Catherine J Moran1, Bragi Sveinsson1,2, Brady Quist2, Marcus T Alley1, Bruce Daniel1, and Brian A Hargreaves1
1Radiology, Stanford University, Stanford, CA, United States, 2Electrical Engineering, Stanford University, Stanford, CA, United States
Synopsis
The double echo steady state (DESS) acquisition has the potential to provide high-resolution and distortion-free T2 and diffusion-weighted images in the breast. Initial investigations of DESS in the breast have been limited by the presence of ghosting artifacts. Respiratory-induced B0 variation is one source of these artifacts. A method utilizing an in-vivo time-varying off-resonance estimate along with an image entropy metric to assess the level of artifact is described and investigated for correction of ghosting due to respiration-induced B0 variation in DESS in the breast.Introduction and Purpose
In applications
of the Double Echo Steady State (DESS) in the breast, phase errors due to
respiration-induced B0 variation result in ghosting artifacts that
degrade image quality and limit further clinical investigation of the method
[1]. Correction of these ghosting artifacts with a DC navigator has shown
promising but intermittent results [Figure 1, ref. 2]. In this work we describe a new method in
which a set of possible phase errors due to respiration-induced B0 variation
are simulated and an image entropy metric is utilized to assess the efficacy of
each set of phase errors to correct the ghosting artifacts.
Methods
Time Varying Off-resonance Estimate:
Previous
intermittent success using a DC navigator for correction of ghosting in DESS in
the breast indicates that certain DESS coil/echo DC navigator data effectively
captures phase variation due to respiration induced off-resonance variation (Figure 1). Actual
motion of the breast tissue is very limited in breast MRI particularly in the
medial regions of the breast due to the prone position of the patient. The
relative phase variation of DESS echo 1 navigator data from 4 medial channels
of the breast coil were averaged and used to estimate the off-resonance
variation per TR. The legitimacy of these estimates was verified by comparison
with previous measurements of respiration induced B0 variation in the breast.
Simulation of candidate phase
propagations:
This in-vivo time-varying off-resonance estimate can be utilized with known imaging
parameters (TR, TE, gradient area) and established tissue parameters in the
breast (T1,T2 and ADC) to simulate the expected phase errors in echo 1 and 2
of the DESS acquisition. However, both the amplitude of the off resonance variation as well as B1 can significantly vary across the breasts. To effectively capture these variations in our Bloch
simulations of phase error due to the estimated off resonance variation, we
simulated data for a range of off-resonance amplitudes (0.5, 1.0, and 2.0 x estimated
off-resonance amplitude) and flip angles (0.7, 1.0 and 1.4 x prescribed flip
angle). A total of 9 different
time-varying phase error estimates were simulated. Relative phase from these estimates was
rewound in the acquired k-space data providing 9 image datasets with varying
levels of artifact correction.
Image Entropy Metric for Assessment of
Artifact Reduction:
Respiration
induced B0 variation results in coherent ghosting artifacts across the breasts
as well as across the region with no tissue between the breasts in our axial
acquisition. Based on these
artifact characteristics, an image entropy metric [3] was utilized to
quantify the level of artifact for each corrected image. The metric was calculated on a
slice-by-slice basis for each echo/coil combination. An uncorrected dataset and
a set of images corrected directly from the in vivo DC navigator phase
measurements were also assessed with image entropy
metric. For each slice/coil/echo combination, the corrected image with minimum
image entropy was selected as that with lowest level of artifact and combined
to create a final set of corrected images. A flowchart summarizing the proposed method is shown in Figure 2.
Results
The acquired and simulated phase data from a
volunteer acquisition for the specified range of off-resonance
amplitude variations and flip angles is shown in Figure 3. Correction of the
acquired image data with the candidate simulated phase errors did result in
varying levels of artifact reduction and the image entropy metric effectively quantified
the level of artifact in the images (Figure 4). The proposed DC navigator and image entropy metric based
method reduced the presence of ghosting artifacts in both echoes of DESS in the
breast in comparison to direct application of phase errors in the acquired DC
navigator data (Figure 5).
Discussion and Conclusion
In this initial investigation, the proposed method is successful at
reducing ghosting artifact in both the first and second echo DESS breast
images. The various levels of
correction achieved with different simulated phase errors supports the use of a set of simulated phase errors and also demonstrates that the chosen
amplitude and flip angle ranges effectively capture the in-vivo variation. The
most tenuous aspect of this method is the choice of DC navigator data from which to estimate the off-resonance variation. Future work will include a more systematic method of
determining “good” navigator data on which to base the estimate. Finally, this
method does not provide a voxel-by-voxel off resonance correction but the improvement in image quality may still facilitate further investigation of this technique in the
clinic.
Acknowledgements
Research Support from GE Healthcare and NIH/NIBIB R01 EB009055-06, NIH/NIAMS R01 AR063643-01, NSF GRFP: DGE-114747.References
1. Granlund KL, et al., MRI. 2014; 32:330-41.
2. Moran CJ, et al., 23rd annual Meeting and Exhibition of the ISMRM, 2015, p. 498.
3. Atkinson D, et al., MRM. 1999; 41:163-170.