Synopsis
We aimed
to apply IVIM imaging to estimate blood oxygen saturation in kidneys with extraction of glomerular filtrate fraction by using fluid-attenuated
inversion recovery diffusion-weighted imaging (FLAIR DWI). A 3-compartment
model (renal tissue, blood and water) was considered for the estimation. Combination
of DWI and FLAIR DWI provided renal tissue T1. Images were acquired
with two TEs to provide blood T2 which allows for estimation of
blood oxygen saturation. Our model generated estimation
of glomerular filtrate fraction, blood R2 and blood oxygen
saturation. FLAIR DWI has a potential to estimate blood oxygen saturation with
extracting glomerular filtrates fractions in kidneys.PURPOSE
Potential challenges for
application of intravoxel incoherent motion (IVIM) imaging for kidneys includes
differentiation of blood perfusion-related component from glomerular filtrate.
Also there is a need to consider proton relaxation which alters signal
intensity in estimating local blood volume. We propose fluid-attenuated
inversion recovery diffusion-weighted imaging (FLAIR DWI) to acquire IVIM
imaging where the component from glomerular filtrate is eliminated. Moreover,
combination of DWI and FLAIR DWI can provide renal tissue T
1.
Acquisition with two echo times may provide blood T
2 which can be
converted to blood oxygen saturation fraction. The purpose of this study is the
use of IVIM imaging to estimate blood oxygen saturation in kidneys with exclusion of glomerular filtrate (water) fraction.
METHODS
Three
healthy volunteers were scanned with a 3.0T MR system (Achieva dStream,
Philips Healthcare, Best, the Netherlands) with 32ch ds-Torso and 32ch
ds-posterior coils. FLAIR-DWI and spin-echo (SE)-based DWI with short and long TE
were imaged in coronal plane with free breathing. Basic imaging parameters
were: slice thickness = 8 [mm], number of slice = 1 , acquisition matrix =
67x112 , FOV = 230 [mm], number of signal averaged=15, b=0, 250, 500 [s⋅mm-2]. Parameters of FLAIR-DWI were: TR/ TEshort/TElong
= 5500/59/84 [ms] and TI =2000[ms]. Parameters for SE-based DWI were:
TR/TEshort/TElong = 3000/59/84 [ms]. A 3-compartment
(renal tissue, blood and glomerular filtrates) model was considered1.
Definition
of parameters
Longitudinal
relaxation rates: R1t for renal tissue, R1b = 0.58 [s-1]
for blood2, R1w = 0.234 [s-1] for water3.
Transverse
relaxation rates: R2t for renal tissue, R2b for blood, R2w
for water.
Apparent
diffusion coefficients: Dt for renal tissue, D* for
blood.
Constants
proportional to proton density: m0t for renal tissue, m0b
for blood, m0w for water.
Signal
obtained with DWI can be expressed as: SDWI(m0, R2,
R1, D, K, b) = m0 Er2 Eb (1 + (1 – 2eR1
TE/2) Er1), where Er2 = e–R2 TE,
Eb
= e–bD + K bbDD/6, Er1 = e–R1 TR. Signal for
IRDWI: SIRDWI(m0, R2, R1, D, K, b)
= m0 Er2 Eb (1 – 2αe–R1 TI + (1 –
2α– 2 (1 – 2α) eR1 TE/2) Er1), where α is the efficiency
of inversion (0.95 was assumed). Because m0 Er2 always
appears, we define m2 = m0 Er2.
We adopt
a 3-compartment model so that the signal obtained with DWI can be expressed as:
SDWIb= SDWI(m2w, R1w, Dw,
0, b) + SDWI(m2b, R1b, D*, 0, b) +
SDWI(m2t, R1t, Dt, Kt,
b), and that with IRDWI: SIRDWIb = SIRDWI(m2w, R1w,
Dw, 0, b) + SIRDWI(m2b, R1b, D*,
0, b) + SIRDWI(m2t, R1t, Dt, Kt,
b).
Data
processing
FLAIR DW
images and DW images with long TE were registered to DW images with short TE
using affine transformation. Then, the following processing was performed.
1.Eight image
data sets (DW images and IRDW images with short TE and long TE, b = 250[s⋅mm-2] and 500 [s⋅mm-2] each image) were used to calculate m0, R1t, R2t and Dt; K was set as 0;
signals from moving blood and glomerular filtrates were ignored at b≥250[s⋅mm-2].
2.With parameters
obtained from the above 1, 4 image data set (DW images and IRDW images with
short TE and long TE, b = 0 for each image) were used to calculate m0w,
m0b and R2b.
3.Blood oxygen saturation was measured using a
formula for τCPMG = 20 ms by Lu, et al4, where R2b
was as a function of oxygen saturation (Y) and hematocrit (Hct) at four
different τCPMG (5, 10, 15, 20ms).
RESULTS
Estimation
of glomerular filtrate (water) fraction.
Figure 1 shows. a map
and histogram of water fraction in bilateral kidneys.
Estimation of R2b.
Figure 2
shows a map and histogram of R2b in bilateral kidneys.
Estimation of oxygen saturation (Y).
Figure 3 shows a map and histogram of oxygen saturation
(Y) in bilateral kidneys. Distribution of oxygen saturation indicates that
renal cortex is more saturated than medulla.
DISCUSSION
Transverse relaxation time was estimated with two-dimensional DWI of two echo times (59/84 ms) although a
CPMG sequence is desirable. R
2 values maybe overestimated
due to the effect of diffusion on transvers relaxation especially on long-TE
acquisition. Consequently, estimated oxygen saturation may be underestimated. Further
study is warranted to modify protocols for better R
2 estimation.
Nonetheless, the images show heterogeneous distribution of oxygen saturation
between cortex and medulla, indicating its potential use for evaluating local
oxygenation in normal and diseased kidneys.
CONCLUSION
FLAIR
DWI has a potential to estimate blood oxygen saturation with extracting glomerular
filtrates fractions in kidneys.
Acknowledgements
No acknowledgement found.References
1. Tamura
H, Mugikura S, Komori Y, et al. Estimation and removal
of partial volume effects of cerebrospinal fluid in intravoxel incoherent
motion (IVIM) imaging. ISMRM Proc 2015:2990.
2. Shimada
K, Nagasaka T, Shidahara M, et al. In vivo measurement of longitudinal
relaxation time of human blood by inversion-recovery fast gradient-echo MR imaging
at 3T. Magnetic Resonance in Medical Sciences 2012;11:265–71.
3. Abbas
Z, Gras V, Möllenhoff K, et al. Analysis of proton-density bias corrections
based on T1 measurement for robust quantification of water content in the brain
at 3 Tesla. Magn Reson Med 2014;72:1735–45.
4. Lu H,
Xu F, Grgac K, et al. Calibration and validation of TRUST MRI for the
estimation of cerebral blood oxygenation. Magn Reson Med 2012;67:42–9.