Christopher M Rank1, Thorsten Heußer1, Andreas Wetscherek1, Martin T Freitag2, Heinz-Peter Schlemmer2, and Marc Kachelrieß1
1Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany
Synopsis
To
allow for MR acquisition times as short as 1 minute per bed, we propose a new
method for PET/MR respiratory motion compensation (MoCo), which is based on
strongly undersampled radial MR data. We acquired simultaneous PET/MR data of
the thorax of three patients. Motion vector fields were estimated with a
newly-developed algorithm, which alternates between MR image reconstruction and
motion estimation. Subsequent 4D MoCo PET reconstructions employing
the motion vector field derived from strongly undersampled MR data yielded a considerable
visual and quantitative improvement compared to standard 3D PET and 4D gated PET
reconstructions.Purpose
Respiratory motion leads to
motion blurring of PET images and thus an underestimation of the reconstructed
activity. Recent approaches for PET/MR motion compensation use 3 to 10 minutes
of MR acquisition time per bed position
1. To allow for MR acquisition times as short as 1
minute per bed, we propose a new method for PET/MR respiratory motion
compensation (MoCo) that employs strongly undersampled MR data.
Methods
Simultaneous PET/MR data covering the thorax of three
free-breathing patients diagnosed with lung cancer were acquired at a Biograph
mMR system (Siemens Healthcare, Erlangen, Germany). Data acquisition and
evaluation was in accordance with the local ethics committee and informed
consent was obtained from each patient. We applied a vendor-provided radial
stack-of-stars sequence with golden angle radial spacing and sagittal slice
orientation: field-of-view: 400×400×396 mm
3, voxel size: 1.6×1.6×4.5
mm
3, TR/TE = 3.75/1.70 ms, 88 slices (55% slice resolution, 6/8
partial Fourier), flip angle: 10°, fat supression activated. For
PET imaging the radionuclide fluorodeoxyglucose (
18F-FDG) was used
and the acquisition time was 300 s per bed. MR data and PET list-mode data were
sorted retrospectively into 20 overlapping respiratory motion phase bins with a
width of 10% employing an intrinsic MR motion surrogate signal. For motion
estimation, only MR data measured within the first 60 s of the data acquisition
were used (Fig. 1). These data corresponded to 36 radial spokes per slice and
motion phase and an undersampling factor of 11.2. Motion vector fields
(MVFs) were estimated with a newly-developed algorithm, which alternates
between MR image reconstruction and motion estimation. To
increase the robustness, deformable registrations were carried out between
adjacent motion phases in a cyclic manner, and these were further regularized by cyclic
constraints
2. Standard 4D gated gridding and 4D MoCo
MR reconstructions were performed, with the latter incorporating the estimated
MVFs.
In addition, 3D PET and 4D gated PET images were generated
with a standard OSEM algorithm. For 4D MoCo PET reconstructions, MVFs derived
from MR were incorporated into the system matrix
3. For all PET
reconstructions, 3 iterations with 21 subsets were used and a Gaussian
smoothing filter (FWHM = 3.2 mm) was applied at the end of each iteration.
Results
For demonstration purposes, different motion phases of 4D gated
gridding MR and 4D MoCo MR reconstructions are shown in Fig. 2. 4D gated
gridding MR images revealed severe streak artifacts and high noise levels. Both
artifacts and noise were reduced considerably in the 4D MoCo MR images while the
motion information was fully preserved. Figure 3 presents different PET image
reconstructions of an end-exhale motion phase. 3D PET showed an increased
lesion size due to motion blurring while 4D gated PET yielded an increased
noise level. 4D MoCo PET images were only slightly affected by
these two sources of uncertainty. For quantitative evaluation, SUV
mean
values were measured for three lung lesions of the patients. Compared to 3D
PET, SUV
mean values of 4D gated PET and 4D MoCo PET were 3.0% and 9.1%
larger on average while standard deviations were 81.0% and 9.6% larger on
average, respectively. These findings demonstrate that 4D MoCo PET is able to
reduce the underestimation of activity due to motion blurring compared to 3D
PET and to increase signal-to-noise ratio and contrast-to-noise ratio compared
to 4D gated PET.
Conclusion and Discussion
In this study, we
proposed a new respiratory motion compensation for PET images. It employs strongly
undersampled MR data, which can be acquired within 1 minute. 4D MoCo PET reconstructions achieved a considerable
visual and quantitative improvement compared to standard 3D PET and 4D gated PET
reconstructions. By improving PET quantification and image
quality, the new method can potentially increase the diagnostic value of
clinical PET/MR. In addition, the remaining MR acquisition time per bed (about
4 minutes) is not needed for motion estimation and is thus available for
clinical MR sequences.
Acknowledgements
No acknowledgement found.References
1. Grimm R, Fürst S, Souvatzoglou M, et al. Self-gated
MRI motion modeling for respiratory motion compensation in integrated PET/MRI. Med.
Image Anal. 2015;19(1): 110-120.
2. Brehm
M, Paysan P, Oelhafen M, et al. Self-adapting cyclic
registration for motion-compensated cone-beam CT in image-guided radiation
therapy. Med. Phys. 2012;39(12):7603-7618.
3. Qiao F, Pan T, Clark J, Mawlawi, OR. A
motion-incorporated reconstruction method for gated PET studies. Phys. Med.
Biol. 2006;51(15):3769-3783.