Paul T. Weavers1, Shengzhen Tao1, Kiaran McGee1, Joshua Trzasko1, Yunhong Shu1, Erik Tryggestad2, Ken-Pin Hwang3, Seung-Kyun Lee4, Thomas KF Foo4, and Matt Bernstein1
1Mayo Clinic, Rochester, MN, United States, 2Radiation Oncology, Mayo Clinic, Rochester, MN, United States, 3MD Anderson, Houston, TX, United States, 4GE Global Research, Niskayuna, NY, United States
Synopsis
Radiation therapy, especially proton beam therapy requires
exacting spatial accuracy to deliver a sterilizing dose of ionizing radiation to
the target volume with confidence. The
superior soft tissue contrast afforded by MRI vs. CT has increased interest in
using MRI for treatment planning.
However, gradient non-linearities reduce the spatial accuracy of
MRI. We have developed a fiducial
phantom based calibration procedure to map these gradient nonlinearities on a
system-specific basis and generate up to 9th order spherical
harmonic coefficients for correction.
These coefficients show improved spatial accuracy vs. standard 5th
order, especially at distances >400mm from magnet and gradient isocenter.Background and Significance
Magnetic resonance imaging (MRI) based radiation treatment
planning has gained wide clinical acceptance due to the superior soft tissue
contrast when compared to x-ray computed tomography (CT). MRI is particularly attractive for proton
beam therapy because of the need to clearly demarcate the target from
surrounding radiosensitive organs at risk. There is also interest in using MR
information only for both photon and proton radiation therapy planning [1], as well as for attenuation
correction for PET/MR systems[2,3]. The main technical challenge in utilizing
this information lies in determining the accurate and precise geometric
precision of the tissues that are to be targeted or avoided during radiation
therapy.
A potential source of error that can affect the geometric
precision of an MR image is spatial distortion induced by spatial gradient
nonlinearity (GNL). Imaging gradients
are designed to vary linearly across the field-of-view (FOV), but the Maxwell
equations and engineering limitations dictate that there will be some residual
GNL, usually manifesting as a “pincushion” type of effect. This effect is more pronounced at the edges
of the FOV when a large FOV is prescribed, with position errors of up to 5 mm
from truth potentially impairing the accuracy of treatment planning. Previous work targeting a small FOV imaging
device utilized up to 10th order GNL correction with spherical
harmonic coefficients obtained via an iterative calibration procedure and use
of the ADNI [4]. This work extends that idea to whole body
systems with large FOV imaging.
Methods
This work utilizes a large fiducial phantom [5] as shown in Figure 1. The phantom consists of a matrix of MRI
visible paintballs uniformly distributed throughout. Each paintball is separated by 2.5 cm from its neighbor in each
direction. Images of the phantom were
acquired with a fast gradient-echo sequence with 61.4 cm square FOV, 1.2 mm
isotropic voxels and ±125 kHz receiver bandwidth utilizing the body coil for
transmit and receive on a 3T scanner (GE 750w, DV25.0, GE Healthcare, Waukesha
WI) with a 50cm imaging diameter
spherical volume (DSV) and 70cm patient bore aperture.
A 3D Hough-transform based position tracking software was
developed in house to match both the MRI data with a CT reference scan to map GNL-induced
spatial distortions. These positions
were compared, and used to generate high order (up to 9th)
corrections utilizing the iterative gradient nonlinearity estimation framework [4]. We utilized this framework to generate 3rd,
5th, 7th, and 9th order coefficients, and then
applied these coefficients to reconstruct a series of corrected images. Finally, the error of each fiducial marker
interrogated by the fitting routine was plotted to generate position dependent
accuracy figures.
Results
Figure 2 shows spatial accuracy in the form of RMSE vs. the
CT position reference for a 280mm diameter spherical volume (DSV), comparing
the fitted 5th, 7th, and 9th order
corrections. Additionally, Figure 3
shows the fitted 5,7,9th order corrections for a 500mm DSV.
Figure 4 shows axial and coronal planes at approximately 14
cm inferior to isocenter. Note the
better depiction of planar “paintballs” in the 7th and 9th
order case, vs 5th order.
Figure 5 shows the RMSE comparison as a function of
spherical harmonic model order used. Note
the 9th order showing minimum RMSE.
Discussion
We have demonstrated an imaging-based method for
characterizing the system specific gradient non-linearity of whole-body
scanners with large FOV acquisition capabilities. This reduces the RMSE error substantially
from standard 5th order GNL correction, potentially allowing for a
higher level of confidence in utilizing MR images for radiation therapy
planning. Note in Figure 4 the increased
background noise to the identically window/leveled images with the higher order
correction. This artifact is due to an
increase compensation term from “spreading out” image intensity in the
image-based correction for gradient nonlinearity. An integrated gradient nonlinearity
correction and image reconstruction [6] is expected to further reduce
this effect.
The outcome of this work is an expected increase in
geometric accuracy and precision confidence in proton beam therapy planning. In the future, multiple systems will be
calibrated using the described procedure to account for system-specific
deviations from the standard electromagnetic field simulated coefficients in
order to examine the significance of system-by-system deviation.
Acknowledgements
This work was supported in part by NIH grant
RO1EB010065.References
[1] Paulson ES, Erickson B, Schultz C,
Allen Li X. Comprehensive MRI simulation methodology using a dedicated MRI
scanner in radiation oncology for external beam radiation treatment planning.
Med Phys 2015;42.
[2] Hofmann M, Steinke F, Scheel V,
Charpiat G, Farquhar J, Aschoff P, et al. A Novel Approach Combining Pattern
Recognition and Atlas Registration. J Nucl Med 2008:1875–83.
doi:10.2967/jnumed.107.049353.
[3] Schulz V, Torres-Espallardo I, Renisch
S, Hu Z, Ojha N, Börnert P, et al. Automatic, three-segment, MR-based
attenuation correction for whole-body PET/MR data. Eur J Nucl Med Mol Imaging
2011;38:138–52. doi:10.1007/s00259-010-1603-1.
[4] Trzasko J, Tao S, Gunter J, Shu Y,
Huston III J, Bernstein MA. Phantom-Based Iterative Estimation of MRI Gradient
Nonlinearity. ISMRM Annu. Meet., Toronto: 2015, p. 3735.
[5] Hwang K, McKinnon G, Lorbieki J, Maier
J. Spatial Accuracy Quantification of an MR System. AAPM Annu. Meet., 2012, p.
WE – G – 217A – 6.
[6] Tao S, Trzasko JD, Shu Y, Huston III J,
Bernstein MA. Integrated image reconstruction and gradient nonlinearity
correction. Magn Reson Med 2015;74:1019–31. doi:10.1002/mrm.25487.