Towards direct detection of natural abundance glycogen C2-C6 resonances by localized 13C MRS at 7T.
Eulalia Serés Roig1,2, Lijing Xin3, and Rolf Gruetter4,5

1Laboratory of Functional and Metabolic Imaging (LIFMET), Ecole Polytechnique Fédérale de Lausanne (EPFL), Lausanne, Switzerland, 2Image Guided Interventions Laboratory, Radiology Department, University of Geneva (UNIGE), Geneva, Switzerland, 3Centre d'Imagerie Biomédicale - Animal and Imaging Technology (CIBM-AIT), Lausanne, Switzerland, 4Department of Radiology, University of Lausanne (UNIL), Lausanne, Switzerland, 5Department of Radiology, University of Geneva (UNIGE), Geneva, Switzerland

Synopsis

Glycogen is the main energy store in the human body, while its concentration is particularly abundant in muscle tissue. In vivo 13C MRS is unique for investigating glycogen metabolism, as it allows the non-invasive measurement of natural abundance glycogen C1 resonance in the human muscle. However, spatial localization is desirable to detect glycogen C2-C6 resonances due to their overlap with the glycerol C2 and C1, C3 resonances. In this study, we designed a pulse-acquire sequence for localized 13C MRS using ISIS-1D and OVS schemes. The localization performance of the sequence was validated in vitro and in vivo in human muscle at 7T.

Introduction

In vivo 13C MRS provides unique insight into carbohydrate metabolism, such as the detection of natural abundance glycogen C1 resonance in human muscle [1]. Nonetheless, the un-localized 13C spectra are dominated by intense 13C signals from subcutaneous adipose tissue [2], introducing baseline distortion and rendering difficult the detection of small 13C signals, such as the glycogen C2-C6 resonances due to their overlap with the glycerol C2 and C1,C3 resonances [3]. Localized 13C MRS allows to remove the unwanted signal while the use of high magnetic field strength benefits from an improved signal to noise ratio (SNR) and spectral resolution, leading to an increased sensitivity and separation of carbon resonances [4]. Therefore, the aim of this study was to design a pulse-acquire sequence for localized 13C MRS using OVS and ISIS-1D schemes based on earlier reports [5, 6], and test the efficiency of distinct detection of glycogen C2-C6 resonances in vitro and in vivo in human muscle at 7T within SAR limitation.

Methods

All experiments were performed on a 7T human MR scanner (Siemens Erlangen / Germany) with a 13C-linear / 1H-quadrature RF surface coil (6cm diameter 13C-loop combined with 9cm diameter 1H-loops). A small sphere (7mm diameter) containing 99% 13C-enriched formic acid was placed in the centre of the 13C-linear coil as an external reference. A pulse-acquire sequence for localized 13C MRS was designed with OVS and ISIS-1D [5, 6] schemes (Figure 1). In order to reduce the chemical shift displacement error, localization was applied using hyperbolic secant (HS) pulses of 5ms duration resulting in a 7kHz bandwidth (Figure 2), larger than the chemical shift range of the glycogen C2-C6 resonances (1.3 kHz). Localization was followed by symmetric adiabatic 13C excitation using two adiabatic half passage (AHP) pulses (2ms, 650Hz symmetric bandwidth) with inverted phases in alternate scans [7]. The WALTZ16 scheme [8] was applied during 13C signal acquisition for broadband 1H-decoupling (90° pulse duration=0.9ms, resulting in a decoupling bandwidth of 600Hz), while the NOE scheme was applied prior to localization to achieve 13C signal enhancement (10 NOE pulses (pulse duration/delay=1/100ms)). The localization performance of the sequence was validated in vitro on a two-compartment phantom containing 800mM natural abundance glycogen (inner-compartment) and sunflower oil (outer-compartment). Three separate in vitro experiments were performed: 1) without localization, 2) with ISIS-1D localization in a slice parallel to the 13C coil plane, and 3) with ISIS-1D and three OVS double-bands ((VOI=70x30x30mm3), TR=1.19ms, 1000 averages, vector size=2048, BW=20kHz, decoupling duration=21ms, acquisition time=102ms). To validate the performance of the sequence, localized 13C MR spectra were acquired in vivo on the human calf of two healthy male volunteers who gave informed consent according to the procedure approved by the local ethics committee. Localization was applied in vivo using ISIS-1D (coronal slice parallel to the 13C coil plane) and two OVS double-bands ((2 transversal and 2 sagital), VOI=80x60x40mm3, TR=1s, 3000 averages, vector size=2048, BW=20kHz, decoupling duration=21ms, acquisition time=102ms). All in vitro and in vivo 13C spectra were acquired using the glycerol C2 resonance at the centre of the spectrum. First and second order shim were adjusted using FAST(EST)MAP [9] with VOI=80x60x40mm3.

