Saikat Sengupta1, David S Smith1, and Edward Brian Welch1
1Department of Radiology, Vanderbilt University, Nashville, TN, United States
Synopsis
In
this abstract, we present an approach for rapid mapping of the whole body B1
field at 3 Tesla. We use a combination of dual interleaved TR B1 mapping with continuously
moving table imaging to measure the B1 field in the whole body in 90 seconds. Considerable variation is observed in the B1 distribution in different sections
of the body. This measurement will serve as the basis for dynamic B1 field
shimming with the moving table for whole body imaging in future work.
PURPOSE
Many
clinical studies such as whole-body water/fat quantification
1, peripheral
vascular angiography
2, detection of tumor metastases
3, and multi-contrast
anatomical imaging
4 demand either whole-body coverage or a field of view
(FOV) larger than the uniform z extent of the scanner bore. Continuously Moving Table (CMT) MRI is an
imaging technique that enables rapid whole-body MR examination demanded in such
studies
5,6. In CMT MRI, data are acquired with concurrent uniform velocity z
direction motion of the patient table. However, unlike the alternative of multi-station
whole-body scanning, a limitation of CMT MRI is that the pre-scan RF power calibrations
are performed at one location of the body and
maintained throughout the scan for the full FOV. This results in non-optimal RF
transmit power settings in other sections of the body and RF shading artifacts
that vary across the length of the FOV. To mitigate these artifacts, a rapid
and robust measurement of the whole-body B
1+ field (or flip angle
α assuming proportionality) is necessary. In this
abstract, we present an approach for rapid whole-body measurement of
α that combines golden angle radial CMT MRI and dual TR actual flip angle imaging at 3 Tesla
7.
METHODS
Acquisition: Whole-body CMT MRI was implemented on a
Philips Achieva 3 Tesla scanner (Philips Healthcare, The Netherlands) with a
2-channel transmit and 2-channel receive body coil and table velocity of 20
mm/s. Data were acquired from an adult volunteer with a dual TR golden
angle radial imaging sequence with TR1 = 3.5 ms and TR2 =
17.5 ms. The other scan parameters were: zFOV = 1800 mm, in-plane FOV =
400 x 400 mm2, in-plane resolution = 1.56 x 1.56 mm2, TE
= 1.08 ms, excited slice thickness = 12 mm, flip angle of 40o and a total acquisition time of 90 seconds. The RF power calibration and other
preparation scans were performed at the level of a landmarked position at the abdomen.
Reconstruction: Complex image volumes of the two TRs
were reconstructed offline in Python (Anaconda version 3.4.2, TX) using
64 profiles per reconstructed axial slice. The reconstructed slice thickness was 25.6 mm (64 * 20 mm/s * 21 ms) with a slice separation of 1.56 mm matching the in-plane resolution. GA image reconstruction involved correction for k-space shifts and sampling density
prior to gridding, 2D FFT and roll-off compensation.
Flip
angle (α)
map estimation: Whole-body
α was estimated as described in Ref 7, i.e.,
$$\alpha = \arccos\left[\frac{(S_{2}/S_{1})(TR_{1}/TR_{2})-1}{(TR_{1}/TR_{2})-(S_{2}/S_{1})}\right]$$
where S1 and S2 were the magnitude intensity values from the two interleaved
scans. The flip angle map was estimated only for a masked region of interest
extracted by thresholding the TR1 image.
RESULTS
Figure 1 shows a coronal slice of the whole-body
TR
1 image and estimated flip angle map. Example axial slices are shown in
Figure 2 (anatomical images: a-c, flip angle maps d-f), illustrating the
variation of the B
1+ field along the different sections
of the body. RF shading artifacts are apparent in the anatomical slices, which
correlate spatially with the flip angle maps. While the prescribed flip angle
was 40
o, the estimated α in several locations of the body is seen to
deviate significantly from the nominal angle.
DISCUSSION
The
golden angle radial acquisition scheme allows flexible retrospective
combination of an arbitrary number of profiles to reconstruct slices (and hence
flip angle maps) without significant undersampling artifacts. This in turn
allows required flexibility in the choice of TR
1 and TR
2 . Our future work
will focus on understanding the effects of the moving table on the flip angle
measurements, establishing the fidelity of the flip angle maps and pursuing
other B
1 mapping methods such as dual angle
8 and Bloch-Siegert
9 . Rapid mapping the B
1+
field in the whole-body is the first step in mitigating the
intensity artifacts originating from transmit RF inhomogeneity and anatomical
location dependent miscalibration. With the estimated whole-body flip angle
maps, strategies for dynamic, table-position-specific RF shimming for
continuously moving table MRI can be devised based on traditional multichannel
RF shimming methods
10,11.
Acknowledgements
R21 DK096282 from NIDDK/NIH, UL1 TR000445 from
NCATS/NIH and NCI/NIH K25CA176219References
1. Berglund, J
et al, MRM 63. 1659 (2010) 2. Kruger, D
et al, MRM 47. 224 (2002) 3. Lauenstein TC. Radiology. 233. 139
(2004) 4.
Lauenstein TC.JMRI. 24 .489 (2006) 5. Börnert et al. JMRI 28. 1 (2008) 6. Sengupta,S et al, MRM (2014) 7. Yarnkh,V et al, MRM 57. 192 (2007) 8. Stollberger WP, et al MRM 35, 246
(1996) 9. Sacolick LI, et al MRM, 63
1315 (2010). 10. Katscher U, et al MRM 49, 144 (2003) 11. Mao W, et al MRM, 56 918 (2006).