Frédéric Gretsch1, Tobias Kober2,3,4, Maryna Waszak2,3,4, José P. Marques5, and Daniel Gallichan1
1CIBM, EPFL, Lausanne, Switzerland, 2Advanced Clinical Imaging Technology (HC CMEA SUI DI BM PI), Siemens Healthcare AG, Lausanne, Switzerland, 3Department of Radiology, University Hospital (CHUV), Lausanne, Switzerland, 4LTS5, EPFL, Lausanne, Switzerland, 5Donders Centre for Cognitive Neuroimaging, Radboud University, Netherlands
Synopsis
FID
navigators (FIDNavs) and a few lines of highly accelerated
double-echo 3D fat navigators (FatNavs) were measured each TR of a
high-resolution 3D-GRE acquisition. High temporal resolution $$$B_0$$$
variations and motion parameters could be estimated by the FIDNavs
using the lower
temporal resolution FatNavs to derive
calibration data. The cardiac cycle
pattern emerged clearly
in these estimates and retrospectively
corrected images were of clearly
improved quality, thereby demonstrating the
potential for this hybrid approach.Introduction
MRI
at very high spatial resolution is challenging
due to several limitations, including
lower SNR, increased sensitivity to motion and longer scan-times. For
T2*-weighted imaging of the brain there is the further complication
of artifacts arising due to changes in the $$$B_0$$$-field with the
respiratory cycle. Fat navigators (FatNavs) have already demonstrated
excellent motion tracking and retrospective correction at 7T [1] and
we have also shown their sensitivity to $$$B_0$$$-field tracking [2].
Extremely rapid FID navigators (FIDNavs) have previously demonstrated
to be sensitive to $$$B_0$$$ changes [3] and to motion [4,5], but the
mapping between the FID signals and the $$$B_0$$$ and motion
parameters requires per-session calibration.
Here we propose a GRE
sequence with hybrid integration of segmented dual-echo FatNavs and
FIDNavs which allows the low temporal resolution dual-echo FatNavs to
calibrate the FIDNav mapping for each session. This allows the estimation
of high temporal-resolution motion (6 parameters) and zeroth and
first-order $$$B_0$$$ variations (4 parameters) which can be used for
retrospective correction of the T2*-weighted GRE k-space.
Method
Using
a 7T head-only MR scanner (Siemens
Healthcare, Germany) and a 32-channel RF coil
(Nova Medical Inc.), we scanned a healthy volunteer who was asked to
breathe deeply to amplify the $$$B_0$$$ variations. His cardiac and
respiratory traces were recorded by the external monitoring unit
during the high-resolution slab-selective 3D-GRE. An FID readout
(FIDNav) was inserted between the excitation pulse and the phase
encoding gradients. After each host
sequence TR, three lines of a highly accelerated double-echo FatNav
were acquired, resulting in an entire FatNav volume every 36 TRs
(~2s). Figure 1 summarizes the sequence
design and protocol parameters used. The effective TR (time between
two host excitation pulses) was 56.2 ms. GRAPPA reference lines for
FatNavs were acquired in
a 2.3s pre-scan.
Total acquisition time: 32min25s, whereas
navigator-free would have been 23min2s (although SNR efficiency is
similar due to increased effective TR for water).
FatNavs were retrospectively co-registered using SPM to obtain
motion-estimates, and a $$$B_0 = \beta_0 + \vec \beta \cdot \vec x$$$
fit (constrained to the host sequence FOV) was computed. Each FIDNav
was assigned to the nearest FatNav, a linear mapping $$M = \Lambda F
$$ was determined, where $$$M$$$ are the motion and field parameters
and $$$F$$$ the mean FIDNav signals for the given FatNav volume for
each receive channel. This linear mapping is expected to hold for
sufficiently small variations [4]. Once $$$\Lambda$$$ was found it was
reapplied to each individual FIDNav to obtain higher temporal
resolution motion estimates.
