Weiqiang Dou1, Simone Mastrogiacomo2, Olga Koshkina3, Andor Veltien1, Mangala Srinivas3, X. Frank Walboomers2, and Arend Heerschap1
1Radiology, Radboud University Medical Centre, Nijmegen, Netherlands, 2Biomaterials, Radboud University Medical Centre, Nijmegen, Netherlands, 3Tumor Immunology, Radboud University Medical Centre, Nijmegen, Netherlands
Synopsis
To enhance the MR
image contrast of calcium phosphate cement (CPC) to bone, it was tagged with a 19F
loaded nanoparticle as a contrast agent.
19F zero echo time (ZTE) MRI was applied for the first time for 19F
imaging of this CPC in bone at 11.7T. The T1 and T2*
relaxation times of the 19F nanoparticles were measured for optimal
settings of 19F ZTE MRI. We demonstrate overlaid 1H+19F
ZTE images of in vitro and ex vivo CPC/bone samples with hyperintense 19F
signals allowing a qualitative and quantitative evaluation of these samples. Introduction
Calcium phosphate
cement (CPC), a material used for bone defect restoration, has poor
MRI contrast relative to bone, due to similar T
1 and T
2
relaxation properties.
1 It is thus problematic to use MRI to assess CPC and surrounding
bone tissue. To enhance image contrast, CPCs have been tagged with conventional
gadolinium- or iron-oxide-based contrast agents, but these have unfavorable effects
on the properties of CPC and the MR evaluation of iron-oxide particles suffers
from a blooming effect.
1 As the benefits of
19F MRI to
follow fluorinated tracers are well known,
2,3 we explored if it was
possible to obtain
19F MR images of a
19F loaded
nanoparticle
3 as an alternative contrast agent in
1H/
19F
MRI of CPC.
The aim of
this study was to optimize
19F MRI of
19F labeled CPC.
For this purpose we first determined the
19F relaxation properties
of the
19F nanoparticle.
Materials and Methods
CPC consists
of 59.1wt% alpha-tricalcium phosphate, 1.5wt% carboxymethyl cellulose and
39.4wt% cryo-grinded poly-lactic-co-glycolic
acid (PLGA) particles. As 19F tracer we selected perfluoro-15crown-5-ether
(PFCE)2 embedded into PLGA nanoparticles through a single
emulsion reaction3. This was incorporated
into CPC at a concentration of 16.7%.
In vitro samples
A vertical hole
(diameter:3mm; height:3mm) was drilled in the upper
surface of a pig jaw block, which was filled with mixed material including CPC
(~50mg) and PFCE&PLGA (~10mg). The prepared pig bone was then embedded in a falcon
tube with 5% gelatine. Additionally, 20mg PFCE&PLGA particles were dissolved
into 500μm distilled water in an eppendorf tube as a control sample for T1 and T2* evaluation
and quantification analysis.
Ex vivo sample
A cylindrical
defect (diameter:2.5mm; depth:3mm) was drilled
in each leg of a Wistar rat cadaver. Defects were then filled with injectable
CPC in right leg and CPC+PFCE&PLGA in left leg.
Experiments
Measurements were performed on an 11.7T MR-system (Biospec, Bruker)
using a 1H-19F dual-channel volume coil with 40mm inner
diameter. 19F 3D-ultra-short TE (3D-UTE; TR=4ms, image resolution=1.56mm3)
was applied with six flip angles (FAs) from 2°-8° for T1 estimation
and with eleven TEs from 8-608µs with increments of 20-140µs
for T2*
evaluation.
1H/19F ZTE images for the bone phantom were
acquired with TR=2ms/4ms, image resolution=1.56mm3, 1average/16averages
and FA=2°/4° in 27seconds/13mins 53seconds. Additionally, the 1H/19F
ZTE imaging for rat legs were performed with TR=2ms, image resolution=0.31mm3/1.25mm3,
1averages/32averages and FA=2°/4° in less than 7mins/14mins.
Data Analysis
T1 values were calculated based on the
slope $$$e^{\frac{-TR}{T_{1}}}$$$ from the equation $$$\frac{S\left(FA\right)}{\sin\left(FA\right)}=e^{-\frac{TR}{T_{1}}}\times
\frac{S\left(FA\right)}{\\tan\left(FA\right)}+M_{0}\times\left(1-e^{\frac{-TR}{T_{1}}}\right)
\times e^{\frac{-TE}{T_2^*}}$$$4. A mono-exponential model $$$ S\left(TE\right)=S_{0}\times
e^{\frac{-TE}{T_2^*}}$$$ was used for T2*
fitting. The 19F signals in the bone phantom and control were
quantified in MRIcro software. Regions of interest (ROIs) were manually
outlined. 19F signal per ROI was quantified through multiplying the mean
pixel intensity by the pixel number of the ROI and slice thickness. The total 19F
amount was then summed over slices and expressed in arbitrary units (a.u.).
Results
From three measurements of the 19F nanoparticles, we
determined a mean T1 of 1269±38ms and a mean T2*
relaxation time of 76.5±8μs (Fig.1; R-squared goodness of fit > 0.99). From this T1 value, Ernst angle
settings were determined for 19F MRI. Because of the short T2*
of 19F spins, a ZTE sequence was selected for optimal SNR in 19F
MRI.
On 1H-ZTE
images of the in vitro sample, CPC cannot be distinguished from pig bone (Fig.2A).
However, in 19F-ZTE images the PFCE in CPC becomes visible and overlaying the 19F- on 1H-ZTE images enables a proper sample evaluation (Fig.2B).
The 19F amount
was quantified as 156911 a.u. for the bone phantom
and 319422 a.u. for the reference sample. The 19F quantity
ratio of bone phantom to reference is 49.1% (156911/319422). As the true
ratio is 50% (10mg/20mg), this means that an excellent quantification accuracy (98%)
can be achieved with 19F-ZTE imaging.
Similar to findings
for the in vitro samples, CPC in rat legs cannot be differentiated on 1H-ZTE
images (Fig.3A,C). However, also applying 19F-ZTE imaging the PFCE in
the CPC allows visualization of this material (Fig.3B,D).
Discussion and Conclusion
In this study we
have optimized the parameter settings for the combination of
1H and
19F
ZTE MR imaging of
19F labeled CPC material administered to bone
defects. CPC cannot be distinguished from bone in
1H MRI due to
similar MR relaxation properties,
1 but in combination with
19F
MRI of
19F labeled CPC it is possible to analyze the fact of this
material qualitatively and quantitatively. Further studies are ongoing to
demonstrate that the PLGA&PFCE particles do not have serious adverse
effects on CPC.
Acknowledgements
This work is supported by FP7-PEOPLE-2013-ITN
for the project (607868) iTERM. References
1. Sun Y,
Ventura M, Oosterwijk E, et al. Zero echo time magnetic resonance imaging of contrast-agent-enhanced
calcium phosphate bone defect fillers. Tissue Eng Part C Methods. 2013;19(4):281-7.
2. Srinivas M, Heerschap A, Ahrens
ET, et al. (19)F MRI for
quantitative in vivo cell tracking. Trends
Biotechnol. 2010;28(7):363-70.
3. Srinivas M, Cruz L, Bonetto F, et al. Customizable, multi-functional
fluorocarbon nanoparticles for quantitative in vivo imaging using 19F MRI and optical
imaging. Biomaterials.
2010;31(27):7070-7.
4. Fram E, Herfkens R, Johnson G, et al. Rapid calculation of T1 using
variable flip angle gradient refocused imaging. Magn Reson Imaging. 1987;5(3):201-8.