Yu Y. Li1, Wolfgang Loew1, Ronald Pratt1, Randy Giaquinto1, Stephanie Merhar1, Jean Tkach1, and Charles Dumoulin1
1Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, United States
Synopsis
The presented work investigates
a real-time imaging approach to chest MRI on a 21.8cm bore scanner dedicated
for neonatal examination within the neonatal intensive care unit (NICU). This
approach is based on the development of a 10-channel small coil array, a fast
imaging pulse sequence, and an iterative image reconstruction algorithm. It is
experimentally demonstrated that real-time imaging can provide high-quality
cardiac and pulmonary images for improved clinical diagnosis in premature
babies within the NICU.Purpose
Pulmonary and cardiac
diseases are the leading causes of morbidity and mortality in premature babies.
Hence, chest MRI has the potential to improve clinical diagnosis and patient
care in the neonatal intensive care unit (NICU). However, physiological motion
poses a challenge to MRI data acquisition in neonates who have higher heart and
respiration rates than adults. In our institution, we are investigating chest
MRI on a small-bore (21.8cm) MRI scanner dedicated for neonatal examination
within the NICU
1,2. A real-time imaging technique is developed to
collect neonatal chest MRI data faster than cardiac and respiratory motion. It
is demonstrated that our new approach can provide high-quality cardiac and lung
images for functional and structural chest examination in premature babies.
Methods
The real-time imaging technique for neonatal chest MRI
consists of the following three functional units:
1) A 10-channel receive coil array for parallel imaging
(Figure 1a)3: The coil housing consists of two identical
polycarbonate shells designed using fused deposition modeling (FDM) technology
(Stratasys Ltd., Valencia, CA). When combined, the two shells form a thin
cylinder designed to comfortably hold a swaddled baby. Each shell has five coil loops with a diameter
of ~5 cm. The five elements provides a 2D field
of view (FOV) coverage of ~8x10cm. The two shells are placed within an 18cm transmit-only volume
coil during the scan.
2) A pulse sequence for real-time data acquisition
(Figure 1b): A single-shot 2D steady-state free precession (SSFP) sequence is
used to collect k-space data. The entire k-space is divided into multiple regions. Each region is uniformly
undersampled in the phase encoding direction and the undersampling factors
increase from the center to the outer k-space regions. This undersampling strategy
typically gives a net acceleration factor of 8 for an acquisition matrix of 128x128,
providing a temporal resolution of ~50 milliseconds for real-time data
acquisition.
3) An iterative algorithm for image reconstruction
(Figure 1c): A technique developed in our previous works4,5, correlation imaging, is used to
reconstruct images from undersampled data. The image reconstruction is performed
region-by-region from the center to the outer k-space based on the
undersampling trajectory used in data acquisition. In each region, the
reconstructed data are used as a feedback to improve the calibration of correlation imaging in
the subsequent iteration. The image reconstruction is run until all of k-space is
covered.
To demonstrate real-time imaging for neonatal chest
MRI within the NICU, 5
neonatal patients (2 weeks to 2 months of age and <2.5kg of weight) were
scanned with free-breathing and without sedation. Short-axis cardiac imaging data were
collected in real-time for ~10 seconds with a temporal resolution of 48
milliseconds (FOV 16x16cm, matrix 128x128, TR/TE 3.7/1.1 ms, slice thickness 5 mm, flip angle 45°). A
set of cine images collected with k-space segmentation (8 views per segment) was used
as references. Real-time lung images were collected for ~40 seconds with a
temporal resolution of 123 milliseconds (FOV 16x16cm, matrix 192x192, slice thickness 6mm, flip angle 10°). A set of static lung images collected using a
3D fast gradient echo sequence was used as
references.
Results
Figure 2 shows a
cardiac imaging example from a subject with a heart rate of >150 beats per
minute. Compared with k-space segmentation, real-time imaging gives better
dynamic contrast. The temporal trajectory also demonstrates that real-time
imaging gives a higher speed than respiratory and cardiac motion. Figure 3
shows a lung imaging example from a subject with a respiratory rate of ~80 breaths
per minute. Compared with static imaging, real-time imaging gives
higher signal in lung parenchyma. Using the Fourier decomposition approach described
in a previous work
6, ventilation- and perfusion-weighted
images can be generated from the temporal-dimension Fourier transform of the real-time dynamic
image series.
Discussion
The neonates in our ongoing study have a respiratory
rate of 30-90 breaths per minute and a heart rate of 100-200 beats per minute.
In addition, cardiac and respiratory motion shows non-periodic behaviors. As a
result, k-space segmentation introduces considerable data inconsistency that may
cause a loss of dynamic contrast (Figure 2). In lung imaging, static imaging typically
needs ~3 minutes for data acquisition and motion may introduce a considerable
signal loss. By collecting data faster than motion, real-time imaging can effectively
reduce this loss, providing a signal gain over static imaging within the lung
parenchyma (Figure 3).
Conclusion
It is experimentally
demonstrated that real-time imaging can provide high-quality cardiac and
pulmonary images for improved clinical diagnosis and patient care in premature babies within the
NICU.
Acknowledgements
This work is supported by NIH/NICHD R21HD071540.References
1. Tkach, J et al., Pediatric
Radiology 2012; 42(11), 1347-1356.
2.
Tkach, J et al., Pediatric Radiology
2014; 44(8), 1011-1019.
3.
Loew, W et al., Proc Intl Soc
Magn Reson Med 2015; 23:3180.
4. Li, Y et al., MRM 2012; 68:2005-2017.
5. Li, Y et al., MRM 2014; Doi: 10.1002/mrm. 25546.
6. Bauman, G et al., MRM
2009; 62: 656-664.