Thomas Amthor1, Peter Koken1, Karsten Sommer1, Mariya Doneva1, and Peter Börnert1
1Philips Research Europe, Hamburg, Germany
Synopsis
We propose an acceleration technique for MR
Fingerprinting measurements based on Cartesian or otherwise segmented sampling.
We show that quick repetition of MR Fingerprinting sequences leads to
stationary signal responses. Instead of waiting for the spin system to relax
completely before the start of the next fingerprint sequence, we take the
evolution of the spin system into account and calculate a dictionary of steady-state
fingerprints. We present a Cartesian fingerprint measurement with a SENSE
factor of four. Using our method, additional acceleration by more than a factor
of two could be achieved, without reducing the match accuracy.Purpose
Magnetic Resonance Fingerprinting (MRF) is a promising
technique for quantitative multi-parametric measurements and tissue
characterization [1, 2]. A train of RF pulses produces an MR signal that serves
as a fingerprint for a certain tissue type. This fingerprint can later be
compared with a dictionary of known signals. Many previously reported
approaches combine a long fingerprint sequence with spiral or radial sampling
for fast acquisition. However, these sequences may not be available on
general-purpose MRI systems. Therefore, it would be desirable to also use
Cartesian sampling in combination with parallel imaging for accelerated
acquisition. In such a case, the MRF pulse train must be applied several times
to cover k-space. In order to make
all MRF signals start with a well-defined spin state, one possible approach is
to insert a long delay time between the MRF pulse trains to allow the spin
system to relax to its equilibrium state, which can be very time-consuming. In
this work, we present a steady-state MRF approach, where the delay between MRF
pulse trains can be very short and a stationary solution for the fingerprint
signal is calculated. In this way, Cartesian or otherwise densely sampled MRF
scans can be accelerated significantly without compromising matching accuracy.
The findings are in line with earlier observations for accelerating inversion
recovery experiments [4].
Methods
The fingerprint sequence used in this study is based
on a spoiled gradient-echo sequence [2]. It consists of a train of 200 flip
angles, preceded by an inversion pulse. The repetition times (TR) are constant
at 15ms to minimize B0 artifacts,
leading to a total time of 3 seconds for the 200 steps. For dictionary
generation, the sequence is repeated several times, and the expected signal of
the complete pulse train is calculated using an Extended Phase Graph formalism
[3] (Figure 1). The fingerprint signal reaches a stationary shape. Measurements
have been performed on a Philips Achieva 1.5T system using an 8-channel head
coil with a SENSE factor of 4 and a phantom equipped with known gel samples (Diagnostic
Sonar, Eurospin II). The MRF dictionary consists of 18 entries for the different
samples (227ms ≤
T1 ≤ 1646ms;
48ms ≤
T2 ≤ 369ms),
12 of which are present in the phantom, and one entry for the background.
Results and Discussion
When the delay between the repetitions is short, the
spin system will not have relaxed completely before the next inversion pulse.
While the signal responses of the short-
T1 system (Figure
1, middle) are almost identical for each repetition, the long-
T1 system (Figure 1, bottom) exhibits
different signal responses. However, after three repetitions, even this system
reaches a stationary fingerprint. In the first measurement, the delay between
the MRF sequence repetitions is large (5s), so that a fully relaxed spin system
can be assumed each time. Using a dictionary calculated with this assumption, all
12 sample numbers are matched correctly (Figure 2, left). When the delay is reduced to 0.5s, the same dictionary leads to
wrong matches, because the assumption of a relaxed spin system is violated (Figure 2, middle). When calculating a train of three sequence repetitions and
using the steady-state fingerprint signal for the dictionary, all samples are
matched correctly again (Figure 2, right). The first measurement took 4:17 min, the second one only 1:53
min.
Conclusions
Quick repetition of MR fingerprinting sequences leads
to stationary signal responses. With dictionaries based on these steady-state
fingerprints, acquisition can be accelerated significantly. Using our method,
we could accelerate a Cartesian SENSE measurement by more than a factor of two
without reducing the match accuracy. The method is also applicable for
segmented radial or spiral sampling. However, shortening the MRF repetition
period (and thus the distance of the inversion pulses) may also lead to
decreased differentiation of long
T1 values, so
that a tradeoff is necessary for each application.
Acknowledgements
No acknowledgement found.References
[1] D. Ma et al., Nature 495,
187 (2013)
[2] Y. Jiang et al., Magn. Reson. Med., doi: 10.1002/mrm.25559
(2014)
[3] K. Scheffler, Concepts
Magn. Reson. 11, 291 (1999)
[4] D. Canet et al., J. Magn. Reson. 18, 199
(1975)