19F ZTE MR Imaging of 19F labeled calcium phosphate cement at 11.7T
Weiqiang Dou1, Simone Mastrogiacomo2, Olga Koshkina3, Andor Veltien1, Mangala Srinivas3, X. Frank Walboomers2, and Arend Heerschap1

1Radiology, Radboud University Medical Centre, Nijmegen, Netherlands, 2Biomaterials, Radboud University Medical Centre, Nijmegen, Netherlands, 3Tumor Immunology, Radboud University Medical Centre, Nijmegen, Netherlands

Synopsis

To enhance the MR image contrast of calcium phosphate cement (CPC) to bone, it was tagged with a 19F loaded nanoparticle as a contrast agent. 19F zero echo time (ZTE) MRI was applied for the first time for 19F imaging of this CPC in bone at 11.7T. The T1 and T2* relaxation times of the 19F nanoparticles were measured for optimal settings of 19F ZTE MRI. We demonstrate overlaid 1H+19F ZTE images of in vitro and ex vivo CPC/bone samples with hyperintense 19F signals allowing a qualitative and quantitative evaluation of these samples.

Introduction

Calcium phosphate cement (CPC), a material used for bone defect restoration, has poor MRI contrast relative to bone, due to similar T1 and T2 relaxation properties.1 It is thus problematic to use MRI to assess CPC and surrounding bone tissue. To enhance image contrast, CPCs have been tagged with conventional gadolinium- or iron-oxide-based contrast agents, but these have unfavorable effects on the properties of CPC and the MR evaluation of iron-oxide particles suffers from a blooming effect.1 As the benefits of 19F MRI to follow fluorinated tracers are well known,2,3 we explored if it was possible to obtain 19F MR images of a 19F loaded nanoparticle3 as an alternative contrast agent in 1H/19F MRI of CPC. The aim of this study was to optimize 19F MRI of 19F labeled CPC. For this purpose we first determined the 19F relaxation properties of the 19F nanoparticle.

Materials and Methods

CPC consists of 59.1wt% alpha-tricalcium phosphate, 1.5wt% carboxymethyl cellulose and 39.4wt% cryo-grinded poly-lactic-co-glycolic acid (PLGA) particles. As 19F tracer we selected perfluoro-15crown-5-ether (PFCE)2 embedded into PLGA nanoparticles through a single emulsion reaction3. This was incorporated into CPC at a concentration of 16.7%.

In vitro samples

A vertical hole (diameter:3mm; height:3mm) was drilled in the upper surface of a pig jaw block, which was filled with mixed material including CPC (~50mg) and PFCE&PLGA (~10mg). The prepared pig bone was then embedded in a falcon tube with 5% gelatine. Additionally, 20mg PFCE&PLGA particles were dissolved into 500μm distilled water in an eppendorf tube as a control sample for T1 and T2* evaluation and quantification analysis.

Ex vivo sample

A cylindrical defect (diameter:2.5mm; depth:3mm) was drilled in each leg of a Wistar rat cadaver. Defects were then filled with injectable CPC in right leg and CPC+PFCE&PLGA in left leg.

Experiments

Measurements were performed on an 11.7T MR-system (Biospec, Bruker) using a 1H-19F dual-channel volume coil with 40mm inner diameter. 19F 3D-ultra-short TE (3D-UTE; TR=4ms, image resolution=1.56mm3) was applied with six flip angles (FAs) from 2°-8° for T1 estimation and with eleven TEs from 8-608µs with increments of 20-140µs for T2* evaluation.

1H/19F ZTE images for the bone phantom were acquired with TR=2ms/4ms, image resolution=1.56mm3, 1average/16averages and FA=2°/4° in 27seconds/13mins 53seconds. Additionally, the 1H/19F ZTE imaging for rat legs were performed with TR=2ms, image resolution=0.31mm3/1.25mm3, 1averages/32averages and FA=2°/4° in less than 7mins/14mins.

