Kimberly L Chan1,2,3, Karim Snoussi2,3, Richard AE Edden2,3, and Peter B Barker2,3
1Biomedical Engineering, Johns Hopkins School of Medicine, Baltimore, MD, United States, 2Radiology and Radiological Science, Johns Hopkins School of Medicine, Baltimore, MD, United States, 3F.M. Kirby Center for Functional Brain Imaging, Kennedy Krieger Institute, Baltimore, MD, United States
Synopsis
Glutathione (GSH), a redox metabolite, and lactate, a
product of anaerobic energy metabolism, can
both be detected in the human brain using J-difference editing. Editing each will usually co-edit the other
to some degree, as the GSH editing target spin is at 4.56 ppm and the lactate spin
is at 4.1 ppm. In this abstract, we
investigate optimal simultaneous detection of both metabolites, using a
combination of simulations, and phantom and in vivo experiments. We demonstrate a new acquisition protocol
applying 10 ms editing pulses at 4.35 ppm, which successfully edits both GSH
and lactate signals with near maximal efficiency.Purpose
To optimize the MEGA-PRESS pulse sequence for
the simultaneous detection of glutathione and lactate.
Introduction
In brain MR spectroscopy (MRS), spectral editing is often used to selectively detect signals from lower concentration metabolites that overlap with those of other, more abundant compounds. Usually, one compound is observed at a time, however there have been examples of editing sequences designed to simultaneously detect 2 compounds which can allow for shorter scan times [1,2]. An optimization of the ‘MEGA-PRESS’ editing sequence to simultaneously detect glutathione (GSH, a major anti-oxidant) and lactate (Lac, an indicator of non-oxidative glycolysis) in a single acquisition is presented. The impairment of the function of both metabolites has been implicated in the pathophysiology of various brain pathologies such as schizophrenia and bipolar disorder [3,4,5,6].
Theory
J-difference editing involves the acquisition of two
slightly different experiments, labeled ON and OFF, which are
subtracted to resolve a signal of interest from overlying signals. In the ON
acquisition, frequency-selective editing pulses are applied to refocus the
evolution of a coupling within the molecule of interest. In the OFF
acquisition, these editing pulses are not applied and the coupling is allowed
to evolve. Subtracting the two gives a difference spectrum, which
only contains those signals that are affected by the editing pulses. Typically, these editing pulses are applied
on resonance: 4.56 ppm for GSH and 4.1 ppm for Lac. In this implementation of J-difference
editing, however, a less-selective editing pulse is applied at 4.35 ppm in the
ON acquisition (between the two resonant frequencies), to detect both with near
maximal sensitivity in a single acquisition.
Methods
Phantom experiments and numerical simulations of PRESS and
MEGA-PRESS were performed to determine the optimum TE and editing pulse
frequencies for simultaneous GSH and Lac editing. Phantom spectra were acquired in a phantom
containing 50 mM GSH and 25 mM sodium lactate in phosphate-buffered saline (pH
= 7.2) on a Philips Achieva 3T scanner using a 32-channel receive head coil. In
the PRESS and MEGA-PRESS experiments, the optimal echo time (TE) was found by
varying the echo time from 70 ms to 240 ms in 10 ms increments (TE1 = 13.4 ms,
TE2 = TE – TE1). Frequency-modulated
slice selective refocusing pulses (‘fm_ref07’, 2.2 kHz bandwidth) were used in
order to minimize chemical shift displacement effects. At an echo time of 140 ms, the optimal editing
pulse frequency in the MEGA-PRESS experiment was found by varying the editing
pulse frequency of the sinc-Gaussian editing pulses from 4.1 to 4.56 ppm in
increments of 0.05 ppm for both 20 ms (75 Hz bandwidth) and 10 ms pulses (150
Hz) (‘off’ pulse at 10 ppm). Spatially resolved
simulations of the MEGA-PRESS experiment for all positions on a 19x19
two-dimensional array across a (3 cm)3 voxel were performed in MATLAB,
using the chemical shifts and coupling constants given by (3) and the following
parameters: 2-kHz spectral width, 2048 points, 3-Hertz exponential filter and
zero-filling to 8192 datapoints. Three
MEGA-PRESS implementations were performed in
vivo in one healthy individual with a 4x4x3 cm3 voxel placed in
the midline parietal region: one with 75 Hz editing pulses at 4.1 ppm (on
resonance for Lac), one with 75 Hz editing pulses at 4.56 ppm (on resonance for
GSH), and one with 150 Hz pulses at 4.35 ppm.
Results
Figure 1a shows that in both phantom experiments and
simulations, the maximum signal intensity for both GSH and Lac in the PRESS
experiments occurs at an echo time of 150 – 160 ms. This optimal echo time holds true for
MEGA-PRESS GSH phantom experiments and simulations as well as for MEGA-PRESS
Lac simulations. Figure 2 shows that the
optimal editing frequency to detect GSH and Lac equally occurs at about 4.35
ppm for both simulations and phantom experiments. GSH/Lac editing efficiency is 65% with 75 Hz
editing pulses and is increased to 90% with 150 Hz editing pulses. Figure 3 shows the in vivo spectra from one
subject showing good detection of both GSH and Lac comparable to the two
separate Lac and GSH acquisitions with editing pulses placed on resonance.
Discussion
Simultaneous editing of Lac and GSH is possible at 3T.
Compared to sequential measurements, simultaneous editing results in a 50%
reduction in the scan time while retaining nearly the same sensitivity. Note
that the optimization performed here did not consider T2 relaxation
effects in vivo, which will tend to shorten the optimal TE. Detection of both metabolites can be improved
by using an editing pulse with a squarer profile to perfectly invert the GSH
and Lac spins.
Acknowledgements
NIH: R01EB016089, P41EB015909References
1.
Terpstra
M, Marjanska M, Henry PG, Tkac I, Gruetter R. Detection of an antioxidant
profile in the human brain in vivo via double editing with MEGA-PRESS. Magn
Reson Med 2006;56:1192-1199.
2.
Snoussi, K, Pradhan, S, Harris, AD, Edden, RA, Barker, PB, Optimization of
MEGA-PRESS for the Simultaneous Detection of Glutamate and Glutamine, and GABA.
In: Proceedings of the 23rd Annual Meeting of ISMRM, Toronto, Ontario, Canada, 2015.
3.
Rosa AR,
Singh N, Whitaker E, de Brito M, Lewis a M, Vieta E, Churchill GC, Geddes JR,
Goodwin GM. Altered plasma glutathione levels in bipolar disorder indicates
higher oxidative stress; a possible risk factor for illness onset despite
normal brain-derived neurotrophic factor (BDNF) levels. Psychol Med 2014;44:2409–2418.
4.
Wood SJ, Berger GE, Wellard RM, Proffitt T-M,
McConchie M, Berk M, McGorry PD, Pantelis C. Medial temporal lobe glutathione
concentration in first episode psychosis: a 1H-MRS investigation. Neurobiol Dis
2009;33:354–357.
5.
Dager SR, Friedman SD, Parow A, et al. Brain metabolic alterations in
medication-free patients with bipolar disorder. Arch Gen Psychiatry
2004;61:450-458
6.
Beasley CL, Dwork AJ, Rosoklija G, et al. Metabolic abnormalities in
fronto-striatal-thalamic white matter tracts in schizophrenia. Schizophr Res
2009;109:159-166.