Results

Shimming resulted in a water line-width of 25 Hz. Un-localized 13C MRS in vitro (Figure 3-a) revealed peak resonances in agreement with the literature [2, 3]. Localized 13C MRS in vitro (Figure 3-c) demonstrated efficient suppression of glycerol C2 (69.5ppm) and C1,C3 (61.4ppm) resonances using both OVS and ISIS-1D schemes, allowing simultaneous detection of well resolved carbons of glycogen C4 (78ppm) and C6 (61ppm), and clustered carbons of glycogen C3 (74ppm) and C2,C5 (72.2ppm), in contrast to that using only ISIS-1D scheme (Figure 3-b) in which localization was incomplete. Furthermore, the baseline appeared to be improved when using localization. Localized 13C MRS in vivo revealed well resolved glycogen C2-C6 resonances (Figure 4), while some glycerol C2 and C1,C3 remained in the muscle, ascribed to the presence of intra- and extra-myocellular lipids in muscle and interstitial adipose tissue, respectively [6]. In addition, the localization sequence efficiently suppressed the intense lipid resonance at 130 ppm, as no residual lipid signal was observed on the 13C localized in vivo spectrum (Figure 5).

Conclusion

We conclude that localized 13C MRS is feasible at 7T using an optimized pulse-sequence including both OVS and ISIS-1D schemes for 13C localization followed by symmetric adiabatic 13C excitation, resulting in efficient suppression of glycerol C2 and C1,C3 resonances, as well as intense lipid resonance at 130 ppm, and this will allow further extension of this technique for 13C MRS measurements such as in human brain.

Acknowledgements

This study was supported by Centre d’Imagerie BioMédicale (CIBM) of the UNIL, UNIGE, HUG, CHUV, EPFL and the Leenaards and Jeantet Foundations.

References

[1] Avison M.J. et al, PNAS 1988; 85(5):1634-6.

[2] Heerschap A. et al, NRM Biomed. 1989; 2(3):124-32.

[3] Gruetter R. et al, MRM 1994; 31:583-588.

[4] Gruetter R. et al, NMR Biomed 2003;16:313-338.

[5] Tkac I. et al, App. Magn. Reson. 2005; 29:139-157.

[6] Oz G. et al, Appl. Magn. Reson. 2005; 29, 159-169.

[7] Serés Roig E. et al, 2014 ISMRM/ESMRMB.

[8] Shaka A.J. et al, JMR 1983; 53:313-340.

[9] Gruetter R. et al, MRM 1993; 29, 804-811.

Figures

Figure 1: Pulse-acquire sequence for localized 13C-MRS with OVS and ISIS-1D schemes [5, 6] using HS pulses (5ms). Localization is followed by symmetric 13C adiabatic excitation using two AHP pulses (2ms) with inverted phases applied in alternate scans [7]. The WALTZ16 [8] and NOE schemes are applied for broadband 1H-decoupling.

Figure 2: Bloch simulations showing characteristics of a 5ms hyperbolic secant (HS) pulse: longitudinal magnetization (Mz) versus pulse power (a) and frequency offset (b). Simulations were performed with a self-written Matlab script, using the 13C gyromagnetic ratio, while neither T1 nor T2 relaxation times were included in the Bloch equations.

Figure 3: In vitro 13C spectra of a two-compartment phantom containing 800mM natural abundance glycogen (inner) and sunflower oil (outer). Spectra were acquired without (a) and with localization using ISIS-1D (b) plus 6 OVS bands (c) (VOI=70x30x30mm3). Peak assignments are in agreement with the literature [2, 3] (cf. Table).

Figure 4: In vivo localized 13C spectrum from the calf muscle using ISIS-1D and 4 OVS bands (2 transversal and 2 sagital). Peak assignments are in agreement with the literature [2, 3] (cf. Table). Inset: anatomical transversal image of the calf muscle including the location of the VOI (80x60x40mm3).

Figure 5: In vivo 1H decoupled 13C spectra from the calf muscle showing detection of intense lipid resonance at 130 ppm (CH group of the aliphatic chain), acquired without localization (a), and with localization using ISIS-1D and OVS schemes (b).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3971