Hannes Michel Wiesner1, Xiao-Hong Zhu1, Kamil Ugurbil1, and Wei Chen1
1CMRR, Radiology, University of Minnesota Medical School, Minneapolis, MN, United States
Synopsis
In vivo 17O MRS imaging provides a valuable tool for
quantitatively imaging the cerebral rate of oxygen metabolism, cerebral blood
flow and oxygen extraction fraction from a brief inhalation of 17O-isotope
labeled oxygen gas. In this study, we conducted a pilot test to examine the 17O
sensitivity using the human head size water phantom in the world's first 10.5T
whole-body human scanner at the CMRR, then compared it with that at 7T. We
found approximately doubled 17O sensitivity at 10.5T after careful
consideration signal and noise contributions at both fields.Purpose
To investigate the sensitivity gain of
17O
MRS imaging at ultrahigh field of the world's first 10.5T whole-body human MRI
scanner as compared to 7T.
Introduction
In vivo
17O MRS imaging provides a valuable tool for
quantitatively imaging the cerebral rate of oxygen metabolism, cerebral blood
flow and oxygen extraction fraction from a brief inhalation of
17O-isotope
labeled oxygen gas. Due to the unique NMR properties of short
17O
quadrupolar relaxation time and field independence, the NMR sensitivity of the
17O
water signal was found to increase quadratically with magnetic field strength (B
0)
in the phantom test and animal brain
1, 2. In this study, we
conducted a pilot test to examine the
17O sensitivity using ahuman head size water phantom in the world's first 10.5T whole-body human scanner
at CMRR and compared it with that at 7T.
Methods
Chemical
shift imaging (CSI) based on the Fourier-series window technique was performed
on two human MR-scanners (Siemens Healthcare, Erlangen, Germany) at the field
strengths of 7T and 10.5T, respectively. A 2 liter spherical glass phantom (~16 cm diameter similar to human head size) containing
natural abundance
17O saline solution (with 50 mM NaCl and 25 mM inorganic
phosphate) was scanned at fully relaxed conditions (TR 100ms, TR>>5*T
1
of
17O water) with the following sequence parameters: 0.5ms TE, 0.5ms
hard pulse width, 18x18x15cm
3 FOV, 9x9x7 matrix size, 23ms
acquisition time and 30kHz spectral width. Signal was acquired using a home-built
quadrature surface-coil of two loops (diameter 16cm) tuned and matched to the
resonance frequencies of 40.292MHz and 60.611 MHz at 7T and 10.5, respectively.
The reference RF voltage was optimized to obtain maximal signals in the region
of interest at two fields. Two low noise high-gain pre-amplifiers and
T/R-switches with similar circuitry but different operation frequencies from
the same manufacturer (Stark Contrast, Erlangen, Germany) were used at both
field-strengths. Signal-to-noise ratio (SNR) was measured by dividing the peak
intensity by the peak-to-peak noise level.
In addition, we also measured and compared the extra
magnet room noise by comparing the difference of the noise levels measured under
two conditions: T/R switch was connected with a 50 ohm terminator versus the RF quadrature
coil as a room noise pickup coil.
Results
The
phantom shape was clearly resembled in the transversal slices of 3D
17O
water CSI data as shown in Fig. 1, and the phantom position was confirmed to be
in the same position in both scanners with the iso-center as the center CSI voxel.
Highest signal intensity occurred in the central voxels and tilted towards to the
quadrature coil (Fig 1). For the voxels, the averaged SNR was 545 at 10.5T and 383
at 7T, resulting in an SNR ratio of 1.42 between the two fields or a SNR gain
of 42% at 10.5T. This gain was less than the estimated SNR ratio of near 2.0
based on the field dependence relationship
1-2 of SNR µ B
07/4. Interestingly, we
also observed a higher level of approximately 40% in the 10.5T magnet room
noise based on the comparison of noise measurements between 50 ohm terminator
and RF coil. After considering and correcting the extra noise observed in the
10.5T magnet room, it resulted in an SNR ratio of gain of near 2 that was close
to the prediction.
Discussion
From this first study and preliminary results,
we found a significant SNR gain of
17O MRS imaging at 10.5T as
compared to 7T. We also found a relatively large magnet room noise for the new
10.5T magnet room and it was under investigation nor for fixing. Nevertheless,
after the correction of the extra noise, the net SNR gain was consistent with
the prediction. In addition, we also found a larger RF power is needed at 10.5T
for reaching the same RF pulse flip angle
3,4,5,6.
Conclusion
The preliminary results suggest a possibility
to double sensitivity for in vivo
17O
MRS imaging at 10.5T as compared to 7T. This sensitivity gain will advance the
technology and utility of in vivo
17O
MRS imaging for studying brain metabolism and perfusion.
Acknowledgements
NIH
grants of R24 MH106049, RO1 NS070839,
S10 RR029672, P41 EB015894 and P30 NS076408; and technical
assistance from Drs. Pierre-Francois Van de Moortele and Gregor Adriany. References
1 Zhu et. al., 17O relaxation time and NMR sensitivity of cerebral water and their field dependence, MRM 2001, 45, 543-549.
2 Lu et. al., In vitro and in vivo studies of 17O NMR sensitivity at 9.4 and 16.4 T, MRM 2013, 69, 1523-1527.
3 Collins et. al., Different excitation and reception distributions with a single-loop transmit-receive surface coil near a head-sized spherical phantom at 300 MHz, MRM 2002, 47, 1026-1028.
4
Perman et. al., Methodology of in vivo human sodium MR imaging at 1.5 T., Radiology 1986, 160:811-820.
5 James et. al., Optimization and Characterization of Sodium MRI Using 8-channel 23Na and 2-channel 1H RX/TX Coil, 13th International Conference on Biomedical Engineering, 2009, 23, 138-141.
6 Wiesner et. al. Quantitative Study of TX/RX-efficiency of X-Nuclear MRS/MRI at High/Ultrahigh Field, ISMRM Milan, Italy 2014, #810.