Jiazheng Wang1, Bing Wu2, and Yongchuan Lai3
1Radiology Department, University of Cambridge, Cambridge, United Kingdom, 2GE healthcare MR Research China, Beijing, China, People's Republic of, 3GE heathcare China, Beijing, China, People's Republic of
Synopsis
Usually aper-volume or even
per-slice basis reference scan is usually needed for correcting the phase
inconsistency between odd and even shots attributed to eddy current in EPI. The phase correction is usually first order. In this work, the concept of multiplxed SENSE (MUSE) is extended for EPI N/2 ghost calibration. Introduction
Combined
with the use of parallel imaging, single shot EPI is the most common choice for
diffusion weighted imaging and functional MRI studies. A per-volume or even
per-slice basis reference scan is usually needed for correcting the phase
inconsistency between odd and even shots attributed to eddy current. Not only
the reference scan is time consuming compared to the EPI acquisition itself,
the phase variation may be temporally varying that calls for dynamic reference
scan. Furthermore, the reference scan is limited to first order correction,
whereas EPI acquisitions are often confounded with high order phase
contamination in cases such as oblique scans. In this work, we extend the
concept of multiplexed SENSE [1,2] and further incorporate the between shot
phase inconsistency as part of coil sensitivity weighting and to eliminate such
inconsistency simultaneously with the SENSE correction.
Method
The
proposed method is illustrated in flow chart in Fig.1. Firstly, for an R-acceleration factor SENSE acquisition (R
could be 1), the acquired k-space is split into two parts, namely the even
k-space lines and the odd k-space lines. Conventional SENSE reconstruction is
first performed separately on the two k-spaces leading to two reconstructed N/2
ghosted images. Then the residual phases in the two reconstructed images are
collected and treated as the phase error attributed to eddy currents. A
low-pass filtering is then performed on the two phase maps to remove noises or
phase attributed to tissue susceptibility. The obtained phase maps are then
used to update the original coil sensitivity map to form a set of composite
coil sensitivity maps that contains N/2 phase errors. Notice at this point, the
reconstruction matrix is doubled (Equ.(2), compared to regular SENSE recon in
Equ.(1)) as we are dealing with twice as many virtual coils, however given the effective
acceleration factor R the resulting g-factor penalty remains the same. In this
way, the phase variations caused by eddy currents are intrinsically eliminated
in the reconstructed image. In addition, this multiplexed SNESE phase
correction may deal with higher order phase variations as determined by the
form of the phase map filtering performed.
[MxR] =
[MxR] [Rx1] (1)
[2MxR]
= [2MxR] [Rx1]
(2)
(1)
regular SENSE recon
(2)
multiplexed SENSE recon
M:
number of receiver coils
R :
SENSE acceleration factor
Experiment
Both
phantom and in-vivo experiments were performed to verify the performance of the
proposed method using a whole body 3T scanner (GE Discovery 750) equipped with
an 8-cahnnel brain coil. A uniform sphere phantom (for enhanced visibility of
ghosting) and healthy volunteer (with consent obtained) were scanned with a
single shot SE-EPI and DW-EPI (b=600) sequence with a SENSE acceleration factor
of 2. Acquisition matrix size used was 128x128 and FOV was 22cm. Oblique plane
scan was performed. Standard reference scan corrections and the proposed
multiplexed SENSE corrections were performed to correct for the N/2 ghost on
the same data set retrospectively.
Results
Fig. 2 compares the N/2 ghost corrections
using standard reference scan and the proposed multiplexed SENSE method. It is
seen that in the phantom results, the residual ghosting is very obvious using
the standard reference scan correction due to eddy current contamination at
oblique plane. It is also noticed that the ghosting was 4 fold despite a SENSE
factor of 2, as the N/2 ghost is further multiplexed with SENSE aliasing. The
ghosting is much reduced in the images obtained using the multiplexed SENSE. A
stronger level of residual ghosting is also seen with the SE-EPI in-vivo data
with standard correction, whereas the residual ghosting is much less visible in
the DW-EPI acquisition as the diffusion gradient helps to weaken the ghosting
signal. It is also seen that there is no additional SNR penalty using the
proposed method as expected.
Discussion and conclusion
In
this work, the N/2 ghost in EPI is corrected based on multiplexed SNESE that
treats the phase variations between the even-odd lines as part of the coil
sensitivity. Oblique plane phantom and in-vivo experiment showed much reduced residual
ghosting level with the proposed method as compared to standard reference scan,
which is potentially due to the high order phase variations that are
uncorrected in the latter case. This method may also be beneficial for
multi-phase acquisitions such as fMRI where the phase inconsistency could be
temporally varying.
Acknowledgements
No acknowledgement found.References
[1]
N. Chen et al, Neuroimage, 2013
[2]
HC. Chang et al, Neuroimage, 2015