Chien-Yuan Eddy Lin1,2, Ai-Chi Chen3, David Yen-Ting Chen3, Ying-Chi Tseng3, and Chi-Jen Chen3
1GE Healthcare, Taipei, Taiwan, 2GE Healthcare MR Research China, Beijing, China, People's Republic of, 3Department of Radiology, Taipei Medical University - Shuang Ho Hospital, New Taipei City, Taiwan
Synopsis
Pseudo-continuous
arterial spin labeling (pCASL) has been recently used for investigating
cerebral hemodynamic change on the patient receiving carotid artery stenting (CAS)
because it permits repeated measurement of absolute cerebral blood flow in a
short interval without MR contrast agent or radioactive material. However,
labeling efficiency of pCASL has been proved to be dependent on B0
inhomogeneity. Labeling position may need to be carefully applied after CAS.
The aim of this study was to experimentally determine the optimal labeling
position for pCASL with minimal frequency shift caused by stent in exploring
cerebral perfusion in the patient with CAS treatment.Purpose
Carotid artery stenting (CAS)
is being increasingly used as an alternative treatment to carotid endarterectomy for carotid artery stenosis
1. Variable imaging modalities, including
conventional angiogram, dynamic susceptibility contrast perfusion MR, and positron
emission tomography, have
been applied to investigate cerebral hemodynamic change after CAS and have revealed hemodynamic improvement following CAS
2,3. However, it remains unclear which hemodynamic changes
occur in certain areas of the brain following CAS. Pseudo-continuous arterial spin labelling (pCASL) is a recent emerging MRI
method for studying cerebral perfusion using radiofrequency (RF) pulses for the noninvasive labeling of endogenous water protons in the blood
4 instead of injecting
exogenous contrast agent or radioactive material. This technique is capable of
detecting absolute cerebral blood flow (CBF) at the brain tissue level (ml/100g per minute) permits repeated measurement of CBF in a short interval, which is useful in patients
receiving CAS
5. However, the
labeling process of pCASL is not strictly an adiabatic inversion and the
labeling efficiency may be dependent on B
0 inhomogeneity or other
factors
6. Therefore, labeling position may need to be carefully
applied when patient with CAS. The aim of this study was to experimentally
determine the optimal labeling position for pCASL with minimal frequency shift
caused by stent in exploring cerebral perfusion in the patient with CAS
treatment.
Methods
Patients with symptomatic
internal carotid artery (ICA) stenosis receiving CAS were included in the
study. All MRI acquisitions were performed on a 3T clinical scanner (Discovery
MR750, GE Healthcare, Milwaukee, USA) using an HD neurovascular array coil as
the signal detection and whole body coil for RF transmission. Perfusion study
was conducted using a pCASL with a 3D background suppressed fast-spin-echo
stack-of spiral readout module, a TR=5327 ms, an TE=10.5 ms, FOV=24×24 cm, labeling duration=1.5 s, post labelling
delay=2525 ms, no flow-crushing gradients, matrix=128×128,
NEX=2, slice thickness =4 mm. In addition, labelling
plane was 3.35-mm thick and placed 2.18
cm inferior to the lower edge of the scanning coverage. The CBF maps were generated on an Advantage
Windows workstation using Functool software
(version 9.4, GE
Healthcare). For
identifying stent position, three-slab time-of-flight magnetic resonance
angiography (TOF-MRA) were acquired with TR/TE=21 ms/2.2 ms, flip angle=20°, FOV=23×20 cm, matrix=320×192, slice
thickness=1.4 mm. B0 map for evaluating
the field inhomogeneity was acquired using WAter Saturation Shift Referencing
(WASSR) method with 0.5 μT, 2×80 ms fermi pulses. To determine the
optimal labeling position, the distance between stent and labeling plane was varied
to be 2, 2.5, 3, 3.5, 3.75, 4, and 5 mm. Labeling efficiency is the important
matter for CBF quantification and is dependent on B0 inhomogeneity. However, the field homogeneity in the labeling plane was expected to be perturbed by the presence of the carotid stent. The local field inhomogeneity
at the labeling plane of the pCASL pulse with flowing spin moving at velocity V along the Z direction can be modeled as a
constant shift plus a linear Z gradient, and refer to them as the “off-resonance” (ΔB0) and “off-resonance gradient” (ΔG), respectively. In a cycle of repeated pattern of RF and
slice-selective gradient pulses used for pseudo-continuous inversion, the amount of error introduced into the phase accumulation (Δφerror) between two
RF pulses
(duration between pulses denoted as δ) for the pCASL pulse sequence can be calculated6:
$$\triangle\phi \scriptsize{error}=\gamma\triangle B_{0}\delta+\frac{1}{2}\gamma V\triangle G\delta^{2}$$
Results
The stent position was
determined by TOF-MRA according to the loss of TOF signal caused by
susceptibility artifact (Figure 1a). Sagittal brain anatomy with various color
lines was used to demonstrate the labeling position for pCASL (Figure 1b). The estimated
CBF map (Figure 1c) showed the lower CBF at ipsilateral hemisphere and back to
normal compared to contralateral side as the labeling position of pCASL away
from stent. Quantitative analysis revealed that the CBF value at contralateral
side stays normal (52-66 ml/100g/min) as the change of labeling position, while the value
at ipsilateral hemisphere decreased to 27.5 ml/100g/min at
the distance of 2 mm and was gradually increase and reached to normal level at
3.75 mm of the distance between stent and labeling position (Figure 1d).
Discussion and conclusion
Noncontrast pCASL
method is very useful for evaluating cerebral hemodynamic change on the patient
after CAS. However, prominent frequency shift due to the presence of stent
(Figure 2) would lower the labeling efficiency for pCASL if the labeling plane at
or close to the stent position. Consequently, this unexpected loss of labeling
efficiency results in significant quantification errors. In this study, we
demonstrated that perfusion signal will not be compromised when labeling
position is 3.75 mm away from stent position. It is advised that the distance
may vary with different kind of stent materials (cobalt alloy used in this
case). Nonetheless, once the optimal distance has been determined in the beginning
for the material, the same condition can be applied to CAS patient with the
same stent.
Acknowledgements
No acknowledgement found.References
1. Brott TG, Hobson 2nd RW, Howard G, et al., “Stenting
versus endarterectomy for treatment of carotid-artery stenosis,” N Engl J Med, 363(1):11-23,
2010
2. Chang TY, Liu HL, Lee TH, et al., “Change
in cerebral perfusion after carotid angioplasty with stenting is related to
cerebral vasoreactivity: a study using dynamic susceptibility-weighted
contrast-enhanced MR imaging and functional MR imaging with a breath-holding paradigm,”
AJNR Am J Neuroradiol, 30:1330–1336, 2009.
3. Yun TJ, Sohn CH, Han MH, et al., “Effect carotid artery stenting on
cerebral blood flow: evaluation of hemodynamic changes using arterial spin
labeling,” Neuroradiology, 55:271–281, 2013.
4. Wang JJ, Alsop DC, Li L, et al., “Comparison of quantitative
perfusion imaging using arterial spin labelling at 1.5 and 4.0 tesla,” Magn
Reson Med, 48(2):242-54, 2002.
5. Chen YT, et al., “Loss of labelling efficiency caused by carotid
stent in pseudocontinuous arterial spin labelling perfusion study,” Clinical
Radiology, http://dx.doi.org/10.1016/j.crad.2015.10.004.
6. Jahanian H, et al., “B0 field inhomogeneity considerations in pseudo-continuous
arterial spin labeling (pCASL): effects on tagging efficiency and correction
strategy,” NMR in Biomed, 24:
1202–1209, 2011.