Kimberly L Chan1,2,3, Nicolaas AJ Puts2,3, Karim Snoussi2,3, Ashley D Harris2,3,4,5,6, Peter B Barker2,3, and Richard AE Edden2,3
1Biomedical Engineering, Johns Hopkins School of Medicine, Baltimore, MD, United States, 2Radiology and Radiological Science, Johns Hopkins School of Medicine, Baltimore, MD, United States, 3F.M. Kirby Center for Functional Brain Imaging, Kennedy Krieger Institute, Baltimore, MD, United States, 4Radiology, University of Calgary, Calgary, AB, Canada, 5Hotchkiss Brain Institute and Alberta Children's Hospital Research Institute, University of Calgary, Calgary, AB, Canada, 6CAIR Program, Alberta Children's Hospital Research Institute, University of Calgary, Calgary, AB, Canada
Synopsis
Glutathione is involved in
maintaining redox balance, and can be detected in vivo in
brain tissue using MEGA-PRESS editing. In literature to-date, echo times
from 68 to 131 ms have been stated as optimal; in this abstract, the
TE-dependence of MEGA-edited GSH signals is investigated using simulations, and
phantom and in vivo experiments. It is shown that, in vivo, there is a
moderate (15%) benefit of detecting GSH at TE 120 ms over 68 ms. We also
demonstrate that the longer echo time allows the use of higher-bandwidth, more
rectangular slice-selective refocusing pulses, giving a further 57% gain in
signal.Purpose
Glutathione (GSH) abnormalities have
been implicated in the pathophysiology of several neurological disorders (1-3).
Since the GSH spectrum overlaps with stronger signals at 3T, spectral-editing
is often used for its selective detection. However, it is difficult to estimate
the optimal echo time (TE) for J-difference editing of GSH, due to strong
coupling and T
2-relaxation effects. Consequently, a large range of echo times,
from 68 ms to 131 ms has been reported for
in vivo GSH editing (3-6). The aim of this study was to investigate the
TE-dependence of the edited GSH signal, taking into account the effects of
scalar couplings and T
2 relaxation, using density-matrix
simulations, and phantom and
in vivo experiments.
Methods
Spatially resolved MEGA-PRESS simulations were performed for
TEs 70 ms-240 ms in 10 ms increments with parameters: simulated B0=3T;
20 ms sinc-Gaussian editing pulses (ON 4.56 ppm); GTST refocusing pulses (7,8);
2 kHz spectral width; 2048 points; 2.5-Hertz exponential line-broadening; zero-filling
to 8192 datapoints. The first PRESS spin-echo
duration (TE1) was fixed at 13.4 ms. Matched
experiments were performed on a 50 mM GSH phantom using a Philips Achieva 3T
scanner.
Siv(TE), the in vivo signal intensity as a function
of TE, can be estimated from phantom data:
Siv(TE) = Sp(TE) exp(-TE/T2,iv) exp(TE/T2,phantom)
where Sp(TE) is the signal in phantom
experiments, T2-corrected for
T2,phantom and T2,iv . Siv(TE) was
calculated for estimated T2,iv values of 67 ms and 89 ms. The
phantom T2 was assumed to be 260 ms (based on a similar GABA phantom
at 3T (9)). Edited measurements in 5
healthy adults were performed at TEs of 120 ms, and the commonly used 68 ms in
a (3.6 cm)3 parietal region using VAPOR water suppression. For shorter-TE (68 ms) measurements, “GTST”
refocusing pulses were used (bandwidth=1300 Hz). For the longer TE, measurements were made either
using GTST or longer-duration fmref07 refocusing pulses (bandwidth=2200 Hz; (7-10)).
Other parameters matched the phantom experiments except TE1=26.6 ms in
measurements using fmref07 pulses. Gannet (11,12) was used to frequency-and-phase-correct
transients based on the NAA methyl peak, and the GSH-edited difference signal at
~2.95 ppm was integrated.
Results
Spatial
dependence of the multiplet (red vs. green regions in Figure 2a), due to
chemical shift displacement effects in the second refocusing pulse direction of
the OFF spectra, result in difference spectra (DIFF) with no signal in the red
regions. Figure 2b shows how the
multiplet modulates with TE in the second refocusing pulse dimension. The OFF
spectra modulate with TE (as intended) in green, but not red, regions. Figure 3a shows the simulated TE-modulation
of the voxel-sum multiplet. Figure 3b
shows the spectra integrals. The ON multiplets vary subtly (blue). The difference spectra have a peak intensity at
TE 160-170 ms. Figure 4a,b show the TE-dependence of the spectra for the
phantom, in good qualitative agreement with the simulations. Figure 4c shows the difference curve from
Fig. 4b corrected to remove the phantom T
2-weighting (T
2=260 ms). Reintroducing estimated in vivo T
2
relaxation, (67 ms, 89 ms) the maximum signal shifts to TE 120 ms from 160 ms. Figure 5 shows DIFF spectra from one subject
(a) and average integral from five subjects (b). The signal at TE 120 ms with higher-bandwidth
refocusing pulses is 57% larger than at TE=68 ms. The integral is 15% larger at
TE=120 ms than at TE=68 ms when using lower bandwidth refocusing pulses.