Data Analysis

T1 values were calculated based on the slope $$$e^{\frac{-TR}{T_{1}}}$$$ from the equation $$$\frac{S\left(FA\right)}{\sin\left(FA\right)}=e^{-\frac{TR}{T_{1}}}\times \frac{S\left(FA\right)}{\\tan\left(FA\right)}+M_{0}\times\left(1-e^{\frac{-TR}{T_{1}}}\right) \times e^{\frac{-TE}{T_2^*}}$$$4. A mono-exponential model $$$ S\left(TE\right)=S_{0}\times e^{\frac{-TE}{T_2^*}}$$$ was used for T2* fitting. The 19F signals in the bone phantom and control were quantified in MRIcro software. Regions of interest (ROIs) were manually outlined. 19F signal per ROI was quantified through multiplying the mean pixel intensity by the pixel number of the ROI and slice thickness. The total 19F amount was then summed over slices and expressed in arbitrary units (a.u.).

Results

From three measurements of the 19F nanoparticles, we determined a mean T1 of 1269±38ms and a mean T2* relaxation time of 76.5±8μs (Fig.1; R-squared goodness of fit > 0.99). From this T1 value, Ernst angle settings were determined for 19F MRI. Because of the short T2* of 19F spins, a ZTE sequence was selected for optimal SNR in 19F MRI.

On 1H-ZTE images of the in vitro sample, CPC cannot be distinguished from pig bone (Fig.2A). However, in 19F-ZTE images the PFCE in CPC becomes visible and overlaying the 19F- on 1H-ZTE images enables a proper sample evaluation (Fig.2B).

The 19F amount was quantified as 156911 a.u. for the bone phantom and 319422 a.u. for the reference sample. The 19F quantity ratio of bone phantom to reference is 49.1% (156911/319422). As the true ratio is 50% (10mg/20mg), this means that an excellent quantification accuracy (98%) can be achieved with 19F-ZTE imaging.

Similar to findings for the in vitro samples, CPC in rat legs cannot be differentiated on 1H-ZTE images (Fig.3A,C). However, also applying 19F-ZTE imaging the PFCE in the CPC allows visualization of this material (Fig.3B,D).

Discussion and Conclusion

In this study we have optimized the parameter settings for the combination of 1H and 19F ZTE MR imaging of 19F labeled CPC material administered to bone defects. CPC cannot be distinguished from bone in 1H MRI due to similar MR relaxation properties,1 but in combination with 19F MRI of 19F labeled CPC it is possible to analyze the fact of this material qualitatively and quantitatively. Further studies are ongoing to demonstrate that the PLGA&PFCE particles do not have serious adverse effects on CPC.

Acknowledgements

This work is supported by FP7-PEOPLE-2013-ITN for the project (607868) iTERM.

References

1. Sun Y, Ventura M, Oosterwijk E, et al. Zero echo time magnetic resonance imaging of contrast-agent-enhanced calcium phosphate bone defect fillers. Tissue Eng Part C Methods. 2013;19(4):281-7.

2. Srinivas M, Heerschap A, Ahrens ET, et al. (19)F MRI for quantitative in vivo cell tracking. Trends Biotechnol. 2010;28(7):363-70.

3. Srinivas M, Cruz L, Bonetto F, et al. Customizable, multi-functional fluorocarbon nanoparticles for quantitative in vivo imaging using 19F MRI and optical imaging. Biomaterials. 2010;31(27):7070-7.

4. Fram E, Herfkens R, Johnson G, et al. Rapid calculation of T1 using variable flip angle gradient refocused imaging. Magn Reson Imaging. 1987;5(3):201-8.

Figures

Example mono-exponential T2* fitting for PFCE&PLGA control

Representative 1H (A) and 1H + 19F (B) ZTE images of the pig bone sample filled with CPC+PFCE&PLGA in a falcon tube with gelatine. 19F is shown in false colour.

Representative 1H (A,C) and 1H + 19F (B,D) ZTE images of the rat legs in axial and coronal directions. The left leg was filled with CPC+PFCE&PLGA and the right one only with CPC. The 19F imaging time was less than 14mins. The presence of 19F is shown in false colour.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3877