fiducial phantom [5] as shown in Figure 1. The phantom consists of a matrix of MRI visible paintballs uniformly distributed throughout. Each paintball is separated by 2.5 cm from its neighbor in each direction. Images of the phantom were acquired with a fast gradient-echo sequence with 61.4 cm square FOV, 1.2 mm isotropic voxels and ±125 kHz receiver bandwidth utilizing the body coil for transmit and receive on a 3T scanner (GE 750w, DV25.0, GE Healthcare, Waukesha WI) with a 50cm imaging diameter spherical volume (DSV) and 70cm patient bore aperture.

A 3D Hough-transform based position tracking software was developed in house to match both the MRI data with a CT reference scan to map GNL-induced spatial distortions. These positions were compared, and used to generate high order (up to 9th) corrections utilizing the iterative gradient nonlinearity estimation framework [4]. We utilized this framework to generate 3rd, 5th, 7th, and 9th order coefficients, and then applied these coefficients to reconstruct a series of corrected images. Finally, the error of each fiducial marker interrogated by the fitting routine was plotted to generate position dependent accuracy figures.

Results

Figure 2 shows spatial accuracy in the form of RMSE vs. the CT position reference for a 280mm diameter spherical volume (DSV), comparing the fitted 5th, 7th, and 9th order corrections. Additionally, Figure 3 shows the fitted 5,7,9th order corrections for a 500mm DSV.

Figure 4 shows axial and coronal planes at approximately 14 cm inferior to isocenter. Note the better depiction of planar “paintballs” in the 7th and 9th order case, vs 5th order.

Figure 5 shows the RMSE comparison as a function of spherical harmonic model order used. Note the 9th order showing minimum RMSE.

Discussion

We have demonstrated an imaging-based method for characterizing the system specific gradient non-linearity of whole-body scanners with large FOV acquisition capabilities. This reduces the RMSE error substantially from standard 5th order GNL correction, potentially allowing for a higher level of confidence in utilizing MR images for radiation therapy planning. Note in Figure 4 the increased background noise to the identically window/leveled images with the higher order correction. This artifact is due to an increase compensation term from “spreading out” image intensity in the image-based correction for gradient nonlinearity. An integrated gradient nonlinearity correction and image reconstruction [6] is expected to further reduce this effect.

The outcome of this work is an expected increase in geometric accuracy and precision confidence in proton beam therapy planning. In the future, multiple systems will be calibrated using the described procedure to account for system-specific deviations from the standard electromagnetic field simulated coefficients in order to examine the significance of system-by-system deviation.

Acknowledgements

This work was supported in part by NIH grant RO1EB010065.

References

[1] Paulson ES, Erickson B, Schultz C, Allen Li X. Comprehensive MRI simulation methodology using a dedicated MRI scanner in radiation oncology for external beam radiation treatment planning. Med Phys 2015;42.

[2] Hofmann M, Steinke F, Scheel V, Charpiat G, Farquhar J, Aschoff P, et al. A Novel Approach Combining Pattern Recognition and Atlas Registration. J Nucl Med 2008:1875–83. doi:10.2967/jnumed.107.049353.

[3] Schulz V, Torres-Espallardo I, Renisch S, Hu Z, Ojha N, Börnert P, et al. Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data. Eur J Nucl Med Mol Imaging 2011;38:138–52. doi:10.1007/s00259-010-1603-1.

[4] Trzasko J, Tao S, Gunter J, Shu Y, Huston III J, Bernstein MA. Phantom-Based Iterative Estimation of MRI Gradient Nonlinearity. ISMRM Annu. Meet., Toronto: 2015, p. 3735.

[5] Hwang K, McKinnon G, Lorbieki J, Maier J. Spatial Accuracy Quantification of an MR System. AAPM Annu. Meet., 2012, p. WE – G – 217A – 6.

[6] Tao S, Trzasko JD, Shu Y, Huston III J, Bernstein MA. Integrated image reconstruction and gradient nonlinearity correction. Magn Reson Med 2015;74:1019–31. doi:10.1002/mrm.25487.

Figures

Figure 1: Fiducial phantom similar to that used in this work. Approx. 55cm S/I extent, containing water-filled capsules uniformly spaced 25 mm apart.

Figure 2: Color-coded root-sum-squared distance from CT position reference for a 280mm diameter spherical volume of the "paintball" phantom. Note the maximum on this scale is 1.4 mm error. Sub figure a, b, c are the 5th,7th,9th order image-based fitting results.

Figure 3: 500 mm DSV results, note larger scale in RSS error reported. Increased accuracy observed for 9th order correction.

Figure 4: Axial and coronal views of the fiducial phantom approximately 14 cm from isocenter. Note improved capture of the planar nature of the phantom with the higher order spherical harmonic coefficients.


Figure 5: RMSE vs. the CT reference data for various diameter spherical volumes (DSV) interrogated by the fitting procedure. Note the RMSE increases as more fiducial markers that are far from isocenter are included, but the 9th order correction still shows improvement in these large FOV experiments.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
4307