Discussion
Good
qualitative agreement in the TE-modulation was found between phantom
experiments and simulations , suggesting that the simulation approach used, and
the spin-system parameters, are accurate.
Estimating in vivo T2
biases, these results predict that the optimal TE shifts to 120 ms; compared to the commonly used TE=68 ms, TE=120
ms will have 55% more signal if assuming in vivo T2=89 ms. In vivo data
also show that TE=120 ms gives slightly more signal than 68 ms (15%), but less than
phantom data suggested. One possible
explanation is that the in vivo T2
is shorter than predicted. Another explanation
arises from the different GSH lineshapes at the two TEs. At the shorter TE, the multiplet has negative
outer lobes (Fig.3a, Fig.4a), which might lead to misidentification of the
baseline in quantification, considering the signal as a positive signal within
a baseline well. In vivo data show that the clearest benefit of the longer TE is the
accommodation of better slice-selective RF pulses, which give more signal due
to the higher bandwidth and more rectangular slice profile. Since these pulses
typically also have a long duration, it is easier to implement them at longer TE
values.
Acknowledgements
NIH
grants P41 EB015909, R01 EB016089, and T32EB01002.References
1.
Lavoie S, Murray MM, Deppen P, Knyazeva M, Berk
M, Boulat O, Bovet P, Bush AI, Conus P, Copolov D, et al. Glutathione Precursor,
N-Acetyl-Cysteine, Improves Mismatch Negativity in Schizophrenia Patients. Neuropsych.
2008;33:2187–2199.
2.
Rosa AR, Singh N, Whitaker E, de Brito M, Lewis
a M, Vieta E, Churchill GC, Geddes JR, Goodwin GM. Altered plasma glutathione
levels in bipolar disorder indicates higher oxidative stress; a possible risk
factor for illness onset despite normal brain-derived neurotrophic factor
(BDNF) levels. Psychol Med 2014;44:2409–2418.
3.
Matsuzawa D, Obata T, Shirayama Y, Nonaka H,
Kanazawa Y, Yoshitome E, Takanashi J, Matsuda T, Shimizu E, Ikehira H, et al.
Negative correlation between brain glutathione level and negative symptoms in
schizophrenia: a 3T 1H-MRS study. PLoS One 2008;3:e1944
4.
Terpstra M, Henry P-G, Gruetter R. Measurement
of reduced glutathione (GSH) in human brain using LCModel analysis of
difference-edited spectra. Magn Reson Med 2003;50:19–23.
5.
An L, Zhang Y, Thomasson DM, Latour LL, Baker
EH, Shen J, Warach S. Measurement of glutathione in normal volunteers and
stroke patients at 3T using J-difference spectroscopy with minimized
subtraction errors. J Magn Reson Imaging 2009;30:263–70.
6.
Nehzad FS, Anton A, Deakin B, Williams S.
Glutathione cannot be quantified reliably from short echo PRESS spectra. In:
Proceedings of the 23rd Annual Meeting of ISMRM, Toronto, Ontario, Canada, 2015.
7.
Edden RAE, Barker PB. If J doesn’t evolve, it
won't J-resolve: J-PRESS with bandwidth-limited refocusing pulses. Magn Reson
Med 2011;65:1509–1514.
8.
Murdoch JB, Lent AH, Kritzer MR.
Computer-optimized narrowband pulses for multislice imaging.
1987;74(2):226–263.
9.
Edden
RAE, Intrapiromkul J, Zhu H, Cheng Y, Barker PB. Measuring T2 in vivo
with J-difference editing: application to GABA at 3 Tesla. J Magn Reson Imaging
2012;35:229–234.
10. Henning
A, Fuchs A, Murdoch JB, Boesiger P. Slice-selective FID acquisition, localized
by outer volume suppression (FIDLOVS) for 1H-MRSI of the human brain at 7 T
with minimal signal loss. NMR Biomed. 2009;22(7):683–696
11. Edden
RAE, Puts NAJ, Harris AD, Barker PB, Evans CJ. Gannet: A batch-processing tool
for the quantitative analysis of gamma-aminobutyric acid-edited MR spectroscopy
spectra. J Magn Reson Imaging 2014;1452:1445–1452.
12. Near
J, Edden R, Evans CJ, Paquin R, Harris A, Jezzard P. Frequency and phase drift
correction of magnetic resonance spectroscopy data by spectral registration in
the time domain. Magn Reson Med 2014;50:44–50.