Martin Gajdošík1,2, Lukas Hingerl1, Petr Šedivý3, Miloslav Drobný3, Monika Dezortová3, Michael Krebs4, Siegfried Trattnig1,2, Wolfgang Bogner1, and Martin Krššák1,2,4
1High-field MR Centre, Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Vienna, Austria, 2Christian Doppler Laboratory for Clinical Molecular MR Imaging, Vienna, Austria, 3MR Unit, Institute for Clinical and Experimental Medicine, Prague, Czech Republic, 4Division of Endocrinology and Metabolism, Department of Internal Medicine III, Medical University of Vienna, Vienna, Austria
Synopsis
Omega-3
(n-3) fatty acids are essential to human health. However, the knowledge about
their tissue concentrations is not well established. The spectral dispersion at
7T yields a sufficient spectral separation of n-3 (n-3 CH3) and non n-3
methyl (CH3) groups. Large differences in the T2
relaxation and J-modulation of outer
triplet lines allowed for excellent separation of the n-3 CH3 signal
and for estimation of its relative concentration in adipose tissue for the
detection by 1H MR PRESS with TE of 1000ms.Introduction
Omega-3 (n-3) fatty acids (FA) are essential to
human health. However the knowledge about their tissue
in vivo concentrations is not
well established to this day. Proton (
1H) MR Spectroscopy (MRS) allows non-invasive detecting
of methyl groups of n-3 and non n-3 FA chains at two different frequencies,
0.98 and 0.90ppm respectively. The methyl signals are triplets with
characteristic J-coupling modulations of their two outer signals. The detection
of the left outer n-3 methyl signal (1.08ppm) with PRESS and echo time (TE) of
540ms at 1.5T was already shown
1. Škoch et al. proposed the
utilization of J-difference editing via oMEGA-PRESS
2 that allowed for
the detection of n-3 methyl left outer triplet signal at 1.05ppm with TE of
199.5ms at 3T. However at 7T, the left outer triplet signal of the n-3 methyl
group resonates at 1.01 ppm and due to the limitations of B
0
shimming performance it can hardly be resolved from its central line in
vivo. On the other hand, the spectral dispersion at 7T yields better
resolution of both central signals of n-3 (n-3 CH
3) and non n-3
methyl groups (CH
3) (Figure 1). The concentrations of n-3 FA are
very low (approximately 1:75 in adipose tissue
2). Besides the
J-modulation, the n-3 CH
3 has a longer T
2 relaxation time
than CH
31, which could be an advantage for its detection
with long echo times. In this work we assessed the differences in T
2
relaxation times of CH
2, CH
3 and n-3 CH
3
signals in oils and
in vivo and aimed
for detecting the n-3 FA with PRESS and ultra-long TE of 1000ms in thigh
adipose tissue at 7T.
Materials and Methods
MRS
measurements were performed on 7T Magnetom scanner (Siemens Healthcare, Erlangen, Germany)
with 28 channel knee coil (Quality Electrodynamics, Mayfield Village, OH, USA).
In order to minimize the RF power requirements and chemical shift displacement
error of the PRESS sequence the RF pulses were modified for 7T application: Hermite
2.6ms excitation pulse and Hermite 3.0ms refocusing pulse. The bandwidth was
set to 3000Hz. The oil measurements were performed on three samples with
different n-3 FA concentrations (corn, soybean and fish oil, Sigma-Aldrich, St.
Louis, MO, USA). The T
2 times were calculated from sixty TE’s (30…1200ms) signal decays (TR=5s,
NA=6). The n-3 FA contents of the oils were compared with the literature data
(USDA, National
Nutrient Database).
In vivo
measurements were performed on 4 healthy volunteers (1 female, age 29.8±2.9y,
BMI 22.8±3.1kg.m
-2). The VOI of PRESS (TE=1000ms, TR=5s,
NA=32) was set to 2x2x2cm
3
and placed in subcutaneous adipose tissue in the lower part of the thigh. T
2
times were calculated on three volunteers (nine TEs : 30…1200ms; TR=5s; NA=16).
The T
2 of n-3 CH
3 was calculated from TE range of 350ms
to 1200ms. The MRS signals were fitted in jMRUI with AMARES algorithm
3 using
Gaussian lineshape. The T
2s were calculated in MATLAB (MathWorks, Natick, MA) with
monoexponential function. The content of n-3 FA was calculated as ratio of n-3 CH
3
to the sum of both CH
3 signals after T
2 relaxation
correction.
Results
The
J-modulated n-3 CH
3 signal from the soybean oil measured with PRESS and
TE of 1000ms is depicted in Figure 2A and B. The T
2 times measured
in the oils and
in vivo are in Table
1. The comparison of the n-3 FA content in different oils and literature is
depicted in Figure 3. The results of
in vivo
measurements are provided in Figure 4A and B. n-3 CH
3 resonance can
clearly be resolved with another advantage of the suppression of the
strong CH
2 signal, which would otherwise bias the
quantification of the CH
3 signals. The mean content of n-3 FA
in vivo was 0.021±0.002 a.u..
Discussion
In
oils, T
2 times of n-3 CH
3 were found to be twice as long
as for the CH
3 signal. Similarly we measured 1.8-times longer n-3 CH
3
T
2 time
in vivo. Applying
this spin-spin relaxation correction yields expected n-3 FA fraction of approximately
2% [4]. Further
in vivo measurements and biochemical validation from biopsy
samples of broader population is necessary.
Conclusion
The
antiphase J-modulation of CH
3 outer triplet signals observed with TE
of 1000ms and the differences in T
2 relaxation times between n-3 CH
3
and CH
3 signal allowed for the identification of well resolved n-3
CH
3 resonance making ultra-long TE PRESS a method suitable for detection
of n-3 FA
in vivo at 7T.
Acknowledgements
The
authors wish to thank Dr. Rolf Pohmann for programming of the Hermite pulses, Dr.
Vladimír Mlynárik for help with the oil experiments and Anniversary Fund of
Austrian National Bank (P15363), as well as BMWFW WTZ Mobility (CZ11-2015) for financial
support.References
[1] Lundbom et al. PRESS echo time behavior of triglyceride resonances at 1.5T: detecting omega-3 fatty acids in adipose tissue in vivo. J Magn Reson. 2009; 201(1):39-47
[2] Skoch et al. The in vivo J-difference editing MEGA-PRESS technique for the detection of n-3 fatty acids. NMR Biomed. 2014; 27(11):1293-9
[3] Vanhamme et al. Improved method for accurate and efficient quantification of MRS data with use of prior knowledge. J Magn Reson. 1997; 129(1):35-43
[4] Lundbom et al. Characterizing human adipose tissue lipids by long echo time 1H-MRS in vivo at 1.5 Tesla: validation by gas chromatography. NMR Biomed. 2010; 23(5):466-72