Maysam M Jafar1, Christopher E Dean1, Malcolm J Birch1, and Marc E Miquel1
1Clinical Physics, Barts Health NHS Trust, City of London, United Kingdom
Synopsis
Major
hardware-related geometric distortions in MRI arise from gradient field
non-linearity and static field inhomogeneity. For an accurate mapping of
geometrical distortion in 3D, the number of control points must be sufficiently
large to provide a comprehensive mapping of the spatial variation of distortion
and the positional accuracy of these control points must be ensured. In this
study, the spatial accuracy of coordinates in 3D space is assessed across six
clinical MRI scanners at both 1.5T and 3T field strengths using a previously
reported 3D-printed grid phantom. This is a cost-effective approach to
determine the spatial accuracy of control points.Purpose
Magnetic resonance
imaging (MRI) offers superior soft-tissue contrast compared with computed
tomography (CT) but suffers from inherent geometric distortions. Major
hardware-related geometric distortions arise from gradient field non-linearity
and static field inhomogeneity. Geometric distortions occurring due to
non-linearity in the gradient fields outweigh those occurring due to static
field inhomogeneity as modern superconducting MRI systems are equipped with
active and passive shimming technologies. For an accurate mapping of
geometrical distortion in 3D, two conditions must be satisfied. First, the
number of control points must be sufficiently large to provide a comprehensive
mapping of the spatial variation of distortion and second the positional
accuracy of these control points must be ensured [1]. In this study, the
spatial accuracy of coordinates in
3D space is assessed using a previously reported 3D-printed grid phantom [2]
across six clinical MRI scanners at both 1.5T and 3T field strengths. This is
achieved by quantifying machine-related MR distortion by comparing the
locations of corresponding features in both MR and CT data sets.
Methods
A rigid
170mm×170mm×250mm mesh cuboid insert with
a 2mm wireframe was designed, 3D-printed and fixated inside a 240mmϕ × 250mm cylinder [2]. Off-the-shelf
mineral (baby) oil (Johnson&Johnson, NJ) was used to fill the phantom
container to avoid wavelength-induced artefacts [3,4]. Orthographic projections
of the 3D model (a,b) and printed insert (c,d) are shown in Fig.1.
Each
point of intersection between the mesh lines in 3D space is treated as a
control point with an approximate spacing of 20.0mm in each dimension. The
insert contains a total of 9×9×11 (891) control points. With
the axis of the cylinder aligned along the B0 field, MR images were
acquired using one Philips (Philips Healthcare, Best, the Netherlands) and five
Siemens (Siemens Healthcare, Erlangen, Germany) scanners using the body coil
and a 3D gradient-echo sequence. A single k-space
line was encoded per TR with a mean acquisition time of 25 minutes per
acquisition. Imaging parameters for the 6 clinical MRI scanners are shown in
Table1 (Philips [A: Achieva 3T], Siemens [B: Verio 3T, C: Avanto 1.5T, D:
Prisma 3T, E: Aera 1.5T, F: Aera 1.5T]). In order to define the true,
undistorted control point positions, a corresponding CT scan of the phantom was
generated using a GE LightspeedRT16 system (GE Healthcare, Milwaukee,WI, USA)
using a voxel size of 0.71mm×0.71mm in-plane and 1.25mm through-plane.
All of the 891 control points were
detected using 3D normalized cross correlation (NCC) using a MATLAB (The
Mathworks, Natick, MA) program [2] and were manually inspected. This was
achieved by defining two templates of 7×7×7 voxels extracted from the
iso-center of the CT and MR acquisitions. We assumed that this point in space
was free of distortions in the MR acquisition. Each template was then convolved
with its originating volume yielding a volume of correlation coefficients for
both the CT and MR acquisitions. Thresholding was then used to determine the
control points corresponding to the highest correlating points. Connected
components were then computed of the thresholded correlation coefficients and
the centroids of each of the control points were determined. A distortion-free distance
map was generated from CT. We used the Euclidean distance metric to determine
the distance between the two distance maps corresponding to MR and CT control
points. The distortion-free
distance map was compared to an artificially created volume in MATLAB, which
resembled the control point locations in the 3D-printed insert, and a mean
error of 0 was found.
Results
Measured distances between the control points in
the CT dataset were accurate to within 0.05mm, better than the manufacturer
stated value of 0.15mm. Overall mean error across the entire MR volume for each
of the 6 scanners employed in this study is shown in Table2. For each of the
six scanners, mean error across the volume was less than 2mm. Mean L2-norm
errors for the axial and sagittal control point planes is given in Fig.2 (a, b)
respectively. Distortion in the central planes for all scanners was less than
2mm. Three scanners (A, C, D) achieved an error of less than 1.6mm across the
volume.
Discussion
& Conclusions
Detecting spatial accuracy of coordinates in 3D space is
essential for determining geometric distortion to aid in MR/CT planning. Design
improvements can be made in the 3D-printed insert whereby a larger feature can
represent a control point. This will then alleviate the need for manual
inspection of the 3D normalized
cross correlation, as geometric distortions will often make thresholding of the
NCC coefficients difficult. Nevertheless, this is a cost-effective approach to
determine the spatial accuracy of control points.
Acknowledgements
No acknowledgement found.References
[1] Wang et
al., A novel phantom and method for comprehensive 3-dimensional measurement and
correction of geometric distortion in magnetic resonance imaging, Magnetic
Resonance Imaging. (2004), 22, 529-542
[2] Jafar et
al., A Regularly Structured 3D Printed Grid Phantom for Quantification of MRI
Image Distortion, ISMRM (2015), ePoster 3727
[3] Baldwin et
al., Characterization, prediction, and correction of geometric distortion in 3
T MR images, Medical Physics 34(2):388-399
[4] Price et
al., Quality Assurance Methods and Phantoms For Magnetic-resonance-imaging -
Report of AAPM Nuclear-magnetic-resonance Task Group No-1, Medical Physics
17:287-295