To observe cardiac related variations in these estimates, each
FIDNav and its time-to-peak was mapped to the closest cardiac peak
(from external monitoring), and a binning (150 bins) of the global
time-to-peak range allowed a bin-wise mean of the estimated motion
and $$$B_0$$$ field estimates.
For
retrospective correction, the motion parameters and field
coefficients were temporally smoothed via a sliding-window of 7
FIDNavs. Hence the effective temporal resolution of these estimates
was approximately 400 ms. The correction itself was performed
coil-wise using the NUFFT adjoint operator [6].
Results and discussion
Temporally
smoothed FIDNav-estimated motion parameters are shown in Figure 2.A.
One can readily observe the clear correlation between the cardiac
peaks (dashed vertical line) and the motion estimates. Figure 2.B
shows the binning result for translations and linear $$$B_0$$$
coefficients. As expected, the maximal cardiac effect appears
slightly before the peak in the physiological data, as the pulse was
measured on the right index finger where its arrival time is
later than for the brain.
Comparison
to a similar measurement of the ballistocardiogram obtained from a
camera/marker system [7] is encouraging, although we observe the
strongest effect on x and y axes in this example – rather than the
z direction. It is likely that the direction and amplitude of the
effect is strongly subject-dependent, and also affected by the
position of the subject in the scanner. Further studies will allow
quantification of the associated variability.
The
quality of the corrected image is greatly improved, as exemplified in
Figures 3 and 4 (zoomed section), where high-resolution improvements
in features are clearly identifiable, implying that the motion
estimates are of reasonable accuracy.
Global
improvement is especially visible in a minimal intensity projection
contrast over a 9mm slab (Figure 5) where the diffuse blotches
associated with respiratory artifacts are greatly reduced.
Conclusion
FIDNavs
calibrated by segmented FatNavs were able to track both head motion
and field variation at high temporal resolution. Retrospective
corrections based on to these estimates showed considerable
improvement in image quality thereby demonstrating the potential of
this new hybrid approach.
Acknowledgements
This work was supported by
CIBM of the UNIL, UNIGE, HUG, CHUV, EPFL and the Leenaards and
Jeantet Foundations, as well as SNSF project number 205321_153564.References
[1]
Gallichan D. et al., “Retrospective correction of involuntary
microscopic head movement using highly accelerated fat image
navigators (3D FatNavs) at 7T: 3D FatNavs for High-Resolution
Retrospective Motion Correction”, Magnetic Resonance in Medicine,
Apr. 2015.
[2]
Gretsch F. et al, "Investigating the potential of highly
accelerated FatNavs for dynamic shimming", ISMRM
23rd Annual Meeting, 2015.
[3]
Splitthoff D. N. and Zaitsev M., “SENSE shimming (SSH): A fast
approach for determining B 0 field inhomogeneities
using sensitivity coding,” Magnetic Resonance in Medicine,
vol. 62, no. 5, pp. 1319–1325, Nov. 2009.
[4]
Babayeva M. et al., “Accuracy and Precision of Head Motion
Information in Multi-Channel Free Induction Decay Navigators for
Magnetic Resonance Imaging”, IEEE Transactions on Medical Imaging,
vol. 34, no. 9, pp. 1879–1889, Sep. 2015.
[5]
Kober T. et al, “Head motion detection using FID navigators”,
Magnetic Resonance in Medicine, vol. 66, no. 1, pp. 135–143, Jul.
2011.
[6]
Fessler J. A. and Sutton B. P., “Nonuniform fast Fourier
transforms using min-max interpolation,” Signal Processing, IEEE
Transactions on, vol. 51, no. 2, pp. 560–574, 2003.
[7]
Maclaren J. et al., “Measurement and Correction of Microscopic Head
Motion during Magnetic Resonance Imaging of the Brain.”, PLoS ONE 7
no. 11, 2012.