Development and Application of a 23Na Elliptical Body Coil for 7 Tesla
Tanja Platt1, Thomas M Fiedler1, Armin M Nagel1,2, Andreas K Bitz1, Peter Bachert1, Mark E Ladd1,3, and Reiner Umathum1

1Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Diagnostic and Interventional Radiology, University Medical Center Ulm, Ulm, Germany, 3Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Essen, Germany

Synopsis

Up to now only a few 23Na abdominal MRI studies have been performed at 7T. In this work, a versatile 23Na transceive elliptical-shaped body coil for 7T with a large FOV and a high transmit/receive efficiency was designed, simulated, and implemented on a 7T scanner. The setup was applied for in-vivo imaging of the human torso. The obtained 3D 23Na image data set contains 23Na signals from heart to pelvis. High regional 23Na signals are especially present in the areas of kidney, liver, cartilage, vessels, vertebral disks, spinal canal, and heart.

Introduction

Over the last decades sodium (23Na) MRI has evolved into a valuable biomedical application which can give additional information about the stage of a disease or about therapy response1. Only a few 23Na abdominal MRI studies have been performed at 7T, mostly using radiofrequency (RF) coils constructed for special applications with a medium field of view (FOV)2,3. In this work, a versatile 23Na RF transceive body coil with a large FOV and a high transmit/receive efficiency was designed, simulated, and implemented on a 7T scanner. B1+-maps inside a phantom were obtained from RF simulations and measurements to evaluate coil efficiency. Subsequently, the setup was applied for in-vivo imaging of the human torso.

Methods

Hardware design:

A close-fitting low-pass elliptical birdcage coil was designed that provides a high filling factor and allows positioning in the isocenter of the MR scanner (Fig.1a). For comfortable subject positioning, the birdcage is separated into two halves (Fig.1b).

Cylindrical quadrature birdcage coils provide the advantage of a circularly polarized B1+-field. In an elliptical birdcage rotational symmetry is violated, making the generation of a circularly polarized B1+-field challenging. To improve quadrature performance of the elliptical birdcage, legs were spaced 360°/12=30° apart around the periphery of the ellipse4. Due to the split design, the left and right feeding legs of the birdcage were shifted to the upper half (Fig.1b), resulting in an additional violation of symmetry.

Four feeding ports (top, left, bottom, right) were implemented that enable manipulation of amplitude and phase at four legs. For preliminary measurements and simulations, equal amplitudes and phase differences of 0°, -90°, -180°, and -270° were applied. In further studies, B1+-homogeneity can be improved by applying optimized phase and amplitude settings.

Coil simulations:

RF simulations were performed in CST Studio Suite 20155. B1+-maps were calculated for a phantom (εr’=54 , σ=0.43S/m at 78.6MHz, 23Na-content: w(NaCl)/wges=1.8%). A heterogeneous male body model (34y, 1.77m, 72.4kg)6 was used to obtain specific absorption rates (SAR) for in-vivo application.

Imaging:

Measurements were performed on a MAGNETOM 7T system7. Images were acquired using a density-adapted 3D radial sequence (DA-3DPR)8. Image reconstruction and post-processing were performed in MATLAB9.

B1+-maps inside the phantom were obtained using the dual angle method10. Sequence parameters: Nominal isotropic resolution=10mm, reconstructed FOV=(400 mm)3, FA: α1=55°, α2=110°, tpulse(55°)=1ms, tpulse(110°)=2ms, TE/TR=(1.05ms/250ms), #projections=10000, TAQ(per sequence)=41min40s.

High-resolution 3D in-vivo 23Na image data of the human torso of a free-breathing healthy volunteer (26y, 1.80m, 75kg) were obtained. The maximum time-averaged input power of the RF coil was determined from the simulated SAR in the heterogeneous body model.

Sequence parameters: Nominal isotropic resolution=4mm, reconstructed FOV=(304mm)3, FA: α=44°, tpulse=2ms, TE/TR=(1.05ms/20ms), #projections=18200, avg=5, TAQ=30min20s.

