Synopsis
Functional
information of cell and tissue viability can be obtained via dynamic 17O-MRI. The
quantification of H217O-concentration as a turnover product of oxidative
phosphorylation in combination with 17O2-inhalation enables mapping of the cerebral
metabolic rate of oxygen consumption (CMRO2). Due to low MR-sensitivity of the
17O-nucleus, spatial as well as temporal resolution is limited. Induced partial
volume effects hinder accurate and stable signal quantification which is
essential for dynamic studies. High resolution 17O-data with (4.5mm)3 was
acquired with a Golden Angle acquisition scheme which allowed reconstruction of
arbitrary time-frames. A partial volume correction algorithm was applied to
test correction capability of data of different temporal resolution.PURPOSE
Tissue
and cell viability information can be accessed via evaluation of the energy
balance. An interesting pathway is the oxygen metabolism which can be analyzed
via oxygen-17-MRI (
17O-MRI). With additional
17O
2-inhalation,
the measured variation of H
217O-concentration in time
(nat. abundance 0.037%), which is the turnover product of oxidative
phosphorylation, allows localized mapping of the cerebral metabolic rate of oxygen consumption (CMRO
2)
1. However, the in-vivo
signal of the
17O-nucleus is reduced by 10
5 compared to
protons (
1H). Additionally extremely fast transverse relaxation is
caused by the electrical quadrupole interaction (I=5/2). Thus pulse-sequences
that enable ultra-short echo-times and high SNR-efficiency such as 3D density
adapted radial (3D-DAPR)
2
or twisted projection imaging
3
are required. Spherical acquisition schemes and T
2*-relaxation
additionally enlarge the full-widths-at-half-maximums (FWHM) of the point-spread-functions (PSF) and thus
induce strong partial-volume (PV) effects which diminish the accuracy of signal
evaluation of time-resolved
17O-MRI. Recently, a partial volume correction
(PVC) algorithm
4 was successfully
applied
5,6 to non-proton
MRI. Spatial resolution, as well as temporal resolution for a CMRO
2-experiment
is essential for accurate determination of functional information. Here, PVC
17O-data with high spatial and temporal resolution, with additional B
1-correction
is presented as pre-evaluation for future CMRO
2-experiments.
METHODS
A
custom-built quadrature
17O/
1H-head-coil was optimized
for in-vivo measurements. Imaging was conducted on a 7T MR-system (Magnetom 7T,
Siemens AG).
17O-data of a healthy volunteer (age 26) with nominal
resolution of (4.5mm)
3 was acquired with a 3D-DAPR-sequence with golden
angle (GA)
7 projection
acquisition scheme (Fig.1B) which enables reconstruction of arbitrary time-frames.
Additional data were acquired to calculate B
1-maps (Fig.1A) with a
phase-sensitive method
8 (TR/TE=72ms/0.56ms,
8000 projections, Θ=75°, resolution: (4.5mm)
3).
17O-data was reconstructed with a SNR-enhancing Hamming filter (FOV:
243x243x243mm
3) and B
1-corrected. The
1H-channel
was used for shimming and acquisition of basic anatomical reference-data (3D-GRE-sequence
(TR/TE=8.1ms/4.88ms, Θ=10°, (1mm)
3)) that was used as registration
basis. Anatomical masks for CSF, grey (GM) and white matter (WM) were obtained
from high resolution anatomical data (Fig.1C) that was acquired with a 24-channel
1H-head-coil . Oxygen-data of different temporal resolution was then PV corrected to
quantify the water-content considering GM, WM and two CSF compartments (lateral
ventricles (CSF
i)/sulci (CSF
o)). Previously determined
T
2*-values for GM (T
2*=2.5ms), WM (T
2*=2.8ms), CSF (T
2*=3.9ms) and H
2O
(T
2*=5.3ms) where used for PSF-simulation. Tubes filled with pure H
2O
where used as external reference.
RESULTS
Quantification
accuracy of
17O-signal was improved for all considered temporal resolutions
where discrepancy between CSF
i and CSF
o (ΔCSF) was
reduced. Before PVC a mean mismatch of Δ
CSF=18% was cut back to Δ
CSF=3% after
correction. Both considered brain matter compartments (GM and BM) experienced
upward signal correction, where grey matter was shifted from 52±1% to 82±1% and
white matter from 50±1% to 61±2%. For low SNR-data (Δt=1min/2min, SNR=7/10)
similar results with little deviation to high-SNR data (Δt=30min, SNR=36) was obtained.
Signal and correction stability was evaluated considering several time steps,
where
17O-signal was corrected individually (Fig.2). Corrected water-content
remained stable within ±2% for brain matter compartments and ±5% for CSF with
maximal fluctuation of 8%. The measured B
1-map allowed localized
signal correction and was capable of adjusting the signal ratio R
tubes of left and right
reference (Fig.1B, 1&2) tube from R
tubes=0.77±0.03
to Rtubes=1.02±0.05.
DISCUSSION
Signal
quantification of in-vivo
17O-data is strongly influenced by PV effects. With
chosen GA projection acquisition scheme it was possible to evaluate data-sets with
a nominal resolution of (4.5mm)
3 with variable temporal resolution. The
applied PVC algorithm is capable of recovering water-content values close to
literature
9 with even low SNR-data (Δt=1min), showing good signal and
correction stability. If no correction was applied GM and WM values were
underestimated by up to 35%; after correction signal mismatch was reduced to
<10%. However, GM and WM values are still underestimated by 4-10%, most
likely due to not fully corrected transverse relaxation. It was seen that
B
1-mapping with a phase sensitive method is feasible for
17O-MRI, showing good
signal adaption for reference tubes where B
1 heavily influences
homogeneity. Additional information obtained from the built in
1H-channel
improved registration accuracy of high resolution anatomical data and allowed
shimming. General signal stability of non-moving calibration tubes and brain
matter compartments was very good, with maximal fluctuation of 3-4%. Variation
for CSF compartments was stronger due to higher sensitivity to motion which was
not considered in correction. Separate anatomical mask for each time step would improve correction
stability.
CONCLUSION
In
this study
17O-data was evaluated considering spatial and temporal aspect
of signal quantification which is of interest for dynamic
17O-studies. Applied
PVC allowed correct, stable signal quantification of considered brain
compartments for high spatial and temporal resolution with additional capable
B
1-correction.
Acknowledgements
References
1. Atkinson et al., NeuroImage
2010 (2): 723-733, 2. Nagel et al., Magn Reson Med 2009 (62):1565-73, 3. Boada et al., Magn Reson
Med (1997); 37: p. 706-715, 4. Rousset
et al., J Nucl Med 1998(5):904-911,
1998, 5. Niesporek et al., NeuroImage, 2015(112): 353–363 6. Hoffmann et al., MAGMA 2014(27):579-87, 7. Chan,
R.W. et al., Magn Reson Med 2009(61): p.
354–363, 8. Morell, Magn Reson Med 2008 (60):889-94, 9. Neeb et al., NeuroImage, 2006 (31): 1156–1168