Repeatability and accuracy of a novel, MRI-based, semi-automated analysis method for quantifying abdominal adipose tissue and thigh muscle volumes
Michael Simca Middleton1, William Haufe1, Jonathan Hooker1, Magnus Borga2,3,4, Olaf Dahlqvist Leinhard2,3,5, Thobias Romu2,3,4, Patrik Tunon2, Nickolas Szeverenyi6, Gavin Hamilton6, Tanya Wolfson6,7, Anthony Gamst6,7, Rohit Loomba8, and Claude B. Sirlin1

1Department of Radiology, UCSD, San Diego, CA, United States, 2Advanced MR Analytics AB (AMRA), Linkoping, Sweden, 3Center for Medical Image Science and Visualization, Linkoping University, Linkoping, Sweden, 4Department of Biomedical Engineering, Linkoping University, Linkoping, Sweden, 5Department of Medicine and Health, Linkoping University, Linkoping, Sweden, 6Radiology, UCSD, San Diego, CA, United States, 7Computational and Applied Statistics Laboratory (CASL), UCSD, San Diego, CA, United States, 8Department of Medicine, UCSD, San Diego, United Kingdom

Synopsis

Current MRI methods to estimate body tissue compartment volumes rely on manual segmentation, which is laborious, expensive, not widely available outside specialized centers, and not standardized. To address these concerns, a novel, semi-automated image analysis method has been developed. Image acquisition takes about six minutes, and uses widely available MRI pulse sequences. We found that this method permits comprehensive body compartment analysis and provides high repeatability and accuracy. Current and future clinical and drug development studies may benefit from this methodology, as may clinical settings where monitoring change in these measures is desired.

Introduction

Magnetic resonance imaging (MRI) can be used to estimate adipose tissue and muscle volumes 1,2. However, current MRI methods to estimate body tissue compartment volumes rely on manual segmentation, which is laborious, expensive, not widely available outside specialized centers, and not standardized. Thus, currently implemented MRI-based tissue volume assessment techniques are impractical for many research studies, and not feasible for clinical care. Rapid, inexpensive, standardized methods are needed to reduce analysis time and cost, and to enable the use of accurate and precise imaging biomarkers for research and, perhaps in the future, clinical practice. To address the problems inherent in manual analysis methods, a novel, semi-automated image analysis method has been developed 3-6. Image acquisition takes about six minutes and requires overlapping stacks of MRI fat-water separated images, now widely available as product sequences. Thus, the aim of this study was to assess the repeatability and accuracy of a semi-automated analysis method to estimate abdominal adipose tissue and thigh muscle volumes.

Methods

In this single-site, prospective, cross-sectional, observational study of 20 adults, an axial, 3D, dual spoiled-gradient-echo, fat-water separation MRI sequence was used to estimate adipose and thigh muscle tissue volumes on a 3T scanner (GE Signa EXCITE HDxt, GE Medical Systems, Milwaukee, WI). Two MRI exams were performed. In the first MRI exam, the sequence was acquired twice (i.e., repeats 1 and 2). Then, the subject was taken off the table, placed back on the table, and the second MRI exam was performed in which that sequence was acquired again (i.e., repeat 3). Hence, this sequence was acquired three times for each subject.

Source images were reconstructed offline to generate water, and calibrated fat images. Each composite image stack (repeats 1, 2, and 3) was segmented into tissue compartments using a novel, semi-automated, 3D, non-rigid, multi-atlas segmentation method 3-6. Sixteen labeled atlases were registered to the image volumes. A voting scheme was then used to combine the 16 labels into a 3D segmentation of each compartment. A subset of 20 images obtained in repeat 1 for each subject was selected for assessment of volume estimation accuracy. As a reference standard, these 20 images were also segmented manually (sliceOmatic software package; Tomovision, Ontario, Canada).

Intra- and inter-examination repeatability was assessed with Bland-Altman analysis, intra-class correlation (ICC), and coefficient of variation (CV) for paired data. Accuracy was assessed by linear regression using the manually-segmented tissue volume measurements as reference.

Results

Cohort characteristics are summarized in Table 1; subjects spanned a broad range of body habitus. Examples of semi-automated adipose and muscle tissue segmentations are shown in Figure 1.

Intra- and inter-examination volume estimation repeatabilities based on all images in the composite image stacks are summarized in Tables 2 and 3, respectively. Tissue volume repeatability was excellent, with intra- and inter-examination ICCs ranging from 0.996 to 0.998, and CVs ranging from 1.5 to 3.6%.

