Fabian Kording1, Hendrik Kooijman2, Jin Yamamura1, Manuela Tavares3, Mathias Kladeck1, Kurt Hecher3, Gerhard Adam1, and Bjoern Schoennagel1
1Department of Diagnostic and Interventional Radiology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany, 2Philips Medical Systems, Hamburg, Germany, 3Department of Obstetrics and Fetal Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany
Synopsis
Intrauterine growth
restriction is associated with a decreased oxygen availability. In clinical practice,
indicators such as Doppler changes of the ductus venosus and umbilical artery
are measured and a direct measurement using MR oximetry would be desirable. The
relationship between T2 relaxation time and oxygen saturation (sO2) is
determined in in-vitro fetal blood samples and the feasibility of a noninvasively
determination of fetal sO2 is evaluated. The calculated fetal sO2 highly correlated
with measured sO2 values and determined parameters were successfully evaluated
in-vivo in the left fetal ventricle.Introduction
Intrauterine
growth restriction (IUGR) is associated with an increased perinatal mortality
and morbidity and hence is of significant clinical impact. (1). One reason for
fetal growth abnormalities originates from placental abnormalities or maternal
disease which is associated with an decreased oxygen delivery to the fetus (2).
With the relationship between T2 relaxation time and oxygen
saturation (sO2), magnetic resonance oximetry represents a valuable method for
a direct noninvasive determination of fetal oxygen saturation (3). The purpose
of this work was to establish a relationship between fetal sO2 and T2
relaxation time based on the Luz and Meiboom model (4) using in-vitro fetal
blood samples. Parameters describing the T2 relaxation of fetal
blood were consecutively validated in-vivo.
Materials and
Methods
In
order to measure T2 signal intensities insensitive to inflow effects
or turbulences, a three dimensional balanced steady-state free precession (SSFP)
sequence in combination with T2 preparation pulses (5) was applied at
1.5 Tesla (Achieva, Philips Healthcare, Best, The Netherlands). The sequence was
used in a multishot manner whereas the interval between T2 preparation
and SSFP shot was used for dephasing. The TE-times of the T2-preparation
ranged from 30 to 150 ms, 7 TE-times were measured. The SSFP shot had the
following parameters: (TE/TR: 2.4 ms / 4.8 ms, field of view: 150 mm,
acq. Matrix: 96 x 96, voxel size: 156 x 156 x 8.00, slices: 8). Fetal blood for
in-vitro measurements was derived from the umbilical cord during abdominal
delivery from three different fetuses. The blood from each fetus was heparinized
and divided into 5 samples ( each 7.5 ml) with different oxygen saturation
levels ranging from 30 % to 100%. For MRI, the samples were placed in a customized
holder into a waterbath tempered to 36°, whereas the position of the samples
was turned after each scan to prevent sedimentation. As the used sequence is
critically dependent on locally correct pulse angles, a B1 map based on the two
angle method was used to ensure a correct transmit field. If necessary, the
radiofrequency (RF) field was adjusted using the scaling factor of the transmit
RF field. In order to ensure a correct T2 weighting, calculated T2
values using the SSFP sequence were compared to a conventional multi echo spin
echo sequence. The relationship between T2 and blood can be
expressed as a function of fully oxygenated blood (T20), hematocrit (hct),
susceptibility of hemoglobin (a) and exchange time (tex) ( S=A*exp(-TE/T20+hct
(1-hct)*a*tex)x(1-O2/100)^2*f(tex,TE/4). Measured
signal intensities were fitted to the equation using hct and sO2 values
measured using an blood gas analyzer (Radiometer, ABL system). The determined relationship
between T2 and sO2 based on fitted values of tex, a, and
T20 was consecutively applied to one subject (34 gestation week) in
the left ventricle using the same SSFP sequence. Furthermore, as the hct value
is unknown for in-vivo measurements, the impact of the hct value (ranging from
0.3 to 0.6) was calculated and compared to measured sO2 values.
Results
The
validation of the SFFP sequence with the conventional T2 multi echo sequence
resulted in a high correlation of r = 0.99 and an offset of 3 ms (Fig.1). The
blood samples from each fetus were fitted simultaneously to the equation
(Fig.2), yielding mean parameters for T20: 160±10 ms, tex: 4.5 ms±1.2
and a: 0.032±0.002 103sec-1. The mean parameters were retrospectively
used to calculate the sO2 for each sample for verification (Fig.3), yielding a
high agreement between measured and fitted sO2 for each sample ( r = 0.9 for
all blood samples). Results of measured signal intensities in the left fetal ventricle
using the mean calculated parameters resulted in a sO2 value of 98 %. However, only
three TE values (30 ms,60 ms,90 ms) were considered due to motion
during the remaining measurements (Fig.4). The influence of hct on the sO2 calculation
is shown in Fig. 5. The maximum difference between measured and calculated sO2 over
the range of all hct values was 5 %.
Discussion and Conclusion
The
T2-prepared SSFP sequence was
successfully validated using a conventional multi-echo T2 sequence.
Calculated parameters to determine the relationship between T2 and
sO2 were similar for each fetal blood sample and could be used to
calculate the sO2 value for each sample with a high accuracy compared to
measured sO2 values and were successfully evaluated in-vivo. In conclusion, MR
oximetry is a promising method for a noninvasively determination of oxygen
saturation. In future, the calculated parameters have to be validated using a
higher number of blood samples and the application has to be validated in a
larger patient population.
Acknowledgements
No acknowledgement found.References
(1)
Maršál K. Intrauterine growth restriction. Current opinion in Obstetrics
and Gynecology 2002;14(2):127-135.
(2)
Resnik R. Intrauterine growth restriction. Obstetrics & Gynecology
2002;99(3):490-496.
(3)
Thulborn KR, Waterton JC, Matthews PM, Radda GK. Oxygenation dependence
of the transverse relaxation time of water protons in whole blood at high
field. Biochim Biophys Acta 1982;714:265–270.
(4)
Luz Z, Maiboom S. Nuclear magnetic resonance study of protolysis of trimethylammonium
ion in aqueous solution: order of the reaction with respect to the solvent. J
Chem Physics 1963;39:366–370
(5)
Wedegärtner U, Kooijman H, Yamamura J, Frisch M, Weber C, Buchert R, Huff
A, Hecher K, Adam G. In vivo MRI measurement of fetal blood oxygen saturation
in cardiac ventricles of fetal sheep: a feasibility study. Magnetic Resonance
in Medicine 2010;64(1):32-41.