Results

Phantom study:

Simulated and measured B1+-maps correspond very well (Fig.2). Mean B1+ and standard deviation (SD) determined in the delineated ROI are 12.2µT (SD=1.4µT) in simulation and 14.1µT (SD=1.4µT) in measurement. B1+-homogeneity can further be improved by applying optimized phases. First simulations (Fig.3) with optimized phases (0°, -73°, -180°, -287°) demonstrate a B1+-distribution which is symmetric with respect to the coil center. Additionally, mean B1+ increases (14.3µT) and SD decreases (1.1µT) compared to the standard phase settings.

In-vivo study:

Fig.4 shows exemplary slices of a high-resolution 3D 23Na image data set of the human torso. The large FOV of (304mm)3 covers a region from heart to pelvis. High regional 23Na signals are especially present in the areas of kidney, liver, cartilage, vessels, vertebral disks, spinal canal, and heart. 23Na signal distribution of left kidney in Fig.4a-c indicates great potential of 23Na MRI to characterize renal physiology. Although the heart is located at the outer part of the FOV, the left and right ventricles can be discriminated in time-averaged 23Na signal distributions (Fig.4d-e).

Discussion and conclusion

Simulated and measured mean B1+ correspond very well with the target value of 13.57µT (FA=55°). Values of B1+ standard deviations are in an acceptable range. B1+-maps show the expected non-uniformity due to the asymmetry of this elliptical birdcage with two vertically shifted feeding legs. For further studies, B1+-uniformity can be improved by adjusting the phase differences between the feeding ports.

High-resolution 23Na images show the great potential of 23Na MRI at 7T. The developed body coil especially offers the possibility to examine 23Na signal distribution at 7T in organs of the human torso like heart, breast, liver, kidney, and prostate. Additionally, retrospective gating can be used to separately reconstruct different phases of cardiac cycle or respiratory motion11.

Acknowledgements

This work was funded by the Helmholtz Alliance ICEMED - Imaging and Curing Environmental Metabolic Diseases, through the Initiative and Networking Fund of the Helmholtz Association.

References

1) Madelin G and Regatte RR. Journal of Magnetic Resonance Imaging 38.3: 511 (2013)

2) Graessl A, et al. Journal of Cardiovascular Magnetic Resonance 15: W14 (2013)

3) Moon CH, et al. European Radiology 24.6: 1320 (2014)

4) Kurczewski R, et al. Proc Intl Soc Mag Reson Med: 4025 (1992)

5) CST AG, Darmstadt, Germany

6) Christ A, et al. Physics Med Biol 55(2): N23 (2010)

7) Siemens Healthcare, Erlangen, Germany

8) Nagel AM, et al. Magnetic Resonance in Medicne 62: 1565 (2009)

9) The MathWorks Inc., Natick, MA, USA

10) Insko EK and Bolinger L. Journal of Magnetic Resonance, Series A, 103: 82 (1993)

11) Resetar A, et al. Magnetic Resonance Imaging 33.9: 1091 (2015)

Figures

Fig.1: (a) Elliptical low-pass birdcage coil providing coverage of a large part of a human body. (b) Coil consists of two halves to enable comfortable subject positioning. Electrical contacts connect both coil halves during examination.

Fig.2: Simulated and measured B1+-maps correspond very well. Mean B1+ and standard deviation in the marked ROI are 12.2µT and 1.4µT in simulation and 14.1µT and 1.4µT in measurement. (Reconstruction of measured data: zero filling factor=2 , Gaussian filter: 6mm. Simulation: Input power from measurement is used.)

Fig.3: Simulated B1+-map with optimized phase of 0°, -73°, -180°, and -287° demonstrates an improved uniformity in comparison to Fig.2. Optimized B1+-distribution is symmetric with respect to the coil center. Mean value is increased (14.3µT) and standard deviation is decreased (1.1µT) compared to the standard phase settings.

Fig.4: High-resolution in-vivo 23Na images of human torso of a free-breathing healthy 26-year-old volunteer at 7T. 23Na signals especially derive from left kidney(LK), right kidney(RK), liver(L), cartilage(C), vessels(V), vertebral disks(VD), spinal canal(SC) and heart with left(LV) and right ventricles(RV). (Nominal isotropic resolution=4mm, reconstructed FOV=(304mm)3, TAQ=30min20s, zero filling factor=2 , Hamming-filter)



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3977