Tissue volume accuracy indices for the semi-automated method, using manually-determined measurements as reference for the 20 selected images, are summarized in Table 4. Abdominal adipose and thigh muscle estimated volumes were close to the corresponding manually-determined reference measurements. 95% confidence intervals (CIs) for all regression slopes included 1, and 95% CIs for all intercepts included 0. Regression bias was small for all measures, in the range 28.8 to 100.5 cm3.

Discussion and Conclusion

One limitation of this study was that only a single 3T MR scanner from a single manufacturer was used. Another limitation is that we assessed accuracy in only a subset of non-randomly selected images, but accuracy of this method has been demonstrated previously 4,7. We did not address intra- or inter-operator repeatability using the semi-automated analysis method, but the tested method requires minimal operator supervision, and intra- and inter-reader agreement was previously shown to be high 7.

In summary, a novel, semi-automated tissue segmentation and volume estimation method permits comprehensive body compartment analysis and provides high repeatability and accuracy. We are not aware of other segmentation methods with comparable levels of repeatability. Current and future clinical and drug development studies may benefit from this methodology, as may clinical settings where monitoring change in these measures is desired. Future studies might address ways to further develop, qualify, and validate semi-automated tissue volumes as possible biomarkers of clinical endpoints.

Acknowledgements

This study was supported by a grant from Pfizer, Inc. and by NIH grant R01 DK088925.

References

1. Artz NS, Haufe WM, Hooker CA, Hamilton G, Wolfson T, Campos GM, Gamst AC, Schwimmer JB, Sirlin CB, Reeder SB. Reproducibility of MR-based liver fat quantification across field strength: Same-day comparison between 1.5T and 3T in obese subjects. J Magn Reson Imaging. 2015 Sep;42(3):811-7. doi: 10.1002/jmri.24842. Epub 2015 Jan 23

2. Valentin S, Yeates TD, Licka T, Elliott J. Inter-rater reliability of trunk muscle morphometric analysis. J Back Musculoskelet Rehabil, 2015; 28:181-190

3. Thomas MS, Newman D, Leinhard OD, Kasmai B, Greenwood R, Malcolm PN, Karlsson A, Rosander J, Borga M, Toms AP. Test-retest reliability of automated whole body and compartmental muscle volume measurements on a wide bore 3T MR system. Eur Radiol. 2014 Sep;24(9):2279-91. doi: 10.1007/s00330-014-3226-6. Epub 2014 May 29.

4. Karlsson A, Rosander J, Romu T, Tallberg J, Grönqvist A, Borga M, Dahlqvist Leinhard O. Automatic and quantitative assessment of regional muscle volume by multi-atlas segmentation using whole-body water-fat MRI. J Magn Reson Imaging. 2015 Jun;41(6):1558-69. doi: 10.1002/jmri.24726. Epub 2014 Aug 11.

5. Andersson T, Romu T, Karlsson A, Norén B, Forsgren MF, Smedby Ö, Kechagias S, Almer S, Lundberg P, Borga M, Leinhard OD. Consistent intensity inhomogeneity correction in water-fat MRI. J Magn Reson Imaging. 2015 Aug;42(2):468-76. doi: 10.1002/jmri.24778. Epub 2014 Oct 30.

6. Leinhard OD, Johansson A, Rydell J, Smedby O, Nystrom F, Lundberg P. Quantitative Abdominal Fat Estimation Using MRI. Int Conf Pattern Recognition. 2008; 19 :2137-2140.

7. Borga M, Thomas EL, Romu T, Rosander J, Fitzpatrick J, Dahlqvist Leinhard O, Bell JD. Validation of a fast method for quantification of intra-abdominal and subcutaneous adipose tissue for large scale human studies. NMR in Biomedicine (in press), 2015.

Figures

Table 1. Population and cohort characteristics

Table 2. Semi-automated abdominal adipose and thigh muscle volume intra-examination repeatability

Table 3. Semi-automated abdominal adipose and thigh muscle volume inter-examination repeatability

Table 4. Tissue volume quantification accuracy indices for the semi-automated analysis method, with manually-determined volume measurements as reference (20 selected slices per subject)

Figure 1. Coronal, sagittal, and axial reconstructions of segmented image stacks for each acquisition (repeats 1, 2, 3). SCAT (blue), VAT (aqua), and thigh muscle volumes (multiple other colors) are shown.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3917