Fast PCr Imaging of Rat Calf Muscles Using Spiral-In/Out bSSFP
Yuchi Liu1, Yun Jiang1, Charlie Yi Wang1, Mark Alan Griswold1,2, and Xin Yu1,2,3,4

1Biomedical Engineering, Case Western Reserve University, Cleveland, OH, United States, 2Radiology, Case Western Reserve University, Cleveland, OH, United States, 3Physiology and Biophysics, Case Western Reserve University, Cleveland, OH, United States, 4Case Center for Imaging Research, Case Western Reserve University, Cleveland, OH, United States

Synopsis

In this study, a fast phosphocreatine (PCr) imaging method was developed by combining spectrally selective bSSFP with single-shot spiral-in/out encoding for measuring PCr distribution in vivo. 2 min acquisition yielded an SNR of ~9 in rat calf muscles with high spatial resolution (1.25 mm×1.25 mm×5 mm) at 9.4T. The SNR was increased to 25.5 with 30 min acquisition. This method also has the potential for imaging other metabolites such as ATP with a different carrier frequency, dynamic PCr imaging in exercise-recovery or ischemia-reperfusion studies, and quantification of the absolute metabolite concentration with appropriate T1, T2, and flip angle calibration/correction.

Target Audience

Researchers interested in hetero-nuclei imaging and non-Cartesian acquisition scheme.

Introduction

31P MR spectroscopy has been used to quantify non-localized high-energy phosphate in vivo. Spectroscopic imaging offers both spatial and spectral encoding. However, due to the low metabolite concentration, 31P spectroscopic imaging with high spatial resolution requires long acquisition time that renders the technique impractical. Alternatively, imaging using spectrally selective excitation allows only one metabolite to be imaged using conventional imaging sequence. Further acceleration can be achieved by using a balanced steady-state free precession (bSSFP) approach. In addition, the relatively long T2 relaxation time of phosphocreatine (PCr) provides more favorable T2/T1 contrast in bSSFP1. In this study, a fast PCr imaging method was developed by combining spectrally selective bSSFP with single-shot spiral-in/out encoding for measuring PCr distribution in vivo.

Methods

A user-defined Gaussian pulse with 4 ms duration was designed to spectrally select PCr. For image acquisition, a single-shot, zero moment compensated spiral trajectory was designed using the minimum-time gradient method2. The spiral waveform was executed twice within one repetition time (TR) in a spiral-in/out fashion and the center k-space was acquired at the middle of each TR. The spiral-in/out trajectory was measured manually3 and images were reconstructed with NUFFT4 using the measured trajectory. 2D imaging was performed by combining the spectrally selective pulse and outer volume saturation without slice selection gradient.

All experiments were performed on a Bruker 9.4T horizontal scanner (Bruker Biospin Co., Billerica, MA) using an in-house built 31P transmit/receive saddle coil. The selectivity of the Gaussian pulse was evaluated on a 500 mM inorganic phosphate (Pi) phantom by sweeping the carrier frequency from -5 to 5 ppm with a step size of 0.5 ppm. The Gaussian pulse combined with the spiral-in/out bSSFP sequence was validated on two phantoms with 40 mM phosphocreatine (PCr) and 20 mM adenosine triphosphate (ATP), respectively.

In vivo PCr imaging of the calf muscle was performed on a 10 month-old Fischer rat. An M2M volume coil was used for proton imaging. The B1 fields of the 31P coil and the 1H coil were placed orthogonal to minimize the coupling between the two coils. Localized shimming using PRESS was performed before the imaging experiments. Reference 1H image was acquired using FLASH sequence. Acquisition parameters were: TR, 500 ms; TE, 7 ms; flip angle, 40°; FOV 4×4 cm2; matrix size 256×256; slice thickness, 5 mm. PCr image was acquired using the proposed spiral-in/out bSSFP sequence with outer volume saturation. Acquisition parameters were: TR, 20 ms; TE, 10 ms; flip angle, 25°; FOV 4×4 cm2; matrix size 32×32; slice thickness, 5 mm. The saturation module was implemented at every 32 TRs, followed by 10 dummy scans. A delay of 350 ms was also inserted between every 32 TRs to prevent gradient from overheating. A total of 48000 repetitions were acquired in 30 min. Image reconstruction used an average of 1600, 3200, 8000, 16000, 24000, 32000, 40000, and 48000 repetitions, corresponding to an acquisition time of 1, 2, 5, 10, 15, 20, 25, and 30 min, respectively.

Results

Designed and measured trajectory is shown in Fig. 1. Fig. 2 shows the selectivity of the 4-ms Gaussian pulse. Residual signal at 2.5 ppm, where γATP resonance occurs, was 5.8%. This profile allowed the excitation of PCr only. Fig. 3 shows the reference anatomy image (Fig. 3a) and PCr selective image (Fig. 3b) using the proposed method on PCr and ATP phantoms. In vivo PCr images were reconstructed using different number of signal averages, corresponding to an acquisition time of 1, 2, 5, 10, 15, 20, 25, and 30 min, respectively, and the corresponding signal to noise ratio (SNR) was calculated (Fig. 4). 2 min acquisition yielded an SNR of ~9. The SNR was increased to 25.5 with 30 min acquisition. Representative images with 5, 10, 20, and 30 min acquisition time are shown in Fig. 5. PCr images on phantom and in vivo were zero padded to matrix size 64×64, and FOV of 2cm×2cm is shown.

Discussion & Conclusion

This study shows a fast PCr imaging method with spiral encoding. PCr map of rat calf muscles with high spatial resolution (1.25 mm×1.25 mm×5 mm) was achieved in acceptable acquisition time. With a different carrier frequency, this method can also be used for imaging other metabolites such as ATP. Quantification of the absolute metabolite concentration can be achieved with appropriate T1, T2, and flip angle calibration/correction. This method also has the potential for dynamic PCr imaging in exercise-recovery or ischemia-reperfusion studies.

Acknowledgements

This work was supported by National Heart, Lung, and Blood Institute Grants R21 HL126215.

References

1. Speck O et al. MR. 2002.

2. Lee et al. MRM. 2003.

3. Duyn et al. JMRI. 1998.

4. Fessler and Sutton. IEEE Trans. Signal Process. 2003.

Figures

Figure 1. Designed single-shot spiral trajectory (a), measured spiral-in (b) and spiral-out (c) trajectories.

Figure 2. Excitation profile of the Gaussian pulse. Signal intensity was normalized to that of the on resonance signal.

Figure 3. 1H reference image of ATP and PCr (a) and PCr selective image (b).

Figure 4. SNR of in vivo PCr images as a function of acquisition time.

Figure 5. PCr maps of rat calf muscles superimposed on 1H anatomy image. Data acquired in 5min (a), 10min (b), 20min(c), 30min (d).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3941