Niranchana Manivannan1, Bradley D. Clymer1, Anna Bratasz2,3, and Kimerly A. Powell2,4
1Department of Electrical and Computer Engineering, The Ohio State University, Columbus, OH, United States, 2Small Animal Imaging Shared Resources, The Ohio State University, Columbus, OH, United States, 3Davis Heart & Lung Research Institute, The Ohio State University, Columbus, OH, United States, 4Department of Biomedical Informatics, The Ohio State University, Columbus, OH, United States
Synopsis
In most small animal imaging studies both long axis (coronal or
sagittal) and short axis (axial) images of the region of interest are obtained.
The goal of this study is to explore whether combining two orthogonal views obtained
with different slice-select directions could reduce the motion artifacts and
improve image quality. The advantages of this method are that no a priori
knowledge or external hardware is needed and it doesn’t increase the acquisition
time. The results show that it is advantageous to use the available orthogonal
image(s) to improve the image quality by reducing ghosting artifacts caused by
motion.Introduction
The modulation in
k-space data which causes
the motion artifacts in MRI occurs only in one direction (i.e., the phase encode direction). Previous
works have shown that combining two MRI images acquired over the same
field-of-view (FOV) with their readout and phase encode directions interchanged
can reduce motion artifacts.
1,2 However,
these studies used the same slice-select directions. The concept of using two
MRI images acquired over the same region of interest using different
slice-select directions (i.e., orthogonal views) has not been explored. In most
small animal imaging studies both long axis (coronal or sagittal) and short
axis (axial) images of the region of interest are obtained. It can be useful to
study whether combining these orthogonal images will help in reducing the
motion artifacts. The goal of this study is to explore whether combining two
orthogonal views obtained with different slice-select directions could improve
the quality of the image by reducing artifacts due to motion. The advantages of this method are that no a priori
knowledge about the type of motion is required for their implementation and no
external hardware is needed and it doesn’t increase the acquisition time over
conventional scans. The evaluation of this motion correction algorithm was performed in phantom and
in vivo mouse images.
Methods
The long axis and short axis images have motion
artifacts in different (phase encoding) directions, and have different in-plane
FOVs and slice-select directions. The
flowchart for the implementation of the algorithm is shown in the Figure
1. Consider
axial image stack $$$I_a(X,Y,z)$$$ and coronal image stack $$$I_c(x,Y,Z)$$$ acquired orthogonally to each other. To
correct for motion in the axial view, the coronal view is upsampled along the
x-axis and the axial view is upsampled along the z-axis. $$$\hat{I_a}(X,Y,Z)$$$ and $$$\hat{I_c}(X,Y,Z)$$$ represents
the upsampled axial and coronal image stacks respectively. The complex
conjugate product is obtained using the two upsampled views. The square root of
the magnitude of the complex conjugate can then be downsampled in either
direction to obtain a motion corrected image.
Simulated motion in biological
phantom: A paraformaldehyde fixed
E17.5 embryo was used as a biological phantom as it possesses anatomic structures similar
to those observed in live animals but does not suffer from motion artifacts
observed for in vivo imaging. Coronal and axial views were acquired using T1-weighted
FLASH imaging sequence (TR=519.5ms, TE=4ms, FA=30.0, FOV =2.2*2.2cm, slice
thickness=0.26mm, matrix=512*512, navgs=1, acquisition time=3min, number of
contiguous slices=46) in 500 MHz 11.7T magnet (Bruker Biospin). Artificial motion as observed in in vivo case is simulated along the phase
encode direction in both the images.3 The bulk misalignment is simulated in these images by rotating the coronal image by 3°.4
In vivo experiment: For quantification
of fat volume, coronal and axial image stacks of the abdominal region were
acquired using respiratory triggered RARE T1 weighted sequence (TR=1200 ms, TE=7.5ms,
FA=180.0, FOV=3*3cm, matrix=256*256, slice thickness=0.75 mm, acquisition
time=12 min) in 400 MHz 9.4T magnet. Both qualitative (visual inspection)
and quantitative measures are used for the evaluation of the algorithm.
Results
and Discussion
Simulated
motion in biological phantom: In Figure 2a motion artifact caused by the
simulated motion is evident in spinal cord region of coronal image. Figure 2c shows
the result of motion correction algorithm which has less ghosting artifacts
when compared to Figure 2a.
In
vivo
experiment: The
coronal image of in
vivo mice
abdomen with visible motion and the result of the correction algorithm are
shown in Figure 3. The ghosting caused by motion (highlighted in the coronal
image by red arrow) is reduced by the correction algorithm. Due to the presence of motion in the coronal
view, the anatomical details inside the kidneys are blurred and those details
are visible in the upsampled and rotated axial image stack, and by using the correction
algorithm, some of these structures are recovered
(highlighted by white arrow). The correction algorithm improved signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and mean edge width in both the
experiments [Table 1,2].
The effective
resolution achieved in the corrected image is lower than the original image.
This resolution compromise is due to the combination of the in-plane view from
an image stack with the through-plane view of the orthogonal image stack. This
motion correction algorithm is useful when orthogonal images are already acquired
as part of the normal workflow of imaging studies, and the original image data
have moderate motion artifacts. The results show that it is advantageous to use
the available orthogonal image(s) to improve the image quality by reducing
ghosting artifacts caused by motion.
Acknowledgements
No acknowledgement found.References
1) Kruger D G, Slavin G S, Muthupillai R, et al. An orthogonal
correlation algorithm for ghost reduction in MRI. Magn Reson Med. 1997 October; 38(4):678-86.
2) Welch E B, Felmlee J P, Ehman R L, et al. Motion correction using the
k-space phase difference of orthogonal acquisitions. Magn Reson Med. 2002 July;
48(1):147-56.
3) Herbst M, Maclaren J, Lovell-Smith C, et al. Reproduction
of motion artifacts forperformance analysis of prospective motion correction in
MRI. Magn Reson Med. 2014 January; 71(1):182-90.
4) Bones P J and Maclaren J R. Improved
bulk rotation detection and correction in MRI. In Conf Proc IEEE Eng Med Biol
Soc. 2007; 2106-9.